  253   Mon Mar 23 15:48:15 2020 LaszloCalibrationrun105 - Xray90 calib Am241, distance 167.5mm, wrong? Detector: 90
Source: Am241 
Distance: 167.5mm
  
  254   Mon Mar 23 16:17:58 2020 LaszloCalibrationrun106 - garbage! run106 was clicked by accident, please delete!   
  255   Mon Mar 23 16:21:33 2020 LaszloCalibrationrun107 - Xray90 calib Am241, distance 167.5mm Detector: 90
Source: Am241 
Distance: 167.5mm
  
  256   Mon Mar 23 17:32:42 2020 LaszloCalibrationrun108 - Xray90 calib Ba133, distance 167.5mm - high rate - deadtime too high! Detector: 90
Source: Ba133 - strong source
Distance: 167.5mm
  
  257   Mon Mar 23 17:40:33 2020 LaszloCalibrationrun109 - Xray90 calib Ba133, distance 167.5mm -low rate Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
  
  259   Mon Mar 23 18:17:27 2020 LaszloCalibrationrun110 - Xray90 calib Ba133, distance 167.5mm -low rate - LESS NOISE Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
  
  267   Wed Mar 25 14:26:22 2020 LaszloCalibrationrun111 - Xray90 calib Am241, distance 167.5mm - LESS NOISE Detector: 90
Source: 241Am
Distance: 167.5mm
  
  260   Mon Mar 23 19:32:53 2020 LaszloCalibrationrun112 - Xray90 calib Am241, distance 200mm - LESS NOISE Detector: 90
Source: 241Am
Distance: 200mm
  
  261   Mon Mar 23 20:52:37 2020 LaszloCalibrationrun113 - Xray145 calib Am241, distance 305mm - LESS NOISE Detector: 145
Source: 241Am
Distance: 305mm
  
  262   Mon Mar 23 21:47:40 2020 LaszloCalibrationrun114 - Xray145 calib Ba133, distance 305mm - LESS NOISE Detector: 145
Source: 133Ba - strong source
Distance: 305mm
  
  263   Mon Mar 23 22:25:41 2020 LaszloCalibrationrun115 - Xray35 Ba133 d=334mm, Xray90 Lead source, d=167.5mm Simultaneous measurement of Xray35 and Xray90.
The dead-time is <3%
  
  264   Tue Mar 24 00:11:35 2020 LaszloCalibrationrun116 - Xray35 calib AM241, distance 334mm Detector: 35
Source: 241Am
Distance: 334mm
  
  123   Sat Mar 21 12:06:50 2020 LaszloGeneralrun61 - SCRAPER moved out we want to see any change on the Silicon count
rate, because we will have now also Rutherford
scattering. 
  
  5   Fri Apr 5 05:46:20 2019 JanDAQrunning DRASI for (p,g) We now have a drasi DAQ running for our (p,g)
experiments.
The RIO we have it installed right now is:
  
  251   Mon Mar 23 14:32:03 2020 JanAnalysisruns_100to104 NO-SCRAPER RESULTS Without scraper, we have very few (p,g) counts,
hard to get numbers quickly, maybe ~50.
 124Xe_Si_map.png124Xe_Si_projection.png124Xe_x-ray_35.png124Xe_x-ray_90.png124Xe_x-ray_145.png 
  244   Mon Mar 23 13:26:59 2020 JanAnalysisruns_90to99 SCRAPER RESULTS With scraper, we have roughly 150-200 (p,g)
counts now.
 124Xe_Si_map.png124Xe_x-ray_35.png124Xe_x-ray_90.png124Xe_x-ray_145.png124Xe_Si_projection.png 
  30   Thu Dec 5 10:57:57 2019 Jan, LaszloDetectorssetting up the DSSD with beam This is a brief how-to about getting the DSSSD
set up with stored beam and target in the
ESR.
  
  360   Fri May 21 01:17:37 2021 LaszloAnalysissimple analysis code2021//a simple code (template) for offline analysis
//made by Laszlo, serves as a simple demonstration
for enthusiastic shifters
  
  22   Wed Sep 25 13:03:46 2019 LaszloAnalysissimulations on 118Te+p I have made some simulations at 8AMeV, 7AMeV
and 4AMeV energies for the main 118Te + p
reaction channels. For the
 11x 
  64   Thu Mar 19 15:09:29 2020 JanGeneralstarting off with DSSSD here is my short guide how to start with the
measurement as soon as we have a beam available.
  