ID |
Date |
Author |
Category |
Subject |
135
|
Thu Apr 7 10:10:31 2022 |
Ragan, Yuri, Danyal | Accelerator | Problem in UNILAC |
The power supply of the UNILAC magnet is broken. The repair will take up to 6-7 hours. |
96
|
Tue Apr 5 12:17:12 2022 |
Ragan, Rui-Jiu | DAQ | Instructions to start and stop NTCAP |
Attached are the instructions to start and stop data taking using the NTCAP. |
154
|
Fri Apr 8 09:13:38 2022 |
Ragan, Danyal | Accelerator | No beam |
Cathode is being changed at the ion source. So no beam for the moment.
Sergey has been contacted to change from two ion bunches to a single ion bunch in the ESR. |
151
|
Fri Apr 8 06:53:29 2022 |
Ragan | Runs | NTCAP data taking |
NTCAP is taking data since 07.04.2022 23:53:47
Carrier Frequency: 242.2 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
SC_2022-04-07_23-53-47
\IQ_2022-04-07_23-53-47 |
155
|
Fri Apr 8 09:32:32 2022 |
Ragan | Runs | NTCAP stopped |
Since there is no beam, NTCAP is stopped.
Carrier Frequency: 242.2 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
SC_2022-04-07_23-53-47
\IQ_2022-04-07_23-53-47 |
157
|
Fri Apr 8 09:39:04 2022 |
Ragan | Runs | NTCAP started |
Carrier Frequency: 242.2 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
SC_2022-04-08_09-39-33
\IQ_2022-04-08_09-39-33 |
161
|
Fri Apr 8 12:04:13 2022 |
Ragan | Runs | NTCAP started |
NTCAP is started for the data taking.
Carrier Frequency: 242.02 M
IQ Rate (S/s): 10 M
Reference level (dBm): -30
SC_2022-04-08_12-04-28
\IQ_2022-04-08_12-03-50 |
218
|
Tue Apr 12 14:11:52 2022 |
Ragan | Detectors | Sideband in Schottky |
> Sometimes we see a broad sideband at lesser Schottky frequencies to the high-intensity thorium beam.
>
> Pic 1 shows the relative broad sideband on the left.
>
> Pic 2 depicts the vanishing of the sideband after ~1-2 minutes.
These are not the sidebands. Maybe some short-lived fragments survive even after the scrapers are used? |
128
|
Thu Apr 7 05:44:40 2022 |
Pierre-Michel, Ragan | DAQ | NTCAP settings |
Carrier Frequency: 244M
IQ Rate: 10M
reference Level: -30
RF - Attenuation: 40
IF - Attenuation: 19.9288
Masterpath p:\E142\sc
Masterpath p:\E142\IQ
Display: No Display
NTCAP data-taking is stopped at 5:44.
IQ_2022-04-07_02-40-05
SC_2022-04-07_02-40-05 |
210
|
Mon Apr 11 14:45:11 2022 |
Paul, Konstantin, Sarper, David | Runs | Afternoon shift runs 11.04 |
We continue scan the new wavelength scan range is 322.41 - 322.00 nm
Time | SIS-intensity | cooler-HV | LMD-DAQ1 file nr | LMD-DAQ2 file nr | scan range(nm) | step width(nm) | dwell time(s) | particles counted | comments
|
14:45 | 1.9e9 | 209618 | 0189 | 7924 | 322.41 - 322.00 | 0.002 | 5s | 9100 |
|
18:54 | 1.83e9 | 209616 | 0191 | 7926 | 322.41 - 322.00 | 0.002 | 5s | 7300 |
|
19:34 | 1.7e9 | 209615 | 0192 | 7927 | 322.41 - 322.00 | 0.002 | 5s | 7110 |
|
20:13 | 1.8e9 | 20961X | 0193 | 7928 | 322.41 - 322.00 | 0.002 | 5s | 6730 |
|
20:53 | 1.8e9 | 209614 | 0194 | 7929 | 322.41 - 322.00 | 0.002 | 5s | 7550 | During the scan the ecooler voltage changed to approximately 209613
|
21:43 | 1.7e9 | 209614 | 0195 | 7931 | 322.41 - 322.00 | 0.002 | 5s | 7000 |
|
22:30 | 1.6e9 | 209612 | 0197 | 7932 | 322.41 - 322.00 | 0.002 | 5s | 7300 |
|
23:10 | 1.4e9 | 209614 | 0198 | 7933 | 322.41 - 322.00 | 0.002 | 5s | 6100 | For this measurement, ESR_Ring went red, we realized only after completing the measurement. The sections went red are Dipole Correction Points (GE01KP17-18-19-20, 4 in total) @~23:45 and fixed @01:35...
|
01:52 | 1.6e9 | 209614 | 0199 | 7934 | 322.41 - 322.00 | 0.002 | 5s | 7151 |
|
02:32 | 1.7e9 | 209616 | 0200 | 7935 | 322.41 - 322.00 | 0.002 | 5s | 8368 |
|
03:22 | 1.8e9 | 209617 | 0201 | 7936 | 322.41 - 322.00 | 0.002 | 5s | 7005 |
|
04:01 | 1.9e9 | 209616 | 0202 | 7937 | 322.41 - 322.00 | 0.002 | 5s | | Ecooler voltage increased 4 volts during the run
|
|
138
|
Thu Apr 7 14:04:36 2022 |
Paul Indelicato | General | no beam until 17:00 |
10:10 There is a problem with the UNILAC.
The repair will take several hours. (4 hrs?)
14:00 We just enquired about the status and when beam could be back.
The current estimate is 17:00.
More work on laser aligment is required.
Estimated time: 1 hour |
203
|
Sun Apr 10 22:47:05 2022 |
Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, Kristian | Runs | Runs night shift 10.04 + morning shift 11.04 |
These are the measurement runs of the night shift (10/11.04.).
Wavelength range is still 322.81 - 322.40 nm.
Time | SIS-intensity | cooler-HV | LMD-DAQ1 start-end | LMD-DAQ2 start-end | scan range(nm) | step width(nm) | dwell time(s) | particles counted | comments
|
23:05 | 1.75e9 | 209620 | 0165 - 0165 | 7901-7901 | 322.81 - 322.40 | 0.002 | 5s | 8500 |
|
23:57 | 1.?e9 | 209620 | 0166 - 0166 | 7902-7902 | 322.81 - 322.40 | 0.002 | 5s | 8500 | The HV of PMT south was changed from 2.5kV to 3kV
|
00:35 | 1.4e9 | 209620 | 0167 - 0167 | 7903-7903 | 322.81 - 322.40 | 0.002 | 5s | 4100 |
|
01:18 | 1.75e9 | 209620 | 0168 - 0168 | 7904-7904 | 322.81 - 322.40 | 0.002 | 5s | 4400 |
|
01:58 | 1.4e9 | 209620 | 0169 - 0169 | 7905-7905 | 322.81 - 322.40 | 0.002 | 5s | 7700 | The ESR went red after this run. The on-call service was called because of some power supply failure, ETA: 45min.
|
04:12 | 1.75e9 | 209620 | 0171 - 0171 | 7906-7906 | 322.81 - 322.40 | 0.002 | 5s | 7600 | With the start of this run, the wavemeter, labview and the python script were restarted. The python script was slightly adjusted.
|
04:50 | 1.6e9 | 209620 | 0172 - 0172 | 7907-7907 | 322.81 - 322.40 | 0.002 | 5s | 8300 |
|
05:31 | 1.6e9 | 209619 | 0173 - 0173 | 7908-7908 | 322.81 - 322.40 | 0.002 | 5s | 8100 |
|
06:09 | 1.6e9 | 209615 | 0174 - 0174 | 7909-7909 | 322.81 - 322.40 | 0.002 | 5s | 8000 |
|
06:52 | 1.8e9 | 209615 | 0175 - 0175 | 7910-7910 | 322.81 - 322.40 | 0.002 | 5s | 8800 | at 07:08 the cooler voltage jumped from 209615 V to 209619 V
|
07:30 | 1.9e9 | 209619 | 0176 - 0176 | 7911-7911 | 322.81 - 322.40 | 0.002 | 5s | forgotten | at 07:50 the cooler voltage jumped from 209619 to 209610 V and back again at to 209619 V
|
08:18 | 1.8e9 | 209619 | 0177 - 0177 | 7912-7912 | 322.81 - 322.40 | 0.002 | 5s | 9100 | 08:47 the cooler voltage jumped from 209618 to 209614 V and back again at to 209618 V
|
08:55 | 1.75e9 | 209619 | 0178 - 0178 | 7913-7913 | 322.81 - 322.40 | 0.002 | 5s | 8300 |
|
09:36 | 1.6e9 | 209619 | 0179 - 0179 | 7914-7914 | 322.81 - 322.40 | 0.002 | 5s | 8194 |
|
The SIS intensity is taken directly before injection to ESR. Not the SIS peak intensity as in yesterday's table. |
232
|
Thu Apr 14 22:44:40 2022 |
Patrick, Simon, Danyal, Max | Runs | Runs night shift 14./15.04. |
Runs of the night shift. The wavelength range is 320.62 - 320.10 nm.
Time | SIS-intensity | cooler-HV | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step width(nm) | dwell time(s) | particles counted | comments
|
22:38 | 9.5e8 | 209621 | 0305 | 8018 | 320.62 - 320.10 | 0.002 | 5s | 5000 | laser scan was started 1 min late
|
23:38 | 9.8e8 | 209622 | 0306 | 8019 | 320.62 - 320.10 | 0.002 | 5s | 5550 |
|
00:32 | 1.0e9 | 209621 | 0307 | 8020 | 320.62 - 320.10 | 0.002 | 5s | 4000 | After ~4000 counted particles (second scan, 1:04 am) the e-cooler went off.
|
01:21 | 1.1e9 | 209622 | 0308 | 8021 | 320.62 - 320.10 | 0.002 | 5s | 6700 |
|
02:12 | 1.2e9 | 209623 | 0309 | 8022 | 320.62 - 320.10 | 0.002 | 5s | 6400 |
|
03:03 | 1.1e9 | 209624 | 0310 | 8023 | 320.62 - 320.10 | 0.002 | 5s | 5500 |
|
03:55 | 1.3e9 | 209622 | 0311 | 8024 | 320.62 - 320.10 | 0.002 | 5s | 7700 |
|
04:47 | 1.2e9 | 209622 | 0312 | 8025 | 320.62 - 320.10 | 0.002 | 5s | 5300 |
|
05:36 | 1.2e9 | 209620 | 0313 | 8026 | 320.62 - 320.10 | 0.002 | 5s | 6666 |
|
06:25 | 1.2e9 | 209619 | 0314 | 8027 | 320.62 - 320.10 | 0.002 | 5s | 6500 | change of shift crew (Danyal & Max)
|
07:16 | 1.2e9 | 209618 | 0315 | 8028 | 320.62 - 320.10 | 0.002 | 5s | 5230 |
|
08:13 | 1.2e9 | 209620 | 0316 | 8029 | 320.62 - 320.10 | 0.002 | 5s | 6600 |
|
09:04 | 1.1e9 | 209619 | 0317 | 8030 | 320.62 - 320.10 | 0.002 | 5s | 6660 |
|
09:54 | 1.2e9 | 209619 | 0318 | 8031 | 320.62 - 320.10 | 0.002 | 5s | 5950 |
|
10:49 | 1.1e9 | 209619 | 0319 | 8032 | 320.62 - 320.10 | 0.002 | 5s | 6580 |
|
11:38 | 1.1e9 | 209617 | 0320 | 8033 | 320.62 - 320.10 | 0.002 | 5s | 5380 |
|
|
|
236
|
Sat Apr 16 04:19:05 2022 |
Patrick, Simon, Alex, Sarper | Runs | Runs night + morning shifts 16.04. |
Runs of the night shift 16.04. after changing the wavelength range. The new range is 320.12 - 319.859 nm. The step size was decreased from 0.002 nm to 0.001 nm.
The goal was to have the same number of scans as before (261), however, the scan range in the labview program could not be set accordingly. We decided to go with 262 scans after the first run.
Time | SIS-intensity | cooler-HV | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step width(nm) | dwell time(s) | particles counted | comments
|
04:24 | 1.28e9 | 209616 | 0338 | 8051 | 320.12 - 319.86 | 0.001 | 5s | 6750 | This run had only 260 scans
|
05:18 | 1.20e9 | 209615 | 0339 | 8052 | 320.12 - 319.859 | 0.001 | 5s | 6300 | First run with 262 scans
|
06:09 | 1.36e9 | 209615 | 0340 | 8053 | 320.12 - 319.859 | 0.001 | 5s | 6910 | laser scan started 2 min late at 650 counts
|
07:03 | 1.24e9 | 209615 | 0341 | 8054 | 320.12 - 319.859 | 0.001 | 5s | 6330 |
|
07:59 | 1.23e9 | 209615 | 0342 | 8055 | 320.12 - 319.859 | 0.001 | 5s | 5712 |
|
08:51 | 1.1e9 | 209615 | 0343 | 8056 | 320.12 - 319.859 | 0.001 | 5s | 4380 |
|
09:44 | 1.28e9 | 209615 | 0344 | 8057 | 320.12 - 319.859 | 0.001 | 5s | 5135 |
|
10:39 | 1.25e9 | 209615 | 0345 | 8058 | 320.12 - 319.859 | 0.001 | 5s | 5860 |
|
11:30 | 1.2e9 | 209615 | 0346 | 8059 | 320.12 - 319.859 | 0.001 | 5s | 6500 |
|
12:22 | 1.23e9 | 209618 | 0347 | 8060 | 322.41 - 322.00 | 0.002 | 5s | 5200 | Laser wavelength scan range is changed to 322.41 - 322.00 nm and increment to 0.002 nm.
|
13:04 | 1.2e9 | 209616 | 0348 | 8061 | 322.41 - 322.00 | 0.002 | 5s | 4411 |
|
13:50 | 1.2e9 | 209616 | 0349-0351 | 8062-8063 | 322.41 - 322.00 | 0.002 | 5s | 10651 | last scan / 10 loops ~ 3 hours of scan -- Ecooler failed, realized at 16:05 we were on loop8, laser is turned off at 16:10
|
|
59
|
Fri Apr 1 23:12:37 2022 |
Patrick | Detectors | Scraper positions "only primary beam" |
Scraper positions where there is only primary beam for detector tests. Use only one outer and one inner scraper at the same time.
Move detectors out before new beam!
outer:
GE02DD2_G: 0mm
GE01DD2AG: -10mm
inner:
GE01DD2IG: 40mm |
117
|
Wed Apr 6 22:29:55 2022 |
Patrick | Accelerator | ECooler voltages |
ECooler voltages:
06.04:
22:30: 209676 V
00:14: 209647 V
07.04.:
00:46: 209647 V
04:06: 209648 V |
118
|
Thu Apr 7 00:15:11 2022 |
Patrick | Laser | Laser timing adjusted |
We have adjusted the laser timing using PMT-middle. Bunching freq. is 3.90362 MHz.
PD NO: 7.79
PD SO: 43.37
mean: 25.08
PMT middle: 25.02
Delta = 0.6 bins * 3.3 ns = 0.2 ns
New injection:
PMT middle: 25.48 / 101.91
ParDet south: 74.31 / 151.31
ParDet south - PMT middle (furthest / closest distance) = 161.1 ns / 91.1 ns |
120
|
Thu Apr 7 01:19:05 2022 |
Patrick | Laser | Laser scanning |
We did some laser scans. The wavemeter shows spikes during scans. This may be just while the frequency changes. See picture. |
122
|
Thu Apr 7 03:06:15 2022 |
Patrick | DAQ | DAQ file sizes |
The file sizes of DAQ1 (229Thor) and DAQ2 (229Th, "Carstens DAQ") are different.
During 20min of laser scanning DAQ1 saves 4 files while DAQ2 saves 2 files which was 3.3 GB and 2 GB, respecively.
The file sizes of DAQ2 should always be greater than those of DAQ1, because of extra ADC data!
The problem seems to be solved after the recabling ID: 125. |
124
|
Thu Apr 7 03:44:21 2022 |
Patrick | DAQ | Measurement procedure |
Measurement procedure
- The laser hits the screen 'NO' (atppc017) right side of the burn spot
- Be ready to start a laser scan
- Scan laser goes into wavemeter!
- Have mbs connected
- Have go4 connected
Preparation done
- 'open file (...)' for DAQ1 & DAQ2
- request a new beam
- check scraping
- 'sta ac' for DAQ1 & DAQ2
- start laser scan
laser scan started
- 'sh ac' for DAQ1 & DAQ2
- check that acquisition is RUNNING and file is OPENED
- fill in table information in the ELOG.
laser scan complete
- 'sto ac' for DAQ1 & DAQ2
- 'close file'
- check if preparation is still fulfilled
- repeat from Preparation done |