137
|
Thu Apr 7 14:01:22 2022 |
Sebastian, Max | Detectors | Mesytec reset |
The Mesytec crate was turned off and on again.
It is again possible to communicate with the XUV detector via python to apply voltage and current. |
136
|
Thu Apr 7 11:11:31 2022 |
Max, Sebastian | Laser | dye changed, pulse energy measurements |
We changed the dye at 10am.
We increased the pump power for the dye Laser to 480mJ (before 450mJ). New calibration factor is 29 for Quanta Ray monitor signal.
We get now 17mJ UV pulse energy after SHG. Pulse energy before was 14mJ. |
135
|
Thu Apr 7 10:10:31 2022 |
Ragan, Yuri, Danyal | Accelerator | Problem in UNILAC |
The power supply of the UNILAC magnet is broken. The repair will take up to 6-7 hours. |
134
|
Thu Apr 7 09:58:55 2022 |
Ragan, Yuri, Danyal | Accelerator | Kicker problem solved |
The kicker problem is solved and the beam is back in the ESR. |
133
|
Thu Apr 7 09:23:18 2022 |
Sebastian | DAQ | DAQ1 & DAQ2 - 1kHz and delayed cs removed |
Questionable 1kHz and delayed common stop signals has been removed.
These two cablings were for some previous tests. |
132
|
Thu Apr 7 08:43:19 2022 |
Danyal, Yuri, Ragan | Accelerator | again no injection in ESR - ESR kicker problem |
we had no injections in the ESR again
it turned out that the injection kicker of the ESR has a problem |
131
|
Thu Apr 7 08:14:03 2022 |
Yuri & Danyal | General | 229Th89+ line in Schottky |
GLOBAL calculations for the "content" of the Th line in the Schottky spectrum
(380 MeV/u exit energy, equilibrium charge state distribution)
Actinium 9.9% - bare
Thorium 42.4% - H-like
Protactinium 44.8% - He-like
Uranium 2.7% - Li-like |
130
|
Thu Apr 7 07:40:09 2022 |
Danyal, Max, Sebastian | DAQ | No beam and DAQs crashed |
since shortly after 7 AM we have no beam in the ESR
(reason unclear)
Both DAQs also crashed.
Beam back in ESR at 7:20 or so
Problems with restarting DAQ1
Problems with restarting DAQ2
Files names 7801 and 7804 and 0047 and 0050 cannot be used |
129
|
Thu Apr 7 05:55:17 2022 |
Ken Ueberholz | Detectors | XUV-Detector Mesytec crash |
When the XUV-detector has to shut down unexpectedly, it can happen that the Mesytek power supply crashes and can't be
communicated with. If this happens the Mesytec PS needs to be restarted from hand inside of the ESR. This can be done by
restarting the crate located at the bottom of the XUV-detector.
This crash happened during the night shift today and the Mesytec should be restarted today as soon as possible. For
example when the laser stabilization gets fixed.
In normal operation this should never happen if no mistakes are made. |
128
|
Thu Apr 7 05:44:40 2022 |
Pierre-Michel, Ragan | DAQ | NTCAP settings |
Carrier Frequency: 244M
IQ Rate: 10M
reference Level: -30
RF - Attenuation: 40
IF - Attenuation: 19.9288
Masterpath p:\E142\sc
Masterpath p:\E142\IQ
Display: No Display
NTCAP data-taking is stopped at 5:44.
IQ_2022-04-07_02-40-05
SC_2022-04-07_02-40-05 |
127
|
Thu Apr 7 05:23:50 2022 |
Patrick | DAQ | How To Start MBS for DAQ2 (Carstens DAQ) |
(MBS server should be running at the LXG0155 machine. If not you will get an error message. Then start the server - last command in console)
Just enter the following commands in the console of any PC
ssh stoe_exp@r4l-68
*** Password on a sheet of paper ***
resl
cd /esr/usr/stoe_exp/E142Brandau/SCA... (auto complete with tab)
mbs
@startup
connect rfio lxg0155 -disk
open file lxg0155:/data.local2/E142/229Th -auto -rfio
sta ac
To stop MBS enter the following commands
sto ac
close file
disconnect rfio
@shutdown
exit
resl |
126
|
Thu Apr 7 04:48:58 2022 |
| | ToDo's for 07.04. |
Some ToDo's for the next shift:
- Work on Laser stabilisation 2nd stage in the ESR
- Mesytec power supply of the xuv-detector crashed and needs to be restarted in the ESR. Simply turning it the crate off and on again)!!
- Fix wavemeter fluctuates
- Compare MBS event size of DAQ 1 ("by laser group") and DAQ 2 ("by Carsten"), DAQ 1 data size was ~2 times larger than DAQ 2 data size.
- Use lmd-files measured last night to test Go4-Analysis.
- Practice measurement procedure with beam, see Message ID: 124. |
125
|
Thu Apr 7 04:10:18 2022 |
Ken Ueberholz | DAQ | Corrected cabling DAQ 2 |
A lot of cables were wrongly or not
connected to the DAQ 2.
Channels are now for both DAQ:
0 XUV-Back
1 XUV-Anode
2 XUV-Anode-delayed
3 PMT-South
4 PMT-Middle
5 PMT-North
6 PMT-Cu
7 Part-det-south
8 Part-det-north
9 1kHz
10 -
11 -
12 Delay-common-stop
13 PD-NO
14 PD-SO
15 Q-switch
All data taken by the DAQ2 has to be
considered as faulty up to now.
Carsten was furious. |
Attachment 1: 20220407_043337.jpg
|
|
124
|
Thu Apr 7 03:44:21 2022 |
Patrick | DAQ | Measurement procedure |
Measurement procedure
- The laser hits the screen 'NO' (atppc017) right side of the burn spot
- Be ready to start a laser scan
- Scan laser goes into wavemeter!
- Have mbs connected
- Have go4 connected
Preparation done
- 'open file (...)' for DAQ1 & DAQ2
- request a new beam
- check scraping
- 'sta ac' for DAQ1 & DAQ2
- start laser scan
laser scan started
- 'sh ac' for DAQ1 & DAQ2
- check that acquisition is RUNNING and file is OPENED
- fill in table information in the ELOG.
laser scan complete
- 'sto ac' for DAQ1 & DAQ2
- 'close file'
- check if preparation is still fulfilled
- repeat from Preparation done |
123
|
Thu Apr 7 03:09:36 2022 |
Carsten | DAQ | NTCAP running since 02:40 |
|
122
|
Thu Apr 7 03:06:15 2022 |
Patrick | DAQ | DAQ file sizes |
The file sizes of DAQ1 (229Thor) and DAQ2 (229Th, "Carstens DAQ") are different.
During 20min of laser scanning DAQ1 saves 4 files while DAQ2 saves 2 files which was 3.3 GB and 2 GB, respecively.
The file sizes of DAQ2 should always be greater than those of DAQ1, because of extra ADC data!
The problem seems to be solved after the recabling ID: 125. |
121
|
Thu Apr 7 02:32:29 2022 |
all on shift | Runs | First Scans 322-320nm |
Laser scanning from 322-320 steps 0.01nm/s
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th
Time | SIS-intensity | cooler-HV | LMD-DAQ1 | LMD-DAQ2
|
02.30 | ? | ? | Th227781-Th227784 | 0027-0028
|
03.03 | 1.75e9 | 209648 | 7785-7787 | 0029-0030
|
03:28 | 1.7e9 | 209648 | 7788 | 0031 (stopped during scan)
|
Break - fixing cables, Elog ID: 125
|
04:00 - problems with UNILAC, stopping for now.
|
Apparently not so bad problems. we can continue.
|
Mesytek PS of xuv-detector crashed, can only be restarted from inside the ESR by turning the crate on and off.
|
05.20 | 1.23e9 | 209646 | 7791-7794 | 0035-0038
|
05.52 | 1.5e9 | 209648 | 7795-7797 | 0039-0042
|
06.24 | 1.5e9 | 209642 | 7798-7800 | 0043-0046 | 1 sec waiting at each laser frequency (dwell time 30)
|
07.36 | xx e9 | 209 xx | 7803-7804 | 0047-0050 | problems with DAQ 1 and DAQ2
|
07.52 | xx e9 | 209646 | 7805-7809 | 0051-0055 | DAQs started again | 5 sec waiting at each laser frequency (dwell time 150)
|
08.30 | e9 | 209 xx | 7810-78 | 0056-00 | DAQs started again | 5 sec waiting at each laser frequency (dwell time 150)
|
|
|
120
|
Thu Apr 7 01:19:05 2022 |
Patrick | Laser | Laser scanning |
We did some laser scans. The wavemeter shows spikes during scans. This may be just while the frequency changes. See picture. |
Attachment 1: 2022-04-07_Wavemeter_during_scan.png
|
|
Attachment 2: 2022-04-07_Wavemeter_during_scan2.png
|
|
Attachment 3: 2022-04-07_Wavemeter_during_scan3.png
|
|
Attachment 4: 2022-04-07_Wavemeter_during_scan4.png
|
|
119
|
Thu Apr 7 00:36:36 2022 |
Uwe | DAQ | Threshholds and shaping Mesytec for ADC |
|
Attachment 1: snapshot3.png
|
|
118
|
Thu Apr 7 00:15:11 2022 |
Patrick | Laser | Laser timing adjusted |
We have adjusted the laser timing using PMT-middle. Bunching freq. is 3.90362 MHz.
PD NO: 7.79
PD SO: 43.37
mean: 25.08
PMT middle: 25.02
Delta = 0.6 bins * 3.3 ns = 0.2 ns
New injection:
PMT middle: 25.48 / 101.91
ParDet south: 74.31 / 151.31
ParDet south - PMT middle (furthest / closest distance) = 161.1 ns / 91.1 ns |
Attachment 1: 220407-Laser-ion-timing.png
|
|
Attachment 2: 220407-Laser-ion-timing-full-period.png
|
|
Attachment 3: 220407-ParDetSO_PMT-middle.png
|
|