Sat Jun 28 15:28:22 2025 |
Carsten., Uwe | Runs | Runs with new scan range | 22:10 SIS Trafo GS09DT scale was changed to 4e8 (not just the lassie monitor)
23:50 Ausfall ESR: GE01 KP01 bis KP05 und KP22 bis KP24; Korrekturelement: GE02KX6 und GE01KX2; beam loss during measurement
01:45 reser press air Scrapers: GE02DD2_G (sc8,11); GE02DS03HG(sc7,8,11); GE01DD2IG(was sc7,10,11)(now7,8,11)
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
15:25 | >20000 | 2.5e9 | 30 | - | - | - | 260 | 234 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: detector started too late
15:52 | 15200 | 2.5e9 | 30 | - | - | - | 261 | 235 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: SIS injection fluctuating
16:09 | 6000 | 3.3e9 | 30 | - | - | - | 262 | 236 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: accumulation curve in ESR much flatter
16:31 | 5800 | 3.0e9 | 30 | - | - | - | 264 | 237 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: data not in online DAN; lmd-file: DAQ2_0263.lmd is empty
16:47 | 10000 | 3.0e9 | 30 | - | - | - | 265 | 238 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: okay
17:10 | 15500 | 3.0e9 | 35 | - | - | - | 266 | 239 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: okay
HHT access
17:20 beam back, but problems with SIS intensity
17:35 intensity good, again. 3x10**9 particles in SIS
17:45 | 16000 | 3.2e9 | 35 | - | - | - | 267 | 240 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | comment: okay
17:45 | 00 | 3.0e9 | 40 | - | - | - | 267 | 240 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 | comment: okay
|
During whole night shift scraping right peak did not work consistently. Remaining intensity varied from similar peak height compared to thorium down to about 20%. |
70
|
Sat Jun 28 06:29:58 2025 |
Stefan, Simon, Konstantin | Runs | Morning shift 28.06. | 6:30 Problem with ion source, on-call service is called
7:30 There is a short in the ion source and it needs to be changed. Expected down time about 2h
10:30 Ion source was successfully changed, but while ramping up the source a power supply failed. Next on-call service is on their way to GSI.
13:00 ions are back
We changed the dye as the laser intensity was quite low. That didnt help, so Konstantin came in to change the dye cell. That did help.
No scans during morning shift. |
69
|
Fri Jun 27 22:14:35 2025 |
Anton, Finn, Carsten | Runs | Runs night shift 27.06. 28.06 | 22:10 SIS Trafo GS09DT scale was changed to 4e8 (not just the lassie monitor)
23:50 Ausfall ESR: GE01 KP01 bis KP05 und KP22 bis KP24; Korrekturelement: GE02KX6 und GE01KX2; beam loss during measurement
01:45 reser press air Scrapers: GE02DD2_G (sc8,11); GE02DS03HG(sc7,8,11); GE01DD2IG(was sc7,10,11)(now7,8,11)
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
22:08 | 17000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 205 | 233 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 |
22:27 | 7200 | 4e8 | 30 | 104524 | 104527 | 104528 | 206 | 234 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 ! small peak on right side, not good scraped |
23:19 | 6000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 207 | 235 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
23:35 | 15000 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 208 | 236 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
23:47 | 00 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 209 | 237 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \ scan not compete, beam loss during measurement |
00:17 | 6200 | 2.7e9 | 30 | 104524 | 104527 | 104528 | 210 | 238 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
00:45 | 6400 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 211 | 239 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:00 | 5100 | 2.75e9 | 30 | 104524 | 104527 | 104528 | 212 | 240 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:15 | 5200 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 213 | 241 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:45 | 4900 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 214 | 242 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \bigger peak on right side, badly scraped |
01:58 | 12600 | 2.25e9 | 30 | 104524 | 104527 | 104528 | 215 | 243 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:13 | 12000 | 2.4e9 | 30 | 104524 | 104527 | 104528 | 216 | 244 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:28 | 13900 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 217 | 245 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:48 | 17000 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 218 | 246 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
02:58 | 13500 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 219 | 247 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
03:13 | 14900 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 220 | 248 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
03:27 | 5200 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 221 | 249 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
03:42 | 13000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 222 | 250 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
04:08 | 14300 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 223 | 251 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
04:23 | 15400 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 224 | 252 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
04:38 | 13800 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 225 | 253 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:03 | 12800 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 226 | 254 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:23 | 3900 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 227 | 255 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:38 | 11200 | 3.e9 | 30 | 104524 | 104527 | 104528 | 228 | 256 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
05:53 | 3900 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 229 | 257 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
06:08 | 11700 | low.intensity | 30 | 104524 | 104527 | 104528 | 230 | 258 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
|
During whole night shift scraping right peak did not work consistently. Remaining intensity varied from similar peak height compared to torium down to about 20%. |
68
|
Fri Jun 27 14:23:39 2025 |
Nils, Wilfried, Carsten, Julian | Runs | Runs afternoon shift 27.06. | 15:30 We have a break due to an issue at UNILAC (quadrupole failed). Also radiation safety interlock was triggered at the fragment separator. This also pushed a faradaycup into the ESR (GE01DC3VP).
Just before the issue during the last bunching, the peak was misaligned to the marker in the Schottky. The marker was sitting on the right flank. So the electron-cooled 229Th did not match with the bunching frequency anymore. We are preparing to adjust the cooler voltage to fix this. For some reason the frequency recovered and is back to its original position. However, our accumulations do not seem to work as well as before, we stagnate much earlier and at a much lower level.
15:50 Beam is back, Carsten is still investigating the drop in intensity we saw before the UNILAC failure.
17:15 TE slits were on -8 and +8, opening to -11 and +11, no change to acquisition pattern, something is still wrong
18:00 We were locked out by HHT people for 30+ mins
18:27 We had to call in ESR support
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
14:24 | 15800 | 2.74 e9 | 30 | 104524 | 104527 | 104528 | 196 | 224 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Carsten optimizing beam at HKR with ESR colleagues
|
14:24 | 6800 | 1.16 e9 | 30 | 104524 | 104527 | 104528 | 197 | 225 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Carsten is back in measurement hut
|
16:30 | 1100 | 1.1 e9 | 40 | 104524 | 104527 | 104528 | 198 | 226 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Searching for reason for small particle number
|
17:00 | 1200 | 0.5 e9 | 30 | 104524 | 104527 | 104528 | 199 | 227 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
19:57 | 7800 | 1e9 | 30 | 104524 | 104527 | 104528 | 200 | 228 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
20:16 | 6800 | 1.5e9 | 30 | 104524 | 104527 | 104528 | 201 | 229 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
20:41 | 7200 | 1.5e9 | 30 | 104524 | 104527 | 104528 | 202 | 230 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
21:28 | 6000 | 2.65e9 | 30 | 104524 | 104527 | 104528 | 203 | 231 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
21:47 | 6600 | 2.7e9 | 30 | 104524 | 104527 | 104528 | 204 | 232 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
22:08 | 17000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 205 | 233 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7
|
|
| |
67
|
Fri Jun 27 11:21:25 2025 |
Esther, Konstantin | DAQ | Timing | We cross-checked the timing. Laser pulse is detected at bins 136 (PD-North) and 171 (PD-south), respectively. Therefore, the laser puls is in the electron cooler at 153.5. The revolution period corresponds to 3e8Hz / 1.54315MHz = 194.
The ion bunch must travel 2.5m more than the half of the circumference (108.3m). Taking this into account, the ion bunch must be detected (0.5 + 2.5 / 108.3) * 194 = 101.5 bins later (meaning less) than 153.5.
Finally, we are happy that the timing is still perfect, since the ion bunch arrives at 52. |
66
|
Fri Jun 27 06:55:45 2025 |
Esther. Konstantin | Laser | Laser-ion overlap | We cross-checked the laser-ion superposition by using the T-scrapers. We did it only after the scans in order to do not change the duty cycle by much. Attachements are in the order of the measurements compiled in the table.
motor | ion beam position (mm) from inside |
GECEDS1VU | -0,5 |
GECEDS1HA | -4 |
GECEDS2VU | 0.9 | |
GECEDS2VU | 0 |
|
Still looks fine to us. |
Attachment 1: 2025-06-27_06-52-01-807.png
|
|
Attachment 2: 2025-06-27_07-12-27-135.png
|
|
Attachment 3: 2025-06-27_07-32-28-599.png
|
|
Attachment 4: 2025-06-27_07-51-52-620.png
|
|
Draft
|
Fri Jun 27 06:13:02 2025 |
Esther, Konstantin | Runs | Early shift - data taking | Notes:
1. TE slits at -10 and +10
2. SIS intensity unstable until about 9:00 after that it became much more stable
3. brief interruption between 09:30 and 10:03 due to interlock issue at SIS
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
06:15 | 16300 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 172 | 199 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
06:30 | 13500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 173 | 200 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
06:44 | 4000 | 1.5-2 e9 | 30 | 104522 | 104526 | 104526 | 174 | 201 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped vertically with scraper 1 after end of laser scan (see entry 66), scraper left in but tilted to allow for injection
|
07:05 | 14800 | 2-3.5 e9 | 30 | 104522 | 104526 | 104526 | 175 | 202 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 1 after end of laser scan (see entry 66), scraper 1 moved back out
|
07:25 | 14000 | 1.1-2.6 e9 | 30 | 104522 | 104526 | 104526 | 176 | 203 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 2 after end of laser scan (see entry 66), scraper 2 left in but tilted to allow for injection
|
07:44 | 4200 | 1-2.5 e9 | 30 | 104522 | 104526 | 104526 | 177 | 204 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 2 after end of laser scan (see entry 66), scraper 2 moved back out
|
08:12 | 5700 | 1-2.5 e9 | 30 | 104522 | 104526 | 104526 | 178 | 205 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | !!file in DAQ1 not closed after end of scan!!; ~5 injections missing because nothing came from SIS
|
08:27 | 16000 | 1.5-3 e9 | 30 | 104522 | 104526 | 104526 | 178 | 206 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | !!file in DAQ1 still open from last scan!!
|
08:44 | 11700 | 1.5-3 e9 | 30 | 104522 | 104526 | 104526 | 179 | 207 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
09:00 | 15600 | 2.6-3 e9 | 30 | 104522 | 104526 | 104526 | 180 | 208 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
09:24 | 6500 | 2.8 e9 | 30 | 104522 | 104526 | 104526 | 181 | 209 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:13 | 14600 | 2.2 e9 | 30 | 104522 | 104526 | 104526 | 182 | 210 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:31 | | | | | | | 183 | 211 | | | | | | | BAD FILES: ESR not stopped in SC9, injection during laser scan
|
10:39 | 14500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 184 | 212 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:55 | 15400 | 2.3 e9 | 30 | 104522 | 104526 | 104526 | 185 | 213 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:11 | 15500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 186 | 214 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:27 | 15800 | 2.5 e9 | 30 | 104522 | 104526 | 104526 | 187 | 215 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:41 | 5200 | 1.4-2.6 e9 | 30 | 104522 | 104526 | 104526 | 188 | 216 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:56 | 16500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 189 | 217 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:12 | 15000 | 2.9 e9 | 30 | 104522 | 104526 | 104526 | 190 | 218 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:27 | 15000 | 2.8 e9 | 30 | 104522 | 104526 | 104526 | 191 | 219 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:44 | 15800 | 2.9 e9 | 30 | 104522 | 104526 | 104526 | 192 | 220 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:21 | 5400 | 2.3 e9 | 30 | 104522 | 104526 | 104526 | 193 | 221 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:42 | 4800 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 194 | 222 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:59 | 6200 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 195 | 223 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| |
64
|
Thu Jun 26 22:51:04 2025 |
Hendrik, Finn | Runs | Nightshift data runs | After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted.
23:10 Quadrupole magnet stopped working, lost beam in ESR. Contacted on-call ESR service
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
0:15 | 8000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 150 | 177 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
0:28 | 20900 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 151 | 178 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
0:59 | 19000 | 2.3-2.7 e9 | 30 | 104522 | 104526 | 104526 | 152 | 179 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:14 | 14500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 153 | 180 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:30 | 19800 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 154 | 181 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:45 | 18000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 155 | 182 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:01 | 6600 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 156 | 183 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:18 | 14500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 157 | 184 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:36 | 14700 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 158 | 185 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:45 | 6000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 159 | 186 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:00 | 5500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 160 | 187 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:17 | 5200 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 161 | 188 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:34 | 14300 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 162 | 189 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:50 | 17200 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 163 | 190 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:07 | 5000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 164 | 191 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:22 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 165 | 192 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:39 | 14000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 166 | 193 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:55 | 17000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 167 | 194 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:12 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 168 | 195 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:26 | 14000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 169 | 196 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:43 | 13100 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 170 | 197 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:56 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 171 | 198 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
| |
63
|
Thu Jun 26 21:27:34 2025 |
Carsten, Jan, Simon | Calibration | Timing | We checked the timing (before the start of the measurement cycle documented below).
Timing and measurement window are attached. |
Attachment 1: Bildschirmfoto_timing_new_2025-06-26_19-00-04.png
|
|
62
|
Thu Jun 26 19:01:46 2025 |
Carsten, Jan, Simon | Runs | Late shift 26.06. |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
19:15 | 23000 | 1.5-3.5 e9 | 33 | 104522 | 104526 | 104526 | 142 | 170 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
19:41 | 19000 | 3.3e9 | 30 | 104522 | 104526 | 104526 | 143 | 171 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -7 and +7
|
20:08 | 17500 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 144 | 172 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | -
|
20:24 | 7000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 145 | 173 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | low Part. Det. Int maybe due to wrond position of part. det.
|
21:19 | 7000 | 3.5e9 | 25 | 104522 | 104526 | 104526 | 146 | 174 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -8 and +8
|
21:40 | 14000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 147 | 175 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | adjusted particle detector position
|
22:04 | 19500 | 3.3e9 | 40 | 104522 | 104526 | 104526 | 148 | 176 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted. |
61
|
Thu Jun 26 18:13:17 2025 |
Jan, Carsten | Detectors | PMT voltages & thresholds | current settings of PMT voltages and thresholds at the CAEN CFD:
PMT South [HV ch2: 2700V] [CFD ch0: 8]
PMT Middle [HV ch1: 2850V] [CFD ch1: 8]
PMT Nord [HV ch3: 2850V] [CFD ch2: 9] |
60
|
Wed Jun 25 05:53:54 2025 |
Rodolfo | Laser | FCU | Cobra's FCU was recalibrated. Table saved as FCU_547-554_RS_2024-06-25.fcu. |
59
|
Tue Jun 24 22:24:36 2025 |
Carsten, Finn, Hiroshi, Esther, Rodolfo | Runs | Data taking (night shift 24.06.) |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Notes:
1. NTCAP was only started after the fourth scan. The file timestamp is '2025-06-25_00-02-08'.
2. Beam was lost at 03:45 and could not be recovered by the operators. On-call service was called but they have to replace a part, current estimate is 2-3 hours.
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
22:39 | 13500 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 128 | 157 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 60 injections, cooling time (SC 7) 28s -> 32s
|
23:00 | 11500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 129 | 158 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 45 injections
|
23:20 | 10500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 130 | 159 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
23:44 | 9100 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 131 | 160 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
00:05 | 14000 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 132 | 161 | 551.91 - 549.41 | 501 | 15 | 1 | yes | back to 60 injections
|
00:28 | 14300 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 133 | 162 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 75 injections, mesytec channel threshold 0-3 set 255
|
01:26 | 13200 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 134 | 163 | 551.91 - 549.41 | 501 | 15 | 1 | yes | there was an interruption at the source, took some time to get the beam back
|
01:48 | 10500 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 135 | 164 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:10 | 12400 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 136 | 165 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:33 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 137 | 166 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:55 | 10000 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 138 | 167 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
03:18 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 139 | 168 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
|
58
|
Tue Jun 24 18:49:18 2025 |
Carsten, Rima, Nils, Simon | Runs | Second attempt (Late shift 24.06.) |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
19:03 | 14800 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 118 | 147 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -30 +30
|
19:22 | 14900 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 119 | 148 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
19:40 | 14500 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 120 | 149 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -20 +20
|
19:58 | 13800 | 2.0e9 | 60 | 104510 | 104514 | 104514 | 121 | 150 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:18 | 10800 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 122 | 151 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:37 | 10100 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 123 | 152 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:56 | 10500 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 124 | 153 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
21:15 | 9000 | 2.2e9 | 60 | 104510 | 104514 | 104514 | 125 | 154 | 551.91 - 549.41 | 501 | 15 | 1 | yes | after this we change to 45 injections to give more cave access to HHT
|
21:37 | 11500 | 2.3e9 | 45 | 104510 | 104514 | 104514 | 126 | 155 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT
|
22:13 | 13400 | 2.5e9 | 45 | 104510 | 104514 | 104514 | 127 | 156 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT, cooling time in SC7 increased from 20s to 28 s
|
|
57
|
Tue Jun 24 17:39:13 2025 |
Carsten, Stefan, Simon, Rima, Nils | Detectors | PMT behaviour - noise / PMT voltages / CFD Threshholds / CAEN Power supply (broken channel resetted) | PMT North and South have been changed:
CFD CH0: Middle Thr 7 / CAEN HV Ch1: 2750 V
CFD CH1: South Thr 8 / CAEN HV Ch2: 2700 V
CFD CH2: North Thr 8 / CAEN HV CH3: 2800 V
MESY CFD Ch4: Thr.17
MESY CFD Ch5: Thr.16
MESY CFD Ch6: Thr.20
Quote: | In position North there is still the "old" 9423B PMT that draws quite some current.
On this detector threre is also some quite substantial noise. Very likely the base is not ok any more. Noise was much lesss for the other two PMTs.
Since position NORTH is the position with the largest sold angle, if there is another break from the accelerator we should exchange it against SOUTH.
It was also difficult to ramp up PMT north ; and I thought that againa channel broke (showing UnV not OVC (Overcurrent)). However, the PS is resetted if
you switch of the whole crate (aka switch the PS off). Before doing so, please ramp down the other channels slowly as intended.
Please leave the PMTs (besides you go in to exhchange the PMTs).
If you ramp pt PMT North (or the same detector at a different position), start with 2400 V and then slowly increase by steps of 50V.
Checked theresholds (but during injections, so in presence of electronic noise from injection). A further adjusted might be done later when we have measurement conditions.
(Or if we have time to exchange the PMTs).
Presently we have the following settings for CAEN:
CH1: Middle 2750 V, Thr 9
CH2: South 2700 V, Thr 7
CH3: North 2800 V, Thr 9 |
|
56
|
Tue Jun 24 15:44:06 2025 |
Konstantin | DAQ | Phase shift in DAQ1 | In the SetParameters.C file is a parameter called fPar->piPhaseOffset. It was shifted from 100 to 50 in order to see ion bunches in the pmt signal and laser pulses in the photodiode signal during the same revolution in the TDC spectrum. Note that this parameter must be taken
into account when setting the window conditions of the arrival time in the SetConditions.C file. |
55
|
Tue Jun 24 12:25:34 2025 |
Carsten | Detectors | PMT behaviour - noise / PMT voltages / CFD Threshholds / CAEN Power supply (broken channel resetted) | In position North there is still the "old" 9423B PMT that draws quite some current.
On this detector threre is also some quite substantial noise. Very likely the base is not ok any more. Noise was much lesss for the other two PMTs.
Since position NORTH is the position with the largest sold angle, if there is another break from the accelerator we should exchange it against SOUTH.
It was also difficult to ramp up PMT north ; and I thought that againa channel broke (showing UnV not OVC (Overcurrent)). However, the PS is resetted if
you switch of the whole crate (aka switch the PS off). Before doing so, please ramp down the other channels slowly as intended.
Please leave the PMTs (besides you go in to exhchange the PMTs).
If you ramp pt PMT North (or the same detector at a different position), start with 2400 V and then slowly increase by steps of 50V.
Checked theresholds (but during injections, so in presence of electronic noise from injection). A further adjusted might be done later when we have measurement conditions.
(Or if we have time to exchange the PMTs).
Presently we have the following settings for CAEN:
CH1: Middle 2750 V, Thr 9
CH2: South 2700 V, Thr 7
CH3: North 2800 V, Thr 9 |
54
|
Mon Jun 23 18:36:39 2025 |
Wilfried, Simon | Detectors | PMT test with LED | We tested the PMTs again, this time with an UV LED that is mounted at a window close to the XUV detector.
PMT south showed a very clear increase in count rate from about 1 to 50 Hz.
PMT mid also showed a small increase by a few Hz.
For PMT north there was no great change between LED off and on, but we think that the count rate also very slightly increased. Note, that PMT north is also the furthest away from the LED.
Also we saw a difference between single- and multi-photon events at PMT south. |
53
|
Mon Jun 23 10:59:36 2025 |
Julian, Lukas, Zsolt, Nadine, Konstantin | Runs | Scan routine | -Start new injection
->wait for 60 injections to be complete
->note down SIS intensity (at extraction)
->observe Schottky to ensure proper bunching
->check laser position (should be well-centered)
-open file, start Go4
->do '@openfile' in both mbs windows
->start acquisition (green arrow) in both Go4 instances (check indicator 'Current Ev/s' at the bottom, should be green, not red)
->start laser scan
->reset Ch3 of the particle detector (in the electronics hut)
->check laser position (disappears at start of scan then drifts through the image from the left)
-when laser scan is finished:
->stop Go4 acquisition (big red 'STOP' button in the top row)
->do '@closefile' in both mbs windows
->check counts on particle detector
-update elog
-repeat |
51
|
Mon Jun 23 10:52:34 2025 |
Julian, Lukas, Zsolt, Nadine, Konstantin | Laser | Frequency conversion unit | We exchanged the old BBO of the Cobra's FCU with a new one. The output power of the Cobra was around 120mJ. In the end, we achieved a UV power of about 21mJ. In addition, we optimized the fiber-coupling to the wavemeter. |
|