Entry  Thu Jun 6 19:47:01 2024, Carsten, DAQ, Exchange of Mesytec CFD/shaper - noise on signal lines from ESR magnets 
Jan exchanged the Mesytec CFD/shper since it could not be "remote controlled" anymore. The new one has different shaping times and delays.
It is noted the threshold range is rather small. 
With the new  shaper only the x100 signals could meaningfully discriminated etc.
Entry  Thu Jun 13 14:56:09 2024, Stefan Schippers, Accelerator, Estimated beam lifetime: About 150 s 
Estimate of storage lifetime using my code 'hydrocal':

 ion mass in u ...................................: 229
Entry  Thu Apr 25 16:34:05 2024, Rodolfo, Laser, Efficiency of Fluorescein27 Fluorescein27_Efficiency.png
The figure below shows the efficiency of Fluorescein27 vs. Wavelength. Each point was averaged over 500 laser shots.

Dye Laser: Sirah, Cobra Strech. 
Entry  Thu Jun 6 19:51:11 2024, Carsten, Accelerator, ESR crew left after persisting problems with ESR magnets 
No beam today.--.
Entry  Wed Jun 5 14:17:36 2024, Carsten, Accelerator, Distance Cooler Detector 
Calculation from Helmut Weick (copied from E142)
Distance e-cooler to particle detectors chamber:
“This is half of ESR plus distance target - detector. This even is on the web page”
Entry  Fri Jun 14 13:14:29 2024, Konstantin, Bernhard, Peter, Lukas, General, DAQ crashes at :10 
10 Minutes past a full hour (XX:10) the windows PCs (sometimes) loose connection.
This causes the labview server & the laser remote ctrl  to crash.
Entry  Sat Jun 1 18:14:01 2024, Carsten, Analysis, DAQ and analysis directories 
[U][B]DAQ1 (original):[/B][/U]
-user atplaser
-runs on lxg1297
Entry  Fri Jun 14 11:44:23 2024, Carsten Brandau, Accelerator, Cooling Time in SC7 increased from 9s to 12s 
Cooling Time in SC7 increased from 9s to 12s
Entry  Thu Jun 13 18:03:36 2024, Afternoon shift, Runs, Continuation: First scans 2024_06_13_229Th_Lifetime_at_190MeV_afterscraping.png2024-06-13_23-13-55-242.png
[TABLE border="1"]
Time | Part. Det. Int | ESR-intensity| cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|-
Entry  Sat Jun 1 19:00:57 2024, Carsten, DAQ, Change of (offline) frequency for common stop (CS) of VUPROMs (and 2nd TDC) and scaling factor 
Temporarily, for tests of the second DAQ I changed the frequency from the frequency generator and scaling factor  that is used for the common stop signal
to a value that is close to the expected value in the experiment.
Reason: For the second TDC (CAEN 25ps) TDC the time between to CS events must be of the order of 45 us. The TDC can store events up to about 51,x us and
Entry  Wed Jun 12 22:48:55 2024, Carsten, , Beam Halo 
The seamingly short live time reported in [URL=https://elog.gsi.de/esr/E0052/53]entry 53[/URL] is due to a halo from a different isotope (?). It can be
scraped away from the outside. The 229Th89+ primary beam can be scaped away from the inside and the halo stays. There is some spacial headroom between
the two components.  
Entry  Fri Jun 14 14:44:51 2024, Afternoon shift, Runs, Afternoon shift 
Pizza shift

[B]the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
Entry  Sun Jun 2 00:01:46 2024, Carsten, DAQ, 25ps TDC CAEN V1290N DAQ2 lead to crash in DAQ2 due to a hardware issue 
On the CAEN V1290N TDC the piggy board for the lower 8 channels is not fixed well to the motherboard.
That prevented to the whole DAQ2 to trigger correctly or even crash every once in a while. Long attempt sot find the error in hard and in software.
Entry  Thu Jun 13 10:32:29 2024, Konstantin, Accelerator, (Proper) Cooler settings 
We change the acceleration voltage of the electron cooler and observe the platform high voltage and the count rate at the particle detector south.
Cooler current at 100mA.