HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 15 of 29  Not logged in ELOG logo
Entry  Fri Apr 30 09:57:57 2021, Jan Glorius, Calibration, run0005 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 09:27:27 2021, Jan Glorius, Calibration, run0004 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Thu Apr 29 20:34:09 2021, Jan Glorius, Calibration, run0003 - Xray90 calib Pb210 d=167.5mm, 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Thu Apr 29 20:32:52 2021, Jan Glorius, Calibration, calibration sources, 2021 
We use the following list of sources for calibration of the Xray detectors:

- 210Pb [40.1 kBq (4%), 01.10.2020 12:00 UTC] [URL=https://elog.gsi.de/esr/X-Ray+Lab/201006_131944/20-006_210Pb_40kBq.pdf]SpecSheet[/URL]
Entry  Wed Apr 28 09:12:47 2021, Jan, Yuri, Detectors, UI-diagram, 2021 UI_meas_curve.pdf
The UI-curve of the detector
U I
10 0.11
Entry  Thu Apr 22 10:56:20 2021, Jan, Detectors, target-chamber det. distances, 2021 E127_xray_distances.jpg
This is the lookup table for the distance of each x-ray detector to the target.
See attached foto for explanation.
Entry  Wed Apr 14 09:42:50 2021, Jan, Detectors, BaF HV gain matching, 2021 BaF_Na22_gain_match_v1.pngbaf_gain_match_v1.dat
The preliminary gain matching settings for a full ADC range of ~12MeV:

[TABLE border="1"]
Entry  Wed Mar 31 13:57:27 2021, Jan, Detectors, BaF HV settings, 2021 
The setting below is a start value for final gain matching with a full ADC range of 12MeV:

[TABLE border="1"]
Entry  Wed Mar 31 13:52:04 2021, Jan, Detectors, BaF Na22 Testing, 2021 
On 30. and 31.03.2021 we did some testing with the 6 BaF detectors.

Trigger & threshold:
Entry  Tue Feb 2 13:59:00 2021, Jan, Detectors, DSSD installation and alignment, 2021 
The DSSD has been exchanged and aligned in 
November 2020. After bakeout at 140°C 
(externally, 120°C at internal temp. Sensor) 
Entry  Tue Nov 24 15:35:49 2020, Jan, Calibration, DSSD X/Y channel mapping, 2021 
During the detector test [URL=https://elog.gsi.de/esr/E127/284]test measurements[/URL] with alpha source the allocation of the 16 X- and 16 Y-channels has
been checked.
Entry  Fri Oct 23 12:58:12 2020, Jan, Detectors, Test of 2nd DSSSD (gen2), 2021 Si2_run001.jpgSi2_run002.jpg
This is the documentation of the source tests with the 2nd micron DSSSD of 2nd generation (label 3288-17, thickness 529um)

The detector is put into vacuum (~5e-6 mbar) in our test chamber. The source is positioned a few cm above (see fotos).
    Reply  Fri Nov 20 15:53:37 2020, Jan, Analysis, Test of 2nd DSSSD (gen2), 2021 quick_si_plots.ce127.hSi2_run001.ana.rootSi2_run002.ana.root
Here is the analysis of the test runs with the 2nd DSSD (gen2) using the quick_si_plots.c script attached.

All 32 Si channels are working with acceptable performance. 
Entry  Fri Jun 5 14:06:15 2020, Laszlo, Detectors, DSSSD and SCRAPER position estimate for Xe and Te experiments, 2020 SP54120060514100.pdfDSSSD_new1.pdfIMG_20200321_025644.jpg
We don't know the exact absolute positions of the detector (+scraper) and the beam. However, what we have to know is only these two relative positions respect
to each other. To get this distance I use two methods:
1, combining the infos from the set position during the beamtime + the measured pg peak position on the detector. The pg peak position is defined only
Entry  Tue May 19 16:48:15 2020, Laszlo, General, E127 beamtime overview,  E127_beamtime_overview.pngE127_beamtime_overview.pdf
Here is a representation how was the time management during E127. The time, what we could spend with measuring the 118Te(pg), was ~20% comparing to the
given 6days.
Entry  Wed Apr 29 16:06:28 2020, Laszlo, Calibration, efficiency fit - 90degree, combined dataset,  efficiency_90degree-1an2datasets_phen_lin.pngefficiency_90degree-1an2datasets_phen_lin_bigRange.png
For the efficiency vs E fit of the 90degree Xray detector I have used the following phenomenological funciton:

f(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d)
    Reply  Thu Apr 30 17:40:57 2020, Laszlo, Calibration, efficiency fits,  90degree_efficiency.zip145degree_efficiency.zip35degree_efficiency.zip
For all 3 detector the calibration data sets were combined to include the systematics in the fit results directly. Combining means not a weighted average,
just simple all data points were included into the fit --> doubled efficiency value for most of the energies.
The g(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d) function was used to describe the behavior of the Germanium detectors for the whole range of energies (global
       Reply  Thu Apr 30 22:55:45 2020, Laszlo, Calibration, inverse square law test,  inverse_square_law.zip
I have made also the inverse square law fits. We have data only for 90degre with the 241Am source, but both for the 1. and 2. calibrations. The 1. and 2.
calibration data sets treated separate.
2 peaks were investigated, 59.5keV and 26.3keV, at two distances 184.8mm and 217.3mm. These distances are the sum of 4 distances:
          Reply  Wed May 6 23:02:42 2020, Laszlo, Calibration, efficiency values,  90degree_efficiency_combined.png90degree_efficiency_combined.ps
I have calculated the efficiency values by putting the energies of the K-REC peaks into the phenomenological (empirical) function and into the linear function
(see below). The K-REC peak's position I got from a Gaussian-fit on the peak. There are 5 data sets in total:
With 1. E-calibration parameters:
Entry  Thu Apr 23 19:09:58 2020, Laszlo, Calibration, theoretical K-REC cross sections ,  9x
Find attached Andrey Surzhykov's calculations for the theta angle in respect to the beam direction (in lab. frame) vs. cross section for 124Xe54+ and 118Te52+.
The calculations made for collision with two H atoms with the accuracy of 1%. There are no molecular corrections done, but these corrections are within
1%.
Entry  Tue Apr 21 22:25:00 2020, Jan, Analysis, ,  
 
Entry  Wed Apr 8 23:02:24 2020, Laszlo, General, lmd to measurement pairing list,  gas_jet_mailfunction.png
I use the following list of lmd files combined to a single root file for each measurement.
It is very important to highlight that sometimes the trigger==1 data got corrupted,massive number of trigger==2 like events appear on the detector as trigger==1
signal, because the gas jet target didn't switch off after its normal phase (gas jet remains ON during injection. Maybe some gas jet issue, maybe some
Entry  Wed Apr 8 20:06:10 2020, Laszlo, Calibration, Counts in the K-REC peaks,  124Xe_wScraper_full_spectra_labels_thick.png124Xe_wScraper_90_backgnd.pngKREC-pics.zip
3 datasets were investigated:
-124Xe with Scraper    (2. Xray calibration parameters needed)
-124Xe without Scraper (2. Xray calibration parameters needed)
Entry  Mon Apr 6 23:00:51 2020, Laszlo, Calibration, beam energies for 124Xe and 118Te measurements,  SP54120040623030.pdf
"The e- cooler settings were the same for the 124Xe and 118Te beams" - Sergey --> speed of the ions is the same --> same MeV/u

E_beam = 10.0606MeV/u
ELOG V3.1.5-fc6679b