HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028
  proton-capture on 118Te, Page 13 of 29  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Category Subjectdown Year Text Attachments
  47   Sun Mar 15 17:12:40 2020 Laszlo, JanDetectorsnew Ortec spec amps  We are now using new SpecAmps for the x-ray
detectors from ORTEC, they deliver much improved
resolution but have other strange effects
  
  368   Fri May 21 19:20:54 2021 LaszloGeneralmisalignment in the pattern2021for some reason the event number of the target
ON and OFF was misaligned with the pattern.
This means that the last few runs (run53,
  
  269   Thu Mar 26 12:05:22 2020 LaszloCalibrationmeasurement of collimators The measurements were taken by using a caliper
ruler, with precision of +/-0.05mm
 SP54120032612091.pdfSP54120032612090.pdf 
  6   Fri Apr 5 06:05:55 2019 JanDAQmaking changes to the DAQ  If you need to add or change a module of the
VME crate you need to follow this proceedure:
  
  274   Wed Apr 8 23:02:24 2020 LaszloGenerallmd to measurement pairing list I use the following list of lmd files combined
to a single root file for each measurement.
It is very important to highlight that sometimes
 gas_jet_mailfunction.png 
  439   Tue May 25 08:36:15 2021 JanGenerallessons learned20211. before using stoch. cooling in ESR, boot
and update the windows PC
2. better diagnostics for low energy/intensity
  
  280   Thu Apr 30 22:55:45 2020 LaszloCalibrationinverse square law test I have made also the inverse square law fits.
We have data only for 90degre with the 241Am
source, but both for the 1. and 2. calibrations.
 inverse_square_law.zip 
  134   Sat Mar 21 15:30:34 2020 SergiyGeneralgo into the ESR We are going into the ESR to check the electronics.
Use calibration sources (133Ba etc).
  
  84   Fri Mar 20 16:48:00 2020 LaszloGeneralflipped Si shaper output in x and y Yesterday evening we have found out that the
energy signals from the silicon are negative
--> wrong polarity of the shapers was chosen
  
  233   Mon Mar 23 09:47:05 2020 LaszloGeneralexplanation of an ESR cycle    ESR_cycle.pngIMG_20200323_095144.jpg 
  345   Sun May 16 12:55:48 2021 Jan GloriusDAQevent server lxg1299:60032021new event server is set up at lxg1299 on port
6003
It runs as:
  
  304   Sat May 1 22:34:16 2021 Jan GloriusCalibrationefficiency xray 90deg2021the attached efficiency curve for the 90 deg.
xray detector is made with the runs
e127b_run0008.lmd
 90deg_eff.png90deg_eff.dat 
  317   Thu May 6 13:39:02 2021 Jan GloriusCalibrationefficiency xray 35deg2021the attached efficiency curve for the 35 deg.
xray detector is made with the runs
e127b_run0015.lmd
 35deg_eff.png35deg_eff.dat 
  281   Wed May 6 23:02:42 2020 LaszloCalibrationefficiency values I have calculated the efficiency values by
putting the energies of the K-REC peaks into
the phenomenological (empirical) function
 90degree_efficiency_combined.png90degree_efficiency_combined.ps 
  279   Thu Apr 30 17:40:57 2020 LaszloCalibrationefficiency fits For all 3 detector the calibration data sets
were combined to include the systematics
in the fit results directly. Combining means
 90degree_efficiency.zip145degree_efficiency.zip35degree_efficiency.zip 
  278   Wed Apr 29 16:06:28 2020 LaszloCalibrationefficiency fit - 90degree, combined dataset For the efficiency vs E fit of the 90degree
Xray detector I have used the following phenomenological
funciton:
 efficiency_90degree-1an2datasets_phen_lin.pngefficiency_90degree-1an2datasets_phen_lin_bigRange.png 
  191   Sun Mar 22 14:58:59 2020 LaszloGeneraldropping density of Gasjet after like 20min run, the density of the gasjet
already dropped from 10^14 to 8.7*10^13.
Now it seems to remain in that value
 ga_jet_drop.png 
  97   Sat Mar 21 02:36:44 2020 LaszloGeneraldetector and scraper position For the 124Xe primary beam measurement:
-with the DSSSD we scraped the beam at position
-40
 IMG_20200321_025644.jpg 
  36   Fri Feb 21 11:29:59 2020 JanDetectorsdet. distances at target chamber This is the lookup table for the distance
of each x-ray detector to the target.
See attched foto for explanation.
 Photonen_Kammer.png 
  60   Tue Mar 17 08:15:49 2020 JanGeneralcurrent grids & screen in TE  measured beam diagnosis with 124Xe48+ at 326
MeV/u
 2020-03-16_18-51-50-170.pngIMG_0694.jpg 
ELOG V3.1.5-fc6679b