HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+, Page 4 of 12  Not logged in ELOG logo
Entry  Sat Apr 9 03:57:10 2022, Carsten, Accelerator, New Cooler voltage after the crash of the cooler 
After the restart of the cooler the Ohmlabs divider showed a voltage 16V lower than before. This could also be seen as a slight shift in the Schottky lines
w.r.t to the bunching frequency and the marker. Corrected 
the voltages at injection to DeltaU +3963V and in Sc 5/7 to + 3943V. Now perfect match with the old Schottky frequency.
Entry  Sat Apr 9 03:46:31 2022, Carsten, Accelerator, Slits in TE line 
Changed GTEDSAH Rechts to 10mm
Entry  Sat Apr 9 03:27:47 2022, Patrick, DAQ, Laser scanning does_not_write_data_to_file.png
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
Entry  Sat Apr 9 02:55:38 2022, Carsten, Accelerator, Kühlerausfall und Gründe dazu 
Wir hatten heute Abend eine Kühlerausfall, der uns mehrere Stunden beschäftigt hat, weil  gem. Anzeige in Device Control kein Netzteil  mit „rot“ angezeigt/ausgefallen
 war (siehe unten).
Nach Rufbereitschaft Serge, dann Regina sowie Controls konnte zunächst gefunden werden, dass das Netzteil einer Clearingelektrode (GECEKD2D) keinen korrekten
Entry  Fri Apr 8 23:56:07 2022, Patrick, DAQ, DAQ improvement 
As requested in ID:146 and ID:147, the laserscan tab was improved:
[LIST]
[*] The laserscan tab shows each observable plotted against the predefined wavelength setpoints now.
Entry  Fri Apr 8 22:26:44 2022, Patrick, DAQ, Spike in all detectors 
In the previous shift, there was a spike in all detectors in the same scan step. Apparently, this was NOT a resonance since in the next step it appeared
again but at a different wavelength.
The current hypothesis is, that this is a disruption signal from one of the machines.
Entry  Fri Apr 8 17:57:45 2022, Konstantin, DAQ, Laser scanning Artifacts_in_PMT_MiddleAndNorth.png


[TABLE border="1"]
Entry  Fri Apr 8 17:44:25 2022, Konstantin, Calibration, Final Timing 
We measured the timing for the laser bunch synchronization. A summary is given below.

[TABLE border="1"]
Entry  Fri Apr 8 15:45:24 2022, YL, , Spectrum manipulation 5 IMG_3027.JPG
See attached
Entry  Fri Apr 8 15:42:20 2022, Yuri, Accelerator, Slits in the TE line 
At around 15:00, The slits in the TE line were changed to :
TE4DS5HL -12 mm -> -20 mm
TE5DSAH 0 mm -> +15 mm
Entry  Fri Apr 8 15:39:36 2022, all, Runs, first data after timing set up 
started taking data again

started at 15:40  
Entry  Fri Apr 8 14:46:02 2022, Danyal & Sebastian & Volker & Ken, Detectors, high dark countrate Cu PMT solved IMG_20220408_082804.jpgIMG_20220408_083031.jpg
The problem - "discovered" by Volker and Ken - of the high dark countrate of the Cu PMT has been solved. 

(when the Cu mirror was moved in, the dark countrate was very high ... also without ion beam) 
Entry  Fri Apr 8 14:43:50 2022, Wilfried, DAQ, laser timing - Q-switch start IMG_20220408_143753_resized_20220408_023840170.jpg
below are my notes concerning laser timing / Q-switch start
Entry  Fri Apr 8 14:14:28 2022, Kristian, DAQ, Delay of bunch RF 
The bunch HF is now fed through a delay to shift the stop trigger relative to the PMT signal. The delay generator is placed directly after the CFG and before
splitting the signal for the two DAQs. The relative timing of both DAQs was not changed.
Entry  Fri Apr 8 14:12:57 2022, morning shift, DAQ, timing for h=1 2022-04-08_timing_h1_final.png
The timing for h=1 has been set up 

The time window is now twice as large as before! 
Entry  Fri Apr 8 12:04:13 2022, Ragan, Runs, NTCAP started IMG_0947.jpg
NTCAP is started for the data taking.

Carrier Frequency: 242.02 M
Entry  Fri Apr 8 11:16:57 2022, Kristian, Laser, Wavemeter Wavemeter.PNGWavemeter2.PNG
We changed the fiber to the wavemeter. The third fiber worked best and we could operate the WM at low exposure times, i.e., below the laser repetition rate.
We used the neutral density filter to adapt the laser power that only one laser pulse is measured by the WM each time (Exposure time =34ms)
Entry  Fri Apr 8 11:02:04 2022, DW, YL, RSS, SL, , Change h=2 to h=1 
After changing to h=1
The frequency manipulation #5: f=242.021487MHz (124th harmonic) [no bunching]
Manipulation #7: f=242.021812 MHz [with bunching]
Entry  Fri Apr 8 10:44:55 2022, Ragan, Yuri, Danyal, Sergey, Runs, NTCAP stopped 
NTCAP is stopped. Sergey will now set an ion bunch instead of two in the ESR.

SC_2022-04-08_09-39-33
Entry  Fri Apr 8 09:39:04 2022, Ragan, Runs, NTCAP started 
Carrier Frequency: 242.2 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
ELOG V3.1.5-fc6679b