  62   Thu May 12 15:21:23 2022 Sebastian, SimonLaserFlash lamps changed
We (Sebastian and I) changed the flash lamps of the Quanta Ray Laser and measured the output energy of the laser system afterwards (still during warm up).
All values in [mJ] and over 500 laser-pulses average

DateOut of Quanta Ray @ 532 nmSeed energy for Cobra @ 532 nmOut of Cobra @ 601 nm
10.05.202253348571
11.05.202248448067
12.05.202257850175


The lambda/2 is at 103°. Correction factor for the 532nm-monitor is 11.4 (exchanged broken monitor).
  63   Thu May 12 15:34:47 2022 Max, Kristian, SebastianDAQlaser scan, laser step and laser busy checked
We checked all three signals in the Messhütte. Laser scan and laser step are OK.
The Laser busy signal is for large increments (0.1nm) also fine. For short increments like our usual 0.005nm increment the signal is irregularly.
Likely the laser busy status is too short.
  64   Thu May 12 15:50:33 2022 MaxAcceleratorNo ESR beam
Since 13:30 we have no beam in the ESR. There is a problem with the ESR septum.
The magnets department is investigating.
  65   Thu May 12 16:35:17 2022 Max, Wilfried, ZoranAcceleratorNO BEAM UNTIL TOMORROW MORNING
The problem at the ESR septum is serious. They have to remove the broken part and try to repair it.
Maybe tomorrow in the morning briefing there will be more information about this.

The night shift will not take place!
The morning shift starts at 9:00 or on further notice!

The PMTs are switched off and the laser also.
  66   Fri May 13 08:32:41 2022 ZoranAcceleratorNO BEAM UNTIL TOMORROW MORNING
> The problem at the ESR septum is serious. They have to remove the broken part and try to repair it.
> Maybe tomorrow in the morning briefing there will be more information about this.
>
> The night shift will not take place!
> The morning shift starts at 9:00 or on further notice!
>
> The PMTs are switched off and the laser also.

Update from the morning briefing: the DCCT (power converter) of the injection septum was demounted and the EPS group is trying to fix it. There is no replacement available, so if they do not succeed this beamtime is over. They promised more information in the midday meeting at 12:45.

In the ESR Messhüte is some cake ("Trostkuchen").
  67   Fri May 13 11:46:51 2022 Sebastian, Simon Laser energy
We (Sebastian and I) measured the output energy of the laser system again.
All values in [mJ] and over 500 laser-pulses average

DateOut of Quanta Ray @ 532 nmSeed energy for Cobra @ 532 nmOut of Cobra @ 601 nm
10.05.202253348571
11.05.202248448067
12.05.202257850175
13.05.202263249878


The lambda/2 is at 100°. Correction factor for the 532nm-monitor is 8.4.
  68   Fri May 13 16:55:58 2022 Danyal, Carsten, Phillip, WilfriedGeneralEnd of E128 Beamtime
Dear all,

unfortunately, the E128 experiment must now come to an early end. 
The problem(s) with the injection septum are severe and no work around solution (with acceptable risks) could be found. 
Several things were tried, but without hope. 

Many thanks to everybody involved during the preparation time and the shifts during the beamtime. 

Best wishes,

Wilfried
  69   Mon May 16 14:20:05 2022 Ragan, Danyal, Carsten, Phillip, Wilfried et al.GeneralE128 Beamtime
> Dear all,
> 
> unfortunately, the E128 experiment must now come to an early end. 
> The problem(s) with the injection septum are severe and no work around solution (with acceptable risks) could be found. 
> Several things were tried, but without hope. 
> 
> Many thanks to everybody involved during the preparation time and the shifts during the beamtime. 
> 
> Best wishes,
> 
> Wilfried

There is beam again in the ESR. The plan is to take beam until tomorrow morning. 
  70   Mon May 16 15:57:37 2022 RodolfoLaserWavemeter calibration
I have calibrated our Wavemeter WS7 using the temperature stabilized He-Ne Laser (SIOS EN60 825-1:2014). In the screenshots below you can see the readout of the wavemeter before (vor) and after (nach) calibration.
Attachment 1: WS7_VorKalibrationMitHe-Ne_16-05-2022.PNG
WS7_VorKalibrationMitHe-Ne_16-05-2022.PNG
Attachment 2: WS7_NachKalibrationMitHe-Ne_16-05-2022.PNG
WS7_NachKalibrationMitHe-Ne_16-05-2022.PNG
  71   Mon May 16 17:15:56 2022 CarstenAnalysisDAQ /Analysis - shift of peaks in the accumulated spectra, changes in code
After a phone call with Wilfried I implemented an offset to all hit channels of the stacked / accumulated spectra where one chekcs the timing peaks.

Therfore, I made a copy of the analysis in 208Bi1 (one "1" at the end). And here, in the file TLdaAnlProc.cxx I added an offset of +70 to the hit channel (variable UInt_t hit, around line 335 /336 of the code).
Pleasse note that I just inlcuded one line with a "magic" number here, since I cannnot test teh compiling and was afraid to introduce an error.

So the lines were

UInt_t hit =tdcSE[i][j];

changed to 

UInt_t hit =tdcSE[i][j];
hit = hit + 70;

Plesaede note that you need to recompile and check and set all conditions for the time-preak window.

So , you now use the analysis in 208Bi1
  72   Mon May 16 17:38:38 2022 Max, Danyal, Wilfried, RodolfoGenerallaser bunch timing
We overlapped the laser and ion bunch in the time domain. 
Therefore we delayed the bunching HF in order to see the PMT signal between bin 10 and 22. Before the signal Peak was in channel ~145-5 and difficult to gate on.
Afterwards we also have to shift the laser.
The new values are:

PD NO: 77.5
PD SO: 113
Ion bunch: 16.5 
Attachment 1: timing1605.png
timing1605.png
  73   Mon May 16 18:01:23 2022 mbsMax, Danyal, Wilfried, Rodolfo, Patrick, Kristian, David, Ruijiu Runsruns afternoon/night shift 16.05
These are the runs after the septum was repaired on 16.05.

Time ESR-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
ca. 18:00 ? ~3.0e7 223222 504-505 8121 601.62 - 601.405 0.005 5s 4 DAQ2 late start
18:11 ? ~3.3e7 223222 506 8122 601.62 - 601.405 0.005 5s 4 aborted, ions are rapidly lost after Bahnverlagerung
18:33 4.5e7 ~3.7e7 223222 507-509 8123 601.62 - 601.405 0.005 5s 4 removed all scrapers, beam life time is better, high intensity, slight instability in bunching, ion intensity at the end: 1.4e7, DC Trafo was in saturation so ampl. was turned down
19:32 ? 3.7e7 223222 512 8125 601.62 - 601.405 0.005 5s 4 rf bunching amplitude decreased from 350V -> 280V
20:00 ? 3.6e7 223223 514 8127 601.745 - 601.32 0.01 5s 4 43 steps of laser, therefore 1 step offset from loop to loop, laser already of ?
20:30 ? 2.9e7 223223 520-522 8129 601.760 - 601.320 0.01 5s 6 aborted after 3 loops, laserstabi was off, probably during hole measurement
21:01 ? 3.6e7 223223 524-528 8130 601.760 - 601.320 0.01 5s 6
21:30 ? 3.4e7 223223 529-532 8131 601.760 - 601.320 0.01 5s 6 coasting beam now !!!
22:10 ? 3.7e7 223223 534-537 8132 601.760 - 601.320 0.01 5s 6 Night shift took over.
22:40 4.4e7 3.6e7 223223 539-542 8133 601.760 - 601.320 0.01 5s 6
23:52 3.8e7 3.2e7 223223 545-548 8134 - 8135 601.760 - 601.320 0.01 5s 6 Switched off the nitrogen before this run, stopped acq. after 4 loops (laser gone), PMT north saw a lot of noise? (ID:78) -> larger files.
00:39 4.3e7 3.2e7 223222 552-555 8136 601.760 - 601.320 0.01 5s 6 The secondary DAC always requires 2-3 restarts until it is working.
01:10 3.8e7 3.0e7 223222 557-558 8137 601.760 - 601.320 0.01 5s 6 Laser moved away during the third loop, acq. stopped.
01:33 4.2e7 3.1e7 223222 561-564 8139 601.760 - 601.320 0.01 5s 6 The previous laser scan was still running, restarted the acq. once.
02:00 > 4.0e7 3.4e7 223222 566-569 8140 601.760 - 601.320 0.01 5s 6
02:30 > 4.0e7 3.2e7 223222 573-576 8141 601.760 - 601.320 0.01 5s 6
02:58 4.2e7 3.4e7 223222 579-582 8142 601.760 - 601.320 0.01 5s 6
03:45 4.2e7 3.3e7 223222 585-589 8143 601.760 - 601.320 0.01 5s 6 Back to bunched beam mode
04:15 4.7e7 3.4e7 223222 594-598 8144 601.760 - 601.320 0.01 5s 6
04:43 4.4e7 3.4e7 223222 600-604 8145 601.960 - 601.520 0.01 5s 6 We changed the scan range Big grin
05:10 4.3e7 3.3e7 223221 606-610 8146 601.960 - 601.520 0.01 5s 6 It looks like the "overWavel" plots actually work better now, because the laser steps somehow are counted wrongly, leading to shifts between loops.
05:37 4.2e7 3.7e7 223221 611-615 8147 601.960 - 601.520 0.01 5s 6
06:27 ? ? 223219 616-618 8148 601.960 - 601.520 0.01 5s 6 Since Go4 was still running during startup we were not able to look into the data during the scan. After reduction of particle number, the dc trafo is no longer capable to measure to number of ions.
06:56 ? ? 223219 619-620 8149 601.960 - 601.520 0.01 5s 6
07:25 ? ? 223219 621-622 8150 601.9 - 601.7 0.01 5s 6 Changed the scan range t0 601.9-601.7 and reduced step increment to 0,005nm
07:56 ? ? 223219 623-624 8151 601.9 - 601.7 0.01 5s 6
08:20 ? ? 223219 625-625 8152 601.9 - 601.7 0.01 5s 6 Restarted scan without new injection, DAQ2 was not running during this scan
08:44 ? ? 223220 626-627 8153 601.9 - 601.7 0.01 5s 6 Reduced Loop number to 4 (aborted after 4 completed scans)
09:03 ? ? 223220 628 8154 601.9 - 601.7 0.01 5s 6 Reduced Loop number to 3 (before starting a new scan)
09:17 ? ? 223221 629 8155 601.9 - 601.7 0.01 5s 6
Since
  74   Mon May 16 19:20:55 2022 Max, WilfriedAnalysisstart go4 before laser scan
It is necessary to start go4 before you start the laser scan. Otherwise you get an offset in the laser Step plots.
  75   Mon May 16 21:08:48 2022 Rodolfo & MaxLaserLaser power
We checked the energy of the laser and optimized it a bit

TimeSeed energy for Cobra @ 532 nmOut of Cobra @ 601 nm
20:4141563
20:4850673


The lambda/2 is now at 108°.
  76   Mon May 16 22:19:54 2022 PatrickDAQKeep go4 running (in theory), MBS
To accumulate more data into the opened go4, do not restart MBS, but just

sto ac
clo fi
open file lxg1297:/data.local2/208Bi/208Bi -auto -rfio
sta ac

Note: This does not work, because go4 somehow accumulates ~5 extra laser steps when starting a new scan which results in an x-axis-offset between the runs.
  77   Mon May 16 22:57:40 2022 Patrick, Kristian, David, Ruijiu DetectorsSwitched off nitrogen
Background counts with nitrogen and air.
Less multiphoton events with air (see pictures, histogram extended to 30 with nitrogen).


Read in 4_Lasersteps/Counts

nitrogen:
XUV: 15e4
Middle: 8e4
North: 9e4

air:
XUV: 13e4
Middle: 1.2e4
North: 0.7e4
Attachment 1: 2022-05-16_PMTs_after_nitrogen_off.png
2022-05-16_PMTs_after_nitrogen_off.png
  78   Tue May 17 00:44:01 2022 Patrick, KristianDAQChanged PMT North CFD channel
In run 8134 - 8135 PMT North showed a super high background rate. After the scan we checked the signal cables.

Incoming signal was still good until CFD. We changed channel from 2 to 12. New threshold is 36, was 10.
  79   Tue May 17 06:37:28 2022 YL, KMAcceleratorReduced beam intensity
The resonance in 209Bi has been seen during the night.
Reducing the intensity by a factor~1000, to simulate 208Bi conditions
Running low intensity since ~6:15-6:20
The present intensity is so low that DCCT does not see the beam.
On Schottky the beam is clearly seen as ~30-40 dB over the background.
The real intensity is to be determined from the NTCap, which is running (checked).
  80   Tue May 17 10:38:44 2022 KonstantinDetectorsPMT North changed
PMT located at the northern window was again replaced by the narrow-band PMT (ET Enterprise type 9423 SB, SN: 650).
  81   Fri May 27 10:39:53 2022 KristianDAQPMT North noise
During the last run, PMT North showed some weird noise.

The PMT was replaced by the deep UV PMT with small aceptance range (100-200nm) but there was still a lot of noise on the output channel in the Messhütte, see attached photo. The intensity of the noise signal was as strong as a normal signal count.

I replaced all cables and used other preamplifier channels. The noise is gone and the baseline is flat now. However, the timing of PMT north will be a bit different now, so we should use PMT middle for the laser/ion overlap as we did before.

Graziano opened the valves for the laser and turned off the vacuum gauges and emergency lights. The background of the UV PMT looks normal (Middle 10 Hz, North 1 Hz) so I think we are fine here.