Entry  Thu Mar 24 17:10:33 2022, Carsten, DAQ, DAQ / MBS VUPROM TDC buffer errors 
VUPROM TDC buffer errors potentially due to long TDC common Stop gate (>1ms from div 200).
Signal piped through coinc. unit gicv es shorter sginal (~200ns). This signal used as common stop seems to have cured the TDC buffer errors..
Entry  Thu Mar 31 13:28:17 2022, Jan Glorius, DAQ, DAQ-2 cabling docu e142_daq2_cabling.pdfe142_daq2_cabling_V2.pdfe142_daq2_cabling_V2.ods
newest version V2 - valid since Mo 04.April 2022
Entry  Thu Mar 31 17:56:32 2022, Konstantin, DAQ, Photodiode test signal Timing_Photodiodes.png
We used a signal generator to provide a 4MHz dummy-RF signal. This was divided by a factor of 2 for laser-"bunch" synchronization and by a factor of 200
for to provide the stop-tigger attached to the DAQ. With this Dummy-RF we were able to measure the relative arrival times of the laser at the photodiodes
located in the laser towers. Note that the signal of the southern photodiode arrives first. This is because the scale of the time axis is given in (10/3ns)-bins
Entry  Thu Mar 31 20:07:47 2022, Shahab, DAQ, Schottky Display in AP Messhütte (or anywhere else) 04_PM_schottky.png
[B]UPDATE: 16.05.2022[/B]

[I]Please use ATPPC029 instead of ATPPC019 for the 245 MHz Schottky
Entry  Thu Mar 31 22:19:14 2022, Kristian, DAQ, How To Start MBS 
(MBS server should be running at the LXG1297 machine. If not you will get an error message. Then start the server - last command in console)

Just enter the following commands in the console of any PC
Entry  Fri Apr 1 01:57:12 2022, Kristian, DAQ, Timing Particle-PMT_Mid.pngPMT-PD.pngIMG_20220331_235846.jpg
with bunched primary beam (bunching frequency 1.94070 MHz) we have strong PMT (mid) and particle detector signal. Time between both detectors was measured
(in 3.3ns bins)
Entry  Fri Apr 1 11:00:41 2022, Rui-Jiu Chen, DAQ, Start NTCAP 
Start taking data with NTCAP.
Carrier Frequency: 245MHz
IQ rate:5 MSamples/s
Entry  Fri Apr 1 11:19:57 2022, Rui-Jiu Chen, DAQ, S.M.A.R.T check on NTCAP image001(04-01-11-10-06).jpg
S.M.A.R.T. check has been done on NTCAP on March 28th. All 12 hard disk drivers  are normal.
Entry  Sun Apr 3 12:35:17 2022, Max, DAQ, Go 4 manual go4_1.pnggo4_2.pnggo4_3.png
How to start go4?:

- go to console
Entry  Mon Apr 4 00:19:16 2022, Ken Ueberholz, DAQ, FirstTests 44_AM.png06_AM.png
New measurement:
Start: About 00:12
Stop: About 00:43
Entry  Mon Apr 4 15:27:31 2022, Sebastian, DAQ, PMT over wavelength histogram 
Kristian and I tried to create a "PMT over wavelength" histogram under step2Anl/LabView.
Unfortunately, the code ist still in progress and does not work yet.
Entry  Mon Apr 4 15:32:38 2022, Kristiand & Sebastian, DAQ, Go4 Backup Code 
Due to some code changes (PMT over wavelength historgram) Kristian and I backuped a working Go4 code in the current go4 folder on the atplaser linux account.

It can be found in the subfolder "working go4 code 4.4.2020"
Entry  Mon Apr 4 16:26:04 2022, Carsten, DAQ, Backup Harddisk /External Disk 
Since we cannot write on lustre, as a temproal solution two RAID 1 (mirror) USB hard disks  with cpacity 6 TB each are installed.
The disk on the lxg1297 the disk can be accessd at /meida/sdd2
an on the lxg0155 at /media/sdc2
Entry  Mon Apr 4 17:26:48 2022, Jan Glorius, DAQ, MSCF shaper settings setting_e142.start3
attached are the current shaper settings.

list of currently connected channels:
Entry  Mon Apr 4 19:05:57 2022, Carsten, DAQ, Cooler Frequency Signals Ion current not connected until now 
Up to now the signals for 
Ion Current
Cooler elctron current
Entry  Tue Apr 5 08:00:50 2022, Kristian & Sebastian, DAQ, MBS Channellist (so far) 
MBS Channellist (time 5.4.2022)

0 Anode
Entry  Tue Apr 5 11:14:52 2022, Sebastian, Kristian, Pierre-Michel, DAQ, PMT over wavelength histogram v2 
Today we finalized the PMT over wavelength histogram in:

Analysis/Histograms/Step2_Anl/LabView/laserscan
Entry  Tue Apr 5 12:17:12 2022, Ragan, Rui-Jiu, DAQ, Instructions to start and stop NTCAP Start_data_taking.pdfStop_data_taking.pdfStart_data_taking.pngStop_data_taking.png
Attached are the instructions to start and stop data taking using the NTCAP.
Entry  Tue Apr 5 13:28:06 2022, Kristian, DAQ, Spikes in PMT signal spike.png
Occasionally, we noticed a spike in the PMT signal. No ions in the ring and at PMT Cu was not even supplied with HV. Has to be some electronic pickup
Entry  Tue Apr 5 14:51:51 2022, Kristian, DAQ, dark counts 
We shielded the PMTs, i.e., the N2 tube, which improved the Cu PMT background rate.

Without beam and ESR lights off: