ID |
Date |
Author |
Category |
Subject |
Text |
 |
54
|
Fri Feb 28 17:18:34 2025 |
Tamina | General | Laserwellenlängen zur Orientierung | Minimale Wellenlänge: IR 1028,7 nm --> 257,175
nm im UV (DFB Temperatur sollte ca. be 16,5
°C sein)
|
|
53
|
Fri Feb 28 16:46:51 2025 |
Tamina, Jens | Calibration | Synchronisation von Laserpulsen mit Bunchingfrequenz | Die Syncronisation der Reprate des Lasers
mit der Bunchingfrequenz wird mit einem Oszi
zusammen aufgenommen/gemessen.
|
 |
52
|
Fri Feb 28 15:53:44 2025 |
Konstantin | Accelerator | HF Signal for Bunching | The HF Bunching Frequency is: 1.7838817 MHz
Given (read) by the ESR operators in HKR. |
|
51
|
Fri Feb 28 15:30:41 2025 |
Tamina, Rodolfo | Detectors | Fast PD Nord West | We had an issue with the PD NW. The signal
after the Rate Divider was not OK.
|
|
50
|
Fri Feb 28 13:53:06 2025 |
Tamina, Rodolfo | Runs | Ionenbunch | Signal of PMT-Mitte. Recorded with an unstable
ion bunch. With approx. 3E8 Ions
|
 |
Draft
|
Fri Feb 28 08:57:30 2025 |
| General | | 28.2.25
Morning shift Aufgaben(nach Sebas:
|
|
48
|
Fri Feb 28 05:53:59 2025 |
Denise, Sebastian, Benedikt | Calibration | Laserstrahllage mit Scraper | Laserstrahl mit Hilfe der Blenden eingestellt,
dann mit Scrapern vermessen
|
|
47
|
Fri Feb 28 04:47:39 2025 |
Denise, Sebastian, Benedikt | Calibration | Gentec ESR Kalibration | Unsicherheiten Gentecs im allgemeinen bei
etwa 6.5%
|
|
46
|
Fri Feb 28 00:44:31 2025 |
Denise, Sebastian | DAQ | Shutter auf - Shutter zu Go4 Signale | Shutter auf (TDC Kanal 12 / Go4 Kanal TrS11_Shutter
auf)
Shutter zu (TDC Kanal 13 / Go4 Kanal TrS12_Shutter |
|
45
|
Fri Feb 28 00:35:46 2025 |
Benedikt | General | Anleitung Wellenlängenscan | Es folgt eine Anleitung, wie die Wellenlänge
des Lasersystems mit der PulsedLaserGUI gesteuert
werden kann.
|
|
44
|
Fri Feb 28 00:19:04 2025 |
Konstantin, Sebastian | Detectors | PMT power supply | We checked the PMT signal inside the cave
by using the local HV power supply. If one
needs to repeat the measurement at another |
|
43
|
Fri Feb 28 00:18:16 2025 |
Tamina | General | Kalibration UV Monitor im Laserlabor | Der UV Monitor steht in der Reflexion des
PBSC auf dem Lasertisch im Laserlabor.
|
|
42
|
Thu Feb 27 23:01:26 2025 |
Danyal, Konstantin, Sebastian | Calibration | Delaytime measurement | Yesterday, we forgot to take the transit time
of the PMT's into account when we set up
the delay times of the detectors PMT middle, |
|
41
|
Thu Feb 27 21:23:02 2025 |
Jens | DAQ | Python Data Channels | Python variables and names in DAQ
Because the Name of variables in DAQ doesnt
match the the real variables, here is List |
|
40
|
Thu Feb 27 20:21:31 2025 |
Sebastian | General | 10 MHz Ref Clock (Butis) | 10 MHz Referenz Clock von BUTIS mit AFG als
externl clock verbunden.
Wird benötigt das die Bunch Kavität (DDS |
|
39
|
Thu Feb 27 16:30:57 2025 |
Konstantin und Rodolfo | DAQ | HF Signal | We have exchanged the HF-Signal in the DAQ
Electronic: from Dummy-HF to the real HF
from accelerator. |
|
38
|
Thu Feb 27 11:19:55 2025 |
Danyal | Accelerator | pictures of the ion beam from 26.02.2025 | Hier sind noch ein paar Bilder vom Ionenstrahl
, wie er gestern Abend war. |
  |
37
|
Thu Feb 27 10:52:32 2025 |
Rodolfo | Detectors | Taking PMTs into operation | [B]PMT Nord[/B] the power supply (CH0) is
giving an alarm already by an applied voltage
of -1707 V (operation voltage is -2450 V). |
 |
36
|
Thu Feb 27 09:59:08 2025 |
Konstantin | Calibration | HV-Divider | The readout of the HV divider runs on a RaspberryPi
inside the HV-cave of the electron cooler.
Starting from 2025-02-27 09:42, the divider |
|
35
|
Thu Feb 27 09:38:04 2025 |
Konstantin & Danyal | Detectors | Laufzeit Messungen | Gestern Abend haben wir die Signal Laufzeiten
im gemessen:
|
|