55
Sun Mar 15 19:46:34 2020
Laszlo
Calibration
run042 - Xray35 calib Ba133 - high rate
Detector: 35
Source: Ba133, strong source
Distance: 334mm
56
Sun Mar 15 20:26:08 2020
Laszlo
Calibration
run043 - Xray35 calib Am241
Detector: 35
Source: Am241
Distance: 334mm
58
Sun Mar 15 21:01:23 2020
Laszlo
Calibration
run044 - Xray35 calib Ba133 - a night long run
Detector: 35
Source: Ba133
Distance: 334mm
63
Thu Mar 19 09:43:28 2020
Laszlo
General
Status of the ESR
After two days of tuning the beam, yesterday
evening (~8pm), it was found out that within
one ESR pattern using the new control system
65
Thu Mar 19 19:14:40 2020
Laszlo
General
primary beam measurement
Now our aim is to decelerate the primary beam
(naked 124Xe) down to 10MeV/u. At this energy
we can test our detectors and that how the
66
Thu Mar 19 23:48:56 2020
Laszlo
General
124Xe primary beam at 10Mev TargetOFF
Beam lifetime is estimated to be 11sec at
least (hard to.see with the cursor the corrct
values on the shottky monitor)
67
Fri Mar 20 01:11:58 2020
Laszlo
General
Target - beam jntersection
We have managed to find the target position
with the beam by looking at the xray spectra.
The trick was that at 10mev we
68
Fri Mar 20 01:21:00 2020
Laszlo
General
target is switched on event based
70
Fri Mar 20 01:36:48 2020
Laszlo
General
124Xe primary beam at 10Mev TargetON
We didmt see any effect on the lifetime if
the target is switched on. It is because
we use only hydrogen target (but why
84
Fri Mar 20 16:48:00 2020
Laszlo
General
flipped Si shaper output in x and y
Yesterday evening we have found out that the
energy signals from the silicon are negative
--> wrong polarity of the shapers was chosen
89
Fri Mar 20 22:51:36 2020
Laszlo
General
Target on signal change
From the previous settings:
JetOn
mashine number 11
90
Fri Mar 20 22:54:30 2020
Laszlo
General
First view of the beam
Intensity from Sis 1.7E9
The intensity is too.low at ESR to be visible.
However we could.catch it somehow with Shottky
91
Fri Mar 20 23:43:26 2020
Laszlo
General
Vacuum while moving in the detector at yesterday evening
Vacuum before DSSSD movement: 1.5844E-11
Top value after movement: 3.64378E-11
92
Sat Mar 21 01:25:07 2020
Laszlo
General
Vacuum while moving in the detector for fragment measurement
Stayed below E-10
Start:2.42E-11
Top after moving: 4.144E-11
93
Sat Mar 21 01:44:57 2020
Laszlo
General
from the SIS side they could optimize the
intensity. we gained a factor ~3. Now the
SIS intensity is at
97
Sat Mar 21 02:36:44 2020
Laszlo
General
detector and scraper position
For the 124Xe primary beam measurement:
-with the DSSSD we scraped the beam at position
-40
118
Sat Mar 21 11:25:47 2020
Laszlo
possible misalignement between target and fragment beam
after 9hours of measurement we see only 19
counts on our targetON silicon detector histo.
this low count rate can indicate two scenarios:
119
Sat Mar 21 11:30:04 2020
Laszlo
General
MWPC movement
now we try to move the capture detector (MWPC
in the 1. dipol before our Si setup) inside
the ring while the DAQ was ON
123
Sat Mar 21 12:06:50 2020
Laszlo
General
run61 - SCRAPER moved out
we want to see any change on the Silicon count
rate, because we will have now also Rutherford
scattering.
125
Sat Mar 21 12:32:00 2020
Laszlo
General
Estimated target thickness
Target thickness must be like 5mm (was measured
with a moving thin wire)
The beam thickness is small because of the