ID Date Author Category Subject
  42   Mon May 12 10:09:21 2025 Christopher, NilsAcceleratorIntervention TK Chopper - no beam for one hour
At 11am there will be an intervention for about one hour related to the TK chopper - hence no beam for SIS as well as ESR.
During that time, the pneumatic drives within ESR will be mechanically checked (lubricated).

The ESR-colleagues re-checked the charge states in the ring. 
  43   Mon May 12 14:47:52 2025 ChristopherDAQfrequency counter
FA-2 frequency counter
Attachment 1: BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf
BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf BG7TBL_FA-2_Frequency_Counter_Operator_Manual-62519_English_Ed.pdf
  44   Mon May 12 21:39:49 2025 ZoranRunsDR running again with Bi209
Stronger spectra than before, probably due to a stronger overlap with the electron beam.
Open files starting with run_0022, finished with 0025, each one about 6 min charge breeding and 11 min ramp
ntcap stopped, electron current 500 mA, DC beam
  45   Mon May 12 22:44:00 2025 ZoranRuns209Bi bunched beam
ntcap restarted (05-12_22-22-23), measuring with bunched beam
10 min breeding time
  46   Mon May 12 23:34:39 2025 CarstenAnalysisSample DR Spectra
See attachment - electron current 500 mA.
First spectrum DC,
Second spectrum bunched.
Resolution is slightly worse for bunched beam but not dramatically.
Please note, that the ratio DR/RR is 1:1. 
Spectra are about 4-5 injection cycles each. (and due to the shorter lifetime with 8 scans instead of 15 scans).
ion beam intensity : about 2e7 Li-like
Other settings as before.

-> Resonance is now at the correct energy!
Attachment 1: DR_209Bi_500mA_newESR_settings.png
DR_209Bi_500mA_newESR_settings.png
Attachment 2: DR_209Bi_500mA_newESR_settings_bunched.png
DR_209Bi_500mA_newESR_settings_bunched.png
  47   Tue May 13 00:05:39 2025 KonstantinCalibrationCooler HV
The DR measurement was repeated with proper electron cooler settings. Readout of electron cooler HV is not affected by the Grafana failure and can be displayed in Medusa in the View "Log-Display". Measured voltages in Subchain 8 are:

  • 500mA cooler current, file 25-26 (CR DAQ), NTCAP folder closed
    Ohmlabs 205854
    PTB 3.2 205861
    PTB 3.3 205862


  • 250mA cooler current, file no.27 (CR DAQ), NTCAP folder "IQ_2025-05-13_00-18-48"
    Ohmlabs 205812
    PTB 3.2 205819
    PTB 3.3 205819


The cooling energy is 900V lower, since it is compensated by the drift tube HV.
  48   Tue May 13 00:36:27 2025 Finn, Simon, Konstantin DR measurements. Night shift.
Time File PTB 32 PTB 33 Ohmlabs ESR intensity SC8 Typical SIS injections comments
00:18 27 205819 205819 205812 2.24e7 4e8
01:11 27 205819 205820 205812 2.45e7 5e8
01:40 27 205818 205820 205812 2.89e7 9e8 ion source cathode changed
02:06 27 205819 205820 205812 3.06e7 9e8
02:33 28 205819 205820 205812 2.77e7 9e8 opened new file
03:00 28 205819 205820 205812 2.77e7 9e8 detector went too far in and killed the beam
03:26 29 205819 205820 205813 3.03e7 9e8 opened new file
? 29 205819 205820 205813 2.84e7 7e8
04:21 29 205819 205820 205813 2.65e7 9e8
04:45 29 205819 205820 205813 2.61e7 8e8
05:08 30 205820 205821 205813 2.63e7 9e8 opened new file
05:31 30 205820 205821 205813 2.63e7 8e8 detector went too far in and killed the beam
05:55 31 205819 205820 205813 2.56e7 7e8 opened new file
  49   Tue May 13 00:43:31 2025 KonstantinGeneralConnection interrupt
Connection reset to all NUCs by accident manual power cut.
  50   Tue May 13 01:21:16 2025 Konstantin, SimonAnalysisLongitudinal bunch length
We measured the longitudinal bunch length for all different detectors

PMT north PMT mid particle detector
width (bins)2.912.742.45
width (ns)9.608.988.09
Attachment 1: Bildschirmfoto_2025-05-13_01-24-40.png
Bildschirmfoto_2025-05-13_01-24-40.png
Attachment 2: Bildschirmfoto_2025-05-13_01-33-06.png
Bildschirmfoto_2025-05-13_01-33-06.png
Attachment 3: Bildschirmfoto_2025-05-13_01-38-19.png
Bildschirmfoto_2025-05-13_01-38-19.png
  51   Tue May 13 12:59:49 2025 Carsten, Wilfried, Christopher, NilsGeneralSetup of ESR for 208Bi
In the morning the colleagues from ESR setup the beam for production of 208Bi81+, however the identification (via the schottky) of the correct fragments and charge states is still ongoing.

Issue with the pneumatic drives (scraper/detector) is still not solved yet, however we want to give it a try with the current limitations unless it becomes absolutely necessary.
  52   Tue May 13 15:34:23 2025 YuriCalibrationParticle Identification
Setting on 208Bi81+
Manipulation after accumulation and orbit shift, before slowing down

f(MHz) - A/Q <<< comment
243.031 - 209/80 <<< primary beam charge state
<<< missing @243.068 - 201/77 
<<< missing @243.111 - 206/79 
243.191 - 203/78
243.231 - 208Bi80+ ******
243.274 - 200/77
243.314 - 205/79
<<< missing @243.360 - 197/76 <<< probably under the bump of uncooled beam <<< has been seen in another manipulation
<<< missing @243.396 - 202/78 <<< probably under the bump of uncooled beam <<< has been seen in another manipulation
<<< missing @243.431 - 207/80 <<< probably under the bump of uncooled beam <<< has been seen in another manipulation
<<< missing @243.482 - 199/77 <<< probably under the bump of uncooled beam
243.516 - 204/79 <<< a_pi=-0.17
243.549 - 209/81 <<< primary beam charge state
<<< missing @243.604- 201/78 
243.634 - 206/80 <<< a_pi=-0.168
243.694 - 198/77
243.721 - 203/79 <<< a_pi=-0.170
243.746 - 208Bi81+ ***** 
243.784 - 195/76
243.811 - 200/78 <<< a_pi=-0.172
243.836 - 205/80
243.879 - 192/75
243.904 - 197/77
243.926 - 202/79
243.946 - 207/81 << a_po=-0.170
  53   Tue May 13 17:46:35 2025 CarstenRunsStart of DR measurements with 208Bi
Same ramps as before, 8 repetitions.
ESR is set-up for breeding of 208Bi80+.
In order to reduce unwanted primary at injection to the ESR, in the transfer line scrapers are in use, presently GTE4DS5HL -5.0, GTE4DS5HR +7.0

The 208Bi80+ can be easily scraped since most unwanted species are lost during the deceleration ramp, and only 3 peaks remain. These 3 peaks are quite far away from the 208Bi80+.
Please note, that frequencies at decelerated beam is different than the ones in Yury's ID.

The detector DD1AG is not moving.

For scraping essentially DS3HG and DD1IG are used plus the particle detector.
DS3HG also shows the "overshooting behavior of DD2_G (particle detector, but not so strong).
Scraping of the 208Bi80+ with DS3HG at 0 but at -35 alreadz all other peaks are gone, thus the setting for the scraper DD3_DS3HG is -35.
With the DD1IG we cannot touch the 208Bi80+. Set to 26.
Particle detector presently is at -65 but this might be too far out.

I_el is 500mA, NOT bunched for the first tests.
stopped and restarted NTCAP: IQ_2025-05-13_17-39-30
run032.lmd (detector position not very good).
We had about 1e6 ions at the low energy.
First DR spectrum of 208Bi80+ but with moderate statistics. Going to different detector position as cross check.
18:05 closed run0032.lmd

Changed detector position to from -65 to -55.
Opened file from run0033.lmd
(At the moment 10 injections, 60 x 10s waiting - 10 min)

Intensity is about 1e6 ions at the beginning of the measurement.
After a few scans we see the first DR spectrum of 208Bi80+

Thus, we are sure to have the right beam.
Resolution is good.

The attached (online) spectrum is about 4 or 5 fully cycles (>1 h).

Please keep in mind that these are "small" DR resonances.

The lower spectrum is about 60% of the true counts during this time.
Attachment 1: DR_208Bi_500mA_unbunched(DC).png
DR_208Bi_500mA_unbunched(DC).png
Attachment 2: DR_208Bi_500mA_unbunched(DC)_true_counts.png
DR_208Bi_500mA_unbunched(DC)_true_counts.png
  54   Tue May 13 21:37:09 2025 CarstenRunsDR run with bunched beam / 500 mA /208Bi80+
Just changed settings towards a new scanning range and bunched beam.
Stopped and started NTCAP, now: IQ_2025-05-13_21-01-44
new ADC file
NEW ENERGY RANGE AND TIMING
Scan from 21 eV to 11 eV with 1001 steps and Delta E = 0.01 eV
8 repetitions
SubTriggerTiming:
5800
10200
10400
10600
200
200
200

500 mA / bunched beam; other settings as before

run0035.lmd opened

Please note that the peak now looks differently, in particular in looks broader (dE = 0.01ev instead of dE =0.02 eV))and is at a different position
  55   Wed May 14 00:11:52 2025 Dimitrios, SimonRunsNight shift DR Bi-208
Time File PTB 32 PTB 33 Ohmlabs ESR intensity SC8 Typical SIS injections comments
00:15 36 205860 205861 205853 0 8e8 something killed the beam after deceleration, particle detectors look fine
00:42 37 205860 205861 205853 5e5 7e8 opened new file
01:09 37 205860 205861 205853 5e5 7e8
01:38 37 205860 205861 205853 5e5 8e8
02:04 37 205861 205862 205854 4e5 5e8
02:31 38 205861 205862 205854 4e5 6e8 opened new file
02:56 38 205861 205862 205854 4e5 6e8
03:23 38 205861 205862 205854 4e5 6e8
03:51 38 205860 205861 205854 6e5 5e8
04:15 39 205860 205861 205854 5e5 5e8 opened new file
04:40 39 205860 205861 205854 4e5 5e8
05:08 39 205860 205861 205854 4e5 5e8
05:34 39 205860 205861 205854 3e5 5e8
06:00 40 205860 205861 205854 4e5 4e8 opened new file
06:29 40 205860 205861 205854 0 4e8 particle detector went too far in and killed the beam
Attachment 1: Screenshot_2025-05-14_06-33-45.png
Screenshot_2025-05-14_06-33-45.png
  56   Wed May 14 02:19:46 2025 Dimitrios, SimonAnalysisLongitudinal bunch length
We repeated this measurement for 208-Bi. The bunch lengths are about half as long (due to lower space charge and consequently less heating?)

PMT north PMT mid particle detector
width (bins)1.71.61.24
width (ns)5.65.34.09



Quote:
We measured the longitudinal bunch length for all different detectors

PMT north PMT mid particle detector
width (bins)2.912.742.45
width (ns)9.608.988.09
Attachment 1: Bildschirmfoto_2025-05-14_02-16-42.png
Bildschirmfoto_2025-05-14_02-16-42.png
Attachment 2: Bildschirmfoto_2025-05-14_02-17-58.png
Bildschirmfoto_2025-05-14_02-17-58.png
Attachment 3: Bildschirmfoto_2025-05-14_02-15-05.png
Bildschirmfoto_2025-05-14_02-15-05.png
  57   Wed May 14 06:31:28 2025 Dimitrios, SimonRunsDR scans @300mA
We reduced the electron cooler current to 300mA and continued measuring the DR spectrum.

Time File PTB 32 PTB 33 Ohmlabs ESR intensity SC8 Typical SIS injections comments
6:56 41 205826 205827 205819 5e5 4e8 opened new file
7:23 41 205826 205827 205820 5e5 6e8
7:48 41 205826 205827 205820 5e5 5e8
8:16 41 205826 205827 205820 0 4e8 the beam was killed, could be by GE02DS3HG, although it did not go in too far -
  58   Wed May 14 08:24:39 2025 Danyal, Nils, Hendrik Day shiftRunsDR scans @300mA
Continuing with the measurements of the last shift, also at 300mA. Creating new file, as beam got killed during last run.

Time File PTB 32 PTB 33 Ohmlabs ESR intensity SC8 Typical SIS injections comments
8:42 42 205826 205827 205820 4e5 4e8 opened new file
9:07 42 205826 205827 205820 4e5 4e8
9:34 42 205826 205827 205820 3e5 4e8
9:59 42 205826 205827 205820 0 4e8 DS3HG went to position 20.4, killing the beam
10:24 43 205826 205827 205820 3e5 4e8 opened new file
10:51 43 205826 205827 205820 0 4e8 Beam in the ESR got killed again. DS3HG is at normal position though, so we don't know why
11:17 44 205826 205827 205820 0 4e8 opened new file, DS3HG killed the beam again
11:41 45 205826 205827 205820 2e5 4e8 opened new file
12:07 45 205826 205827 205820 0 4e8 DS3HG killed the beam by going too far for just a second and then going to normal position. Maybe that was the reason some previous beams got killed even though when we looked at the scraper it was in the normal position
11:41 46 205826 205827 205820 2e5 4e8 opened fie
  59   Wed May 14 11:59:39 2025 NilsAnalysisCurrent DR spectra 208Bi
Results so far for the DR scans @300mA e-cooler current.
Attachment 1: DR_208Bi_300mA_raw.png
DR_208Bi_300mA_raw.png
Attachment 2: DR_208Bi_300mA_Time_normalized_.png
DR_208Bi_300mA_Time_normalized_.png
  60   Wed May 14 12:43:43 2025 Danyal, Nils, Hendrik Day shiftRunsDR scans @300mA
Continuing with the measurements of the last shift, also at 300mA. Creating new file, as beam got killed during last run.

[TABLE border="1"]
Time |File | PTB 32 | PTB 33 | Ohmlabs | ESR intensity SC8 | Typical SIS injections | comments |-
8:42 | 42 | 205826 | 205827 | 205820 | 4e5 | 4e8 | opened new file |-
9:07 | 42 | 205826 | 205827 | 205820 | 4e5 | 4e8 |-|-
9:34 | 42 | 205826 | 205827 | 205820 | 3e5 | 4e8 |-|-
9:59 | 42 | 205826 | 205827 | 205820 | 0 | 4e8 | DS3HG went to  position 20.4, killing the beam |-
10:24 | 43 | 205826 | 205827 | 205820 | 3e5 | 4e8 | opened new file |-
10:51 | 43 | 205826 | 205827 | 205820 | 0 | 4e8 | Beam in the ESR got killed again. DS3HG is at normal position though, so we don't know why |-
11:17 | 44 | 205826 | 205827 | 205820 | 0 | 4e8 | opened new file, DS3HG killed the beam again |-
11:41 | 45 | 205826 | 205827 | 205820 | 2e5 | 4e8 | opened new file |-
12:07 | 45 | 205826 | 205827 | 205820 | 0 | 4e8 | DS3HG killed the beam by going too far for just a second and then going to normal position. Maybe that was the reason some previous beams got killed even though when we looked at the scraper it was in the normal position. Stopped measurements so Sergey can maybe fix the issue with the scraper. Slightly adjusting the setpoints of the scraper should fix the problem of it going too far. Last file got closed. |-
[/TABLE]
  61   Wed May 14 19:06:51 2025 CarstenRunsDifferent pattern/ Fixes during the day new runs for 300 mA (increased intensities)
During the day several things were discussed concerning intensity and the pressurized air devicees.
In total 6 to 7 different things are not ok for a proper opreation of the devices (not to be discussed here).

As a workaround on top of other workarounds:

We now have a different pattern with dedicated subchains for moving detector (+ additional possibilities for manipulations that might become handy)

  • Detectors are moved in in SC9 and out in SC12. Due to the 3 internal beam processes (BP1 to BP3) the old SC7 was not suited for that.
  • The waiting machine for measurements is now SC10
  • The ranp file has now a 5 seconds waiting at the beginning to compensate for the longer time between ramping down and scraping.
  • As a byproduct the deceleration decomes more efficient.
  • Since the detector north 02DD2AG seems to drivemore reliable than the scraper 02DS3HG we now use the detecotr north for scraping.
  • Positions in SC9: 01DD2AG: -55 (destroys beam at ~-25); 01DD1AG: 26; 02DD2_G: -60, all out positions (SC12) are 120 (-120)
  • Due to the higher number of beam processes, changes of electron current or the delta voltage have to be performed in both, SC8 and SC10.
  • The first 10 channels or so in the DR spectra might show some weird behaviour - don't worry that is "by design" (aka due to the changes above).
  • We now had 2e6 208Bi80+ for 8e8 SIS current (15 injections, 45 cycles breeding at 10s) - thre still seems to be some headroom.

NTCAP: 2025-05-14_18-47-20

300mA (bunched) , delta U = 880V (-20V without drifttube cooling)

run 0047.lmd started (at least 0046.lmd is empty, check also teh other files sincve the morning shift).
Further files for 300 mA: 0048.lmd und 0049.lmd

Typical ion beam intensities at beginning of scan: up to 3e6 scraped

Spectra below.

The optimal channels for DRALS would be 290 (max.) (F=4) that is shifted for the F=5 by 878 eV (+88 channels) to channel 378. Here the max of the pure HFS excited resonances would appear.

This translates into lab energy voltages of 3348V (F=4 maximum) and 3243 V (F=5 maximum) and -900V cooling.
These voltages are only valid for 300 mA. Voltages for 500mA are very similar (offline analysis). For practical purposes, the same voltages can be used for 300 mA bunched and 500 mA bunched.

Closed NTCAP file
Attachment 1: DR_208Bi_300mA_optimized_ion_beam.png
DR_208Bi_300mA_optimized_ion_beam.png
Attachment 2: DR_208Bi_300mA_optimized_ion_beam_cts.png
DR_208Bi_300mA_optimized_ion_beam_cts.png