Sat Apr 9 03:46:31 2022, Carsten, Accelerator, Slits in TE line
|
Changed GTEDSAH Rechts to 10mm |
Tue Apr 12 13:24:40 2022, Alexandre, Sebastian, Detectors, Sideband in Schottky
|
Sometimes we see a broad sideband at lesser schottky frequencies to the high intensity thorium beam.
Pic 1 shows the relative broad sideband on the left.
|
Tue Apr 12 14:11:52 2022, Ragan, Detectors, Sideband in Schottky
|
> Sometimes we see a broad sideband at lesser Schottky frequencies to the high-intensity thorium beam.
>
> Pic 1 shows the relative broad sideband on the left.
|
Fri Apr 1 00:27:27 2022, Carsten, , Settings Particle Detectors /Tests with primary beam
|
All the beam settings will be given in a separate post.
We have 238U92+ centered in the ring with typically a few times 10⁷ ions.
|
Wed Apr 6 14:52:29 2022, Rodolfo, Runs, Scraping the ion beam
|
We are T-scraping the ion beam. |
Sat Apr 2 18:19:47 2022, Yuri, Accelerator, Scraping
|
|
Sun Apr 3 11:01:57 2022, Rodolfo, Accelerator, Scrapers @ E-Cooler
|
We the mid- and night shift (Wilfried, Simon, Patrick, Ken and me) we not able to manipulate the scrapers in the E-Cooler.
- Carsten contacted Markus concerning this issue.
|
Fri Apr 1 23:12:37 2022, Patrick, Detectors, Scraper positions "only primary beam"
|
Scraper positions where there is only primary beam for detector tests. Use only one outer and one inner scraper at the same time.
[B]Move detectors out before new beam![/B]
|
Thu Mar 31 20:07:47 2022, Shahab, DAQ, Schottky Display in AP Messhütte (or anywhere else)
|
[B]UPDATE: 16.05.2022[/B]
[I]Please use ATPPC029 instead of ATPPC019 for the 245 MHz Schottky
|
Sat Apr 2 16:10:22 2022, Markus, Bernd, Accelerator, SIS setting complete, U90+ mit Be target: 2022-04-02_16-04-00_SIS18_FAST_E142_20220325_110213.C1.txt
|
# Version: 1.2
# Comment: ParamModi Export at 2022-04-02T16:04:10.778077+02:00[Europe/Berlin]
# Context: SIS18_FAST_E142_20220325_110213.C1
|
Fri Apr 1 11:19:57 2022, Rui-Jiu Chen, DAQ, S.M.A.R.T check on NTCAP
|
S.M.A.R.T. check has been done on NTCAP on March 28th. All 12 hard disk drivers are normal. |
Thu Apr 14 22:44:40 2022, Patrick, Simon, Danyal, Max, Runs, Runs night shift 14./15.04.
|
Runs of the night shift. The wavelength range is 320.62 - 320.10 nm.
[TABLE border="1"]
|
Sun Apr 10 22:47:05 2022, Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, Kristian, Runs, Runs night shift 10.04 + morning shift 11.04
|
These are the measurement runs of the night shift (10/11.04.).
Wavelength range is still 322.81 - 322.40 nm.
|
Sat Apr 16 04:19:05 2022, Patrick, Simon, Alex, Sarper, Runs, Runs night + morning shifts 16.04.
|
Runs of the night shift 16.04. after changing the wavelength range. The new range is 320.12 - 319.859 nm. The step size was decreased from 0.002 nm to 0.001
nm.
The goal was to have the same number of scans as before (261), however, the scan range in the labview program could not be set accordingly. We decided |
Sun Apr 10 15:16:19 2022, Carsten, Rodolfo, Sarper, Runs, Runs afternoon shifts 10.04
|
These are the measurement runs of the afternoon shift (10.04.).
Wavelength range is changed from 323.2 - 322.8 nm to 322.81 - 322.40 nm.
|
Sat Apr 2 22:43:36 2022, Carsten, , Ring set-up for 229/89+, scraped and bunched, Iel=500mA at injection and 100mA during measurment
|
The ring SIS/ESR were set up for 229/89+. Pölease note that this not only 1e- Th but ponteially also the isobars of Ac (0e-1)/Pa(2e-, strongest) and U(3e-).
SIS eenrgy plus 3.5 MeV/u w.r.t. to the 238U92+ settings. We nearly don't see any of the primary beam in the ring
Cooling energy is 208998 KeV (read from the ESR multimeter).
|
Fri Apr 8 09:33:52 2022, Dayshift 1 8.4.2022, , Reference Measurements
|
DAQ1: lxg1297:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
Mon Aug 2 10:54:44 2021, Rodolfo, Laser, Quanta Ray - Water leak in the power supply
|
After fixing the broken water connector in the laser head a new water leak appeared in the Quanta-Ray power supply.
We found the leak in the water pump flow line (arrow up). We took the water connectors appart, rebuilt the teflon seeling and put the water connectors |
Mon Jul 26 17:15:47 2021, Rodolfo, Laser, Quanta Ray - Water leak
|
On Thursday afternoon (22.07.2021) a water connector has broken and water has been spilled in the laser head. |
Fri Apr 8 03:54:39 2022, Patrick, Laser, Q-switch window
|
The solution of increasing the Q-switch window from 2us to 5us described in [I]ID:148[/I] fulfills its purpose, but the window is unstable.
We saw the laser jumping across the screen which was because the Laser had a lot of hickups.
The hickups come from the signal which determines the duration of the window.
|