ID Date Author Categorydown Subject
  100   Tue Apr 5 15:25:47 2022 Sebastian, Kristian, Pierre-MichelDAQGo4 Conditions for "TDC_stacked_Sig/Ref" histogram
First version of TDC_stacked histograms implemented.

Under folder "Conditions" you can now integrate over user defined Limits (Xmin and Xmax).
  105   Wed Apr 6 08:06:02 2022 DanyalDAQstopped NTCAP data acquisition
I stopped the NTCAP data acquisition at 8 AM.

Let's hope that the file contains good data.
  108   Wed Apr 6 13:32:57 2022 Max, SebastianDAQLabView Cobra Channel List
channel 0: linux time (epoch time)
channel 1: ms timer
channel 2: E-Cooler Voltage(Ohmlabs)
channel 3: Cobra Set Wavelength
channel 4: WS7 wavelength
channel 5: WS7 frequency
channel 6: WS7 status
channel 7: Status keysight (?)
channel 8: WS7 pressure (?)
channel 9: CobraMotorPosUpperPart
channel 10: CobraMotorPosLowerPart
channel 11: CobraNumberOfLoop
channel 12: StatusCobra
channel 13: WS7 temperature (?)
channel 14: CobraFromWavelength (higher wavelength)
channel 15: CobraToWavelength (lower wavelength) 
channel 16: CobraIncrement
channel 17: CobraStepNr
channel 18: status?
channel 19: PTB divider (not connected to cooler)
channel 20: ?



 
  119   Thu Apr 7 00:36:36 2022 UweDAQThreshholds and shaping Mesytec for ADC
  122   Thu Apr 7 03:06:15 2022 PatrickDAQDAQ file sizes
The file sizes of DAQ1 (229Thor) and DAQ2 (229Th, "Carstens DAQ") are different.
During 20min of laser scanning DAQ1 saves 4 files while DAQ2 saves 2 files which was 3.3 GB and 2 GB, respecively.
The file sizes of DAQ2 should always be greater than those of DAQ1, because of extra ADC data!

The problem seems to be solved after the recabling ID: 125.
  123   Thu Apr 7 03:09:36 2022 CarstenDAQNTCAP running since 02:40
  124   Thu Apr 7 03:44:21 2022 PatrickDAQMeasurement procedure
Measurement procedure


- The laser hits the screen 'NO' (atppc017) right side of the burn spot
- Be ready to start a laser scan
- Scan laser goes into wavemeter!
- Have mbs connected
- Have go4 connected

Preparation done

- 'open file (...)' for DAQ1 & DAQ2
- request a new beam
- check scraping
- 'sta ac' for DAQ1 & DAQ2
- start laser scan

laser scan started

- 'sh ac' for DAQ1 & DAQ2
- check that acquisition is RUNNING and file is OPENED
- fill in table information in the ELOG.

laser scan complete

- 'sto ac' for DAQ1 & DAQ2
- 'close file'
- check if preparation is still fulfilled
- repeat from Preparation done
  125   Thu Apr 7 04:10:18 2022 Ken UeberholzDAQCorrected cabling DAQ 2
A lot of cables were wrongly or not 
connected to the DAQ 2. 

Channels are now for both DAQ:
0 XUV-Back
1 XUV-Anode
2 XUV-Anode-delayed
3 PMT-South
4 PMT-Middle
5 PMT-North
6 PMT-Cu
7 Part-det-south
8 Part-det-north
9 1kHz
10 -
11 -
12 Delay-common-stop
13 PD-NO
14 PD-SO
15 Q-switch

All data taken by the DAQ2 has to be 
considered as faulty up to now. 

Carsten was furious.
  127   Thu Apr 7 05:23:50 2022 PatrickDAQHow To Start MBS for DAQ2 (Carstens DAQ)
(MBS server should be running at the LXG0155 machine. If not you will get an error message. Then start the server - last command in console)

Just enter the following commands in the console of any PC

ssh stoe_exp@r4l-68
*** Password on a sheet of paper ***
resl
cd /esr/usr/stoe_exp/E142Brandau/SCA... (auto complete with tab)
mbs
@startup
connect rfio lxg0155 -disk
open file lxg0155:/data.local2/E142/229Th -auto -rfio
sta ac




To stop MBS enter the following commands

sto ac
close file
disconnect rfio
@shutdown
exit
resl
  128   Thu Apr 7 05:44:40 2022 Pierre-Michel, RaganDAQNTCAP settings
Carrier Frequency: 244M
IQ Rate: 10M
reference Level: -30
RF - Attenuation: 40
IF - Attenuation: 19.9288

Masterpath p:\E142\sc
Masterpath p:\E142\IQ

Display: No Display

NTCAP data-taking is stopped at 5:44.

IQ_2022-04-07_02-40-05
SC_2022-04-07_02-40-05
  130   Thu Apr 7 07:40:09 2022 Danyal, Max, SebastianDAQNo beam and DAQs crashed
since shortly after 7 AM we have no beam in the ESR
(reason unclear) 

Both DAQs also crashed. 

Beam back in ESR at 7:20 or so 

Problems with restarting DAQ1 

Problems with restarting DAQ2 

Files names 7801 and 7804 and 0047 and 0050 cannot be used
  133   Thu Apr 7 09:23:18 2022 SebastianDAQDAQ1 & DAQ2 - 1kHz and delayed cs removed
Questionable 1kHz and delayed common stop signals has been removed.

These two cablings were for some previous tests.
  140   Thu Apr 7 15:18:25 2022 Danyal & SebastianDAQDAQ channels
We checked again the cables for both DAQs 
(on Carsten's kind request) 


Channels for both DAQs are: 

0 XUV-Back 
1 XUV-Anode 
2 XUV-Anode-delayed 
3 PMT-South (new PMT, new threshold)
4 PMT-Middle 
5 PMT-North 
6 PMT-Cu 
7 Part-det-south 
8 Part-det-north 
9 - 
10 - 
11 - 
12 - 
13 PD-NO
14 PD-SO
15 Q-switch 
  146   Fri Apr 8 01:18:51 2022 Ken UeberholzDAQTODO: DAQ improvements
For somebody who knows the DAQ1 code: Go4 resets the histograms with each laser step. Can there be additional histograms where the data gets accumulated?

And possibly a histogram where important detectors are plotted against the laser steps? The read back wavelength of the wavemeter is not very precise. 
  147   Fri Apr 8 01:48:57 2022 PatrickDAQBugged Trending plots?
The wavelengths/frequencies shown in the graph in the labview window do not match the wavelengths/frequencies in go4.
Compare the picture which shows an incomplete laser scan although the scans in the labview window are completed (The measurement was started at an 'elapsed MBS time' of 2700s and took 1700s/28min).
However, the displayed time is correct. Only the data seems to be "stretched".

The same is true for the 'XUV anode' detector. The counter in the electronic module is ahead of the displayed count rate in go4.


Additional Note: The 'Cobra set wavelength' plot shows wavelengths which seem to be measured. However, we would expect deterministic setpoints from the software, see picture.
  153   Fri Apr 8 08:40:06 2022 SebastianDAQDAQ improvement
Concerning ID 146 a new histogram for accumulating data (no cleaning histograms) can be found under

Step2_Anl/TDC/TDC***_accumulated_***
  162   Fri Apr 8 14:12:57 2022 morning shiftDAQtiming for h=1
The timing for h=1 has been set up 

The time window is now twice as large as before! 
This means that we see the same ion bunch twice. 

(150 bins = 1 turn in the ESR) 

The PMT signal is exactly in between two cycles + the extra flight time (3 bins = 10 ns) from target to PMT north. 
  163   Fri Apr 8 14:14:28 2022 KristianDAQDelay of bunch RF
The bunch HF is now fed through a delay to shift the stop trigger relative to the PMT signal. The delay generator is placed directly after the CFG and before splitting the signal for the two DAQs. The relative timing of both DAQs was not changed.
  164   Fri Apr 8 14:43:50 2022 WilfriedDAQlaser timing - Q-switch start
below are my notes concerning laser timing / Q-switch start
  170   Fri Apr 8 17:57:45 2022 KonstantinDAQLaser scanning


TimeSIS-intensity cooler-HV LMD-DAQ1 (R4L-68) LMD-DAQ2 (R4L-41) scan range (nm) step width(nm)dwell time(s) comments
17:28 1.1e9 - 0083 7822 322.3-321.3 0.01 4 Wrong scan range
17:48 1.4e9 209618 0084 7823 322.3-321.3 0.01 4 Wrong scan range
18:10 1.6e9 - 0085 7824 323.2-321.2 0.01 4 acquisition on r4l-68 crashed
18:13 2e9 - 0086 - 323.2-321.2 0.01 4 acquisition on r4l-41 crashed
18:29 1.9e9 209618 0087 7826 323.2-321.2 0.01 4 An artifact is observed at about 644nm <-> 322nm fundamental wavelength
20:17 2e9 209609 0089 7827 323.2-321.2 0.01 4 Cu-mirror and XUV detector are not driven at the moment. It is unclear for how long this was the case. Optimization of the accelerator settings were done before.
21:12 1.6e9 - 0089 7828 323.2-321.2 0.01 4 Laser got wrong pulse widths during this scan and also during the last one.



At about 20:00 the slits were adjusted to the h=1 settings:

NI: 5mm
NA: -3mm
SA: -10mm (also the voltage for the Scitnilator in this pocket was increased to 1500V)
SA_Scraper: +14mm