27
|
Thu Jun 6 19:47:01 2024 |
Carsten | DAQ | Exchange of Mesytec CFD/shaper - noise on signal lines from ESR magnets |
Jan exchanged the Mesytec CFD/shper since it could not be "remote controlled" anymore. The new one has different shaping times and delays.
It is noted the threshold range is rather small.
With the new shaper only the x100 signals could meaningfully discriminated etc.
After start of the ESR (magnets etc.) the signals of the 10x channels see quite some noise that makes it more difficult to adjust the thresholds for the low signals. |
67
|
Thu Jun 13 14:56:09 2024 |
Stefan Schippers | Accelerator | Estimated beam lifetime: About 150 s |
Estimate of storage lifetime using my code 'hydrocal':
ion mass in u ...................................: 229
ion nuclear charge ..............................: 90
ion charge state ................................: 89
beam energy in MeV ..............................: 45501.1
beam energy in MeV/u ............................: 198.695
beam beta .......................................: 0.566309
beam rigidity in T m.............................: 5.4944
ring circumference in m .........................: 108.36
ring acceptance in mrad .........................: 2
ring temperature in °C ..........................: 25
ring vacuum pressure in mbar ....................: 3e-11
residual gas density in m^-3 ....................: 7.28767e+11
residual gas composition
Z %
1 93.43
6 2.23
7 0.75
8 3.29
18 0.3
cooler length in m ..............................: 2.5
electron current in A ...........................: 0.2
magnetic expansion factor .......................: 1
electron-beam diameter in mm ....................: 50.8
electron density in cm^-3 .......................: 3.62844e+06
cooler cathode temperature in meV ...............: 110
cooling energy in eV ............................: 109000
space_charge (0th iteration) in eV ..............: 44.3477
space_charge (1st iteration) in eV ..............: 44.341
transverse temperature in meV ...................: 110
longitudinal temperature in meV .................: 0.2
RR nmin .........................................: 1
RR lmin .........................................: 0
RR nele [no. electrons in subshell (nmin,lmin)] .: 1
RR nmax .........................................: 128
RR enhancement factor ...........................: 2.5
RR alpha in cm^3 s^-1 ...........................: 8.04602e-08
beam lifetime due to multiple scattering in s ...: 1.86e+06
beam lifetime due to single scattering in s .....: 7.51886e+07
beam lifetime due to charge capture in s ........: 3.5516e+06
beam lifetime due to stripping in s .............: 1e+99
beam lifetime due to recombination in cooler in s: 148.466
beam lifetime with electron cooling in s ........: 148.46
beam lifetime without electron cooling in s .....: 1.20121e+06
detector count rates for 1E6 stored ions and a residual gas pressure of 3e-11 mbar:
recombination detector from ecool (Hz) ....: 4575.41
recombination detector from res.gas (Hz) ....: 0.0545666
total rate on recombination detector (Hz) ....: 4575.46
ionization detector from res.gas (Hz) ....: 1.93798e-94 |
3
|
Thu Apr 25 16:34:05 2024 |
Rodolfo | Laser | Efficiency of Fluorescein27 |
The figure below shows the efficiency of Fluorescein27 vs. Wavelength. Each point was averaged over 500 laser shots.
Dye Laser: Sirah, Cobra Strech.
- Solvent: Ethanol + H2O.
- Concentration Oszillator: 0.4 g + 0,2 g/l NaOH.
- Concentration Main Amplifier: (1/8) Concentration Oscillator .
Pump Laser: InnoLas, SpitLight 1500-30
- Seeding energy: 550 mJ / Pulse.
Note. Dye was mixed on 28.04.2022 |
Attachment 1: Fluorescein27_Efficiency.png
|
|
28
|
Thu Jun 6 19:51:11 2024 |
Carsten | Accelerator | ESR crew left after persisting problems with ESR magnets |
No beam today.--. |
18
|
Wed Jun 5 14:17:36 2024 |
Carsten | Accelerator | Distance Cooler Detector |
Calculation from Helmut Weick (copied from E142)
Distance e-cooler to particle detectors chamber:
“This is half of ESR plus distance target - detector. This even is on the web page”
54.28 m + 21.64 m |
76
|
Fri Jun 14 13:14:29 2024 |
Konstantin, Bernhard, Peter, Lukas | General | DAQ crashes at :10 |
10 Minutes past a full hour (XX:10) the windows PCs (sometimes) loose connection.
This causes the labview server & the laser remote ctrl to crash.
reset:
- "sto ac" on first DAQ1
- restart labview server (medusa - run - make data available for mbs off/on)
- "sta ac" on first DAQ1
It can make sense to try to avoid laser scans in this time window if those crashes occur more regularly. |
10
|
Sat Jun 1 18:14:01 2024 |
Carsten | Analysis | DAQ and analysis directories |
DAQ1 (original):
-user atplaser
-runs on lxg1297
-go4 in ~/go4/229Th
-mbs in /esr/usr/atplaser/mbsrun/E0052/DAQ1
-data in lxg1297:/data.local2/229Th/
filenames follow the style 229ThDAQ1_nnnn.lmd (where nnnn is 4 digit integer number)
Data backup on WDBook lxg0155 local:
/media/stoe_exp/WDBook/E0052daq1 (/dev/sdc2)
Since the Go4 Code is still under development (clean up etc.), I still plan to submit regular updates, the last version before is always in ~go4/229Th
It is planned that all essential settings can be done via the automatically loaded script files SetParamters.C and SetConditions.C, so that no recompilation (and change of code) is required (work in progress).
Values in SetParamters.C and SetConditions.C have to be adjusted to the actual run.
DAQ2 (new / enhanced):
additional 25ps TDC, ADC and Vulom Scaler
-user stoe_exp
-runs on lxg0155
-go4 in ~/E0052/go4
-mbs in /esr/usr/atplaser/mbsrun/E0052/DAQ2
-data in lxg0155:/data.lcoal2/E0052/
filenames follow the style ThDAQ2_nnnn.lmd (where nnnn is 4 digit integer number)
Data backup on WDBook lxg1297 local:
/media/stoe_exp/WDBook/E0052daq2 (/dev/sdd2)
The following .scom in the directories will help to run the experiment
@startup in mbs -> Generic startup of all mbs tasks for data acquisition
@shutdown in mbs -> Ceans up everything before leaving mbs
@connectdisk -> connects to the rfio server on the data local disks.
or this, rawDispRFIO64, has to be started (locally __IN__ the data directory).
@openfile -> automatically opens a new file in the data local disks
@closefile -> closes the file
@disconndisk -> disconnects from the rfio server
Please note that the different path and file names are on purpose, so that they can be distinguished after the run. |
74
|
Fri Jun 14 11:44:23 2024 |
Carsten Brandau | Accelerator | Cooling Time in SC7 increased from 9s to 12s |
Cooling Time in SC7 increased from 9s to 12s |
68
|
Thu Jun 13 18:03:36 2024 |
Afternoon shift | Runs | Continuation: First scans |
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
00:01 | | 3e5 | 104438 | 13 | 27-29 | 550.1 - 550.9 | 201 | 2s | 4 | 40 injections, there is still a fast decaying component, may be due to the higher electron current of 200 mA.
Going back to 100 mA in next run and also put a "loose" scraper on the inside.
|
ntcap 2024-06-12_21-47-54
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
00:50 | ? | 3e5 | 104438 | 14 | 28- | 550.1 - 550.9 | 201 | 1s | 6 | 30 injections, e-cooler current at 100mA, scraper GE01GD2ID (north arc, inner scraper) now at position +30mm, medusa crashed when started, laser position was lost and retrieved by resetting the beam position stabilization
|
01:43 | 25700 | 3e5 | 104438 | 15 | 31-32 | 550.1 - 550.9 | 201 | 1s | 6 | e-cooler current 100mA, NA position -30 activated, NI +28, SA -28, DS3HG +10
|
02:32 | 21750 | 4e5 | 104417 | 16 | 33 | 550.1 - 550.9 | 201 | 1s | 3 | reduced iterations due to short life time
|
02:54 | 29200 | 4e5 | 104417 | 17 | 34 | 550.1 - 550.9 | 201 | 1s | 3 |
|
03:16 | (?) | 4e5 | 104417 | 18 | 35 | 550.1 - 550.9 | 201 | 1s | 3 |
|
03:40 | 33700 | 5e5 | 104417 | 19 | 36 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:03 | 31600 | 5e5 | 104417 | 20 | 37 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:25 | 29200 | 4e5 | 104417 | 21 | 38 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:46 | 27500 | 4e5 | 104417 | 22 | 39 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:08 | 30400 | 4e5 | 104417 | 23 | 40 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:31 | 31300 | 4e5 | 104417 | 24 | 41 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:56 | 30000 | 3e5 | 104417 | 25 | 42 | 550.1 - 550.9 | 201 | 1s | 3 |
|
06:18 | 24500 | 4e5 | 104417 | 26 | 43 | 550.1 - 550.9 | 201 | 1s | 3 | Morning shift took over: Constantine, Peter, Lukas, Jonas, Dounia
|
06:43 | | | | 27 | 44 | 550.1 - 550.9 | 201 | 1s | 3 | Medusa crashed when starting the laser scan
|
06:57 | | 4e5 | 104417 | 28 | 45-46 | 550.1 - 550.9 | 201 | 1s | 3 | DAQ1 started a bit later since we needed to restart and connect the file server again; after starting laser scan we noticed that the beam is gone on glass disk, reason was likely user error as medusa was set to send the laser to 500.9nm instead of 550.9nm
|
07:27 | 11800 | 4e5 | 104417 | 29 | 47 | 550.1 - 550.9 | 201 | 1s | 3 |
|
07:51 | 23500 | 4e5 | 104417 | 30 | 48 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:14 | 15000 | 5e5 | 104417 | 31 | 49 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:37 | 35300 | 5e5 | 104417 | 32 | 50 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:58 | 32800 | 5e5 | 104417 | 33 | 51 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 17:32 | 23 500 | 5e5 | 104412 | 35 | 52 | 550.1 - 550.9 | 201 | 1s | 3 | First scan of afternoon shift. 30 Injections. Electron current was set to 30 mA. (Note added 2024.06.14 17:55 - Carsten: To what I remember electron current is 80mA) Dye was changed before this scan. Wavemeter overexposed on the second scan for 2-3 steps, not sure if during the stepping or actual measurement. In the middle of the third scan, medusa crashed.
|
13.06.2024 | xx | 5e5 | 104417 | 36 | 53 | 550.1 - 550.9 | 201 | 1s | 3 | 25 Injections 75Hz particle detector south, medusa crashed immediately, laser controller was locked up. Restarting the laser controller solved the issue.
|
13.06.2024 18:30 | 28 200 | 5e5 | 104412 | 37 | 54 | 550.1 - 550.9 | 201 | 1s | 3 | 25 Injections , DAQ1 no data collected (no start acquisition pressed)
|
13.06.2024 18:50 | 26 000 | 5e5 | 104412 | 38 | 55 | 550.1 - 550.9 | 201 | 1s | 3 | 23 of 25 Injections, two were empty injections from SIS
|
13.06.2024 19:13 | 31 600 | 5e5 | 104412 | 39 | 56 | 550.1 - 550.9 | 201 | 1s | 3 | 29 of 30 Injections, one was empty injections from SIS
|
13.06.2024 19:45 | 34 700 | 5e5 | 104411 | 40 | 57 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections
|
13.06.2024 20:07 | 33 300 | 6e5 | 104411 | 41 | 58 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections. Note: before and after run, the laser was exactly in the middle of the screen, but went to its proper position (slightly left of the middle) during the scan.
|
13.06.2024 20:36 | 29 900 | 7e5 | 104411 | 42 | 59 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections. Note: Before the run ad during the first scan, the laser was above the middle. In the third scan, it was back in its proper position.
|
13.06.2024 20:57 | 28 900 | 7e5 | 104411 | 43 | 60 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 21:18 | 26 000 | 7e5 | 104411 | 44 | 61 | 550.1 - 550.9 | 201 | 1s | 3 | 27 of 30 Injections
|
13.06.2024 21:40 | 18 000 | 6e5 | 104411 | 45 | 62 | 550.1 - 550.9 | 201 | 1s | 3 | out of the first injections 4 were with low energy and 4 did not arrive at all, for the rest, pretty much every second injection is missing. Checking with HKR.
|
13.06.2024 22:02 | 24 900 | 7e5 | 104411 | 46 | 63 | 550.1 - 550.9 | 201 | 1s | 3 | Same thing with injection.
|
13.06.2024 22:23 | 31 700 | 7e5 | 104411 | 47 | 64 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 22:46 | 30 900 | 7e5 | 104411 | 48 | 65 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 23:09 | 33 100 | 7e5 | 104411 | 49 | 66 | 550.1 - 550.9 | 201 | 1s | 3 | Laser shifted to the left
|
13.06.2024 23:33 | 25 200 | 7e5 | 104411 | 50 | 67 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 23:55 | 28 600 | 7e5 | 104411 | 51 | 68 | 550.1 - 550.9 | 201 | 1s | 3 | We moved the laser back to the original position using the MRC
|
14.06.2024 00:52 | 33 100 | 8e5 | 104410 | 52 | 69 | 549.85 - 550.9 | 239 | 1s | 3 | Widen scan range to account for the difference between wavelength vac. vs. wavelength air; Laser shifted to the left
|
14.06.2024 01:20 | 29 800 | 8e5 | 104410 | 53 | 70 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 01:44 | -- | 7e5 | 104410 | 54 | 71 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted; measurement interrupted; connection of MBS to Labview server crashed and MBS and go4 were restarted (DAQ1)
|
14.06.2024 02:03 | 34 700 | 8e5 | 104410 | 55 | 72 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 02:26 | 31 400 | 7e5 | 104411 | 56 | 73 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 02:49 | -- | 7e5 | 104411 | 57 | 74 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted; measurement interrupted
|
14.06.2024 02:51 | -- | 7e5 | 104411 | 58 | 75 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted;
|
|
The Medusa crash also crashed the Cobra Laser. We tried to restart it but afterwards the power of the Laser was more than one order of magnitude lower than before. It is not clear what caused the problem, Rodolfo is on his way to GSI now to fix the issue. |
Attachment 1: 2024_06_13_229Th_Lifetime_at_190MeV_afterscraping.png
|
|
Attachment 2: 2024-06-13_23-13-55-242.png
|
|
12
|
Sat Jun 1 19:00:57 2024 |
Carsten | DAQ | Change of (offline) frequency for common stop (CS) of VUPROMs (and 2nd TDC) and scaling factor |
Temporarily, for tests of the second DAQ I changed the frequency from the frequency generator and scaling factor that is used for the common stop signal to a value that is close to the expected value in the experiment.
Reason: For the second TDC (CAEN 25ps) TDC the time between to CS events must be of the order of 45 us. The TDC can store events up to about 51,x us and allows for overlap of these regions to be able to stitch together the events to a continuous time axis.
Since we plan to run at ~ beta= 0.5566 the revolution frequency is about
f=1.540974 MHz (0.6489 us)
This value is used in the frequency generator.
The scaling factor is set to 68.
=> 68 x 0,6489us = 44,1252 us between to CS.
Of course later the real RF value from the accelerator will be used.
Please leave or, if required to change, mark in ELog. |
55
|
Wed Jun 12 22:48:55 2024 |
Carsten | | Beam Halo |
The seamingly short live time reported in entry 53 is due to a halo from a different isotope (?). It can be scraped away from the outside. The 229Th89+ primary beam can be scaped away from the inside and the halo stays. There is some spacial headroom between the two components.
Scraping from the outside using GE02DS3HG @+6mm starts to scrape the halo in the south arc. Please note that the +6mm means that you move past central orbit (scraper starts from -130mm).
Using this scraper allows for using the particle detector behind the e-cooler (GE02DD2_G). Position PD south in event mode is -30mm (has to be verified with higher ion beam intensity and electron current for larger recombination rate=.
The primary 229Th89+ is touched from the inside with GE01DB2IG @+14mm (north arc).
Position of GE02DS3HG in event mode is +8mm.
Electron current in SC8 is increased from 100mA to 200mA for faster cooling at low energy in order to condition after deceleration and before scraping.
positions particle detectors and scrapers
NA position -30 activated
NI +28
SA -28
DS3HG +10 |
77
|
Fri Jun 14 14:44:51 2024 |
Afternoon shift | Runs | Afternoon shift |
Pizza shift
the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
From hardware side, the laser can only scan in one direction: long wave length to short wave length, independent of what is written in the tables.
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
14.06.2024 12:45 | 38900 | 2e5 | 104411 | 72 | 89 | 549.6 - 550.4 | 201 | 1s | 3 | starting with 50 accumulation
|
14.06.2024 13:13 | 41400 | 2e5 | 104411 | 73 | 90 | 549.6 - 550.4 | 201 | 1s | 3 | there was a :10 crash during the injection, laser scan started normally.
|
14.06.2024 13:41 | 39500 | 2e5 | 104411 | 74 | 91 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 14:38 | 67750 | 2e5 | 104412 | 75 | 92 | 549.6 - 550.4 | 201 | 1s | 3 | afternoon shift takes over. After short break from hkr to optimize beam, proceeding with 30 injections, particle number detector might not have been reset, check next run
|
14.06.2024 15:02 | 34500 | 2e5 | 104412 | 76 | 93 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 15:38 | 27500 | 1e5 | 104412 | 77 | 94 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 16:01 | 25300 | 1e5 | 104412 | 78 | 95 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 16:23 | 28700 | 1e5 | 104412 | 79 | 96 | 549.6 - 550.4 | 201 | 1s | 3 | At around 550.2 nm the wavemeter is overexposed
|
14.06.2024 16:48 | xxx | xx | xx | 80 | 97 | xx | xx | xx | xx | empty file , no laser scan, work on laser, while trying to adjust the wavemeter, we tried to move the laser using medusa, at some point, this caused the cobra to go into error mode again, onlz a restart could solve it. As a consequence, we still do not know if the wavemeter coupling is optimal
|
14.06.2024 17:17 | 30000 | 1e5 | 104412 | 81 | 98 | 549.6 - 550.4 | 201 | 1s | 3 |
wavemeter coupling looks good now
|
14.06.2024 17:41 | 56900 | 1e5 | 104412 | 82 | 99 | 549.6 - 550.4 | 201 | 1s | 3 | ratemeter was probably not reset
|
14.06.2024 18:04 | 8100 | 1e5 | 104412 | 83 | 100 | 549.6 - 550.4 | 201 | 1s | 3 | really bad injection (looked fine (?)), low ion count from SIS, they exchanged a cathode @UNILAC
|
14.06.2024 18:42 | 3200 | 3e4 | 104412 | 84 | 101&102(automatically opened new file without command after 1GB?) | 549.6 - 550.4 | 201 | 1s | 3 | problems with accelerator: after deacceleration the beam is gone. Scraper position has moved(?). The cooling water from GSI is too warm, so we had to shut down the laser. NTCAP restarted. And NTCAP restarted again @20:39 for test data for Shahan.
|
|
14
|
Sun Jun 2 00:01:46 2024 |
Carsten | DAQ | 25ps TDC CAEN V1290N DAQ2 lead to crash in DAQ2 due to a hardware issue |
On the CAEN V1290N TDC the piggy board for the lower 8 channels is not fixed well to the motherboard.
That prevented to the whole DAQ2 to trigger correctly or even crash every once in a while. Long attempt sot find the error in hard and in software.
I fixed/reattached the piggy board, but be careful with the attached Lemo cables. They might loose the board again (it used to fixed with hot glue).
DAQ2 now also seem to run smoothly. |
59
|
Thu Jun 13 10:32:29 2024 |
Konstantin | Accelerator | (Proper) Cooler settings |
We change the acceleration voltage of the electron cooler and observe the platform high voltage and the count rate at the particle detector south.
Cooler current at 100mA.
set voltage | measured voltage | Schottky width |
-40 | | |
80 | | |
-40 | | |
100 | | |
120 | | |
-40 | | |
-60 | | |
-80 | | |
-100 | | |
-120 | | |
-38 | | |
-40 | | |
-42 | | |
-44 | | |
-46 | | |
-48 | | |
-40 | | |
-38 | | |
-36 | | |
-34 | |
|
with scraper driven in (first signal drop in attachment 2):
set voltage | measured voltage | Schottky width |
-40 | | |
-38 | | |
-36 | | |
-34 | | |
-32 | | |
-30 | | |
-20 | | 25kHz |
-10 | | 48kHz |
-0 | | 63kHz |
-40 | | |
-42 | | |
-44 | | |
-46 | | |
-56 | | 7kHz |
-66 | | 19kHz |
-76 | | 40kHz
|
5kHz Shottky width at proper cooling conditions.
Modified bunching HF in beam process 7 and 10 in Paramodi folder "Ring HF" to get coasting beam conditions.
Afterwards manipilated the electron current from 100mA to 200mA to 50mA in BP8. |