Entry  Fri Apr 8 05:46:35 2022, V. Hannen, Detectors, High background rate in Cu PMT points to light leak in ESR or active pressure gauge 
The background rate of the Cu mirror PMT is only about 300 cps when the mirror is in the parking position but ca. 7000 cps when in the beam position, even
if no beam is present in the ESR. This was noticed when we had an "empty injection" into the ring, causing the Cu mirror to drive into the beam position
with no ions present. 
Entry  Fri Apr 8 14:46:02 2022, Danyal & Sebastian & Volker & Ken, Detectors, high dark countrate Cu PMT solved IMG_20220408_082804.jpgIMG_20220408_083031.jpg
The problem - "discovered" by Volker and Ken - of the high dark countrate of the Cu PMT has been solved. 

(when the Cu mirror was moved in, the dark countrate was very high ... also without ion beam) 
Entry  Sun Apr 10 03:56:56 2022, V. Hannen, Detectors, Low rate in PMT South 2022-04-10_PMTs_over_wavel-03_56.png
PMT south shows a much lower count rate than observed in elog entry 170.

Yesterday the voltage on this PMT was increased to 2400 V and the threshold setting of the CAEN CFD to 90 units, (from 70 units before) however the low
Entry  Tue Apr 12 13:24:40 2022, Alexandre, Sebastian, Detectors, Sideband in Schottky IMG_20220412_125530_resized_20220412_125736155.jpgIMG_20220412_125643_resized_20220412_125736537.jpg
Sometimes we see a broad sideband at lesser schottky frequencies to the high intensity thorium beam.

Pic 1 shows the relative broad sideband on the left.
    Reply  Tue Apr 12 14:11:52 2022, Ragan, Detectors, Sideband in Schottky 
> Sometimes we see a broad sideband at lesser Schottky frequencies to the high-intensity thorium beam.

> Pic 1 shows the relative broad sideband on the left.
Entry  Tue Apr 12 14:52:19 2022, Rodolfo, Detectors, PMT Nord, Middle. Datasheets ET-Enterprises_DataSheets_PMT_9423-9422.pdf
Attached are the datasheets for the solar-blind PMTs.

- PMT Nord. Type 9423. SN: 650 
Entry  Tue Apr 12 17:01:30 2022, Konstantin, Detectors, Voltage PMT North 
Voltage of PMT North has been increased from 2500 V to 2600 V. Additionally the threshold was adapted from 25 mV to 12 mV.
Entry  Thu Apr 14 12:27:18 2022, Sebastian, Max, Detectors, Connection to XUV / Cu reestablished! 
The connection to the XUV and Cu were lost during the nightshift.

The XUV detector was in the park position,
Entry  Thu Apr 14 12:38:38 2022, Sebastian, Detectors, XUV / Cu checkups 
Check regularly if the connection to the XUV and Cu detectors are still active and the detectors move in (after injection) and out (after killing the ion
beam). 
Entry  Fri Aug 6 10:23:00 2021, Rodolfo, DAQ, atplaser on r4l-41 
The user [B]atplaser[/B] is now existing on the [B]R4L-XY[/B] nodes for the ESR group.

R4 = [B]R[/B]IO[B]4[/B] Computer
Entry  Fri Aug 6 10:40:54 2021, Rodolfo, DAQ, mbs in atplaser 
The "MBS program" (data acquisition) is now located within atplaser@r4l-41, in the folder "th21":

Entry  Fri Nov 5 17:23:23 2021, Rodolfo, DAQ, Dye trigger out  Cobra_ScanSettings_05-11-2021.PNG
The dye laser offers different scan settings (see pages 65 - 67 of the user manual). We are using the following setting, see attached screenshot, to produce
the status "laser busy" for the DAQ. 
Entry  Thu Mar 10 15:20:26 2022, Jan , DAQ, analog branch signal patches to Messhütte  red_panel_ESR.jpgpulser_signal_ESR.jpgpulser_signal_MH.jpgdelay_ESR-MH-ESR.jpgcable_bundle_mirror_chamber.jpg
The additional signals from PMTs and XUV detector needed for spectroscopic analysis in the second DAQ are patched to AP Messhütte (MH) via the red patch
panel at the inside of the target region. The patch channels F1 to F8 (also labeled A to H in red on white) are used and extended by a cable bundle to
reach below the mirror chamber. Inside the MH the signals arrive at the red patch panel and are forwarded to the blue patch panel close to the laser DAQ
Entry  Wed Mar 16 08:31:52 2022, Kristian, DAQ, Check cables and timing 
Konstantin and I generated a signal in the Messhütte and sent it to the ESR. 
There, we used a 44m BNC cable that was sufficently long to reach all detectors. 
Our signal was connected instead of the PMT/PD. Hence, it was a roundtrip measurement.
Entry  Fri Mar 18 16:51:27 2022, Carsten, DAQ, Main DAQ /MBS error 
MBS error "TDC data ready wait" solved. Common stop trigger (20 kHz) was missing. (according to Sebastian)
Entry  Mon Mar 21 10:46:18 2022, Rodolfo, DAQ, TDC & Scaler description channels Screenshot_TLdaInputDescription_21-03-2022.JPG
For the First Vuprom we have the following channel configuration (Stand:21.03.2021). 

Please note 1.
Entry  Mon Mar 21 16:33:23 2022, , DAQ, Changing code of go4 
We agreed to indicate changes in the go4 analysis code according to the
following scheme
    Reply  Mon Mar 21 17:03:47 2022, , DAQ, Changing code of go4 
Correction: Comments in C/C++ start with // , i.e. forward slash

  [B]//--2022-mm-dd HH:MM comment on the respective change[/B].
Entry  Mon Mar 21 21:48:46 2022, Jan, DAQ, MesyTec Shaper Control 
The MesyTec Shaper & Constant Fraction (MSCF) is used to process the PMT/det signals for the ADC in the second DAQ.
It can be setup on its front panel, but single channel setup is very tedious.
Remote control of the settings is possible using an epics server and client.
Entry  Thu Mar 24 13:27:47 2022, Rodolfo, DAQ, Disk space - LXG1297 DiskSpace_lxg1297.JPG