Entry  Thu Jun 26 22:51:04 2025, Hendrik, Finn, Runs, Nightshift data runs 
After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted.

23:10 Quadrupole magnet stopped working, lost beam in ESR. Contacted on-call ESR service

Time Part. Det. Int SIS-intensity accumulations in ESR cooler-HV PTB K250 cooler-HV PTB red cooler-HV orange LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time (laser shots) n loops checked laser position mesytec rate ok comments
0:15 8000 1.5-3.5 e9 30 104522 104526 104526 150 177 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
0:28 20900 1.5-3.5 e9 30 104522 104526 104526 151 178 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
0:59 19000 2.3-2.7 e9 30 104522 104526 104526 152 179 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
1:14 14500 1.5-3.5 e9 30 104522 104526 104526 153 180 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
1:30 19800 1.5-3.5 e9 30 104522 104526 104526 154 181 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
1:45 18000 1.5-3.5 e9 30 104522 104526 104526 155 182 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
2:01 6600 1.5-3.5 e9 30 104522 104526 104526 156 183 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
2:18 14500 1.5-3.5 e9 30 104522 104526 104526 157 184 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
2:36 14700 1.5-3.5 e9 30 104522 104526 104526 158 185 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
2:45 6000 1.5-3.5 e9 30 104522 104526 104526 159 186 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
3:00 5500 1.5-3.5 e9 30 104522 104526 104526 160 187 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
3:17 5200 1.5-3.5 e9 30 104522 104526 104526 161 188 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
3:34 14300 1.5-3.5 e9 30 104522 104526 104526 162 189 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
3:50 17200 1.5-3.5 e9 30 104522 104526 104526 163 190 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
4:07 5000 1.5-3.5 e9 30 104522 104526 104526 164 191 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
4:22 16500 1.5-3.5 e9 30 104522 104526 104526 165 192 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
4:39 14000 1.5-3.5 e9 30 104522 104526 104526 166 193 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
4:55 17000 1.5-3.5 e9 30 104522 104526 104526 167 194 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
5:12 16500 1.5-3.5 e9 30 104522 104526 104526 168 195 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
5:26 14000 1.5-3.5 e9 30 104522 104526 104526 169 196 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
5:43 13100 1.5-3.5 e9 30 104522 104526 104526 170 197 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
5:56 16500 1.5-3.5 e9 30 104522 104526 104526 171 198 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
Entry  Sun Jun 29 03:03:56 2025, Nadine, Finn, Runs, Night shift runs with laser 
around 22:00 water cooling circuit failure of ESR electron cooler--> ~1h downtime
Wavemeter frequently overexposed around 549,3nm; other wavelengths are fine

Time Part. Det. Int SIS-intensity accumulations in ESR cooler-HV PTB K250 cooler-HV PTB red cooler-HV orange LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time (laser shots) n loops checked laser position mesytec rate ok TE slits comments
23:25 13500 2.5e9 40 - - - 274 247 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
23:42 5000 2.5e9 40 - - - 275 248 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
00:02 5700 2.8e9 40 - - - 276 249 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
00:26 5000 3e9 30 - - - 277 250 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
00:40 5300 2.8e9 30 - - - 278 251 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
00:55 5550 3e9 30 - - - 279 252 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
01:13 16800 3e9 40 - - - 280 253 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
01:32 13300 3e9 50 - - - 281 254 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
01:50 13700 3e9 40 - - - 282 255 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
02:07 14450 3e9 40 - - - 283 256 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
02:24 13800 2.8e9 40 - - - 284 257 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
02:41 12600 2.8e9 40 - - - 285 258 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 not properly scraped; right peak -100dB
02:58 13200 3e9 40 - - - 286 259 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
03:15 13700 2.7e9 40 - - - 287 260 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
03:32 3400 2.7e9 40 - - - 288 261 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
03:45 5000 2.7e9 40 - - - 289 262 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
04:12 3700 3e9 50 - - - 290 263 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
04:30 4100 3e9 40 - - - 291 264 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
04:47 12800 3e9 40 - - - 292 265 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
05:10 2800 2.7e9 60 - - - 293 266 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 file 266 was not closed until next file should be opend! Happens at 5:24 in the Morning Smile
05:28 12200 2.7e9 40 - - - 294 267 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
05:57 15000 2.7e9 40 - - - 295 268 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
06:09 00 2.7e9 40 - - - 296 269 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
Entry  Sun Jun 29 22:06:19 2025, Carsten, Alex, Hendrik, Runs, Night shift 29.06. 
[TABLE border="1"]
Time | Part. Det. Int | SIS-intensity| accumulations in ESR |  cooler-HV PTB K250 |  cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments |-
22:16 | 4800 | 2e9 | 30 | 104525 | - | - | 300 | 327 | 548.51 - 548.01 | 201 | 15 | 2 | yes | only few ions, did not start Go4 |-
22:28 | 9000 | 2e9 | 15 | 104525 | - | - | 301 | 328 | 548.51 - 548.01 | 201 | 15 | 2 | yes | After this run: Shaky SIS intensity, cathode is being changed |-
22:53 | 7000 | 2e9 | 35 | 104525 | - | - | 302 | 329 | 548.51 - 548.01 | 201 | 15 | 2 | yes |-|-
[/TABLE] 
23:20 Failure of GE01KP04, GE01KP05, GE01KP22 (or GE01KP24) devices in the ESR.
Problems with gtemu3-hick (??) , take injections since 
01:49 Bernd tried to make things work by switching off and on some of the dipole correction coils. He was even able to get a quite good intensity by leaving one of the coils switched off (although with a slight shift of frequencies visible in Schottky spectrum) but then the ESR electron cooler has failed probably due to problems with the cooling water. So, the decision was taken to stop the measurement at this point. The laser was switched off.
Entry  Mon Jun 16 21:47:23 2025, Carsten, Wilfried, Simon , Detectors, New threshold values for PMTs 
We adjusted the thresholds for the PMTs and increased the voltage of PMT middle to 2.7kV

PMT Model Operation Voltage Threshold (mV)
South ET9422B Broadband -1750 V 30
Middle ET9423B Smallband -2700 V 18
North ET9423B Smallband -2650 V 10
Entry  Thu Jun 12 16:27:45 2025, Carsten, DAQ, NTCAP directory 
On the NTCAP the directory is changed to 
Q:/E0052
Entry  Sat Jun 28 06:29:58 2025, Stefan, Simon, Konstantin, Runs, Morning shift 28.06. 
6:30 Problem with ion source, on-call service is called
7:30 There is a short in the ion source and it needs to be changed. Expected down time about 2h
10:30 Ion source was successfully changed, but while ramping up the source a power supply failed. Next on-call service is on their way to GSI.
13:00 ions are back


We changed the dye as the laser intensity was quite low. That didnt help, so Konstantin came in to change the dye cell. That did help.

No scans during morning shift.
Entry  Sun Jun 29 06:11:49 2025, Stefan, Konstantin, Runs, Morning Shift 29.06. 


Time Part. Det. Int SIS-intensity accumulations in ESR cooler-HV PTB K250 cooler-HV PTB red cooler-HV orange LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time (laser shots) n loops checked laser position mesytec rate ok TE slits comments
06:09 3300 2.7e9 40 - - - 296 269 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
08:49 16000 2.5e9 40 - - - 297 270 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
09:10 4100 2.6e9 40 - - - 298 271 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 slightly modified the laser position, which was necessary to ensure proper laser-ion superposition after retuning the laser, some bad injections from SIS
09:2X 16000 2.8e9 40 - - - 299 272 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
09:59 2200 2.4e9 40 - - - 300 273 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 One bad injection cycle between the previous measurement and this one. Particle detector reset lately
10:33 3800 2.0e9 40 - - - 301 274 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
10:52 4200 2.1e9 40 - - - 302 275 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
11:17 4100 2.1e9 30 - - - 303 276 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
11:41 4400 2.1e9 30 - - - 304 277 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
11:53 4600 2.1e9 30 - - - 305 279 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 Bad (empty) injections from SIS
12:19 4600 2.5e9 30 - - - 306 279 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 Potentially problems with transfer beam line
12:33 3700 2.1e9 30 - - - 307 280 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
12:50 4300 2.1e9 30 - - - 308 281 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
13:07 18300 2.5e9 30 - - - 309 282 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
13:30 2300 2.5e9 30 - - - 310 283 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5
14:05 1 2.5e9 30 - - - 311 284 550.610 - 548.110 501 15 1 yes yes TE slits -9 +5.5 Particle detector hasn't moved in

We did a recalibration of the FCU. After this (7:40) a SIS power supply failure was detected, and we are waiting now for the on-call service.
HHT cave access around 10:06 during a short break due to a source failure.
Every once in a while we have empty shots from SIS.
Entry  Mon Jun 2 16:36:57 2025, Carsten, Detectors, Mesycontrol MSCF16-F-V.pdf
ssh -Y litv-exp@x86l-204
'mesycontrol'
Entry  Fri Jun 6 14:27:10 2025, Rodolfo, Konstantin, Laser, MRC Beam Stabilization Presettings MRC-Settings_2025_06-06.JPGAFG_Settings_Thorium_2025-06-06.jpgMRC-Controller_Settings_Thorium_2025-06-06.jpg
These are the settings for the Beam Stabilization from MRC.

Laser-beam energy ca. 18 mJ @ 275 nm
Entry  Sat Jun 21 08:48:16 2025, Arya, Julien, Wilfried, Accelerator, Low 229Th particle numbers and 229Th lifetime in ring Bildschirmfoto_timing_full_2025-06-21_09-30-46.pngBildschirmfoto_timing_end_of_manipulation_2025-06-21_09-36-24.pngBildschirmfoto_timing_full_60_2025-06-21_11-25-51.pngBildschirmfoto_PDSouth_lifetimes_2025-06-21_11-26-43.png
With reference to entry 37 Laser timing:

- at last 30 injections accumulation: 10k counts on particle detector (accumulated during manipulation until it saturates, e.g. no more Th in ESR)
- at first 70 injections accumulation: 7.6k counts
- at second 70 injections accumulation: 2.8k counts

did another 2 30 injections accumulations to confirm lower yield: 
- at first 30 injections accumulation: 3.7k counts
  SIS intensities look fine, around 1.2e9
- at second 30 injections accumulation:  6.9k counts
  saved to /data.local2/2025_229Th/2025_229ThDAQ1_0101.lmd with DAQ1
  SIS intensities look fine, around 1.2e9, only users during injection
  "late" contaminants on PD seem to quickly disappear after start of modulation 3 
  (attachment 1: full timing spectrum, attachment 2: timing spectrum at the end of the modulation)
- at second 60 injections accumulation:  0 counts, maybe scrapers overshot,did another one, still not accumulating

     -------------------   Some issue on accelerator side   -------------------
        FSR people had to go into cave, that's why we were the only users, 
        but since nothing got into ESR
        Connection to scrapers was lost, on call service had to do a reset

- at second 60 injections accumulation:  11.9k counts
  saved to /data.local2/2025_229Th/2025_229ThDAQ1_0102.lmd with DAQ1
  again observed that the late contaminant om the particle detector (see attachment 3) is more prominent in at the beginning.
  This also manifests in an early steep fall off on the counts/s, see attachment 4
Entry  Sun Jun 29 18:04:11 2025, Carsten, Simon, Runs, Late shift 29.06. 
Time Part. Det. Int SIS-intensity accumulations in ESR cooler-HV PTB K250 cooler-HV PTB red cooler-HV orange LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time (laser shots) n loops checked laser position comments
18:10 18000 2.8e9 30 104525 - - 287 314 548.51 - 548.01 201 15 3 yes Part. Det Int. after 3 loops
18:26 14000 2.8e9 30 104525 - - 288 315 548.51 - 548.01 201 15 2 yes Part. Det Int. after 2 loops
18:41 16000 2.8e9 35 104525 - - 289 316 548.51 - 548.01 201 15 2 yes -
18:57 21000 2.0e9 40 104525 - - 290 317 548.51 - 548.01 201 15 3 yes -
19:14 13000 2.0e9 40 104525 - - 291 318 548.51 - 548.01 201 15 3 yes -
19:53 15000 2.6e9 35 104525 - - 292 319 548.51 - 548.01 201 15 2 yes -
20:08 15000 2.0e9 40 104525 - - 293 320 548.51 - 548.01 201 15 2 yes -
20:25 19000 2.4e9 40 104525 - - 294 321 548.51 - 548.01 201 15 3 yes -
20:44 3000 2.4e9 45 104525 - - 295 322 548.51 - 548.01 201 15 2 yes very low intensity, so we kept Go4 off and only ran mbs
21:04 23000 2.8e9 30 104525 - - 296 323 548.51 - 548.01 201 15 3 yes
21:21 24000 2.8e9 30 104525 - - 297 324 548.51 - 548.01 201 15 3 yes
21:39 18000 2.8e9 30 104525 - - 298 325 548.51 - 548.01 201 15 3 yes
21:57 6400 2e9 40 104525 - - 299 326 548.51 - 548.01 201 15 2 yes low intensity, no Go4


All Particle detector values are after the respective number of scans.

AT 1700 Bernd came in since we couldn't properly accumulate. Bernd adjusted the deceleration HF and slightly changed the extraction kicker from 1.9mrad to 1.7 mrad. High intensity of 229/89.

At 19:35 again problems with transmission SIS -> ESR. Put the Kciker (GTE3Mu1/HIKICK) back to 1.9 mrad (orignal value).
Entry  Thu Jun 26 19:01:46 2025, Carsten, Jan, Simon, Runs, Late shift 26.06. 

Scan parameters:
scan range551.91 - 549.41
positions501
step width0.005nm
advance after15 rising edges
iterations1
rev. frequency 1.54315e6




Time Part. Det. Int SIS-intensity accumulations in ESR cooler-HV PTB K250 cooler-HV PTB red cooler-HV orange LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time (laser shots) n loops checked laser position mesytec rate ok comments
19:15 23000 1.5-3.5 e9 33 104522 104526 104526 142 170 551.91 - 549.41 501 15 1 yes yes TE slits at -10 and +10, SIS intensity unstable
19:41 19000 3.3e9 30 104522 104526 104526 143 171 551.91 - 549.41 501 15 1 yes yes TE slits at -7 and +7
20:08 17500 3.3e9 25 104522 104526 104526 144 172 551.91 - 549.41 501 15 1 yes yes -
20:24 7000 3.3e9 25 104522 104526 104526 145 173 551.91 - 549.41 501 15 1 yes yes low Part. Det. Int maybe due to wrond position of part. det.
21:19 7000 3.5e9 25 104522 104526 104526 146 174 551.91 - 549.41 501 15 1 yes yes TE slits at -8 and +8
21:40 14000 3.3e9 25 104522 104526 104526 147 175 551.91 - 549.41 501 15 1 yes yes adjusted particle detector position
22:04 19500 3.3e9 40 104522 104526 104526 148 176 551.91 - 549.41 501 15 1 yes yes


After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted.
Entry  Fri Jun 27 06:55:45 2025, Esther. Konstantin, Laser, Laser-ion overlap 2025-06-27_06-52-01-807.png2025-06-27_07-12-27-135.png2025-06-27_07-32-28-599.png2025-06-27_07-51-52-620.png
We cross-checked the laser-ion superposition by using the T-scrapers. We did it only after the scans in order to do not change the duty cycle by much. Attachements are in the order of the measurements compiled in the table.

motorion beam position (mm) from inside
GECEDS1VU -0,5
GECEDS1HA -4
GECEDS2VU 0.9
GECEDS2VU 0


Still looks fine to us.
Entry  Fri Jun 20 15:58:21 2025, Wilfried, Konstantin, Simon, Laser, Laser stepwidth 
From previous Bi-Beamtimes, we expect the width of the transition to be >50GHz (FWHM). The current step width is 0.002nm which corresponds to 8GHz, so the laser would be in resonance for 6 steps.
We should aim for at least 5 steps in resonance.

line width (sigma) >20GHz
FWHM ~50GHz
step width (fundamental) 0.004nm (4GHz)
step width (doubled) 0.002nm (8GHz)
Entry  Sat Jun 21 02:32:24 2025, Alex, Anton, Rodolfo, Hiroshi, Carsten, Laser, Laser scanning setting 
This is the initial scanning range
  • 3 Loops
  • 501 Steps
  • 15 Shots / Step
  • From 550.4 nm to 547.9 nm. In total 2.5 nm @ 0.005 nm

Test file on DAQ2 is 0132.lmd / 0133.lmd
Entry  Sat Jun 21 07:28:35 2025, Arya, Julien, Wilfried, Laser, Laser Timing Bildschirmfoto_IonBunch_PDs_2025-06-21.pngBildschirmfoto_rough_timing_2025-06-21_07-22-59.pngBildschirmfoto_timingcorrected_2025-06-21_08-21-53.pngBildschirmfoto_timing_final_2025-06-21_08-46-18.png
We roughly adjusted the timing based on attachment 1, for ions to be in center of electron cooler (east arc), corrected for PMTs being 2.5m closer to the north on the west arc
then did a new 30 injection accumulation (10.8k counts on particle detector) and readjusted the timing to fine tune this
(attachment 2, first settings right peaks, adjusted towards left)

Then we did a 60 injection accumulation (only 7.6k counts?) to check the settings and shift the timing by half a revolution (324 ns = 98 bins)
(attachment 3, data saved in /data.local2/2025_229Th/2025_229ThDAQ1_0099.lmd with DAQ1).

Checked these settings and particle counts with another injection (2.8k counts)
(attachment 4, data saved in /data.local2/2025_229Th/2025_229ThDAQ1_0100.lmd with DAQ1)


Estimated bunch width: 20bins = 67ns;corresponds to 20/200 of 1 evolution, so about 12m
                                     assuming v+0.5c0: 11m
Entry  Wed Jun 18 11:44:15 2025, Julien, Konstantin, Stefan, Rima, , Jitter CFD 
Pulses on photodiodes in entry 20 

1. seem wide, so we also looked at qswitch signal; also seemes wide.
2. Have a tail

1.:

We investigated this and came to the conclusion, that the CFD is the issue:
- On an oscilloscope, when looking at both HF signal (CFD input) and CFD output, there is a 2ns jitter.
- The same jitter is visible when looking at the CFD output when triggering on a freq. generator signal with same frequency and amplitude as HF signal (1.54263 MHz; 1V peak to peak).
- There is not significant jitter when comparing the CFD output and the qswitch signal

We tried differents CFD (also a leading edge CFD / different thresholds), but this did not significantly improve what we saw on the oscilloscope.
Applying a 10MHz filter to the HF signal only made the jitter worse.
There might be an option to get a logic (NIM) signal from ESR instead of the sin-signal
Alternatively, one could think about using a PLL.
Note: the 2ns might also be limited by the trigger of the oscilloscope. We saw that if the trigger of the oscilloscope is not on the steepest part of the HF signal, the jitter "gets worse".

2.:

We saw on the qswtich signal, that there are instances, when  the q-switch signal is too early; 
Ww will try to find out which NIM module between the CFD and the final qswitch signal causes this
Preliminary: It's coincidance crate
Entry  Wed Jun 18 14:01:40 2025, Julien, Konstantin, Stefan, Rima, , Ion bunch on PMTs Bildschirmfoto_bunch_on_PMT_2025-06-18_14-11-44.png
Apparentlty did not work yesterday evening (no signal), but looks fine now (see attachment)
Entry  Tue Jun 17 16:33:16 2025, Carsten, Simon, DAQ, Gate Q-swtich 
The gate generator of the Q-switch signal for the Vuprom TDC was set to 10ms.
Entry  Mon Jun 23 10:52:34 2025, Julian, Lukas, Zsolt, Nadine, Konstantin, Laser, Frequency conversion unit 
We exchanged the old BBO of the Cobra's FCU with a new one. The output power of the Cobra was around 120mJ. In the end, we achieved a UV power of about 21mJ. In addition, we optimized the fiber-coupling to the wavemeter. 