  80   Mon Mar 30 07:02:15 2020 tinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 20 mA I_cooler = 20 mA

beam: 206Pb81+
  
  41   Sat Mar 28 16:12:25 2020 ruijiu, raganDAQenter ESR and turn on the HV crate.    
  15   Tue Feb 25 15:48:31 2020 ruijiu chenDAQImportant folders of e121 eperiments:1. slow control:
1.1. program file:
litv-exp@ringpi:~/epics/mrcc_ioc/db $ 
  
  16   Tue Feb 25 15:49:58 2020 ruijiu chenDAQchanging scaling factor.How to change scaling factor?
The parameters is in f_user.c file:
1. open the f_user.C with emacs :
  
  99   Mon Mar 30 14:48:02 2020 ruijiu and TinoDetectorsThe leakcurrent of SiPad is over the setting limit.Because the 
leakcurrent of 
6 Sipads is 
 Screenshot_20200330-145551.jpgScreenshot_20200330-145544.jpgIMG_20200330_144915.jpg 
  100   Mon Mar 30 15:03:58 2020 ruijiu Tino, Yuri, Lasoro, Nik and UliDetectorsthe gas bottle is empty.The gas bottle was empty. Tino replaced 
the previous old bottle with 50 bar gas left.
 IMG_20200330_152735_mh1585575780057.jpg 
  37   Sat Mar 28 15:07:40 2020 ruijiuDAQstart run     
  38   Sat Mar 28 15:19:23 2020 ruijiu start run e121_run_0236.lmdstore 206Pb81+ beam inside ESR and measure
its halflife. 
The HV of detector is off.
  
  44   Sun Mar 29 09:40:35 2020 ruijiuDAQrestart go4go4 crush up. 
the problem is 
due to  mbs. 
 IMG_20200329_093354.jpg 
  45   Sun Mar 29 09:57:26 2020 ruijiuDetectorsthe HV of charged particle detector        MHV1     MHV2
ch1:1st Sipad  4th 
Sipad
 IMG_20200329_095014.jpg 
  46   Sun Mar 29 10:09:14 2020 ruijiu start run e121_run_0237.lmdSergey and Yuri are tuning Tl beam. The particle
number in ers is about 1e4. I(SIS)=5e8
run stop: after few minutes (garbage data)
  
  51   Sun Mar 29 13:05:36 2020 ruijiuDAQstart ntcapIQ_2020-03-29_13-05-
56
SC_2020-03-29_13-05-
 IMG_20200329_130438.jpg 
  53   Sun Mar 29 14:01:45 2020 ruijiu beam position on DSSDmove the detector 1cm 
outside.from 60mm to 
70mm.
 IMG_20200329_140003.jpg 
  54   Sun Mar 29 14:29:13 2020 ruijiu e121_run_00240.lmds6 positon at 50mm.   
  55   Sun Mar 29 18:41:31 2020 ruijiuAnalysisRatio between MPWC rate and DSSD rateslit Ratio error
-1 18.64 0.4
-2 18.18 0.38
 ratio.png 
  58   Sun Mar 29 21:02:43 2020 ruijiuAnalysisdecay line of 205Tl beamdecay line of 205Tl beam with Ar gas jet target
on and S6 slit at -8mm
Ar gas density~1e12 atoms/cm2
 main_s6_neg8_202003292059.pngmain_s6_neg8_202003292059.root 
  60   Sun Mar 29 21:39:27 2020 ruijiuAnalysisIonization cross section and capture cross section for h-206Pb + ArIonization cross section = 1320 barn.
Capture cross section = 557+137+44.5=735.5.
sigma_Ion/sigma_cap=1.79
 202003292141.png 
  90   Mon Mar 30 12:12:25 2020 ruijiuGeneralInstruction of NTCAP for shift personTo start a new run with NTCap, we need to
retart NTCAP. 
1. Stop labview program. 
 Instruction_of_NTCAP_for_shift_person.png 
  95   Mon Mar 30 13:40:57 2020 ruijiuDetectorsHV of Inner detector   IMG_20200330_133909.jpg 
  96   Mon Mar 30 13:41:20 2020 ruijiuDetectorsHV of Inner detector   IMG_20200330_133909.jpg 