HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018 E0052_2025
  Dielectronic recombination-assisted laser spectroscopy, Page 6 of 6  Not logged in ELOG logo
Entry  Tue May 31 19:59:49 2022, Carsten, Danyal, Accelerator, change in frequency mismatch 208Bi --> 209Bi 
time = 20:00 

The RF has been adjusted so that it bunches at the correct frequency for 209Bi 
Entry  Tue Apr 26 16:28:14 2022, Carsten, DAQ, Cable of Bunch HF disconnected and 1.995Mhz signal connected for DAQ tests  
see subject.
Entry  Mon May 9 16:58:41 2022, Carsten, Detectors, Voltage particle dectectors NA_SZ und SA_SZ switched on 
Voltage SA_SZ (capture behind cooler): 1500 V (thin plastic scintillator)
Voltage NA_SZ (capture behind target): 750V (thick plastic, short absorption length)
Entry  Mon May 9 17:01:02 2022, Carsten, DAQ, directories / paths / comments / reduction factors /startup.scom files for the 2 DAQs  
[U][B]DAQ1 (original):[/B][/U]
-user atplaser
-runs on lxg1050
Entry  Mon May 9 17:14:02 2022, Carsten, DAQ, NTCAP DAQ 
The NTCap DAQ computer in the HKR (Schottky logbook) is started, and all preparations for data taking (mounting of NAS etc.) are done.
Parameters for NTCAP DAQ will be adjusted as soon as beam is available. 
Entry  Mon May 9 17:16:34 2022, Carsten, Accelerator, Important: New ESR pattern - requies action for XUV detector in/out 
The ESR pattern is enhanced compared to the thorium experiment by additional machines that can be  used for cooling/waiting in order to optimize 208Bi yield.
As a consequence, the sub chain numbers (previous: virtuelle Beschleuniger) are changed. The new numbers still have to be dialed in into the devices in
the Messhütte and the automatic in/out of the XUV-detector tested-
Entry  Mon May 16 17:15:56 2022, Carsten, Analysis, DAQ /Analysis - shift of peaks in the accumulated spectra, changes in code 
After a phone call with Wilfried I implemented an offset to all hit channels of the stacked / accumulated spectra where one chekcs the timing peaks.

Therfore, I made a copy of the analysis in 208Bi1 (one "1" at the end). And here, in the file TLdaAnlProc.cxx I added an offset of +70 to the hit channel
Entry  Fri May 27 20:39:50 2022, Carsten, Analysis, Short Manual for updated Go4 DAQ (version 2022_05_27) Short_Introduction_to_the_208Bi_Go4_Analysis.pdfShort_Introduction_to_the_208Bi_Go4_Analysis.odt
Short descriptions of the most recent version of the GO4 code is attached. I will install the corresponding vesion early on sunday.
Entry  Mon May 30 19:00:19 2022, Carsten, General, Scraper Settings and Parameters for the last measrements with thick target screenshot-2022-05-30_19-12-151030168345517073342_(002).png
old Bunch frequency: 1.98777 MHz

Positions PL devices:
Entry  Mon May 30 20:58:26 2022, Carsten, Accelerator, New Settings for 208Bi Beam  
Old SetParameters.C and old SetConditions.C were saved as SetParameters_209.C and SetConditions_209.C

The ion beam settings were changed to 208Bi82+:
Entry  Thu Jun 30 11:12:21 2022, Carsten, Analysis, Updated Go4 Analysis with small Bugfix (PMTOverWavelength SPectra) 208Bi_20220629_fix.tgz
Here attached is a more recentversion of the Go4 Analysis Code with some smaller bugfixes.
For adaequate *.C files please have alook in 208Bi go4 directory on the atplaser account.
Please note, that there is also a (not fully up-to-date) description/manual somewhere here in the Elog.
ELOG V3.1.5-fc6679b