HOME
ESR
EXPERIMENTS
LABS etc
E121
E127
E132
E125
E143
laser_cooling_2021
E142
Ê128
E146
E0052
E0018
E0028
E0038
Nuclear hyperfine mixing in 229Th89+, Page 10 of 12
Not logged in
Find
|
Login
|
Help
Full
| Summary |
Threaded
-- All entries --
Last day
Last 3 Days
Last week
Last month
Last 3 Months
Last 6 Months
Last Year
-- Category --
Analysis
Calibration
DAQ
Laser
General
Runs
Detectors
Accelerator
233 Entries
Goto page
Previous
1
,
2
,
3
...
9
, 10,
11
,
12
Next
All
ID
Date
Author
Category
Subject
Text
54
Fri Apr 1 15:52:23 2022
Rodolfo
Detectors
UV PMT South
The "Solar Blind PMT" labeled as "PMT South"
has been installed by Simon and Max in the
gas target chamber on a CaF Viewport. The
53
Fri Apr 1 15:27:00 2022
Rodolfo
Detectors
Nitrogen for UV-PMTs
There was a question concerning the nitrogen
we use to rinse the "Solar Blind PMTs".
52
Fri Apr 1 11:19:57 2022
Rui-Jiu Chen
DAQ
S.M.A.R.T check on NTCAP
S.M.A.R.T. check has been done on NTCAP on
March 28th. All 12 hard disk drivers are
normal.
51
Fri Apr 1 11:00:41 2022
Rui-Jiu Chen
DAQ
Start NTCAP
Start taking data with NTCAP.
Carrier Frequency: 245MHz
IQ rate:5 MSamples/s
50
Fri Apr 1 02:29:40 2022
Carsten
ESR Cooler crashed
ESR cooler crashed. We could not restart the
cooler. Since we had several other problems,
we stop now.
49
Fri Apr 1 01:57:12 2022
Kristian
DAQ
Timing
with bunched primary beam (bunching frequency
1.94070 MHz) we have strong PMT (mid) and
particle detector signal. Time between both
47
Fri Apr 1 00:27:27 2022
Carsten
Settings Particle Detectors /Tests with primary beam
All the beam settings will be given in a separate
post.
46
Thu Mar 31 22:41:35 2022
Kristian
Calibration
HV divider
We noticed a difference in the electron cooler
high voltage:
45
Thu Mar 31 22:19:14 2022
Kristian
DAQ
How To Start MBS
(MBS server should be running at the LXG1297
machine. If not you will get an error message.
Then start the server - last command in console)
44
Thu Mar 31 20:07:47 2022
Shahab
DAQ
Schottky Display in AP Messhütte (or anywhere else)
[B]UPDATE: 16.05.2022[/B]
[I]Please use ATPPC029 instead of ATPPC019
43
Thu Mar 31 17:56:32 2022
Konstantin
DAQ
Photodiode test signal
We used a signal generator to provide a 4MHz
dummy-RF signal. This was divided by a factor
of 2 for laser-"bunch" synchronization and
42
Thu Mar 31 15:53:50 2022
Konstantin
Detectors
Control software for XUV and copper mirror detector
> Attached you find a manual for the software
used to control the movement of the XUV detector
and the copper mirror detector.
41
Thu Mar 31 13:28:17 2022
Jan Glorius
DAQ
DAQ-2 cabling docu
newest version V2 - valid since Mo 04.April
2022
40
Thu Mar 31 11:49:54 2022
Konstantin
Detectors
HV-Supply of particle detectors
We cross-checked the cabeling of the particle
detectors north and south. Particle detector
south is connected to channel [B]HV5[/B]
39
Wed Mar 30 09:27:15 2022
Rodolfo
Accelerator
Issue? Scrapers at E-Cooler
Carsten* forwarded me an E-Mail from Markus
yesterday. "Betreff: Scraper im ESR Elektronenkühler"
38
Tue Mar 29 21:37:13 2022
Rodolfo
Laser
Pump Laser
The gold "reflectors and end plates" of our
pump laser (Quanta-Ray PRO) have been exchanged
by new ones.
37
Tue Mar 29 12:47:26 2022
Ken Ueberholz
Parameter setting for XUV and copper mirror detector
Some Default Settings for the XUV
Detector. The copper
mirror should always move fully in.
36
Tue Mar 29 12:01:08 2022
Rodolfo
Laser
Laser targets
I have installed two targets* (size: 2 inches)
to monitor the position of the laser beam.
35
Fri Mar 25 14:31:24 2022
Rodolfo
Laser
Photodiode signal SO in the Messhütte
Measured at the Panel "Messhütte - Laserhütte",
BNC "Fast PD / SO".
34
Thu Mar 24 17:10:33 2022
Carsten
DAQ
DAQ / MBS VUPROM TDC buffer errors
VUPROM TDC buffer errors potentially due to
long TDC common Stop gate (>1ms from div
200).
Goto page
Previous
1
,
2
,
3
...
9
, 10,
11
,
12
Next
All
ELOG V3.1.5-fc6679b