Fri May 14 11:17:48 2021 |
Jan Glorius | Calibration | run0043 - Xray145 calib 210Pb d=305mm | 2021 |
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Threshold slightly increased
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 11:17:23 - 14.05.2021
Stop time: 13:47:42 - 14.05.2021
file name: e127b_run0043.lmd
avrg. rate: 40Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
341
|
Fri May 14 13:58:52 2021 |
Jan Glorius | Calibration | run0044 - Xray145 calib 210Pb d=305mm | 2021 |
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
90deg TFA diff set from 50 to 200.
Also threshold is slightly increased.
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 13:58:33 - 14.05.2021
Stop time: 16:46:00 - 14.05.2021
file name: e127b_run0044.lmd
avrg. rate: 40Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
342
|
Sat May 15 20:21:59 2021 |
Jan Glorius | Calibration | run0045 - Si pulser run | 2021 |
Detector: DSSD
Source: CAEN Detector Emulator
Start time: 20:24:14 - 15.05.2021
Stop time: 21:01:34 - 15.05.2021
file name: e127b_run0045.lmd
avrg. rate: 1000Hz
dead-time: 7.5%
Si bias test cycle:
0V > 50V/200uA > 100V/280uA > 150V/370uA > 120V/320uA > 90V/260uA > 60V/210uA > 30V/160uA > 0V
This cycle is run twice i the first 15 minutes and shall be used for calibration of the scaler channels.
Also ESR trafo is running with beam. |
343
|
Sun May 16 11:25:25 2021 |
Jan Glorius | Calibration | run0046 - Xray E-calib 241Am all dets | 2021 |
Detectors: Xray1 (GEM1800), Xray2 (GLP), Xray3 (GEM3000)
Sources: different Am241 at each detector
Start time: 11:25:04 - 16.05.2021
Stop time: 11:29:45 - 16.05.2021
file name: e127b_run0046.lmd
avrg. rate: 6000Hz
dead-time: 30% |
344
|
Sun May 16 11:54:12 2021 |
Jan Glorius | Calibration | Xray performance in lab | 2021 |
These are the current Xray detector performances in the lab using data from run0046
det | DAQ ch | threshold | resolution @ 60keV
|
GEM1800 | xray1 | 5keV | 495eV
|
GLP2000 | xray2 | 5keV | 430eV
|
GEM3000 | xray3 | 7keV | 525eV
|
|
345
|
Sun May 16 12:55:48 2021 |
Jan Glorius | DAQ | event server lxg1299:6003 | 2021 |
new event server is set up at lxg1299 on port 6003
It runs as:
/data.local1/gstore/mrevserv64 lxg1275 |
347
|
Tue May 18 01:03:59 2021 |
Jan Glorius | Calibration | run0048 - calib all Xray det. | 2021 |
Efficiency calibration in ESR
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detector: all
Source @ 90deg: 210Pb
Source @ 35deg: 133Ba_hi
Source @ 145deg: 241Am_hi
Distance: each
Start time: 01:20:23 - 18.05.2021
Stop time: 11:05:30 - 18.05.2021
file name: e127b_run0048.lmd
avrg. rate: 400Hz
dead-time: 3% |
348
|
Tue May 18 01:21:27 2021 |
Jan Glorius | Calibration | run0047 - crap | 2021 |
|
349
|
Tue May 18 11:22:57 2021 |
Jan Glorius | Calibration | run0049 - calib all Xray det. | 2021 |
Efficiency calibration in ESR
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detector: all
Source @ 90deg: 210Pb
Source @ 35deg: 133Ba_hi
Source @ 145deg: 241Am_hi
Distance: each
Start time: 11:22:36 - 18.05.2021
Stop time: 13:36:55 - 18.05.2021
file name: e127b_run0049.lmd
avrg. rate: 400Hz
dead-time: 3% |
351
|
Wed May 19 13:47:26 2021 |
Jan Glorius | General | GasTarget Operation | 2021 |
We have two tasks to take over for the target operation:
1. We have to refill the blue LN2 bottle on the roof of ESR every morning (see 1st attached foto)
For this we use the small silver dewar on the floor there. In case this dewar is empty we have to refill it in the yard.
No need to stop the measurement for this.
2. We have to monitor the LN2-level warning on Ulis web page: http://ulpc24.gsi.de/mediawiki/index.php
In case there is a warning (~12h remaining), we have to refill the big dewar at the target in ESR.
This means we need ESR on TSG with the gate open.
In case the nozzle is clogging and we do not reach reasonable densities anymore and E1 pressure is going down, we have to heat up the nozzle manually.
This can be done at the control panel on the roof of ESR (see 2nd attached foto).
The proceedure is:
1. push "set point"
2. type new value in Kelvin (e.g. 100K for start of heating)
3. press "enter" to apply or "escape" to correct typo
4. monitor E1 pressure (should go up)
Overview page:
https://popps.org/cgi-bin/jet.py?gasart=Wasserstoff |
563
|
Sun May 30 10:24:20 2021 |
Jan Glorius | Calibration | run0212 - X-calib. 90-pb 145-am | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145
Source @ 90deg: 210Pb
Source @ 145deg: 241Am_hi
Distance: each
Start time: 10:24:55 - 30.05.2021
Stop time: 11:19:29 - 30.05.2021
file name: e127b_run0212.lmd
avrg. rate: 250Hz
dead-time: 2% |
564
|
Sun May 30 11:32:06 2021 |
Jan Glorius | Calibration | run0213 - X-calib. 90-am 145-pb 35-ba | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145 + 35
Source @ 90deg: 241Am_hi
Source @ 145deg: 210Pb
Source @ 35deg: 133Ba_hi
Distance: each
Start time: 11:31:47 - 30.05.2021
Stop time: 13:10:14 - 30.05.2021
file name: e127b_run0213.lmd
avrg. rate: 700Hz
dead-time: 6% |
565
|
Sun May 30 13:23:26 2021 |
Jan Glorius | Calibration | run0214 - X-calib. 90-ba 145-ba 35-am | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145 + 35
Source @ 90deg: 133Ba_lo
Source @ 145deg: 133Ba_hi
Source @ 35deg: 241Am_hi
Distance: each
Start time: 13:22:38 - 30.05.2021
Stop time: 14:28:09 - 30.05.2021
file name: e127b_run0214.lmd
avrg. rate: 1150Hz
dead-time: 8% |
566
|
Sun May 30 14:38:09 2021 |
Jan Glorius | Calibration | run0216 - X-calib. 90-ba 35-pb | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 35
Source @ 90deg: 133Ba_lo
Source @ 35deg: 210Pb
Distance: each
Start time: 14:46:46 - 30.05.2021
Stop time: 20:11:08 - 30.05.2021
file name: e127b_run0216.lmd
avrg. rate: 330Hz
dead-time: 3% |
574
|
Thu Jun 24 09:43:01 2021 |
Jan Glorius | DAQ | Unpacker | 2021 |
The unpacker used for the production runs of E127b is located at
/u/litv-exp/unpacker/unpackexps/e127_TO_2021_with_ECL
It is also aliased for litv-exp user as: E127_unpack
This is for the data structure as documented here:
https://elog.gsi.de/esr/E127/346
As of now, this is compiled for ROOT v6-18-04 located here:
/cvmfs/csee.gsi.de/root/618-04/bin/root |
2
|
Thu Jan 17 08:45:43 2019 |
Jan | Detectors | Micron W1 DSSSD - technical information | |
Attached are the spec sheets and design drawings of the new DSSSDs from Micron.
Additionally, the 3D model made by Laszlo is also attached as stp and dwg. |
5
|
Fri Apr 5 05:46:20 2019 |
Jan | DAQ | running DRASI for (p,g) | |
We now have a drasi DAQ running for our (p,g) experiments.
The RIO we have it installed right now is: r4l-58
It shall be used with the user <litv-exp>
The DAQ-controle shall be done from lxg1275 inside a SCREEN SESSION named "daq"
tab 0 >> drasi
tab 1 >> logger
tab 2 >> rate
tab 3 >> stream server
tab 4 >> file writing (optional)
To check and control running daq:
login on lxg1275
> ssh litv-exp@lxg1275
enter screen session
> screen -x daq
DRASI
go to tab 0
> ctrl-a 0
should be quite, no errors showing, no terminal outputs
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./start.sh (for det. triggered daq)
> ./pulser.sh (for internal pulser trigger)
> ctrl-c (for killing daq)
LOGGER
go to tab 1
> ctrl-a 1
should be quite, no errors showing, no terminal outputs
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./logger.sh
RATE MONITOR
go to tab 2
> ctrl-a 2
rates should update each second
"incr" should be event rate per sec
"send" column should be reasonable value if data is written
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./rate.sh
EMPTY STREAM SERVER
go to tab 3
> ctrl-a 3
empty stream should be running
number of processed events should keep increasing
you should see the number of clients connected
full restart proceedure (on lxg1275)
> empty stream://r4l-58 --server=stream:6002
WRITE A FILE
can be done by litv-exp from any lgx-machine
ROOT FILE:
> e127_unpack --stream=lxg1275 --ntuple=RAW,/data.local1/e127/pre_data/file.root
LMD FILE:
> e127_unpack --stream=lxg1275 --output=/data.local1/e127/pre_data/test.lmd
|
6
|
Fri Apr 5 06:05:55 2019 |
Jan | DAQ | making changes to the DAQ | |
If you need to add or change a module of the VME crate you need to follow this proceedure:
1. stop DRASI
2. add/change module in esrdaq_2018/r4l-58/main.cfg
a. add module name and VME address: CAEN_V830 (0x00A00000) {}
b. use BARRIER keyword between the modules
3. restart DRASI (e.g. ./pulser.sh)
4. check empty unpacker data output
> empty --stream=r4l-58 --print --data
should run without error and now have your new module event data
5. adjust unpacker in ~/e127/unpackexps/esr
a. make sure you have a .spec file defining readout of you module, e.g. "vme_caen_v775.spec"
b. edit main_user_esr.spec
- add module as e.g. "tdc[0] = VME_CAEN_V775(geom=0x2, crate=0);"
- select "geom=0x1" value according to ascending order of all modules (0x1, 0x2, 0x3, ...)
c. edit det_mapping.hh
- add SIGNAL() keyword for your module
- e.g.: SIGNAL( TDC_1, esr.tdc[0].data[0],TDC_32, esr.tdc[0].data[31], DATA12);
- choose data format, e.g. "DATA12", as specified in module .spec-file
d. compile your unpacker:
> make -j
e. try it out:
> ./esr --stream=r4l-58 --print --data
- should run without errors
> ./esr --stream=r4l-58 --ntuple=RAW,/dir/file.root
- root-file should have your new branches with your module data |
7
|
Wed Jun 5 11:29:13 2019 |
Jan | Detectors | position/channel calib. | |
This is the first position calibration done with the detector installed in the holder. The goal is to assign rough positions (top/bottom/left/right) to PreAmp- and Shaper-channels.
PreAmp Channel | polarity | position
|
1-1 (Y-1) | neg. | bottom
|
1-16 (Y-16) | neg. | top
|
2-1 (X-1) | pos. | left (close to beam)
|
2-16 (X-16) | pos. | right (far from beam)
|
|
8
|
Fri Jun 7 10:18:23 2019 |
Jan | Runs | prerun001 - MixedAlpha | |
save point: lxg1275:/datalocal1/e127/predata/prerun001-root
~2h run
DAQ setup:
32 ADC channels
ch. 1-16 (section 0) >> y-strips
ch. 17-32 (section1) >> x-strips
32 TDC channels
ch. 1-16 (section 0) >> y-strips
ch. 17-32 (section1) >> x-strips
32 scaler channels (not used)
16 MDPP channels (not used)
Detector setup:
bias -90V
MixedAlpha source ~in central position |