Thu Jun 13 02:05:04 2024, Carsten, Accelerator, Short Lifetime at 100 mA
|
Lifetime still is short |
Thu Jun 13 18:23:46 2024, Julian, Laser, Restarting the laser controller
|
If the laser controller is locked up again, the little display will show an error message after an attempted medusa scan.
To restart it, follow these steps:
1.) Unlock the beam stabilization.
|
Sat Jun 1 19:19:07 2024, Carsten, Detectors, Reminder: Check signals in scope and in DAQ of detectors with voltages switched off (vs. CFD threshold)
|
When the ion pumps and vacuum gauges are switched off the scaler data of
PMT_south, PMT_middle, PMT_north, PD_NO, PD_SO, PartDet North, PartDet South and (Channeltron) should be checked without voltages applied in dependence
on the CFD thresholds to determine the "cable noise level" at which thre are NO entries other than 0 in hte spectra (minimum background).
|
Sun Jun 9 10:20:38 2024, Carsten/Danyal, Detectors, Recap from yesterdays work with test beam (Detectors / beam /DAQ)
|
In the ESR we had thw whole day (saturday) a 400MeV/u bunched 238U92+ test beam (about 2 MHz). The RF signal is cabled as input to the DAQ (seperate post/entry).
Intensity is artificially reduced to a few 10^6 (with some tests alsodone at 10^5).
In all the tests the both particle detectors were moved in pus the empty pocket on the inside to mimick the background conditions during the run (aka no |
Sat Jun 1 18:00:20 2024, Carsten, DAQ, ReadOut Trigger im Timer Scaler changed because of faulty signal from frequency generator
|
Old readout trigger (~200 Hz) in 1MHz Scaler showed two nearby peaks indicating a jitter of the readout frequency or same faulty output driver of the frequency
generator Metrix GX240 used to produce the readout frequency (see below).
[IMG]elog:9/1[/IMG]
|
Thu Jun 13 12:05:31 2024, Konstantin, Peter, Jonas, Accelerator, RF-amplitude
|
To realize the shortest bunches possible, we check the bunch length as a function of two different bunching amplitudes. With the bucket fill set to 1.0
we observe the spectra compiled in attachment 1. The second attachment is recorded with a bucket fill 0f 0.2.
|
Thu Jun 13 10:42:02 2024, Danyal, General, Question
|
What is the ion beam lifetime, according to the DCCT and Schottky?
Can somebody measure it, please?
Yesterday we could not see a fast drop in the "main" beam.
|
Sat Jun 8 13:11:52 2024, Carsten, General, Put the go4login in atplaser .bashrc
|
Put the go4login in atplaser .bashrc in order to avoid that one always has to go to the go4 directory to set the environment variable.
go4login64 is now executed automaticlally and the environments are now set every time you open a (bash) shell.
A fixed version 6.3 is used. The login file might be changed back after the beamtime. |
Thu Jun 13 22:08:07 2024, Dayshift, Accelerator, Problems with Accelerator?
|
It seems to be that there are some problems with the injections in the ESR.
The last 2 accumulation cycles showed that wheter only a small amount of ions are injected into the ESR or the injection is missing.
|
Wed Jun 26 11:59:12 2024, Ruijiu, Shahab, Carsten, Analysis, Prepare plots for Thomas's presentation 6x
|
data:
(env_OnlineDataAnalysisSystem) litv-exp@lxbk0497:/lustre/astrum/experiment_data/2024-05_E018/OnlineDataAnalysisSystem/data/iq/IQ_2024-06-
12_21-47-54$
|
Thu Jun 6 09:39:23 2024, Carsten, DAQ, Preliminary Channel Table (DAQ1 and DAQ2)
|
|
Sun Jun 9 10:08:24 2024, Carsten/Danyal, Detectors, Particle detectors - Recap of yesterdays work
|
[B]South (behind cooler) - "SA_SZ" - DD2...[/B]
We have put the particle detecotr North to the poistion behind the cooler (north) to replace the defective particle detector.
This detector is an old "Kozhuharov" style plastic szintillator that runs with a voltage of 1400-1500 V.
|
Wed Jun 5 19:22:43 2024, Carsten, Konstantin, Rodolfo, Detectors, PMT tests and preliminary thresholds
|
For all channels of the CAEN CFD the width of the logic signal is set to 255ns (maximum) to mitigate potential double pulses. For the very low count rates
of the experiment this is not problem.
|
Mon Jun 3 18:24:44 2024, Rodolfo, Detectors, PMT arrangement
|
The picture shows the actual configuration of the PMTs.
We are using a 4-Channel High-Voltage supply. The channels are labeled according to the PMTs names.
|
Tue Jun 4 22:54:02 2024, Rodolfo, Detectors, PMT - Manufacture - Test Result Data
|
|
Wed Jun 12 21:21:15 2024, Yuri, Detectors, NTCap started
|
Fcenter - 244.6 MHz
BW - 10 MHz |
Fri Jun 14 06:38:41 2024, Konstantin Mohr, Bernhard Maass, Peter Micke, Lukas Kau, Dounia Boudefla, Runs, Morning runs
|
Very successful shift, no crashes until 11:10
Previous night shift had problems with the pulse laser starting around 3am.
Something was loose at the FCU. Rodolfo was called in and was able to fix it.
|
Mon May 27 13:37:31 2024, Jan, DAQ, MesyTec Shaper Control
|
For the 2nd DAQ the MesyTec Shaper ([URL=https://elog.gsi.de/esr/NucAR+lab/230418_110112/MesyTec_MSCF16-F-V.pdf]MSCF manual[/URL]) is used as a spectroscopy
amplifier and CFD.
It is controlled via the MesyControlGUI, which is installed on SADPC055 with a USB->mesyBUS connection to the MSCF.
|
Thu Jun 13 12:37:14 2024, Ruijiu Chen, David, Analysis, Lifetime of beam in center of mass is 193 s.
|
|
Thu Jun 13 14:41:22 2024, Ruijiu Chen, David, Analysis, Lifetime of beam during June 13 9:00~11:00
|
|