ID Date Author Categoryup Subject
  166   Fri Apr 8 15:39:36 2022 allRunsfirst data after timing set up
started taking data again

started at 15:40  
h=1  

file names 
229Th7821 DAQ 1  607 MB
229Th0082 DAQ 2  820 MB 

stopped at 16:25 


The laser was not really properly scanned. 
This was mainly a test run to see if everything was working. 
  181   Sat Apr 9 06:25:22 2022 Max, KristianRunsmorning shift runs
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th

Shottky marker pos: 243.973MHz

TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)comments
xx:xxx.xe9209xxx00xx78xx32x.x - 32x.x0.01xsketch for filling the table
06:401.7e92096190098-00997836323.2 - 321.20.0024slaser pulse energy is @ ~12mJ after SHG
09:351.4e92096240101-01017838-7838323.2 - 322.80.0023sonly 0.5nm range, WM power + coupling tuned to scan range, Konstantin updated LV code to hold Cobra Set WL value, aborted LV readout problem
11:251.0e92096220102 - 1027839-7839323.2 - 322.80.0023sLV readout problem solved. WM readout is wrong (Cobra set is good)
12:15old beam2096220103 - 1037840-7840323.2 - 322.80.0023s WM readout is wrong
12:502.3e92096170104 - 1047841-7841323.2 - 322.80.0023s couldnt open go4 stream view, stopped after one good scan
13:202.0e92096170104 - 1057842-7842323.2 - 322.80.0023s first scan with LV correct
  183   Sat Apr 9 13:45:42 2022 Max, KristianRunsfirst scans with LV problem solved!
This is the first time with LV buffer problem solved (see Elog ID: 184). Before the PMT signals were out of sync with the wavelength data!
We concluded to scan with step sizes of 0.002nm, scan range of 0.5nm (f
nm), dwelltime = 150(5s), loops = 2 (approx.: 30min measurements).

DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th

Shottky marker pos: 243.973MHz

TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)comments
11:251.0e92096220102 - 1027839-7839323.2 - 322.80.0023sLV readout problem solved. WM readout is wrong (Cobra set is good)
12:15old beam2096220103 - 1037840-7840323.2 - 322.80.0023s WM readout is wrong
12:502.3e92096170104 - 1047841-7841323.2 - 322.80.0023s couldnt open go4 stream view, stopped after one good scan, WM readout is wrong .
13:202.0e92096170104 - 1057842-7842323.2 - 322.80.0023s first scan with LV correct
13:401.2e92096180106 - 1067843-7843323.2 - 322.80.0023s 3 loops
14:201.5e92096120107 - 1077844-7844323.2 - 322.80.0023s 1 loop


Runs after this have the right cobra set value but propably the wrong WM data until 10.04.22 12:44 (see also ID:198)
  185   Sat Apr 9 15:47:21 2022 Carsten, Rodolfo, Sarper, PatrickRunsAfternoon and night shift runs
After the morning shift, we got new increment (nm) for the laser scanning. We are now scanning 2 nm of wavelength range in 5 big steps (5 x 0.4 nm). Started with 323.2 nm - 322.8 nm range. Every big step (0.4 nm) is scanned with 0.002 nm of wavelength increment.

Every scan takes ~ 18 min (0.4 nm range with 0.002 nm steps), we have scanned 2 loops. Total scanning time ~36 min.

DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th

Schottky marker pos: 243.973MHz


TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)comments
xx:xxx.xe9209xxx0xxx78xx32x.x - 32x.x0.002xsketch for filling the table
15:141.5e92096240111-01117847-7847323.2 - 322.80.0025s DAQ 2 no VUPROM TDC Pat south
xx:xx1.5e92096240111-01117847-7847323.2 - 322.80.0025s DAQ 2 changed TDC part south from ch 8 to ch 10
17:052.2e92096240113-01137848-7848323.2 - 322.80.0025s successful scan
19:041.8e92096160116-0116 7850-7850 323.2 - 322.80.0025s Intensity optimization, see below, DAQ 2 TDC part south back to channel 8 - ok, XUV detector and Cu mirror moved correctly, lost the laser beam after the half of the second loop
19:491.8e92096140117-0117 7851-7851 323.2 - 322.80.0025s successful scan
20:282.0e92096220118-0118 7852-7852 323.2 - 322.80.0025s successful scan
21:072.2e92096230119-119 7853-7853 323.2 - 322.80.0025s successful scan
22:011.8e92096210122-0122 7856-7856 323.2 - 322.80.0025s successful scan
22:41?20962001237857323.2 - 322.80.0025s First scan started by night shift
23:211.1e920962001247858323.2 - 322.80.0025s mbs crashed at 'sta ac'
23:321.3e920961801257859323.2 - 322.80.0025s e-cooler turned off
23:501.5e9209xxx01267860323.2 - 322.80.0025s the stop point was not set
23:531.5e920961101277861323.2 - 322.80.0025s DAQ2 must have crashed between 23:30 - 23:50
00:071.9e9209xxx01287862323.2 - 322.80.0025s mbs crashed at 'sta ac'
00:231.7e920962301317864323.2 - 322.80.0025s successful scan. we 'resl' after every measurement from now on. 'open file' is included in '@startup' now (ID:190).
01:081.6e920962301327865323.2 - 322.80.0025s successful scan
01:471.6e920962301337866323.2 - 322.80.0025s successful scan
02:351.3e920962301347868323.2 - 322.80.0025s successful scan
03:192.0e920962201357869323.2 - 322.80.0025s successful scan
04:041.3e920962201367870323.2 - 322.80.0025s successful scan
04:441.0e920962201377871323.2 - 322.80.0025s e-cooler turned off
05:251.9e920962201407875323.2 - 322.80.0025s successful scan. DAQ1 was started ~7s before DAQ2. The files between the last measurement an this one are essentially empty.


Carsten (19:00h): I notized that a lot of beam gets lost when starting scraping. I called Sergey Litvinov to help me to introduce some additional cooling/waiting (10s) after the ramp and before scraping and bunching. It increased the number of ions available for the measurement roughly by a factor of 3-5.

Some runs have wrong WM data. Set Wavelength is good.

Following files are saved differently, the command missed the 229Th in the end (/data.local2/BeamTimeTh22/229Thor -auto -rfio)
7850 (19:32)
7855 (21:56)
7856 (22:30)
  195   Sun Apr 10 06:39:16 2022 Kristian, Marek, MaxRunsruns morning shift 10.04.
These are the measurement runs of the morning shift (10.04.).


TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)comments
06:xx3e82096xx0141-01417876-7876323.2 - 322.80.0025sjust one loop
06:392e92096210142-01427877-7877323.2 - 322.80.0025s
07:182.1e92096160143-01437878-7878323.2 - 322.80.0025s 8000 particles on counter
07:572.2e92096210144-01447879-7879323.2 - 322.80.0025s
08:371.9e92096200145-01457880-7880323.2 - 322.80.0025s 7200 particles on counter
09:242.1e92096210146-01467881-7881323.2 - 322.80.0025s laser pulse energy @ start of run: 488mJ green monitor, 11.6mJ UV, 7700 particles on counte
10:072.1e92096200147-01477882-7882323.2 - 322.80.0025s optimized WM coupling, the WM data is stretched in this run (buffer problem?), also possible for previous runs
11:042.1e92096200148-01487884-7884323.2 - 322.80.0025s restarted WM, still wrong WM data, stopped after 1 run
11:272.1e92096160149-01497885-7885323.2 - 322.80.0025s restarted WM + LV
12:112.0e92096160150-01507886-7886323.2 - 322.80.0025s wrong WM data
12:551.9e92096160151-01517887-7887323.2 - 322.80.0025s changed LV code (ID: 198), WM data works now!
13:371.8e92096200152-01527888-7888323.2 - 322.80.0025s
14:162.1e92096210153-01537889-7889323.2 - 322.80.0025s1 loop
  200   Sun Apr 10 15:16:19 2022 Carsten, Rodolfo, SarperRunsRuns afternoon shifts 10.04
These are the measurement runs of the afternoon shift (10.04.).

Wavelength range is changed from 323.2 - 322.8 nm to 322.81 - 322.40 nm.


TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)comments
16:261.95e92096200155-01557891-7891322.81 - 322.400.0025slaser scan did not work, labview (Thorium_main.vi) communication issue
16:511.8e92096200156-01567892-7892322.81 - 322.400.0025s
17:331.6e92096180157-01577893-7893322.81 - 322.400.0025s 7894 particles on the counter
18:171.6e92096180158-01587894-7894322.81 - 322.400.0025s 8200 particles on the counter
19:061.8e92096180159-01597895-7895322.81 - 322.400.0025s 8650 particles on the counter
19:441.9e92096180160-01607896-7896322.81 - 322.400.0025s 8600 particles on the counter
20:241.7e92096200161-01617897-7897322.81 - 322.400.0025s 4200 particles on the counter
21:041.7e92096190162-01627898-7898322.81 - 322.400.0025s 8380 particles on the counter
21:451.8e92096190163-01637899-7899322.81 - 322.400.0025s 8550 particles on the counter
22:241.8e92096190164-01647900-7900322.81 - 322.400.0025s 8016 particles on the counter
  203   Sun Apr 10 22:47:05 2022 Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, KristianRunsRuns night shift 10.04 + morning shift 11.04
These are the measurement runs of the night shift (10/11.04.).

Wavelength range is still 322.81 - 322.40 nm.


TimeSIS-intensitycooler-HVLMD-DAQ1 start-endLMD-DAQ2 start-endscan range(nm)step width(nm)dwell time(s)particles countedcomments
23:05 1.75e92096200165 - 01657901-7901322.81 - 322.400.0025s8500
23:57 1.?e92096200166 - 01667902-7902322.81 - 322.400.0025s8500The HV of PMT south was changed from 2.5kV to 3kV
00:35 1.4e92096200167 - 01677903-7903322.81 - 322.400.0025s4100
01:18 1.75e92096200168 - 01687904-7904322.81 - 322.400.0025s4400
01:58 1.4e92096200169 - 01697905-7905322.81 - 322.400.0025s7700 The ESR went red after this run. The on-call service was called because of some power supply failure, ETA: 45min.
04:12 1.75e92096200171 - 01717906-7906322.81 - 322.400.0025s7600 With the start of this run, the wavemeter, labview and the python script were restarted. The python script was slightly adjusted.
04:50 1.6e92096200172 - 01727907-7907322.81 - 322.400.0025s8300
05:31 1.6e92096190173 - 01737908-7908322.81 - 322.400.0025s8100
06:09 1.6e92096150174 - 01747909-7909322.81 - 322.400.0025s 8000
06:52 1.8e92096150175 - 01757910-7910322.81 - 322.400.0025s8800 at 07:08 the cooler voltage jumped from 209615 V to 209619 V
07:30 1.9e92096190176 - 01767911-7911322.81 - 322.400.0025s forgotten at 07:50 the cooler voltage jumped from 209619 to 209610 V and back again at to 209619 V
08:18 1.8e92096190177 - 01777912-7912322.81 - 322.400.0025s 9100 08:47 the cooler voltage jumped from 209618 to 209614 V and back again at to 209618 V
08:55 1.75e92096190178 - 01787913-7913322.81 - 322.400.0025s 8300
09:36 1.6e92096190179 - 01797914-7914322.81 - 322.400.0025s 8194


The SIS intensity is taken directly before injection to ESR. Not the SIS peak intensity as in yesterday's table.
  210   Mon Apr 11 14:45:11 2022 Paul, Konstantin, Sarper, DavidRunsAfternoon shift runs 11.04
We continue scan the new wavelength scan range is 322.41 - 322.00 nm



TimeSIS-intensitycooler-HVLMD-DAQ1 file nr LMD-DAQ2 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
14:45 1.9e9 209618 0189 7924 322.41 - 322.00 0.002 5s 9100
18:54 1.83e9 209616 0191 7926 322.41 - 322.00 0.002 5s 7300
19:34 1.7e9 209615 0192 7927 322.41 - 322.00 0.002 5s 7110
20:13 1.8e9 20961X 0193 7928 322.41 - 322.00 0.002 5s 6730
20:53 1.8e9 209614 0194 7929 322.41 - 322.00 0.002 5s 7550 During the scan the ecooler voltage changed to approximately 209613
21:43 1.7e9 209614 0195 7931 322.41 - 322.00 0.002 5s 7000
22:30 1.6e9 209612 0197 7932 322.41 - 322.00 0.002 5s 7300
23:10 1.4e9 209614 0198 7933 322.41 - 322.00 0.002 5s 6100 For this measurement, ESR_Ring went red, we realized only after completing the measurement. The sections went red are Dipole Correction Points (GE01KP17-18-19-20, 4 in total) @~23:45 and fixed @01:35...
01:52 1.6e9 209614 0199 7934 322.41 - 322.00 0.002 5s 7151
02:32 1.7e9 209616 0200 7935 322.41 - 322.00 0.002 5s 8368
03:22 1.8e9 209617 0201 7936 322.41 - 322.00 0.002 5s 7005
04:01 1.9e9 209616 0202 7937 322.41 - 322.00 0.002 5s Ecooler voltage increased 4 volts during the run
  215   Tue Apr 12 04:43:29 2022 Alexandre, Kristian, SebastianRuns322.01 - 321.60 nm
We scan the wavelength scan range is 322.01 - 321.60 nm

before the first scan of the new wavelength range we changed the high voltage divider ratio in Labview from 248514 to 248517, which is the correct value for this voltage. This leads to slightly higher cooler voltages in the log.

TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
04:40 1.9e9 209625 0203 7938 322.01 - 321.60 0.002 5s 8830
05:20 1.7e9 209626 0204 7939 322.01 - 321.60 0.002 5s 7800
06:00 1.6e9 209626 0205 7940 322.01 - 321.60 0.002 5s 8500
06:45 1.7e9 209620 0206 7941 322.01 - 321.60 0.002 5s 9400 cooler HV decreased slowly by a few V, jumped back at ~7:03
07:25 1.8e9 209620 0207 7942 322.01 - 321.60 0.002 5s 9300
08:05 1.75e9 209620 0208 7943 322.01 - 321.60 0.002 5s 9100
08:45 1.7e9 209618 0209 7944 322.01 - 321.60 0.002 5s 9000 cooler HV decreased 5V to 209614 and jumped back to initial value after ~5s
09:25 1.7e9 209618 0210 7945 322.01 - 321.60 0.002 5s 8500
10:05 1.9e9 209618 0211 7946 322.01 - 321.60 0.002 5s 9200 cooler HV decreased steadily to 209611 V during the run
10:45 1.9e9 209611 0212 7947 322.01 - 321.60 0.002 5s 7500 After injecting into the ESR, we saw after cooling some tails left and right of the Shottky peak of Th. Was not there in the runs before. After some minutes it disappeared in the spectrum
11:35 2e9 209611 0213 7948 322.01 - 321.60 0.002 5s it seems to be that much lesser ions are stored in the ESR - counter shows low rate - ABORTED laser didn't start scanning
11:58 1.75e9 209615 0214 7949 322.01 - 321.60 0.002 5s 7700 cooler HV jumped to 209610 at 12:12 and stayed there
12:53 1.85e9 209611 0215 7950 322.01 - 321.60 0.002 5s 7155 cooler HV increased slowly from initial value to ~209618
13:32 1.9e9 209612 0216 7951 322.01 - 321.60 0.002 5s 6638
14:39 1.8e9 209610 0217 7952 322.01 - 321.60 0.002 5s 6815 Voltage was jumping around from 209610 up to 209624 and back
15:20 1.75e9 209612 0218 7953 322.01 - 321.60 0.002 5s 8400 In contrast to the previous measurement, the cooler voltage was drifting by about 3V around 209612
16:00 1.7e9 209612 0219 7954 322.01 - 321.60 0.002 5s 7920
16:58 1.5e9 209612 0220 7955 322.01 - 321.60 0.002 5s 4577 PMT North voltage and threshold modified Entry 221, Ecooler voltage dcreased to 209609 V, weak injection
18:09 1.62e9 209609 0221 7957 322.01 - 321.60 0.002 5s During the last injection the beam suddenly got lost during the scraping, File 7956 was discarded since go4 was still during the startup of mbs
18:49 1.69e9 209609 0222 7958 322.01 - 321.60 0.002 5s 6133
19:49 1.6e9 209610 0223 7959 322.01 - 321.60 0.002 5s 7035
20:29 1.78e9 209611 0224 7960 322.01 - 321.60 0.002 5s 4490
21:12 1.86e9 209622 0225 7961 322.01 - 321.60 0.002 5s 10296 Note that the cooler voltage is increased. Particle number might be overestimated since it was read when the beam was already dumped.
Attachment 1: DriftingVoltage14-40.PNG
DriftingVoltage14-40.PNG
  223   Wed Apr 13 14:29:04 2022 Carsten, Rodolfo, Sarper, David, Pierre-Michel, Max, SebastianRunsAfternoon - morning shift runs 13.+14.04.
Changed the wavelength scan range from 321.62 - 321.10 nm to 321.12 - 320.60 nm after one scan.

TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
14:28 1.9e9 209618 0254 7983 321.62 - 321.10 0.002 5s 7950
15:35 1.7e9 209613 0255 7984 321.12 - 320.60 0.002 5s 7100 Wavelength range has changed starting from this scan
16:21 1.8e9 209612 0256 7985 321.12 - 320.60 0.002 5s 6900
17:11 1.8e9 209619 0257 7986 321.12 - 320.60 0.002 5s 7950
17:57 2.0e9 209618 0258-0259 7987 321.12 - 320.60 0.002 5s 8900
18:48 2.1e9 209620 0260-0261 7988 321.12 - 320.60 0.002 5s 9260 After the scan was complete, we have realized that TCL 1064 computer is crashed.. fixed @ 20:05
20:07 2.1e9 209621 0262 7989 321.12 - 320.60 0.002 5s 7800
20:56 2.0e9 209620 0263 7990 321.12 - 320.60 0.002 5s 8190 After closing file, we replaced the dye from the laser.
22:10 0 209620 0264 7991 321.12 - 320.60 0.002 5s 0 No beam due to failures in Unilac and SIS. NTCAP stopped 01:44, restarted 02:15
02:20 1.5e9 209620 0265 7992 321.12 - 320.60 0.002 5s 0 Electron cooler problem during scan, bad file.
02:52 1.4e9 209624 0266-0267 7993 321.12 - 320.60 0.002 5s 3076 --> CU mirror in all time !!!
03:44 1.6e9 209625 0268-0269 7994 321.12 - 320.60 0.002 5s 5156 --> CU mirror in all time !!!
04:35 1.7e9 209624 0270-0271 7995 321.12 - 320.60 0.002 5s 3430 --> CU mirror in all time !!!
05:27 1.6e9 209621 0272-0273 7996 321.12 - 320.60 0.002 5s 3200 --> CU mirror in all time !!!
06:17 1.6e9 209621 0274-0275 7997 321.12 - 320.60 0.002 5s 3371 low particle count since 22:00, see ID:225, --> CU mirror in all time !!!
07:24 1.5e9 209622 0276-0277 7998 321.12 - 320.60 0.002 5s 3366 --> CU mirror in all time !!!
08:33 1.6e9 209621 0278-0279 7999 321.12 - 320.60 0.002 5s 2940 ecooler jumped to 209615 during measurement, --> CU mirror in all time !!!
09:26 1.5e9 209618 0280 8000 321.12 - 320.60 0.002 5s aborted, request of Carsten
09:46 1.6e9 209621 0283-285 8003 321.12 - 320.60 0.002 5s 2950 small ecooler current changes (300mA injection and 50mA @ measurement), fragments survive in schottky @ lower frequencies (see attached pic1 and pic2), it seems to be that the cooler does not work correctly in "phase 1" (pic 1), --> CU mirror in all time !!!
11:28 1.75e9 209624 0287 8005 321.12 - 320.60 0.002 5s ~1800 ESR cooler back to 450mA (injection), 40mA (measurement), additional manual waiting time of 15s in machine 3 (after injection before ramping); ABORT laserscan on Carstens request after ~1 loop; Sergey arrived and we get some time for ion beam adjustements, --> CU mirror in all time !!!
12:23 1.5e9 209623 0289-0290 8007 321.12 - 320.60 0.002 5s 8100 Cu mirror is again automated and intensity in ESR is back up ; @12:42 increase of the width of the ion beam (pic3) @12:43 again normal width, intensity dropped significantly over time afterwards
13:09 1.55e9 209622 0291-292 8008 321.12 - 320.60 0.002 5s 4852 beam killed @ loop 2 scan nbr 253 (8 steps missed)
14:00 1e9 209622 0293-0294 8009 321.12 - 320.60 0.002 5s 5509
14:48 9.5e8 209622 0295-0296 8010 321.12 - 320.60 0.002 5s 5477
15:47 1e9 209622 0297-298 8011 321.12 - 320.60 0.002 5s 6336
16:42 7e8 209619 0299 8012 321.12 - 320.60 0.002 5s 2424
17:45 9.9e8 209620 0300 8013 321.12 - 320.60 0.002 5s 5133
Attachment 1: coolerProblemsAndFragments.JPG
coolerProblemsAndFragments.JPG
Attachment 2: fragments.JPG
fragments.JPG
Attachment 3: BroaderIonBeam.JPG
BroaderIonBeam.JPG
Attachment 4: againNormalWidth.JPG
againNormalWidth.JPG
  230   Thu Apr 14 17:47:55 2022 Carsten, Rodolfo, Konstantin, SarperRunsAfternoon shift runs 14.04 changing the wavelength scan range
The new scan range is changed from 321.12 - 320.60 nm to 320.62 - 320.10 nm

TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
18:36 9.5e8 209618 0301 8014 320.62 - 320.10 0.002 5s 6270
19:45 9.4e8 209615 0302 8015 320.62 - 320.10 0.002 5s 5400
20:44 8.4e8 209618 0303 8016 320.62 - 320.10 0.002 5s 4210
21:37 8.9e8 209616 0304 8017 320.62 - 320.10 0.002 5s 4550
  232   Thu Apr 14 22:44:40 2022 Patrick, Simon, Danyal, MaxRunsRuns night shift 14./15.04.
Runs of the night shift. The wavelength range is 320.62 - 320.10 nm.

TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
22:38 9.5e8 209621 0305 8018 320.62 - 320.10 0.002 5s 5000 laser scan was started 1 min late
23:38 9.8e8 209622 0306 8019 320.62 - 320.10 0.002 5s 5550
00:32 1.0e9 209621 0307 8020 320.62 - 320.10 0.002 5s 4000 After ~4000 counted particles (second scan, 1:04 am) the e-cooler went off.
01:21 1.1e9 209622 0308 8021 320.62 - 320.10 0.002 5s 6700
02:12 1.2e9 209623 0309 8022 320.62 - 320.10 0.002 5s 6400
03:03 1.1e9 209624 0310 8023 320.62 - 320.10 0.002 5s 5500
03:55 1.3e9 209622 0311 8024 320.62 - 320.10 0.002 5s 7700
04:47 1.2e9 209622 0312 8025 320.62 - 320.10 0.002 5s 5300
05:36 1.2e9 209620 0313 8026 320.62 - 320.10 0.002 5s 6666
06:25 1.2e9 209619 0314 8027 320.62 - 320.10 0.002 5s 6500 change of shift crew (Danyal & Max)
07:16 1.2e9 209618 0315 8028 320.62 - 320.10 0.002 5s 5230
08:13 1.2e9 209620 0316 8029 320.62 - 320.10 0.002 5s 6600
09:04 1.1e9 209619 0317 8030 320.62 - 320.10 0.002 5s 6660
09:54 1.2e9 209619 0318 8031 320.62 - 320.10 0.002 5s 5950
10:49 1.1e9 209619 0319 8032 320.62 - 320.10 0.002 5s 6580
11:38 1.1e9 209617 0320 8033 320.62 - 320.10 0.002 5s 5380
  236   Sat Apr 16 04:19:05 2022 Patrick, Simon, Alex, SarperRunsRuns night + morning shifts 16.04.
Runs of the night shift 16.04. after changing the wavelength range. The new range is 320.12 - 319.859 nm. The step size was decreased from 0.002 nm to 0.001 nm.
The goal was to have the same number of scans as before (261), however, the scan range in the labview program could not be set accordingly. We decided to go with 262 scans after the first run.


TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)particles countedcomments
04:24 1.28e9 209616 0338 8051 320.12 - 319.86 0.001 5s 6750 This run had only 260 scans
05:18 1.20e9 209615 0339 8052 320.12 - 319.859 0.001 5s 6300 First run with 262 scans
06:09 1.36e9 209615 0340 8053 320.12 - 319.859 0.001 5s 6910 laser scan started 2 min late at 650 counts
07:03 1.24e9 209615 0341 8054 320.12 - 319.859 0.001 5s 6330
07:59 1.23e9 209615 0342 8055 320.12 - 319.859 0.001 5s 5712
08:51 1.1e9 209615 0343 8056 320.12 - 319.859 0.001 5s 4380
09:44 1.28e9 209615 0344 8057 320.12 - 319.859 0.001 5s 5135
10:39 1.25e9 209615 0345 8058 320.12 - 319.859 0.001 5s 5860
11:30 1.2e9 209615 0346 8059 320.12 - 319.859 0.001 5s 6500
12:22 1.23e9 209618 0347 8060 322.41 - 322.00 0.002 5s 5200 Laser wavelength scan range is changed to 322.41 - 322.00 nm and increment to 0.002 nm.
13:04 1.2e9 209616 0348 8061 322.41 - 322.00 0.002 5s 4411
13:50 1.2e9 209616 0349-0351 8062-8063 322.41 - 322.00 0.002 5s 10651 last scan / 10 loops ~ 3 hours of scan -- Ecooler failed, realized at 16:05 we were on loop8, laser is turned off at 16:10