HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018 E0052_2025
  Bound beta decay experiment on Tl-205, Page 8 of 9  Not logged in ELOG logo
Entry  Tue Mar 31 20:32:05 2020, Ragan, General, 203Tl80+ 200Hg79+ 205Tl81+ 85AD7203-736C-4531-BA80-D045CC690BED.jpeg
Below in the schottky spectrum, on left is 203Tl80+ contamination, in middel is 200Hg79+ contamination and on right most is 205Tl81+.
These contaminations cannot be seen on CsISiPHOS but can only be seen on schottky.
The goal now is to get rid of these contaminations by putting slits.
Entry  Tue Mar 31 20:50:05 2020, Ragan, General, 200Hg79+ 205Tl81+ 31E7D028-F469-4BFF-B683-203821C174E9.jpeg
203Tl80+ impurity is removed by moving scrapper inside ESR.
Only 220Hg79+ is left now
Entry  Tue Mar 31 21:07:05 2020, Ragan, General, Separation of 203Tl80+ and 200Hg79+ using scrapper in ESR 06B012F5-5435-41D7-AD82-4A5FDADF281E.MOV
It can be seen in the video how 203Tl80+ (on extreme left) and then 200Hg79+ (in middle originally) is separted from 205Tl81+ by using scrapper in ESR.
Now, we want 205Pb81+ mixed with 205Tl81+ to be separated as much as possible in FRS.
Entry  Thu Apr 2 02:36:48 2020, Ragan, General, Stochastic cooling and orbit problem 
On 1 April by 5 pm, we could stop the stochastic cooling after the stacking process and also change the orbit of the Tl beam to centre. 
Thanks to the efforts by the ESR team !
Entry  Thu Apr 2 06:25:13 2020, Ragan, Runs, Template 
beam: 205Tl81+ (no scraping)
energy: 400 MeV/u
storage time: ___ hrs
Entry  Thu Apr 2 15:24:15 2020, Ragan, General, How to know whether electron cooling is working IMG_8238.JPGIMG_8239.JPG
For Alle and Kuehler the bottom left status should look like as shown in pictures below.
Entry  Sat Apr 4 03:06:12 2020, Ragan, General, Instructions if electron cooler goes off during long storage time measurements electron_cooler.pdf
Please read the attachment.
Entry  Mon Apr 6 20:03:50 2020, Ragan, General, Experiment finished )) 
E121 experiment is finished with the last 10 hours measurement at 19:47 ))
Many thanks to everyone for there significant contributions ))
Entry  Thu Apr 9 11:16:35 2020, Ragan, General, Measurements done in the experiment storage11.pdfstorage12.pdfstorage13.pdf
 
Entry  Mon Apr 20 11:17:18 2020, Ragan, General, Logbook logbook.pdfcombined_FRS.pdfcombined_data.pdf
 
Entry  Sun May 30 12:09:00 2021, Ragan, General, NTCAP ntcap_carsten.pdf
Attached is the info given by Carsten.
The instructions to operate NTCAP are given on https://wiki.gsi.de/ESR_EXP/ManualOfNTCAP
Entry  Mon Mar 29 16:31:09 2021, Pierre-Michel, DAQ, cross section Pb81onAr.png
(1.393*^-8 + 3.18*^-13 + 4.707*^-8)/4.707*^-8 = 1.29595
Entry  Sat Mar 28 13:09:11 2020, Mei,Yury, Tino, Ragan, Ruijiu, General, Shift plan 
From Yury,

For now, measure the beam life time with 3 different electron current in the e-cooler,i.e. 200mA(default), 10mA, and 20mA. For each current, 4 sets of
Entry  Fri Feb 21 11:30:52 2020, Laszlo, General, Shift plan 
e127: https://docs.google.com/spreadsheets/d/14m8WcCq1erx6HqWJRK7TLbCx121IkjqGa5l-frf4YjI/edit#gid=0

e121: https://docs.google.com/spreadsheets/d/10MUzA5-Ilf2WzCyNj8u_KSG6yzPg0e4vtZGGPFB3K-k/edit#gid=0
Entry  Thu Sep 13 11:32:17 2018, Jan Glorius, Collaboration, Proposal E121_Litvinov.pdfE121_Approval.pdf
Attached is the beam time proposal submitted and accepted by the G-PAC in 2017.
Entry  Fri Dec 7 10:56:04 2018, Jan Glorius, SlowControl, EPICS ioc on raspberry 
We have an EPCIS server (ioc) running on our raspberry (ext-027377). It talks to our MesyTec modules (MSCF shapers, MHV4 high-voltages).
In principal it should be running all the time, but here is the manual how to check/start it:
Entry  Fri Dec 7 11:39:41 2018, Jan Glorius, SlowControl, GUI for remote control of MSCF+MHV4 
We have a GUI for remote controlling our MesyTec modules MSCF16 and MHV4 for E121 experiment.

Here is the instructions to run it:
Entry  Tue May 28 08:36:14 2019, Jan, DAQ, MBS start & RFIO server 
You need a DAQ-PC and the RIO crate computer.
All of the DAQ proceedures laid out here are operated from inside the DAQ screen session.
Right now for e121 the following DAQ-setup is used: (might change in the future)
Entry  Wed Jun 26 16:20:31 2019, Jan, DAQ, unpacker: lmd > root E121.specdet_mapping.hh
I have written an unpacker for the lmd data taken with the current daq-system for E121.
The crate configuration is as follows:
ADC1 channels 1-32 (Caen 785)
Entry  Sun Mar 29 23:48:12 2020, Jan, Analysis, run0251: E_spectra for CsISiPhos  14x
This is a collection of energy spectra of the CsISiPhos detectors from run0251 with primary beam.

There is no good spectra for n-side of Si-pad 1!
ELOG V3.1.5-fc6679b