  Dielectronic recombination-assisted laser spectroscopy, Page 3 of 6  Not logged in ELOG logo
ID Date Author Category Subject
  42   Wed May 11 08:01:44 2022 ZoranRunsStart laser scanning

Quote:

Questions:
There are many other masses/isotopes/etc. visible in the Shottky spectrum. Is this wanted?
Are the vacuum gauges turned off?
Laser timing correct? +- 3 bins?


No, this is not wanted and yes, the machine is aware of the problem. The decision on Tuesday evening was to leave the ESR in that state in order to start the measurement and continue setting on Wednesday. Otherwise the night shift could not take place.

The additional isotopes are only a few percent and should contribute insignificantly to the background rate.
  43   Wed May 11 14:48:31 2022 KonstantinDAQVoltage range of the Keysight DMM changed from 10V to 1V
During the setup of the ion beam the voltage range of the Keysight DMM was changed from 10V to 1V. Additionally, the name of the high-voltage divider was adjusted (this changes only the filename of the file, which is stored at the local pc decpc003).
Attachment 1: VoltageRangeChanged.png
VoltageRangeChanged.png
  44   Wed May 11 14:59:01 2022 RodolfoLaserLaser status
I have measured the output energy of the laser system. All values in [mJ] and over 500 laser-pulses average

DateOut of Quanta Ray @ 532 nmSeed energy for Cobra @ 532 nmOut of Cobra @ 601 nm
10.05.202253348571
11.05.202248448067
  45   Wed May 11 18:00:40 2022 Wilfried, Phillip, MaxDetectorsdouble PD NO peak problem solved
In entry 36 one finds an additional PD NO peak around bin 100.
This was caused by signal reflection in the cable.
The problem was solved by adjusting the CFD threshold from 10 to 30.
 
  46   Wed May 11 18:13:04 2022 Wilfried, Phillip, MaxDAQlaser scan/step complete
The scaler channel laser scan complete and laser step complete is interchanged in go4. Carsten will fix this on the software side.
Additionally the laser scan complete signal was not there in the first place. After checking the signal on multiple points, the signal was suddenly back in the DAQ.
Possibly there is a bad cable/contact somewhere (on the Labjack?). The signal should be checked in go4 from time to time. 
  47   Wed May 11 20:47:54 2022 ZoranAcceleratorbeam lost in SIS
For some reason the beam was lost in the SIS soon after taking the ESR beam this evening, reason unknown. Solved by sending teh full context to the SIS machine and having 
operators to manually set the voltage of the SIS extraction kicker. Afterwards a slight readjustment of the transfer beamline was necessary:

In the SIS: GTE5 KY1, MU0, QD21, QD22 as well as in the ESR: GE01MU4I (5 -> 6 mrad)

This brought back around 1E7 particle in the ESR after 5 accumulated SIS injections.
  48   Wed May 11 22:04:30 2022 Phillip, Wilfried, MaxGeneralspacial laser - ion overlap verified
After the SIS/ESR optimization we checked the spatial overlap with the T-scrapers again.

The values are:
devicemove in dir.scratch (mm)gone (mm)
GECEDS1VUup -3.0 -2.2
GECEDS1HAright -22.9 -22.0
left -11.0 -13.0
GECEDS2VUup -4.5 -2.7
GECEDS2HAright -21.0 -19.6
left -7.8 -8.4


The beam position is approx. as before. The laser position is still good and we have spatial overlap.
  49   Wed May 11 22:19:17 2022 Max, Wilfried, VolkerLaserlaser bunch timing
We overlapped the laser and ion bunch in the time domain. The delay has to be set on the module above the laser and NOT in the ESR Messhütte!

The values are:

PD NO: 63
PD SO: 98.5
Ion bunch: 150 
Attachment 1: timing10_5_22.png
timing10_5_22.png
  50   Wed May 11 22:45:22 2022 Max, Wilfried, PatrickGeneralscanning wavelength
The scan parameters for the next shift are:

in COBRACONTROL:

increment: 0,005nm

from position: 601,62nm

to position: 601,405nm

These are the wavelengths in air!


The wavelength in vacuum we expect is: 601.832nm

You can check the current wavelength in vacuum on the wavemeter in the LabView GUI.

edit: In the night shift, we checked the scanned vacuum wavelengths in the wavemeter and also checked the expected transition wavelength.
The resonance should be very much in the center of the above mentioned scan range.
The increment corresponds to 4.1 GHz steps. Is this small enough?
  51   Wed May 11 23:34:39 2022 Patrick, CarstenDetectorsParticle Detectors / Scraper positions
Voltages of particle detectors off.

Scraper positions:

GE02DS3HG: -8
GE01DD2AG: -25
GE01DD2IG: 6
GE02DD2_G: -30
  52   Wed May 11 23:36:24 2022 Ruijiu Chen, Carsten, GuyDAQstart NTCAP.
Carrier frequency:244MHz
IQ rate: 10M Samples/seconds
Reference level: -25
RF-Attenuation(dB):40
IF-Attenuation(dB):19.9288
Samples/Record:2^18=262144
Records/File:2^12=4096
Masterpath:p:\208Bi\iq\IQ_2022-05-10_17-06-29
           p:\208Bi\sc\IQ_2022-05-10_17-06-29
Free space of hard disk: 34.9TB.
  53   Thu May 12 00:16:13 2022 Patrick, Volker, Guy, Rui-JiuRunsRuns night shift

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
00:15 ~1.0e8 1.0e7 223219 433 - 435 8094 601.62 - 601.405 0.005 5s 4
00:47 3.7e8 2.0e7 223219 436 - 440 8095 - 8096 601.62 - 601.405 0.005 5s 8 increased to 15 injections, still not saturated
Tried 20 injections, got no injection into ESR, so moved back to 15. The scrapers were not moving for quiet some time, even after fixing them. The problem may have been that two ParamModi windows were opened. The injection into the SIS does not work when there are other users performing special operations with the machines.
02:28 1.0e8 1.0e7 223220 441 - 444 8097 601.62 - 601.405 0.005 5s 6
02:54 2.8e8 1.7e7 223219 445 - 449 8098 - 8099 601.62 - 601.405 0.005 5s 6
03:21 ~3e8 ~2e7 223219 DAQ did not write 8100 - 8101 601.62 - 601.405 0.005 5s 6
The vacuum gauge of the e-cooler failed. It was restarted. Apparently, upon changing/fixing anything related to our accelerator pattern, the scrapers forget to move. Then they have to be disabled and enabled again in ParamModi. Also, the injection does not work very well. Either the transition from SIS to ESR fails or we are just skipped completely. After fixing the injection, again the scrapers failed as expected. We keep the scrapers out now.
04:34 2.8e8 1.8e7 223219 DAQ failed, started late 454 - 458 8102 - 8103 601.62 - 601.405 0.005 5s 6 Set max.counts in PMT middle and north to 15 (In file 'SetConditions.C', applied when starting go4). No scrapers used.
05:02 2.5e8 1.7e7 223219 DAQ late, 461 - 464 8104 - 8105 601.62 - 601.405 0.005 5s 6
05:30 2.5e8 1.8e7 223219 465 - 469 8106 - 8107 601.62 - 601.405 0.005 5s 6 Set max.counts in PMT middle and north to 20.
05:57 2.5e8 1.7e7 223219 470 - 474 8108 - 8109 601.62 - 601.405 0.005 5s 6
06:25 2.6e8 1.7e7 223219 475 - 479 8110 - 8111 601.62 - 601.405 0.005 5s 6
06:50 2.4e8 1.6e7 223219 480 - 484 8112 - 8113 601.62 - 601.405 0.005 5s 6
07:24 2.2e8 1.5e7 223219 485 - 489 8114 - 8115 601.62 - 601.405 0.005 5s 6
07:49 - 6.5e6 223219 490 - 493 8116 601.62 - 601.405 0.005 5s 6 UNILAC tunnel will be open for ~1h -> no new beam from SIS available
08:15 - 2.6e6 223219 494 - 496 8117 601.62 - 601.405 0.005 5s 6 UNILAC tunnel will be open for ~1h -> no new beam from SIS available
  54   Thu May 12 03:25:37 2022 PatrickDAQLaser scan complete signal off
As in ID:46, the 'laser scan complete' again is not coming in.
It is not very useful to repeatedly fix cable contact problems.
Therefore, Carsten is adding a 'number of steps' line to the config file to "repeat" the x-axis.
  55   Thu May 12 10:10:35 2022 Simon, Christopher, Wilfried, SebastianDAQ1ms gated PMT north / middle added to DAQ
We added on DAQ1 and DAQ2 1ms gated PMT North and PMT Middle signals.
Now we "supress" the signal by a factor of ~30.

DAQ1:
Channel 11 --> 1ms gated PMT North
Channel 12 --> 1ms gated PMT Middle

DAQ2:
Channel 12 --> 1ms gated PMT North
Channel 13 --> 1ms gated PMT Middle

Update: the same gate has been added to the XUV anode and anode delayed signals, but on their original TDC channels not on new ones, in order to not only reduce the background in the XUV signals but also reduce the load on the DAQ.
  56   Thu May 12 10:21:30 2022 Ragan, Shahab, Simon, Sebastian, Christopher, WilfriedGeneralNo Beam
There is no beam in the ESR due to some problem in the UNILAC.
Attachment 1: Screenshot_from_2022-05-12_09-21-38.png
Screenshot_from_2022-05-12_09-21-38.png
  57   Thu May 12 11:30:18 2022 Ragan, Shahab, Simon, Sebastian, Christopher, WilfriedGeneralBeam is back
Beam is back in the ESR.
Attachment 1: Screenshot_from_2022-05-12_10-32-24.png
Screenshot_from_2022-05-12_10-32-24.png
  58   Thu May 12 11:31:11 2022 Simon, Christopher, Shahab, Wilfried, Sebastian, RaganGeneralRuns morning shift

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
11:30 - 8,2e6 223219 497 - 500 8119 - 8119 601.62 - 601.405 0.005 5s 6 after this run we had to give the ESR Group the beam
  59   Thu May 12 12:02:56 2022 Ragan, Shahab, Simon, Sebastian, Christopher, WilfriedGeneralNo beam
There is no beam in the ESR again.
Attachment 1: Screenshot_from_2022-05-12_11-02-37.png
Screenshot_from_2022-05-12_11-02-37.png
  60   Thu May 12 12:51:30 2022 Ragan, Shahab, Simon, Sebastian, Christopher, WilfriedGeneralBeam is back
12:50 Beam is back again in the ring.
12:52-until now Kicker problem in the ESR.
Attachment 1: Screenshot_from_2022-05-12_11-51-35.png
Screenshot_from_2022-05-12_11-51-35.png
Attachment 2: Screenshot_from_2022-05-12_12-08-52.png
Screenshot_from_2022-05-12_12-08-52.png
  61   Thu May 12 14:04:09 2022 KonstantinDAQGo4 Laserstep histograms adapted
In order to correctly display the gated PMT signals PMT Middle and PMT North some modifications in the Go4 code were necessary. Affected files are TLdaAnlProc.cxx, TLdaAnlProc.h and TLdaInputDescription.h.
Modifications are indicated according to the scheme "//-- HH:MM mm-dd-yyyy: <description>" (an example is given in the attached screenshot). They were needed to

  1. Create relevant histograms (change of variable MAX_STEP from 10 to 13 and insertion of histogram names "QSWP1MS", "NORTH_GATED" and "MIDDLE_GATED" in the std::string array stepNames.
    Both adaptions were made in TLdaInputDescription.h)

  2. Fill the respective histograms. Changed lines of files TLdaAnlProc.cxx, TLdaAnlProc.h are listed below.

filename modified lines
TLdaAnlProc.h 47
TLdaAnlProc.cxx 258-261, 381-394, 442-443, 484-494, 536-542
Attachment 1: Go4Modifications.PNG
Go4Modifications.PNG