4 Entries
ID
Date
Author
Category
Subject
4
Wed Jun 4 16:18:57 2025
Rodolfo, Simon
Detectors
PMT dark count rates
I adjusted the offsets of the amplifiers in the cave and measured the dark count rates :
PMT
Model
Operation Voltage
Count rate (Hz)
South
ET9422B
Broadband
-1500 V
1300
Middle
ET9423B
Smallband
-2650 V
20
North
ET9423B
Smallband
-2650 V
30
Also the MCP anode of the XUV detector is now connected to the amplifiers and fed to the data cable of "PMT Cu".
Quote:
This is the arrangement of the photomultipiers at the optical detection region at ESR for Thorium 2025
PMT
Model
SN
Operation Voltage
Threshold
South
ET9422B
1042
Broadband
-1500 V
Middle
ET9423B
653
Smallband
-2650 V
10 (preliminary)
North
ET9423B
652
Smallband
-2650 V
12 (preliminary)
3
Wed Jun 4 16:17:47 2025
Simon
Detectors
Nitrogen
The nitrogen supply for the PMTs is turned
ON
.
2
Mon Jun 2 16:36:57 2025
Carsten
Detectors
Mesycontrol
ssh -Y litv-exp@x86l-204
'mesycontrol'
1
Tue May 20 16:14:30 2025
Rodolfo, Simon
Detectors
PMT arrangement at ESR
This is the arrangement of the photomultipiers at the optical detection region at ESR for Thorium 2025
PMT
Model
SN
Operation Voltage
Threshold
South
ET9422B
1042
Broadband
-1500 V
Middle
ET9423B
653
Smallband
-2650 V
10 (preliminary)
North
ET9423B
652
Smallband
-2650 V
12 (preliminary)
ELOG V3.1.5-fc6679b