  118   Thu Apr 2 07:01:32 2020 Ragan, Esther, OliverRunsMeasurement with 4 hrs storage time
beam: 205Tl81+ (no scraping)
energy: 400 MeV/u
storage time: 4 hrs

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):       mm

run start at 01:48 , first file: run0287.lmd
file running:    0287.lmd

_______________________________________________________________________________

time when accumulation starts: 04:45
time when storage starts: 04:55
time when storage ends: 06:55
I_cooler during storage time: 10 mA

I_cooler =  mA


I_SIS: 1.8e9      
ESR particles after stacking: ~e5  
Number of stackings: 20
Target ON density:   4e12
Target on time: 06:56:30
Target off time: 07:06:30
Target on time for measurement: 10 mins (by local mode)       


NTCAp file:
SC_2020-04-01_16-36-43/0000xxx.iq.tdms
\IQ_2020-04-01_16-36-43/0000xxx.iq.tdms
IQ rate: 8 MS/s
(saved in drive : o)

copy and repeat the above (below the line) for every storage time measurement
  119   Thu Apr 2 09:21:19 2020 RuijiuRuns2nd measurement with 2 hrs storage time (repeat)
Because the electron cooler was not switched on during beam accumulation in the previous measurement, we repeat this measurement again.
(In the picture below, it can be seen that the beam cooled. On right most is the Tl beam and rest are the contaminants.)

beam: 205Tl81+ (no scraping)
energy: 400 MeV/u
storage time: 2 hrs

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):       mm

run start at 08:46 , first file: run290.lmd
file running:   run290.lmd


time when accumulation starts: 09:06
time when storage starts:  09:16
time when storage ends:  10:59
I_cooler during storage time: 10 mA

I_cooler =  __ mA


I_SIS: 1.7e9      
ESR particles after stacking: 1e6
Number of stacking: 20
Target ON density:   3.9e12 atoms/cm2
Target on time:  11:00
Target off time: 11:10
Target on time for measurement: 10 
mins (by local mode)       


NTCAp file:
SC_2020-04-02_07-10-40
\IQ_2020-04-02_07-11-48
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the 
line) for every storage time 
measurement
Attachment 1: IMG_20200402_092631.jpg
IMG_20200402_092631.jpg
  122   Thu Apr 2 21:03:40 2020 Ragan, Yuri, Sergey, NikosRunsZero hour measurement (S2:-10/35 mm)
beam: 205Tl81+ (no scraping, Al ~1g/cm2)
energy: 400 MeV/u
storage time: zero hrs
S2:-10/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd (File was started when stacking was ongoing. Maybe not all the stackings could be sen in the analysis.)
file running: 0292.lmd

_______________________________________________________________________________

time when accumulation starts: --
time when storage starts: 19:10
time when storage ends: 19:10
I_cooler during storage time: --

I_cooler =  177 mA


I_SIS: 2.2e9      
ESR particles after stacking: 3.5e6   (offset 1e5)
Number of stackings: ~40 (can be checked on NTCap data)
Target ON density: 3.45e12 
Target on time for measurement: 10 mins (by event mode)       


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
  123   Thu Apr 2 21:10:57 2020 Ragan, Yuri, Sergey, NikosRunsOne hour measurement (S2:-10/35 mm)
beam: 205Tl81+ (no scraping, Al ~1g/cm2)
energy: 400 MeV/u
storage time: 1 hrs
S2:-10/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0293.lmd

_______________________________________________________________________________

time when accumulation starts: 19:23
time when storage starts: 19:36
time when storage ends: 20:36
I_cooler during storage time: -1 mA

I_cooler =  177 mA


I_SIS: 2.2e9      
ESR particles after stacking: 2.6e6   (offset 1e5)
Number of stackings: 30
Target ON density: 3.44e12 
Target on time for measurement: 10 mins (by event mode)       


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
  124   Thu Apr 2 21:12:39 2020 Ragan, Nikos, Pierre-Michel, DimaRunsThree hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 3 hrs
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0295.lmd

_______________________________________________________________________________

time when accumulation starts: 20:37
time when storage starts: 20:51
time when storage ends: 23:51
I_cooler during storage time: -1 mA

I_cooler =  177 mA


I_SIS: 1.75e9      
ESR particles after stacking: 2e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.48e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 5 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 46CBD7E3-0AF1-447C-847E-31BA5224684D.jpeg
46CBD7E3-0AF1-447C-847E-31BA5224684D.jpeg
Attachment 2: 7A404038-88E9-4CAF-91E3-A26A0F8A51EF.jpeg
7A404038-88E9-4CAF-91E3-A26A0F8A51EF.jpeg
  125   Fri Apr 3 00:39:07 2020 Ragan, Pierre-Michel, DimaRunsHalf hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0.5 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0295.lmd

_______________________________________________________________________________

time when accumulation starts: 12:08
time when storage starts: 12:25
time when storage ends: 12:55
I_cooler during storage time: -1 mA

I_cooler =  177 mA


I_SIS: 2e9      
ESR particles after stacking: 8.2e5 (offset 1e5)
Number of stackings: 9
Target ON density: 3.45e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 3 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: CC419497-E2AA-4C71-92AD-7F282775FFC2.jpeg
CC419497-E2AA-4C71-92AD-7F282775FFC2.jpeg
Attachment 2: AB10D6CB-87FC-4D9E-9AF0-AA9279D62066.jpeg
AB10D6CB-87FC-4D9E-9AF0-AA9279D62066.jpeg
  126   Fri Apr 3 01:22:13 2020 Ragan, Pierre-Michel, DimaRunsOne hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 1 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0295.lmd

_______________________________________________________________________________

time when accumulation starts: 01:09
time when storage starts: 01:20
time when storage ends: 02:20
I_cooler during storage time: -1 mA

I_cooler =  174 mA


I_SIS: 1.8e9      
ESR particles after stacking: 2.5e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.48e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 5 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 8E469F1D-992F-4E96-A30A-022B0D1A0E5A.jpeg
8E469F1D-992F-4E96-A30A-022B0D1A0E5A.jpeg
  127   Fri Apr 3 03:18:06 2020 Ragan, Pierre-Michel, DimaRunsZero hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0298.lmd

_______________________________________________________________________________

time when accumulation starts: 02:36
time when storage starts: 02:47
time when storage ends: 02:47
I_cooler during storage time: -1 mA

I_cooler =  175 mA


I_SIS: 2e9      
ESR particles after stacking: 2.7e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.49e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 3 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 3C7167D2-5B67-4A2A-BB16-76DE5DCE6D7F.jpeg
3C7167D2-5B67-4A2A-BB16-76DE5DCE6D7F.jpeg
  128   Fri Apr 3 03:19:46 2020 Ragan, Pierre-Michel, DimaRunsHalf hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0.5 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 19:05 , first file: run0292.lmd 
file running: 0298.lmd

_______________________________________________________________________________

time when accumulation starts: 03:00
time when storage starts: 03:12
time when storage ends: 03:42
I_cooler during storage time: -1 mA

I_cooler =  175 mA


I_SIS: 1.8e9      
ESR particles after stacking: 2.4e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.5e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 :  3 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_19-10-21
\IQ_2020-04-02_19-10-14
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: FD4F77A6-57C8-40EE-A5D6-B8E84D87C0EF.jpeg
FD4F77A6-57C8-40EE-A5D6-B8E84D87C0EF.jpeg
Attachment 2: D9AB6DAC-8CB1-4AA6-8D65-EF0BEB1BC913.jpeg
D9AB6DAC-8CB1-4AA6-8D65-EF0BEB1BC913.jpeg
Attachment 3: 7ECADF49-BD0A-4AB0-82E3-8825C08C81E3.jpeg
7ECADF49-BD0A-4AB0-82E3-8825C08C81E3.jpeg
  129   Fri Apr 3 03:59:14 2020 Ragan, Pierre-Michel, DimaRunsTwo hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 2 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 04:13,  run0300.lmd 
file running: 0300.lmd

_______________________________________________________________________________

time when accumulation starts: 04:21
time when storage starts: 04:32
time when storage ends: 06:32
I_cooler during storage time: -1 mA

I_cooler =  179 mA


I_SIS: 1.75e9      
ESR particles after stacking: 2.1e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.66e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 7 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-03_04-17-44
\IQ_2020-04-03_04-17-51
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 2B409195-AF78-46AA-9B5B-F0990B564238.jpeg
2B409195-AF78-46AA-9B5B-F0990B564238.jpeg
Attachment 2: ED3309B6-82E5-4484-A787-ADF94FD55814.jpeg
ED3309B6-82E5-4484-A787-ADF94FD55814.jpeg
  131   Fri Apr 3 06:35:37 2020 Ragan, Pierre-Michel, Dima, Mei, ShahabRunsHalf hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0.5 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 04:13,  run0300.lmd 
file running: 0301.lmd

_______________________________________________________________________________

time when accumulation starts: 06:50
time when storage starts: 07:01
time when storage ends: 11:01
I_cooler during storage time: -1 mA

I_cooler =  179 mA


I_SIS: 1.75e9      
ESR particles after stacking: 2.1e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.66e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 8 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-02_04-17-44
\IQ_2020-04-02_19-04-17-51
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: A04655E8-EF30-4ECB-A6F4-2B98982502FA.jpeg
A04655E8-EF30-4ECB-A6F4-2B98982502FA.jpeg
Attachment 2: AE7CF1AD-0E03-45AA-8234-D36D34E85252.jpeg
AE7CF1AD-0E03-45AA-8234-D36D34E85252.jpeg
Attachment 3: 95BAD6CD-5C2E-4B93-8CE5-0F7F6E62B4FA.jpeg
95BAD6CD-5C2E-4B93-8CE5-0F7F6E62B4FA.jpeg
  132   Fri Apr 3 10:06:15 2020 ShahabRunsPlots from Scopes
Just two plots from the scopes for documentation:

magnets.png:

yellow: dipole
green and blue: quadrupoles
red: ESR current


kicker.png:

blue: SIS kicker
yellow: ESR kicker

green: SIS current
red: ESR current

trigger: SIS Kicker
Attachment 1: magnets.png
magnets.png
Attachment 2: kicker.png
kicker.png
  133   Fri Apr 3 11:52:27 2020 ShahabRunsStacking 54 times!
Attachment 1: 2020-04-03_11-48-11-781.png
2020-04-03_11-48-11-781.png
  134   Fri Apr 3 11:59:50 2020 Ragan, Pierre-Michel, DimaRunsFour hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 4 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 04:13,  run0300.lmd 
file running: 0301.lmd

_______________________________________________________________________________

time when accumulation starts: 06:50
time when storage starts: 07:01
time when storage ends: 11:01
I_cooler during storage time: -1 mA

I_cooler =  179 mA


I_SIS: 1.75e9      
ESR particles after stacking: 2.3e6 (offset 1e5)
Number of stackings: 30
Target ON density: 3.3e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 5 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-03_04-17-44
\IQ_2020-04-03_04-17-51
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 708F0673-2D65-4EFA-902F-7067FA5885B4.jpeg
708F0673-2D65-4EFA-902F-7067FA5885B4.jpeg
Attachment 2: 8D6B87F7-B622-41B6-8CBC-E6404443987F.jpeg
8D6B87F7-B622-41B6-8CBC-E6404443987F.jpeg
Attachment 3: 2D175AD1-813B-4C6B-93E0-43B90EFE2584.jpeg
2D175AD1-813B-4C6B-93E0-43B90EFE2584.jpeg
  135   Fri Apr 3 12:02:32 2020 Ragan, Mei, ShahabRunsHalf hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0.5 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 11:17,  run0303.lmd 
file running: 0303.lmd

_______________________________________________________________________________

time when accumulation starts: 11:19
time when storage starts: 11:46
time when storage ends: 12:16
I_cooler during storage time: -1 mA

I_cooler =  177 mA


I_SIS: 2e9      
ESR particles after stacking: 3.2e6 (offset 5e5)
Number of stackings: 54   (was set for 60 but we missed a few)
Target ON density: 3.66e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 23 mins (as Shahab wanted to record data with 400 MHz schottky)
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-03_11-20-14
\IQ_2020-04-03_11-20-05
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 44FE81B4-D167-4B4D-9E6E-D0C54CD79F23.jpeg
44FE81B4-D167-4B4D-9E6E-D0C54CD79F23.jpeg
Attachment 2: 6FFE01F8-FC91-4A7A-A8E0-52BFC068DD27.jpeg
6FFE01F8-FC91-4A7A-A8E0-52BFC068DD27.jpeg
Attachment 3: 19EBC2DF-C029-4655-A697-7CD69C6E4ADF.jpeg
19EBC2DF-C029-4655-A697-7CD69C6E4ADF.jpeg
Attachment 4: 68740E5C-468F-476A-9D81-FC3D4995D61E.jpeg
68740E5C-468F-476A-9D81-FC3D4995D61E.jpeg
  136   Fri Apr 3 13:21:16 2020 Ragan, Mei, ShahabRunsZero hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 11:17,  run0303.lmd 
file running: 0304.lmd

_______________________________________________________________________________

time when accumulation starts: 12:58
time when storage starts: 13:16
time when storage ends: 13:16
I_cooler during storage time: -1 mA

I_cooler =  176 mA


I_SIS: 2e9      
ESR particles after stacking: 2.4e6 (offset 5e5)
Number of stackings: 38   (was set for 60 but we missed a few)
Target ON density: 3.54e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 2 mins
(i.e. time for which electron cooler is on after gas jet) 


NTCAp file:
SC_2020-04-03_11-20-14
\IQ_2020-04-03_11-20-05
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 8560F317-7B65-476E-84EF-CFFB9C021878.jpeg
8560F317-7B65-476E-84EF-CFFB9C021878.jpeg
  137   Fri Apr 3 13:39:58 2020 Ragan, Mei, Shahab, LaszloRunsEight hour measurement (S2: 0/35 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 8 hr
S2: 0/35 mm

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 11:17,  run0303.lmd 
file running: 0306.lmd

_______________________________________________________________________________

time when accumulation starts: 13:32
time when storage starts: 13:51
time when storage ends: 21:51
I_cooler during storage time: -1 mA

I_cooler =  179 mA


I_SIS: 2e9      
ESR particles after stacking: 3e6 (offset 0)
Number of stackings: 50   
Target ON density: 3.95e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 5 mins
(i.e. time for which electron cooler is on after gas jet) 

at 18:30 (start was at 13:51) the stored particles: I_ESR = 1.7e6
at 19:00 the stored particles: I_ESR = 1.6e6
at 19:30 the stored particles: I_ESR = 1.48e6
at 20:00 the stored particles: I_ESR = 1.5e6
at 20:30 the stored particles: I_ESR = 1.4e6
at 21:00 the stored particles: I_ESR = 1.34e6
at 21:30 the stored particles: I_ESR = 1.34e6


NTCAp file:
SC_2020-04-03_11-20-14
\IQ_2020-04-03_11-20-05
IQ rate: 8 MS/s
(saved in drive : p)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: A9A3E9A0-85E9-4A47-93FA-5D722730FCFC.jpeg
A9A3E9A0-85E9-4A47-93FA-5D722730FCFC.jpeg
Attachment 2: B24C41F3-BD4D-4400-9A30-B9CEF7928BBE.jpeg
B24C41F3-BD4D-4400-9A30-B9CEF7928BBE.jpeg
Attachment 3: 7B276C07-1E38-4858-924E-DFEC12F10578.jpeg
7B276C07-1E38-4858-924E-DFEC12F10578.jpeg
  138   Fri Apr 3 14:08:41 2020 ShahabRunsStacking
Attachment 1: 2020-04-03_14-07-13-250.png
2020-04-03_14-07-13-250.png
  139   Fri Apr 3 17:55:13 2020 ShahabRunsNice picture from 205-Tl
In this high contrast Schottky recording, the main three frequency peaks are resolved w.r.t. time from left to right:

203 Hg 80 +, 200 Hg 79+ and 205 Tl 81+

There are also some spurs visible, whose nature is not 100% sure. One possible explanation can be:

Since this is taken during the storage time (gas target still off), we have no separation between parent and daughter ions, the daughter 205 Pb 81+ basically has the same revolution frequency as parent. It is a possibility that after the decay, daughter ions are slowly cooled back to the nominal frequency. Daughter ions are highly charged and have a high velocity, so single ion sensitivity is expected. They come from all directions, but Schottky measures only the longitudinal plane. Symmetric spurs can hence be an indication of this.
Attachment 1: Tl-daughter_spurs.PNG
Tl-daughter_spurs.PNG
  140   Sat Apr 4 00:00:20 2020 Ragan, Laszlo, Pierre-Michel, DimaRunsZero hour measurement (S2: 0/35 mm and S6: -20/7 mm)
beam: 205Tl81+ (no scraping , Al ~1g/cm2)
energy: 400 MeV/u
storage time: 0 hr
S2: 0/35 mm
s6: -20/7

purpose: data with TARGET ON

Detector position (CsISiPHOS):  out of the ring
Detector position (MWPC):  -60 mm

run start at 22:11,  run0308.lmd 
file running: 0309.lmd

_______________________________________________________________________________

time when accumulation starts: 23:29
time when storage starts: 23:53
time when storage ends: 23:53
I_cooler during storage time: -1 mA

I_cooler =  176 mA


I_SIS: 1.75e9      
ESR particles after stacking: 2.9e6 (offset 5e5)
Number of stackings: ~53   
Target ON density: 4e12 
Target on time for measurement: 10 mins (by event mode)      
time spend at manipulation 13 : 2 mins
(i.e. time for which electron cooler is on after gas jet) 



NTCAp file:
SC_2020-04-03_22-12-50
\IQ_2020-04-03_22-13-48
IQ rate: 8 MS/s
(saved in drive : O)

copy and repeat the above (below the line) for every storage time measurement
Attachment 1: 0453DD6F-453B-4751-96A3-D9D1CD808CD5.jpeg
0453DD6F-453B-4751-96A3-D9D1CD808CD5.jpeg
Attachment 2: 99B7B973-C12E-4621-BD2B-DDF1D6753DF2.jpeg
99B7B973-C12E-4621-BD2B-DDF1D6753DF2.jpeg
ELOG V3.1.5-fc6679b