  Nuclear hyperfine mixing in 229Th89+ 2024, Page 4 of 4  Not logged in ELOG logo
ID Date Authordown Category Subject
  28   Thu Jun 6 19:51:11 2024 CarstenAcceleratorESR crew left after persisting problems with ESR magnets
No beam today.--.
  29   Fri Jun 7 09:15:31 2024 CarstenDAQUpdated Channel List
Attachment 1: ChannelList.pdf
ChannelList.pdf ChannelList.pdf
Attachment 2: ChannelList.odt
  33   Sat Jun 8 12:02:20 2024 CarstenDetectorsTest Particle Detectors - Det South not working
With the test beam from ESR I started to have a look at the PartDets.:
North outside: Works properly with a voltage of 1450 V
South outside (behind cooler) : no signal.

Later (when Danyal is here) we will check detectors cables etc., and if required put the north detector to position south and mount the untested (newly made, but with "old" material and PMT) spare detecotr to poistion north.
  34   Sat Jun 8 12:45:05 2024 CarstenDAQTemporary change of RF signal
The setting was 1.540974 MHz and division factor 68 (for a 200 MeV/u) beam, and the signal was taken from a signal generator.
I now switch to the "real" RF signal from the accelerator (~400MeV/u), that is 1.9693 MHz and the factor is set to 88.
This should be set back for laser tests or adjusted when we have the real decellerated beam.
  35   Sat Jun 8 13:11:52 2024 CarstenGeneralPut the go4login in atplaser .bashrc
Put the go4login in atplaser .bashrc in order to avoid that one always has to go to the go4 directory to set the environment variable.
go4login64 is now executed automaticlally and the environments are now set every time you open a (bash) shell.
A fixed version 6.3 is used. The login file might be changed back after the beamtime.
  39   Mon Jun 10 09:17:38 2024 CarstenDAQUpdated Channel List
Attachment 1: 2024_06_10_ChannelList.pdf
2024_06_10_ChannelList.pdf 2024_06_10_ChannelList.pdf
Attachment 2: ChannelList.odt
  44   Tue Jun 11 22:59:06 2024 CarstenAcceleratorIdentification / NTCAP started
ESR is set-up in most parts.We are trying to identify our 229Th89+. It is more cumbersome then exspected because of the thick target an d the according energy loss, the pattern has changed compared to the last beamtime.

Just started the NTCAP with center frequency 246,5 MHz, 10 MSamples/s and -25db.
Center frequency will later be changed back to about 244,5 Mhz.

Directory 06-11-22_55_15 (correct time stamp).
  47   Wed Jun 12 09:17:43 2024 Carsten Tests of background with ionization pumps on (from monday)
On monday the ionization pumps were switch on to test the background, and if it is possible to run with the pumps switched on.
Results and comments, see below:

Later some data were recorded with DAQ1 to allow for a more thorough analysis - beam is 238U38+ at 401 MeV/u - file 0013
Please note the data of this file was recorded after the figures below were taken.

In short, we have switched off the the ionization pumps because the conclusion is somehow ambiguous. Background rates from real photons are considerably low at ~5Hz and 10Hz for the small-band PMTs (see figure), Yet, PMT south goes up to 350 Hz and sees a lot of photons from the plasma discharge. The few Hz of the small-band detector would be suppressed by the length of the ion bunch / full lenght of the ring. But what is not known is how much it would actually influence the local vacuum of the ring.

Semjon Strohmenger also checked that the valve to the turbos in front and behind the gas-jet is open (--yes). They are running the whole beam time.
I exchanged the prepump at a pocket located between gas-jet and VUV detection set-up. Here after baking a leak was found, and the pocket was pumped with an "old" prepump.
Vacumm in the pocket was ~1mBar. Installed a more powerful Edwards XDS i10. When we left ESR the pressure in the pocket was already at low 10^-2 mbar and still decreasing.
May be this helps to improve local pressure.
Attachment 1: mit_IZ_trending_sued_alles(blind_durch_IZ).png
mit_IZ_trending_sued_alles(blind_durch_IZ).png
Attachment 2: mit_IZ_trending_mitte_Nord_mit_Strahl.png
mit_IZ_trending_mitte_Nord_mit_Strahl.png
Attachment 3: mit_IZ_trending_mitte_Nord_ohne_Strahl.png
mit_IZ_trending_mitte_Nord_ohne_Strahl.png
Attachment 4: mit_IZ_trending_mitte_Nord_alles.png
mit_IZ_trending_mitte_Nord_alles.png
  48   Wed Jun 12 09:43:57 2024 CarstenAcceleratorStopped NTCAP at 9:43
Setup at ESR is carried on - stopped the long-time NTCAP run from overnight,
  55   Wed Jun 12 22:48:55 2024 Carsten Beam Halo
The seamingly short live time reported in entry 53 is due to a halo from a different isotope (?). It can be scraped away from the outside. The 229Th89+ primary beam can be scaped away from the inside and the halo stays. There is some spacial headroom between the two components.

Scraping from the outside using GE02DS3HG @+6mm starts to scrape the halo in the south arc. Please note that the +6mm means that you move past central orbit (scraper starts from -130mm).
Using this scraper allows for using the particle detector behind the e-cooler (GE02DD2_G). Position PD south in event mode is -30mm (has to be verified with higher ion beam intensity and electron current for larger recombination rate=.

The primary 229Th89+ is touched from the inside with GE01DB2IG @+14mm (north arc).

Position of GE02DS3HG in event mode is +8mm.

Electron current in SC8 is increased from 100mA to 200mA for faster cooling at low energy in order to condition after deceleration and before scraping.


positions particle detectors and scrapers
NA position -30 activated
NI +28
SA -28
DS3HG +10
  57   Thu Jun 13 02:05:04 2024 CarstenAcceleratorShort Lifetime at 100 mA
Lifetime still is short
Attachment 1: 2024_06_13_229Th_Short_Lifetime3.png
2024_06_13_229Th_Short_Lifetime3.png
  53   Wed Jun 12 18:48:43 2024 Afternoon shiftRunsFirst laser scans
Time Part. Det. Int ESR-intensity cooler-HV LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time(s) n loops comments
19:15 38257 xx e x 104416 10 21-23 550.1 - 550.9 201 2s 8 40 SIS injections. Very short lifetime, further investigation is needed. preset_2024-06-12_19-08-23.mps
20:22 53056 xx e x 104416 11 24-25 550.1 - 550.9 201 1s 3 40 SIS injections. Only PDetector behing cooler, retracted by 8 mm. Still short lifetime.
20:58 40500 xx e x 104416 12 26-26 550.1 - 550.9 201 1s 2 40 SIS injections.
Attachment 1: 2024_06_12_229Th_Lifetime_at_190MeV.png
2024_06_12_229Th_Lifetime_at_190MeV.png
  68   Thu Jun 13 18:03:36 2024 Afternoon shiftRunsContinuation: First scans
Time Part. Det. Int ESR-intensity cooler-HV LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time(s) n loops comments
00:01 3e5 104438 13 27-29 550.1 - 550.9 201 2s 4 40 injections, there is still a fast decaying component, may be due to the higher electron current of 200 mA. Going back to 100 mA in next run and also put a "loose" scraper on the inside.


ntcap 2024-06-12_21-47-54

Time Part. Det. Int ESR-intensity cooler-HV LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time(s) n loops comments
00:50 ? 3e5 104438 14 28- 550.1 - 550.9 201 1s 6 30 injections, e-cooler current at 100mA, scraper GE01GD2ID (north arc, inner scraper) now at position +30mm, medusa crashed when started, laser position was lost and retrieved by resetting the beam position stabilization
01:43 25700 3e5 104438 15 31-32 550.1 - 550.9 201 1s 6 e-cooler current 100mA, NA position -30 activated, NI +28, SA -28, DS3HG +10
02:32 21750 4e5 104417 16 33 550.1 - 550.9 201 1s 3 reduced iterations due to short life time
02:54 29200 4e5 104417 17 34 550.1 - 550.9 201 1s 3
03:16 (?) 4e5 104417 18 35 550.1 - 550.9 201 1s 3
03:40 33700 5e5 104417 19 36 550.1 - 550.9 201 1s 3
04:03 31600 5e5 104417 20 37 550.1 - 550.9 201 1s 3
04:25 29200 4e5 104417 21 38 550.1 - 550.9 201 1s 3
04:46 27500 4e5 104417 22 39 550.1 - 550.9 201 1s 3
05:08 30400 4e5 104417 23 40 550.1 - 550.9 201 1s 3
05:31 31300 4e5 104417 24 41 550.1 - 550.9 201 1s 3
05:56 30000 3e5 104417 25 42 550.1 - 550.9 201 1s 3
06:18 24500 4e5 104417 26 43 550.1 - 550.9 201 1s 3 Morning shift took over: Constantine, Peter, Lukas, Jonas, Dounia
06:43 27 44 550.1 - 550.9 201 1s 3 Medusa crashed when starting the laser scan
06:57 4e5 104417 28 45-46 550.1 - 550.9 201 1s 3 DAQ1 started a bit later since we needed to restart and connect the file server again; after starting laser scan we noticed that the beam is gone on glass disk, reason was likely user error as medusa was set to send the laser to 500.9nm instead of 550.9nm
07:27 11800 4e5 104417 29 47 550.1 - 550.9 201 1s 3
07:51 23500 4e5 104417 30 48 550.1 - 550.9 201 1s 3
08:14 15000 5e5 104417 31 49 550.1 - 550.9 201 1s 3
08:37 35300 5e5 104417 32 50 550.1 - 550.9 201 1s 3
08:58 32800 5e5 104417 33 51 550.1 - 550.9 201 1s 3
13.06.2024 17:32 23 500 5e5 104412 35 52 550.1 - 550.9 201 1s 3First scan of afternoon shift. 30 Injections. Electron current was set to 30 mA. (Note added 2024.06.14 17:55 - Carsten: To what I remember electron current is 80mA) Dye was changed before this scan. Wavemeter overexposed on the second scan for 2-3 steps, not sure if during the stepping or actual measurement. In the middle of the third scan, medusa crashed.
13.06.2024 xx 5e5 104417 36 53 550.1 - 550.9 201 1s 3 25 Injections 75Hz particle detector south, medusa crashed immediately, laser controller was locked up. Restarting the laser controller solved the issue.
13.06.2024 18:30 28 200 5e5 104412 37 54 550.1 - 550.9 201 1s 3 25 Injections , DAQ1 no data collected (no start acquisition pressed)
13.06.2024 18:50 26 000 5e5 104412 38 55 550.1 - 550.9 201 1s 3 23 of 25 Injections, two were empty injections from SIS
13.06.2024 19:13 31 600 5e5 104412 39 56 550.1 - 550.9 201 1s 3 29 of 30 Injections, one was empty injections from SIS
13.06.2024 19:45 34 700 5e5 104411 40 57 550.1 - 550.9 201 1s 3 30 of 30 Injections
13.06.2024 20:07 33 300 6e5 104411 41 58 550.1 - 550.9 201 1s 3 30 of 30 Injections. Note: before and after run, the laser was exactly in the middle of the screen, but went to its proper position (slightly left of the middle) during the scan.
13.06.2024 20:36 29 900 7e5 104411 42 59 550.1 - 550.9 201 1s 3 30 of 30 Injections. Note: Before the run ad during the first scan, the laser was above the middle. In the third scan, it was back in its proper position.
13.06.2024 20:57 28 900 7e5 104411 43 60 550.1 - 550.9 201 1s 3
13.06.2024 21:18 26 000 7e5 104411 44 61 550.1 - 550.9 201 1s 3 27 of 30 Injections
13.06.2024 21:40 18 000 6e5 104411 45 62 550.1 - 550.9 201 1s 3 out of the first injections 4 were with low energy and 4 did not arrive at all, for the rest, pretty much every second injection is missing. Checking with HKR.
13.06.2024 22:02 24 900 7e5 104411 46 63 550.1 - 550.9 201 1s 3 Same thing with injection.
13.06.2024 22:23 31 700 7e5 104411 47 64 550.1 - 550.9 201 1s 3
13.06.2024 22:46 30 900 7e5 104411 48 65 550.1 - 550.9 201 1s 3
13.06.2024 23:09 33 100 7e5 104411 49 66 550.1 - 550.9 201 1s 3 Laser shifted to the left
13.06.2024 23:33 25 200 7e5 104411 50 67 550.1 - 550.9 201 1s 3
13.06.2024 23:55 28 600 7e5 104411 51 68 550.1 - 550.9 201 1s 3 We moved the laser back to the original position using the MRC
14.06.2024 00:52 33 100 8e5 104410 52 69 549.85 - 550.9 239 1s 3 Widen scan range to account for the difference between wavelength vac. vs. wavelength air; Laser shifted to the left
14.06.2024 01:20 29 800 8e5 104410 53 70 549.85 - 550.9 239 1s 3
14.06.2024 01:44 -- 7e5 104410 54 71 549.85 - 550.9 239 1s 3 Medusa crashed and was restarted; measurement interrupted; connection of MBS to Labview server crashed and MBS and go4 were restarted (DAQ1)
14.06.2024 02:03 34 700 8e5 104410 55 72 549.85 - 550.9 239 1s 3
14.06.2024 02:26 31 400 7e5 104411 56 73 549.85 - 550.9 239 1s 3
14.06.2024 02:49 -- 7e5 104411 57 74 549.85 - 550.9 239 1s 3 Medusa crashed and was restarted; measurement interrupted
14.06.2024 02:51 -- 7e5 104411 58 75 549.85 - 550.9 239 1s 3 Medusa crashed and was restarted;


The Medusa crash also crashed the Cobra Laser. We tried to restart it but afterwards the power of the Laser was more than one order of magnitude lower than before. It is not clear what caused the problem, Rodolfo is on his way to GSI now to fix the issue.
Attachment 1: 2024_06_13_229Th_Lifetime_at_190MeV_afterscraping.png
2024_06_13_229Th_Lifetime_at_190MeV_afterscraping.png
Attachment 2: 2024-06-13_23-13-55-242.png
2024-06-13_23-13-55-242.png
  77   Fri Jun 14 14:44:51 2024 Afternoon shiftRunsAfternoon shift
Pizza shift

the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
From hardware side, the laser can only scan in one direction: long wave length to short wave length, independent of what is written in the tables.


Time Part. Det. Int ESR-intensity cooler-HV LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time(s) n loops comments
14.06.2024 12:45 38900 2e5 104411 72 89 549.6 - 550.4 201 1s 3 starting with 50 accumulation
14.06.2024 13:13 41400 2e5 104411 73 90 549.6 - 550.4 201 1s 3 there was a :10 crash during the injection, laser scan started normally.
14.06.2024 13:41 39500 2e5 104411 74 91 549.6 - 550.4 201 1s 3
14.06.2024 14:38 67750 2e5 104412 75 92 549.6 - 550.4 201 1s 3 afternoon shift takes over. After short break from hkr to optimize beam, proceeding with 30 injections, particle number detector might not have been reset, check next run
14.06.2024 15:02 34500 2e5 104412 76 93 549.6 - 550.4 201 1s 3
14.06.2024 15:38 27500 1e5 104412 77 94 549.6 - 550.4 201 1s 3
14.06.2024 16:01 25300 1e5 104412 78 95 549.6 - 550.4 201 1s 3
14.06.2024 16:23 28700 1e5 104412 79 96 549.6 - 550.4 201 1s 3 At around 550.2 nm the wavemeter is overexposed
14.06.2024 16:48 xxx xx xx 80 97 xx xx xx xx empty file , no laser scan, work on laser, while trying to adjust the wavemeter, we tried to move the laser using medusa, at some point, this caused the cobra to go into error mode again, onlz a restart could solve it. As a consequence, we still do not know if the wavemeter coupling is optimal
14.06.2024 17:17 30000 1e5 104412 81 98 549.6 - 550.4 201 1s 3 wavemeter coupling looks good now
14.06.2024 17:41 56900 1e5 104412 82 99 549.6 - 550.4 201 1s 3 ratemeter was probably not reset
14.06.2024 18:04 8100 1e5 104412 83 100 549.6 - 550.4 201 1s 3 really bad injection (looked fine (?)), low ion count from SIS, they exchanged a cathode @UNILAC
14.06.2024 18:42 3200 3e4 104412 84 101&102(automatically opened new file without command after 1GB?) 549.6 - 550.4 201 1s 3 problems with accelerator: after deacceleration the beam is gone. Scraper position has moved(?). The cooling water from GSI is too warm, so we had to shut down the laser. NTCAP restarted. And NTCAP restarted again @20:39 for test data for Shahan.