| 274 | Wed Apr  8 23:02:24 2020 | Laszlo | General | lmd to measurement pairing list |  | I use the following list of lmd files combined to a single root file for each measurement.
 It is very important to highlight that sometimes
 |   | 
| 273 | Wed Apr  8 20:06:10 2020 | Laszlo | Calibration | Counts in the K-REC peaks |  | 3 datasets were investigated: -124Xe with Scraper    (2. Xray calibration
 parameters needed)
 |     | 
| 272 | Mon Apr  6 23:00:51 2020 | Laszlo | Calibration | beam energies for 124Xe and 118Te measurements |  | "The e- cooler settings were the same for the 124Xe and 118Te beams" - Sergey --> speed
 of the ions is the same --> same MeV/u
 |   | 
| 271 | Mon Apr  6 04:39:19 2020 | Laszlo | Calibration | 2. Xray energy calibration |  | 30angle:  E[keV] = 0.015898 *ch-1.91 90angle:  E[keV] = 0.0211075*ch-2.39
 145angle: E[keV] = 0.01663  *ch-1.758
 | 9x   | 
| 270 | Sun Apr  5 20:34:42 2020 | Laszlo | Calibration | Xray energies - 1. calibration |  | 30angle:  E[keV] = 0.0158507 *ch-2.07456 90angle:  E[keV] = 0.01950*ch-2.76
 145angle: E[keV] = 0.017988  *ch-1.809
 | 8x   | 
| 269 | Thu Mar 26 12:05:22 2020 | Laszlo | Calibration | measurement of collimators |  | The measurements were taken by using a caliper ruler, with precision of +/-0.05mm
 |    | 
| 268 | Wed Mar 25 15:06:09 2020 | Laszlo | Analysis | quick summary of the run numbers |  |  |   | 
| 267 | Wed Mar 25 14:26:22 2020 | Laszlo | Calibration | run111 - Xray90 calib Am241, distance 167.5mm - LESS NOISE |  | Detector: 90 Source: 241Am
 Distance: 167.5mm
 |  | 
| 266 | Tue Mar 24 10:59:18 2020 | Jan | Calibration | Sources - Specifications |  | We used the following sources: 
 [B]Am241[/B] (OM666) [GSI - Uwe]
 |      | 
| 265 | Tue Mar 24 10:20:26 2020 | Laszlo | Calibration | Pictures of the xray calibration |  |  |      | 
| 264 | Tue Mar 24 00:11:35 2020 | Laszlo | Calibration | run116 - Xray35 calib AM241, distance 334mm |  | Detector: 35 Source: 241Am
 Distance: 334mm
 |  | 
| 263 | Mon Mar 23 22:25:41 2020 | Laszlo | Calibration | run115 - Xray35 Ba133 d=334mm, Xray90 Lead source, d=167.5mm |  | Simultaneous measurement of Xray35 and Xray90. The dead-time is <3%
 
 |  | 
| 262 | Mon Mar 23 21:47:40 2020 | Laszlo | Calibration | run114 - Xray145 calib Ba133, distance 305mm - LESS NOISE |  | Detector: 145 Source: 133Ba - strong source
 Distance: 305mm
 |  | 
| 261 | Mon Mar 23 20:52:37 2020 | Laszlo | Calibration | run113 - Xray145 calib Am241, distance 305mm - LESS NOISE |  | Detector: 145 Source: 241Am
 Distance: 305mm
 |  | 
| 260 | Mon Mar 23 19:32:53 2020 | Laszlo | Calibration | run112 - Xray90 calib Am241, distance 200mm - LESS NOISE |  | Detector: 90 Source: 241Am
 Distance: 200mm
 |  | 
| 259 | Mon Mar 23 18:17:27 2020 | Laszlo | Calibration | run110 - Xray90 calib Ba133, distance 167.5mm -low rate - LESS NOISE |  | Detector: 90 Source: Ba133 - weak source
 Distance: 167.5mm
 |  | 
| 258 | Mon Mar 23 18:15:55 2020 | Laszlo | General | noisy 90degree Xray det |  | Uwe found out, that there was some common ground effect between 90 and 145 degree detectors.
 after moving the grounds, we could record
 |  | 
| 257 | Mon Mar 23 17:40:33 2020 | Laszlo | Calibration | run109 - Xray90 calib Ba133, distance 167.5mm -low rate |  | Detector: 90 Source: Ba133 - weak source
 Distance: 167.5mm
 |  | 
| 256 | Mon Mar 23 17:32:42 2020 | Laszlo | Calibration | run108 - Xray90 calib Ba133, distance 167.5mm - high rate - deadtime too high! |  | Detector: 90 Source: Ba133 - strong source
 Distance: 167.5mm
 |  |