Entry  Mon May 30 08:35:41 2022, Kristian, Simon, Ken, Ragan, Detectors, Changed PMT North 
We did not see any signal at PMT North which was the narrowband PMT. Changed it to the old solar blind PMT. Now we see the resonance. It is about 3 times
larger as PMT middle, most likely because of the larger window. CFD threshold was set. I also increased the CFD threshold for PMT middle slightly.
Entry  Wed Jun 1 08:39:50 2022, Kristian, Konstantin, Calibration, Wavemeter reading 
We recorded files with high wavemeter readout rate and saved it at the atplaser network drive (Q:) - Bismuth - wavemeter folder.

Could be used to calibrate the laser steps. The MBS readout rate is 4 Hz, here 30 Hz, so there are a few more values.
Entry  Mon May 30 16:17:39 2022, Kristian, David, Wilfried, Carsten, Danyal, Rodolfo, Runs, Afternoon shift 30.05 
These are the first runs after changing to the thick stripper foil.

[TABLE border="1"]
Entry  Fri May 27 10:39:53 2022, Kristian, DAQ, PMT North noise IMG_20220527_095501.jpg
During the last run, PMT North showed some weird noise.

The PMT was replaced by the deep UV PMT with small aceptance range (100-200nm) but there was still a lot of noise on the output channel in the Messhütte,
Entry  Wed Jun 1 00:03:08 2022, Konstantin, Simon, Kristian, Runs, 209-Bi runs 
[TABLE border="1"]
Time | SIS-intensity | ESR-intensity after orbit change| cooler-HV | cooler current (mA) | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step
width increment (nm) | dwell time(s) | n loops |comments
Entry  Wed Jun 1 00:16:02 2022, Konstantin, Simon, DAQ, Timing 209Bi thick target Timing_209Bi_final.png
Timing for the final measurements on 209Bi was determined for the new frequency 1.93777MHz. Peak positions are

PD NO 106
Entry  Mon May 30 01:13:40 2022, Konstantin, Max, Patrick, Detectors, XUV-Detector voltage off 
XUV-Detector shows up to 100 kHz and, therefore, was turned off.
Entry  Wed May 11 14:48:31 2022, Konstantin, DAQ, Voltage range of the Keysight DMM changed from 10V to 1V VoltageRangeChanged.png
During the setup of the ion beam the voltage range of the Keysight DMM was changed from 10V to 1V. Additionally, the name of the high-voltage divider was
adjusted (this changes only the filename of the file, which is stored at the local pc decpc003).
Entry  Thu May 12 14:04:09 2022, Konstantin, DAQ, Go4 Laserstep histograms adapted Go4Modifications.PNG
In order to correctly display the gated PMT signals [I]PMT Middle[/I] and [I]PMT North[/I] some modifications in the Go4 code were necessary. Affected files
are [I]TLdaAnlProc.cxx[/I], [I]TLdaAnlProc.h[/I] and [I]TLdaInputDescription.h[/I].
Modifications are indicated according to the scheme "[B]//-- HH:MM mm-dd-yyyy: <description>[/B]" (an example is given in the attached screenshot). They
Entry  Tue May 17 10:38:44 2022, Konstantin, Detectors, PMT North changed 
PMT located at the northern window was again replaced by the narrow-band PMT (ET Enterprise type 9423 SB, SN: 650).
Entry  Mon May 30 09:28:43 2022, Ken, Kristian, Simon, Ragan, Detectors, XUV-Detector 
9:27 XUV-Detector is running again.

-5V cathode, 8A main coil, 2.4A sec.coil
Entry  Tue May 10 22:29:45 2022, Jan Glorius, DAQ, DAQ2 shaper settings setting_e128.v1
Attached are the MSCF shaper settings applied on Tue May 10 at 22:30.
Entry  Tue May 10 22:37:10 2022, Jan Glorius, DAQ, DAQ2 cabling docu e128_daq2_cabling.ods
 
Entry  Fri May 13 16:55:58 2022, Danyal, Carsten, Phillip, Wilfried, General, End of E128 Beamtime  
Dear all,

unfortunately, the E128 experiment must now come to an early end. 
Entry  Fri Jun 3 12:01:17 2022, Danyal, Detectors, switched off XUV detector on 1 June ~17:00  
I have switched off the XUV detector on 1 June at about 17:00. 
The high voltages of the MCP (in ESR laser lab) were turned off and inside the ESR cave the power supplies (coils and cathode) and the crate were turned
off. 
Entry  Tue May 31 19:32:56 2022, Carsten, Wilfried, Max, Calibration, electron cooler calibration voltage - frequency 
[TABLE border="1"]
SetVoltage (DeltaU) | MeasVoltage | freq | Current 
|-
Entry  Wed May 4 22:45:12 2022, Carsten, Sebastian, DAQ, Q-Switch Signal (ch15) fixed  
The "dead" Q-Switch signal was converted 
two times... This explains why in the 
Thorium experiment the Q-Switch signal 
Entry  Mon May 30 23:18:06 2022, Carsten, Rodolfo, Konstantin, Patrick, Ragan, Ken, Kristian, Max, Danyal, Runs, Runs shifts 31.05. 2022-05-31_particles.png
These are the first runs with 208Bi.

[TABLE border="1"]
Entry  Tue May 31 19:37:30 2022, Carsten, Max, Wilfried, Detectors, SASZ and NASZ voltages switched off 
SASZ and NASZ voltages were switched off bevor we went back to 209Bi. Laser beam stabilization turned off.
Entry  Tue May 31 21:08:04 2022, Carsten, Max, Detectors, XUV detector automation off 
We deactivated the automation of the XUV detector. It stays out now.