HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+, Page 1 of 12  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Fri Dec 31 00:00:00 100, Ken Ueberholz, Detectors, XUV-Detector Rates Screenshot_20220404-000343_Gallery.jpgScreenshot_20220404-000327_Gallery.jpgScreenshot_20220404-000411_Gallery.jpg
Count rates were measured for the XUV-
detector for a full circle:
Cathode: -5V, shields: 0V, main coil 8A, 
Entry  Wed Apr 20 10:09:50 2022, Rodolfo, Laser, Optic with spots SHG_Compensator_mitFleck_19-04-2022.jpgUV_Umlenkspiegel_mitFleck_19-04-2022.jpg
The average energy produced by the laser along the beamtime was about 10 mJ/puls. After about two weeks of beamtime some of the optics, especially the optic
placed just after the BBO crystall, got some spots, see pictures below. Here the diameter of the laser beam is about 6 mm.
Entry  Sat Apr 16 04:19:05 2022, Patrick, Simon, Alex, Sarper, Runs, Runs night + morning shifts 16.04. 
Runs of the night shift 16.04. after changing the wavelength range. The new range is 320.12 - 319.859 nm. The step size was decreased from 0.002 nm to 0.001
nm.
The goal was to have the same number of scans as before (261), however, the scan range in the labview program could not be set accordingly. We decided
Entry  Fri Apr 15 20:43:18 2022, Konstantin, Laser, Dye changed 
Dye of the laser system has been changed. UV laser power increased from ~8 mJ to 11 mJ pulse energy.
Entry  Fri Apr 15 12:53:25 2022, Max, Danyal, Laser, checked laser pulse energy after change in wavelength range 
previous scan range: 
the laser pulse energy @ 320.62 nm  was ~11 mJ 
Entry  Fri Apr 15 12:42:18 2022, Max, Danyal, Carsten, Konstantin, Patrick, Simon, General, laser wavelength range changed: 320.12 - 319.60 nm VoltageJump1830.PNG

Runs of the morning/afternoon/night shift. The wavelength range is now 320.12 - 319.60 nm.
Entry  Thu Apr 14 22:44:40 2022, Patrick, Simon, Danyal, Max, Runs, Runs night shift 14./15.04. 
Runs of the night shift. The wavelength range is 320.62 - 320.10 nm.

[TABLE border="1"]
Entry  Thu Apr 14 19:46:46 2022, Sarper & Konstantin & Rodolfo, Laser, Laser power 
We checked the laser power before starting the new scan range.

- Seed energy before Cobra @ 532 nm: 513 mJ (before 499 mJ).
Entry  Thu Apr 14 17:47:55 2022, Carsten, Rodolfo, Konstantin, Sarper, Runs, Afternoon shift runs 14.04 changing the wavelength scan range 
The new scan range is changed from [B]321.12 - 320.60 nm to 320.62 - 320.10 nm[/B]

[TABLE border="1"]
Entry  Thu Apr 14 12:38:38 2022, Sebastian, Detectors, XUV / Cu checkups 
Check regularly if the connection to the XUV and Cu detectors are still active and the detectors move in (after injection) and out (after killing the ion
beam). 
Entry  Thu Apr 14 12:27:18 2022, Sebastian, Max, Detectors, Connection to XUV / Cu reestablished! 
The connection to the XUV and Cu were lost during the nightshift.

The XUV detector was in the park position,
Entry  Thu Apr 14 10:57:11 2022, Carsten, Accelerator, Maintanence UNILAC / Op training 
We had no beam since about 1000. Also, we are still trying to find out why we have such a low intensity in the expereimnt. 
Presently, there is a planned maintenance in the UNILAC, beam is expected back in 25-30 mins. 
The profile grid TE3 in the transfer beamline in front of the target will be used to check if the extracted SIS beam has changed compared to before the
Entry  Thu Apr 14 08:41:42 2022, Carsten, Accelerator, Low intensity at the experiment after series of failures during the night 
After the recovery from the failures during the night, we see much lower intensity at the experiment (measured with the particle detector count rate).
Instead of 8000-10000 cts we see only 3000-4000cts at comparable SIS intensities.
Also ESR cooler seems to be slightly unstable (wiggling lines) during injection. However, it is visually evident from the Schottky that the injected intensity
Entry  Thu Apr 14 06:29:18 2022, Max, Sebastian, Accelerator, lesser ESR injection intensity 2022_04_14-lowCountRate-LesserESRInjection-SchottkyIntensity.JPG
We now have less ESR injection intensity (see screenshot attached).

This can also be seen in the count rate after "UNILAC/SIS failures".
Entry  Wed Apr 13 15:41:29 2022, Rodolfo & Sarper, Laser, Laser status 
We have checked the power of the laser and optimized it "a bit".

- Seed energy for the Cobra @ 532 nm: 508 mJ (before 496 mJ)*
Entry  Wed Apr 13 14:29:04 2022, Carsten, Rodolfo, Sarper, David, Pierre-Michel, Max, Sebastian, Runs, Afternoon - morning shift runs 13.+14.04. coolerProblemsAndFragments.JPGfragments.JPGBroaderIonBeam.JPGagainNormalWidth.JPG
Changed the wavelength scan range from [B]321.62 - 321.10 nm[/B] to [B][B]321.12 - 320.60[/B] nm[/B] after one scan.

[TABLE border="1"]
Entry  Tue Apr 12 21:56:54 2022, Volker, David, Pierre-Michel & Danyal, Alexandre, Sebastian, General, new laser wavelength scan range 321.62 - 321.10 nm Ecooler12VdropOn13_04_22_9_52am.pngEcooler18VEcoolerJumpOn13_04_22_13_18pm.png
New wavelength scan range is [B]321.62 - 321.10 nm[/B]. Before the new scan range (compare old scan range in [URL=https://elog.gsi.de/esr/E142/215] Entry
215[/URL]) was applied,
the cooler settings were slightly modified. Current was changed from [B]50 mA to 40 mA[/B], correction voltage was changed from [B]3943 V to 3948 V[/B].
Entry  Tue Apr 12 17:01:30 2022, Konstantin, Detectors, Voltage PMT North 
Voltage of PMT North has been increased from 2500 V to 2600 V. Additionally the threshold was adapted from 25 mV to 12 mV.
Entry  Tue Apr 12 15:05:56 2022, Rodolfo, Laser, Laser status 
We have checked the laser power at 14:15 *

- Pump energy after Q-Ray @ 532nm: 559 mJ.
Entry  Tue Apr 12 14:52:19 2022, Rodolfo, Detectors, PMT Nord, Middle. Datasheets ET-Enterprises_DataSheets_PMT_9423-9422.pdf
Attached are the datasheets for the solar-blind PMTs.

- PMT Nord. Type 9423. SN: 650 
ELOG V3.1.5-fc6679b