  32   Tue Feb 11 13:20:01 2020 JanDetectorsBaF2 - HV settings with these HV values the 6 channels are roughly
gain matched, such that the 7.83 MeV Po-decay
is around channel 5000.
 BaF2_HV_settings.pngbaf_bg_ch1.png 
  33   Mon Feb 17 08:14:44 2020 JanSimulations118Te K-REC X-ray energies X-ray energies of K-REC for 118Te according
to 
[URL=http://www-ap.gsi.de/Thomas/ap_html_research/energy/index.php]http://www-ap.gsi.de/Thomas/ap_html_research/energy/index.php[/URL]
 Photonen_Kammer.pngx-ray_calculator.txt 
  36   Fri Feb 21 11:29:59 2020 JanDetectorsdet. distances at target chamber This is the lookup table for the distance
of each x-ray detector to the target.
See attched foto for explanation.
 Photonen_Kammer.png 
  37   Tue Mar 10 08:52:34 2020 JanDetectorsDSSSD positioning To find a good Si position, we need to follow
a procedure similar to this:
 optimum_detector_position_2.jpg 
  38   Tue Mar 10 09:41:26 2020 JanDetectorsx-ray background from BaF The BaF2 detectors planned to be installed
at the target induce an additional background
in the x-ray detectors sitting below inside
  
  39   Wed Mar 11 09:08:21 2020 JanDAQDAQ & controls - OVERVIEW For all DAQ & slow control buissines the user

[COLOR=red][SIZE=3]litv-exp[/SIZE][/COLOR]
  
  41   Sun Mar 15 09:42:40 2020 JanCalibrationrun030 - Xray90 calib Am241 measurement with old tennelec specAmps, do
not use for efficiency!
  
  42   Sun Mar 15 13:08:38 2020 JanCalibrationrun031 - Xray90 calib Ba133 hi-rate measurement with old tennelec specAmps, do
not use for efficiency!
  
  43   Sun Mar 15 13:40:11 2020 JanCalibrationrun032 - Xray90 calib Ba133 measurement with old tennelec specAmps, do
not use for efficiency!
  
  44   Sun Mar 15 13:52:05 2020 JanCalibrationrun033 - Xray145 calib Am241 measurement with old tennelec specAmps, do
not use for efficiency!
  
  45   Sun Mar 15 14:41:34 2020 JanCalibrationrun034 - Xray145 calib B133 high rate -SOURCE MOVED! measurement with old tennelec specAmps, do
not use for efficiency!
  
  46   Sun Mar 15 15:08:44 2020 JanCalibrationrun035 - Xray145 calib B133 high rate measurement with old tennelec specAmps, do
not use for efficiency!
  
  59   Mon Mar 16 15:15:22 2020 JanCalibrationGamma- and X-ray intensities for efficiency calib Intensities for special gamma or x-ray lines
needed for efficiency calibration can be
found here:
  
  60   Tue Mar 17 08:15:49 2020 JanGeneralcurrent grids & screen in TE  measured beam diagnosis with 124Xe48+ at 326
MeV/u
 2020-03-16_18-51-50-170.pngIMG_0694.jpg 
  61   Tue Mar 17 11:44:58 2020 JanRunsrun entry - template  beam: 118Te52+
energy: 7.4 MeV/u
  
  64   Thu Mar 19 15:09:29 2020 JanGeneralstarting off with DSSSD here is my short guide how to start with the
measurement as soon as we have a beam available.
  
  71   Fri Mar 20 04:29:45 2020 JanDAQDAQ settings for the first night    e127.trlomain.cfgr3bfuser.cfgsetting.1584673412 
  85   Fri Mar 20 16:54:53 2020 JanDAQBaF OFF downscaled red=32 I introduced a downscaling by a factor of
2^5=32 for the BaF2 during target OFF.
The reason is, that we get a lot reasonable
  
  86   Fri Mar 20 16:58:02 2020 JanDAQXray2 & Xray3 cabling We found that Xray2 and Xray3 the cabling
is not consistent:
  
  87   Fri Mar 20 19:37:43 2020 JanCalibrationXray energies - rough 1. calibration  -- E calibration seems to have changed! ---

This calibration is based on the source data
  
ELOG V3.1.5-fc6679b