HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+, Page 5 of 12  Not logged in ELOG logo
ID Date Author Categorydown Subject Text Attachments
  28   Wed Mar 23 11:20:55 2022 SimonDetectorsPMT operating voltagesDuring the testing of the PMTs the following
operating voltages were applied:
  
  30   Wed Mar 23 16:16:42 2022 V. HannenDetectorsControl software for XUV and copper mirror detectorAttached you find a manual for the software
used to control the movement of the XUV detector
and the copper mirror detector. 
 Manual.pdf 
  31   Wed Mar 23 16:58:33 2022 V. HannenDetectorsStatus copper mirror and XUV detectorsOver the past days, the parabolic copper mirror
and XUV detectors haven been prepared and
tested for the upcoming beamtime (see first
 det-overview.jpglaserlab-crate.jpgtrigger-module.jpg 
  40   Thu Mar 31 11:49:54 2022 KonstantinDetectorsHV-Supply of particle detectors We cross-checked the cabeling of the particle
detectors north and south. Particle detector
south is connected to channel [B]HV5[/B]
  
  42   Thu Mar 31 15:53:50 2022 KonstantinDetectorsControl software for XUV and copper mirror detector> Attached you find a manual for the software
used to control the movement of the XUV detector
and the copper mirror detector. 
 XUV_Cu_GUI.png 
  53   Fri Apr 1 15:27:00 2022 RodolfoDetectorsNitrogen for UV-PMTsThere was a question concerning the nitrogen
we use to rinse the "Solar Blind PMTs".
 Stickstoff_f_PMTs.jpgStickstoff_Flussrate.jpg 
  54   Fri Apr 1 15:52:23 2022 RodolfoDetectorsUV PMT SouthThe "Solar Blind PMT" labeled as "PMT South"
has been installed by Simon and Max in the
gas target chamber on a CaF Viewport. The
 PMT_Sued_Target.jpg 
  55   Fri Apr 1 16:46:33 2022 CarstenDetectorsAll pressuzied air drives are tested now and are working properlyAll pressuzied air drives are tested now and
are working properly
  
  57   Fri Apr 1 18:30:38 2022 CarstenDetectorsPositions / Voltages threshlds for particle detectors[B]DC beam 238U92+ (205kV, e-beam)[/B]
We used the detector NA_SZ to determine the
central primary-beam (238U92+) position by
  
  58   Fri Apr 1 22:01:14 2022 KonstantinDetectorsPMT operating voltages> During the testing of the PMTs the following
operating voltages were applied:
  
  59   Fri Apr 1 23:12:37 2022 PatrickDetectorsScraper positions "only primary beam"
Scraper positions where there is only primary
beam for detector tests. Use only one outer
  
  60   Fri Apr 1 23:47:45 2022 CarstenDetectorsPMT thresholds / PMT north does not allow voltage abow ~ -1kVChecked thresholds of PMT with beam and without
beam.
  
  77   Sun Apr 3 11:40:48 2022 Shahab, YuriDetectors245 MHz Resonator tuned to new frequencyresonance frequency changed from 243.2 MHz
to 244.692 MHz by reducing the value of both
of the plungers from 395 to 325. Now after
 Screenshot_2022-04-03_at_11.24.15.png 
  97   Tue Apr 5 13:24:40 2022 KristianDetectorsPMT CuHV cable was in wrong channel in power supply.
Now it is back to the correct position. Checked
connection at the PMT.
  
  115   Wed Apr 6 19:39:08 2022 Ken Ueberholz DetectorsLED installed The UV-LED of Münster was connected over 
channel 8 to the Messhütte. 6V are 
supplied by the power supply (picture 2).
 20220406_193541.jpg20220406_193439.jpg 
  116   Wed Apr 6 21:03:25 2022 Ken Ueberholz DetectorsXuv-Detektor, further information Some further information regarding the 
xuv-detector and its components.
 xuv_detektor.pdf 
  129   Thu Apr 7 05:55:17 2022 Ken UeberholzDetectorsXUV-Detector Mesytec crashWhen the XUV-detector has to shut down unexpectedly,
it can happen that the Mesytek power supply
crashes and can't be 
  
  137   Thu Apr 7 14:01:22 2022 Sebastian, MaxDetectorsMesytec resetThe Mesytec crate was turned off and on again.

It is again possible to communicate with
  
  139   Thu Apr 7 14:13:27 2022 DanyalDetectorsnew PMT southI have replaced the PMT south (looks at the
gas target). 
in the hope that this will produce less dark
  
  141   Thu Apr 7 16:45:10 2022  DetectorsDetector distancesRelativ distances of the detector   20220407_164345.jpg 