127
|
Thu Apr 7 05:23:50 2022 |
Patrick | DAQ | How To Start MBS for DAQ2 (Carstens DAQ) | (MBS server should be running at the LXG0155
machine. If not you will get an error message.
Then start the server - last command in console)
|
|
145
|
Fri Apr 8 00:06:37 2022 |
Patrick | Laser | Connection to the laser pc | To login to the laser pc use the command
xfreerdp -g 1920x1080 -u atplaser atppc023.campus.gsi.de |
|
147
|
Fri Apr 8 01:48:57 2022 |
Patrick | DAQ | Bugged Trending plots? | The wavelengths/frequencies shown in the graph
in the labview window do not match the wavelengths/frequencies
in go4.
|
|
149
|
Fri Apr 8 03:54:39 2022 |
Patrick | Laser | Q-switch window | The solution of increasing the Q-switch window
from 2us to 5us described in [I]ID:148[/I]
fulfills its purpose, but the window is unstable.
|
|
171
|
Fri Apr 8 22:26:44 2022 |
Patrick | DAQ | Spike in all detectors | In the previous shift, there was a spike in
all detectors in the same scan step. Apparently,
this was NOT a resonance since in the next |
|
172
|
Fri Apr 8 23:56:07 2022 |
Patrick | DAQ | DAQ improvement | As requested in ID:146 and ID:147, the laserscan
tab was improved:
[LIST]
|
|
174
|
Sat Apr 9 03:27:47 2022 |
Patrick | DAQ | Laser scanning | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
|
178
|
Sat Apr 9 04:40:01 2022 |
Patrick | DAQ | Bugged Trending plots? | The problem from yesterday (ID:147) still
persists. All the plots (not only the labview
stuff) comes in too late in go4. However, |
|
180
|
Sat Apr 9 05:26:25 2022 |
Patrick | Laser | Wavemeter | The wavemeter still shows large spikes from
time to time. The exposure time is still
at 34ms. |
|
190
|
Sun Apr 10 02:20:17 2022 |
Patrick | DAQ | New mbs startup guide | Connecting to the rfio server and opening
a file is included in '@startup' now. Since
mbs crashes reliably every 3 or 4 runs, it |
|
192
|
Sun Apr 10 04:07:53 2022 |
Patrick | DAQ | go4 shows wrong frequencies and wavelengths | [B]Update (ID:194):[/B] In labview "Stop All",
close labview, stop python script (ID:197),
close wavemeter. Then restart wavemeter, |
|
193
|
Sun Apr 10 04:37:29 2022 |
Patrick | DAQ | Next wavelength range | Measurements with the next wavelength scan
range should be started roughly around 12:00
noon. The plan is to go to lower wavelengths |
|
204
|
Mon Apr 11 04:26:53 2022 |
Patrick | DAQ | Next wavelength range | Measurements with the next wavelength scan
range should be started roughly around 10:00
am.
|
|
144
|
Thu Apr 7 22:18:06 2022 |
Night shift 7.4-8.4 | | Next scans | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
|
108
|
Wed Apr 6 13:32:57 2022 |
Max, Sebastian | DAQ | LabView Cobra Channel List | channel 0: linux time (epoch time)
channel 1: ms timer
channel 2: E-Cooler Voltage(Ohmlabs)
|
|
136
|
Thu Apr 7 11:11:31 2022 |
Max, Sebastian | Laser | dye changed, pulse energy measurements | We changed the dye at 10am.
We increased the pump power for the dye Laser |
|
225
|
Thu Apr 14 06:29:18 2022 |
Max, Sebastian | Accelerator | lesser ESR injection intensity | We now have less ESR injection intensity (see
screenshot attached).
|
|
181
|
Sat Apr 9 06:25:22 2022 |
Max, Kristian | Runs | morning shift runs | DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
|
|
183
|
Sat Apr 9 13:45:42 2022 |
Max, Kristian | Runs | first scans with LV problem solved! | This is the first time with LV buffer problem
solved (see Elog ID: 184). Before the PMT
signals were out of sync with the wavelength |
|
233
|
Fri Apr 15 12:42:18 2022 |
Max, Danyal, Carsten, Konstantin, Patrick, Simon | General | laser wavelength range changed: 320.12 - 319.60 nm |
Runs of the morning/afternoon/night shift.
The wavelength range is now 320.12 - 319.60 |
|