  235   Fri Apr 15 20:43:18 2022 KonstantinLaserDye changed
Dye of the laser system has been changed. UV laser power increased from ~8 mJ to 11 mJ pulse energy.
  67   Sat Apr 2 17:47:16 2022 Ken Ueberholz  Telefon, Gästehaus
My guesthouse phone number: 06159-71-2957
  73   Sun Apr 3 00:41:45 2022 Ken Ueberholz  Xuv-Detektor Signale
The Xuv-detector cabel channels seem to be 
wrongly labeled in the Messhütte (see 
picture) and are propably also wrongly 
connected. Need to be relabeled and reconnected either in the 
messhütte or in the laserlab. I will look for the best 
solution tomorrow morning. 
  81   Sun Apr 3 15:34:34 2022 Ken Ueberholz  Xuv-Detektor CFD-treshhold
I co controlled all detector signals and 
adjusted the threshold to be just above 
the noise. I will double check this when 
the detector can be moved in and we have a 
beam. 
  115   Wed Apr 6 19:39:08 2022 Ken Ueberholz DetectorsLED installed
The UV-LED of Münster was connected over 
channel 8 to the Messhütte. 6V are 
supplied by the power supply (picture 2). 
The Led is located at the xuv detector 
(picture 1).
  116   Wed Apr 6 21:03:25 2022 Ken Ueberholz DetectorsXuv-Detektor, further information
Some further information regarding the 
xuv-detector and its components.
  84   Mon Mar 24 06:22:21 104 Ken UeberholzDetectorsXUV-Detector Rates
Count rates were measured for the XUV-
detector for a full circle:
Cathode: -5V, shields: 0V, main coil 8A, 
sec coil 2A, 
No beam - pos 0mm: ~10 Hz
Hot beam - pos 0mm: ~5 kHZ
Cool beam - pos 147mm: ~200Hz
  37   Tue Mar 29 12:47:26 2022 Ken Ueberholz Parameter setting for XUV and copper mirror detector
Some Default Settings for the XUV 
Detector. The copper 
mirror should always move fully in.

Possible positions: Between 0mm and 147mm. 
Background was 
very high for anything higher than 100mm 
in the last 
experiment. 50mm was optimal. This should 
be tested again 
with repaired and modified shields. 

Good default values for the XUV detector 
parameters have 
been proven before to be:

Cathode: -5V
Movable shield: 0V
Fixed shield: 0V
Main coil: 8A
Secondary coil: -2.4A

(negative values set in hardware, only 
positive software 
inputs are possible)
  85   Mon Apr 4 00:19:16 2022 Ken UeberholzDAQFirstTests
New measurement:
Start: About 00:12
Stop: About 00:43
LMDfile: 229ThFirstTest7775.lmd-229ThFirstTest7776.lmd
First tries to see the bunched ions with full test experiment circle and detectors.
Bunching visible in xuv-anode, sa_sz, PMT-south, PMT-north.
Bunches seem to be very wide, possible wrong revolution frequency set in the DAQ.
PD NO doesn't see the laser, screen prop. still in front of the window.
7e8 ions in the SIS
Standard XUV-settings




Several DAQ crashes and errors Confused




New measurement:
Start: About 02:00
Stop: About 02:30
LMDfile: 229ThFirstTest7780.lmd
Same test with new freq: 3.90415MHZ and without screen.
Nice Bunches in particle detector. Broad bunches in detectors.
Errors in DAQ
PD not working correctly or laser wrongly setup.
4.3e8 ions in the SIS
Standard XUV-settings
  125   Thu Apr 7 04:10:18 2022 Ken UeberholzDAQCorrected cabling DAQ 2
A lot of cables were wrongly or not 
connected to the DAQ 2. 

Channels are now for both DAQ:
0 XUV-Back
1 XUV-Anode
2 XUV-Anode-delayed
3 PMT-South
4 PMT-Middle
5 PMT-North
6 PMT-Cu
7 Part-det-south
8 Part-det-north
9 1kHz
10 -
11 -
12 Delay-common-stop
13 PD-NO
14 PD-SO
15 Q-switch

All data taken by the DAQ2 has to be 
considered as faulty up to now. 

Carsten was furious.
  129   Thu Apr 7 05:55:17 2022 Ken UeberholzDetectorsXUV-Detector Mesytec crash
When the XUV-detector has to shut down unexpectedly, it can happen that the Mesytek power supply crashes and can't be 
communicated with. If this happens the Mesytec PS needs to be restarted from hand inside of the ESR. This can be done by 
restarting the crate located at the bottom of the XUV-detector. 

This crash happened during the night shift today and the Mesytec should be restarted today as soon as possible. For 
example when the laser stabilization gets fixed.

In normal operation this should never happen if no mistakes are made. 
  146   Fri Apr 8 01:18:51 2022 Ken UeberholzDAQTODO: DAQ improvements
For somebody who knows the DAQ1 code: Go4 resets the histograms with each laser step. Can there be additional histograms where the data gets accumulated?

And possibly a histogram where important detectors are plotted against the laser steps? The read back wavelength of the wavemeter is not very precise. 
  41   Thu Mar 31 13:28:17 2022 Jan GloriusDAQDAQ-2 cabling docu
newest version V2 - valid since Mo 04.April 2022
  62   Sat Apr 2 01:27:25 2022 Jan GloriusRunstest runs 02.04.2022
Beam settings:
primary beam 
copper stripper
I_el=300mA
U_el=205.490kV (Ohm_labs)
3.5e7 ions at injection


MSCF/ADC settings: 
ch2 < Mid PMT x10
ch6 < Mid PMT x100
see attachment (setting_e142....) for detailed MSCF settings

Files/Runs [DAQ2]:

------------------new file-----------------
229Th0022.lmd
bunched beam [ESR in manip. at 7]
beam scraped using detector drives
-SA pos: 0
-NA pos: -10
-Ni pos: 40

no corresponding file in DAQ1

------------------new file-----------------
229Th0023.lmd
bunched beam [ESR in manip. at 7]
beam scraped using detector drives
-SA pos: 0
-NA pos: -10
-Ni pos: 40
laser scan active:
- steps: 21
- cycles: 5

file in DAQ1:
...7768.lmd

------------------new file-----------------
229Th0024.lmd
bunched beam [ESR in manip. at 7]
beam scraped using detector drives
-SA pos: out (-127)
-NA pos: -10
-Ni pos: 40
laser scan active:
- steps: 21
- cycles: 5

file in DAQ1:
...7770.lmd

------------------new file-----------------
229Th0025.lmd
bunched beam [ESR in manip. at 7]
beam scraped using detector drives
-SA pos: 0
-NA pos: -10
-Ni pos: 40
laser scan not:

VUPROM -> delay stop trigger ch12 

file in DAQ1:
no file

no laser scan

------------------new file-----------------
229Th0026.lmd
bunched beam [ESR in manip. at 7]
beam scraped using detector drives
-SA pos: 0
-NA pos: -10
-Ni pos: 40
laser scan active:
- steps: 21
- cycles: 5

VUPROM -> delay stop trigger ch12 

file in DAQ1:
...7772.lmd
  90   Mon Apr 4 17:26:48 2022 Jan GloriusDAQMSCF shaper settings
attached are the current shaper settings.

list of currently connected channels:
ch2: mid x1
ch4: Cu  x1
ch6: mid x2
ch8: Cu  x2
ch9: XUV x1
ch10:XUV x2
  15   Thu Mar 10 15:20:26 2022 Jan DAQanalog branch signal patches to Messhütte
The additional signals from PMTs and XUV detector needed for spectroscopic analysis in the second DAQ are patched to AP Messhütte (MH) via the red patch panel at the inside of the target region. The patch channels F1 to F8 (also labeled A to H in red on white) are used and extended by a cable bundle to reach below the mirror chamber. Inside the MH the signals arrive at the red patch panel and are forwarded to the blue patch panel close to the laser DAQ using channels 1 to 8 (also labeled A to H).
 
The entire connection for these 8 channels from the mirror chamber to the blue panel has been tested with a pulser signal of ~115mV amplitude. After transfer the amplitude has decreased to ~40mV, see also attached fotos.

The signal transport time from the mirror chamber to the blue panel is ~407 ns. This was measured by patching a logic puls from ESR to MH and back to ESR, as show in the delay_ESR... foto attached.  
  27   Mon Mar 21 21:48:46 2022 JanDAQMesyTec Shaper Control
The MesyTec Shaper & Constant Fraction (MSCF) is used to process the PMT/det signals for the ADC in the second DAQ.
It can be setup on its front panel, but single channel setup is very tedious.
Remote control of the settings is possible using an epics server and client.

All of the following is done with the "litv-exp" user:

server: atpnuc001 (in the screen session "E142")
client: lxg1275 

The server should run undisturbed, but in case of a power failure, you can start the server like this:
> ssh litv-exp@atpnuc001
> cd e142
> ./E142_start_ioc.sh

To check server:
> ssh litv-exp@atpnuc001
> screen -x (CTRL-a CTRL-d to leave screen)

To connect the client gui for remote control, do:
> ssh litv-exp@lxg1275
> E142_epics
The panel is more or less self-explaining. 
One thing has to be noted: the Monitor Channel selection is shifted by -1.
I.e., [ch 1] is button [0], etc.

The MSCF settings can be stored and loaded to/from file by using a bash-script.
The settings will be saved/loaded from: 
/u/litv-exp/e142/mscf_settings/
files names are "setting_e142.<linux_timestamp>"

The scripts are also in /u/litv-exp/e142/mscf_settings/
use 
./mscf_save.sh                   for saving the current settings to a new file
use 
./mscf_load_last.sh              to load the last settings file (by timestamp in the filename) 
use 
./mscf_load_file.sh <filename>   to load the settings of a specific file
  156   Fri Apr 8 09:33:52 2022 Dayshift 1 8.4.2022 Reference Measurements
DAQ1: lxg1297:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th

center freq. 242.2M, 10MIQs, -30dBm
Shottky marker pos: 243.973MHz

TimeSIS-intensitycooler-HVLMD-DAQ1LMD-DAQ2scan range(nm)step width(nm)dwell time(s)comments
09:30??20961400797817323.2-321.20.0142 loops
10:152.3e920961100817820321.0-320.00.018just 1 loop, aborted 2nd loop and started from 320.5-320.4 in 0.005 steps and 10s dwell time (300), started again from 320.5-320.4 in 0.005 steps and 10s dwell time (300)


We crosschecked the last measurements from the nightshift for reference measurements.
Now the ESR group start switching to 1 bunch.
  132   Thu Apr 7 08:43:19 2022 Danyal, Yuri, RaganAcceleratoragain no injection in ESR - ESR kicker problem
we had no injections in the ESR again 

it turned out that the injection kicker of the ESR has a problem 
  130   Thu Apr 7 07:40:09 2022 Danyal, Max, SebastianDAQNo beam and DAQs crashed
since shortly after 7 AM we have no beam in the ESR
(reason unclear) 

Both DAQs also crashed. 

Beam back in ESR at 7:20 or so 

Problems with restarting DAQ1 

Problems with restarting DAQ2 

Files names 7801 and 7804 and 0047 and 0050 cannot be used