  proton-capture on 118Te, Page 2 of 29  Not logged in ELOG logo
Entry  Sat May 1 22:34:16 2021, Jan Glorius, Calibration, efficiency xray 90deg, 2021 90deg_eff.png90deg_eff.dat
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Entry  Mon May 3 14:55:44 2021, Jan Glorius, Calibration, Xray energy calib. table, 2021 
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.

E [keV] xray90 (GEM1800) xray35 (GLP) xray145 (GEM3000)
lab (28.04.21) lab (04.05.21) lab (11.05.21)
59.54 2402.45 ch 2399.07 ch 2423.24 ch
26.34 1107.38 ch 1116.65 ch 1127.94 ch
13.95 624.33 ch 638.38 ch 645.39 ch
slope A 0.0256387 keV/ch 0.0258923 keV/ch 0.0256408 keV/ch
offset B -2.05477 keV -2.57638 keV -2.59114 keV
Entry  Tue May 4 14:31:51 2021, Jan Glorius, Calibration, run0012 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:31:24 - 04.05.2021
Stop time:  14:42:13 - 04.05.2021

file name: e127b_run0012.lmd
avrg. rate: 25 Hz
dead-time:  0%


Threshold < 10keV

The scale serial communication is off.

Stopped due to noise rate...
Entry  Tue May 4 14:34:51 2021, Jan Glorius, Detectors, Xray weight control induces signals, 2021 
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command sent.

The solution for now is, that we activate weight control NOT DURING MEASUREMENT!
Entry  Tue May 4 14:46:15 2021, Jan Glorius, Calibration, run0013 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:45:51 - 04.05.2021
Stop time:  15:47:48 - 04.05.2021

file name: e127b_run0013.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 16:04:42 2021, Jan Glorius, Calibration, run0014 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 16:04:21 - 04.05.2021
Stop time:  20:41:45 - 04.05.2021

file name: e127b_run0014.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 21:18:15 2021, Jan Glorius, Calibration, run0015 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 21:17:44 - 04.05.2021
Stop time:   8:42:09 - 05.05.2021

file name: e127b_run0015.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 5 10:02:27 2021, Jan Glorius, Calibration, run0016 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 10:02:07 - 05.05.2021
Stop time:  14:13:23 - 05.05.2021

file name: e127b_run0016.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 5 14:24:57 2021, Jan Glorius, Calibration, run0018 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:28:26 - 05.05.2021
Stop time:  21:05:00 - 05.05.2021

file name: e127b_run0018.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

There are many rate spikes in the first hour of measurement.
Entry  Wed May 5 14:29:02 2021, Jan Glorius, Calibration, run0017 crap run, 2021 
 
Entry  Wed May 5 21:22:27 2021, Jan Glorius, Calibration, run0019 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:22:06 - 05.05.2021
Stop time:   8:55:21 - 06.05.2021

file name: e127b_run0019.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Many rates spikes, check carefully!
Entry  Thu May 6 09:40:37 2021, Jan Glorius, Calibration, run0020 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 9:42:24 - 06.05.2021
Stop time: 14:03:44 - 06.05.2021

file name: e127b_run0020.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Thu May 6 11:03:46 2021, Jan Glorius, DAQ, Cabling Documentation, 2021 e127b_cabling_docu_06_05_2021_.pdfe127b_2021_cabling_documentation.xlsx
 
Entry  Thu May 6 13:39:02 2021, Jan Glorius, Calibration, efficiency xray 35deg, 2021 35deg_eff.png35deg_eff.dat
the attached efficiency curve for the 35 deg. xray detector is made with the runs
e127b_run0015.lmd
e127b_run0016.lmd
e127b_run0017.lmd + e127b_run0019.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Entry  Thu May 6 14:30:35 2021, Jan Glorius, Calibration, run0021 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:21:09 - 06.05.2021
Stop time:  21:06:10 - 06.05.2021

file name: e127b_run0021.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Thu May 6 21:15:06 2021, Jan Glorius, Calibration, run0022 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:14:44 - 06.05.2021
Stop time:  09:34:10 - 07.05.2021

file name: e127b_run0022.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Check for rate spikes
Entry  Tue May 11 14:11:40 2021, Jan Glorius, Calibration, run0024 - Xray145 calib Am241_hi d=305mm, 2021 
Efficiency calibration in the lab

Detector: GEM3000 - 145 deg
Source: 241Am_hi
Distance: 305mm
Start time: 14:10:55 - 11.05.2021
Stop time:  16:40:27 - 11.05.2021

file name: e127b_run0024.lmd
avrg. rate: 210 Hz
dead-time:  2%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Check for rate spikes
Entry  Tue May 11 16:49:23 2021, Jan Glorius, Calibration, run0025 - Xray145 calib Pb210 d=305mm, 2021 
Efficiency calibration in the lab

Detector: GEM3000 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 16:49:03 - 11.05.2021
Stop time:  09:21:13 - 12.05.2021

file name: e127b_run0025.lmd
avrg. rate: 50 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Check for rate spikes
Entry  Wed May 12 10:16:10 2021, Jan Glorius, Calibration, run0026 - Xray145 calib Ba133_hi d=305mm, 2021 
Efficiency calibration in the lab

Detector: GEM3000 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 10:15:51 - 12.05.2021
Stop time:  10:59:35 - 12.05.2021

file name: e127b_run0026.lmd
avrg. rate: 850 Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 12 11:29:23 2021, Jan Glorius, Calibration, run0028 - Xray90 calib Ba133_hi d=167.5mm, 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************

Detector: GEM3000 - 90 deg
Source: 133Ba_hi
Distance: 167.5mm
Start time: 11:32:50 - 12.05.2021
Stop time:  11:44:29 - 12.05.2021

file name: e127b_run0028.lmd
avrg. rate: 2500Hz
dead-time:  17%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.