HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 15 of 29  Not logged in ELOG logo
Entry  Fri Apr 30 09:57:57 2021, Jan Glorius, Calibration, run0005 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 9:58:00 - 30.04.2021
Stop time:  10:01:09 - 30.04.2021

file name: e127b_run0005.lmd
avrg. rate: 50Hz
dead-time:  1%

Still rate peaks, probably the ESR kicker or some other external influence. 
So we have double triggers occasionally. This might slightly affect deadtime determination.
Entry  Fri Apr 30 09:27:27 2021, Jan Glorius, Calibration, run0004 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 9:26:58 - 30.04.2021
Stop time:  9:33:22 - 30.04.2021

file name: e127b_run0004.lmd
avrg. rate: 50Hz
dead-time:  1%

Still rate peaks, further investigations.
Entry  Thu Apr 29 20:34:09 2021, Jan Glorius, Calibration, run0003 - Xray90 calib Pb210 d=167.5mm, 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 20:39:54 - 29.04.2021
Stop time:  08:43:58 - 30.04.2021

file name: e127b_run0003.lmd
avrg. rate: 70Hz
dead-time:  1%

Rate spikes every ~2-3 sec. Need to be checked!
Entry  Thu Apr 29 20:32:52 2021, Jan Glorius, Calibration, calibration sources, 2021 
We use the following list of sources for calibration of the Xray detectors:

- 210Pb [40.1 kBq (4%), 01.10.2020 12:00 UTC] SpecSheet

- 241Am_low [40.5 kBq (3%), 01.10.2020 12:00 UTC] SpecSheet

- 241Am_high [389 kBq (3%), 01.10.2020 12:00 UTC] SpecSheet

- 133Ba_low [40.8 kBq (3%), 01.10.2020 12:00 UTC] SpecSheet

- 133Ba_high [404 kBq (3%), 01.10.2020 12:00 UTC] SpecSheet
Entry  Wed Apr 28 09:12:47 2021, Jan, Yuri, Detectors, UI-diagram, 2021 UI_meas_curve.pdf
The UI-curve of the detector
U I
10 0.11
20 0.13
30 0.15
40 0.17
50 0.19
60 0.21
70 0.22
80 0.23
90 0.25
100 0.27
110 0.29
120 0.31
130 0.33
140 0.35
150 0.37
Entry  Thu Apr 22 10:56:20 2021, Jan, Detectors, target-chamber det. distances, 2021 E127_xray_distances.jpg
This is the lookup table for the distance of each x-ray detector to the target.
See attached foto for explanation.

It is the same values as in 2020: https://elog.gsi.de/esr/E127/36

Detector A (dA) [mm] B (dB) [mm] C (dC) [mm] calculate result [mm]
35° 136.0 (1.0) 20.0 (0.2) 450.0 (0.5) C + B - A 334.0 (?)
90° 76.0 (0.5) 20.0 (0.2) 447.0/2 = 223.5 (1.0) C + B - A 167.5 (?)
145° 165.0 (1.0) 20.0 (0.2) 450.0 (0.5) C + B - A 305.0 (?)

Errors are estimated after measurement, final errorbars have to be double checked!
Entry  Wed Apr 14 09:42:50 2021, Jan, Detectors, BaF HV gain matching, 2021 BaF_Na22_gain_match_v1.pngbaf_gain_match_v1.dat
The preliminary gain matching settings for a full ADC range of ~12MeV:

Det.Nr. ADC/TDC ch HV I comment
1 20 2400V 390uA CAEN HV ch0
2 21 1763V 287uA CAEN HV ch1
3 22 2340V 380uA CAEN HV ch2
4 23 2175V 354uA CAEN HV ch3
5 24 2017.5V ? emetron HV
6 25 1780V ? not yet biased


Attached is a 22Na-spectrum taken with detectors 1 - 5.
The used MSCF settings are also attached.

Detector 6 does not yet have a HV channel available.
Entry  Wed Mar 31 13:57:27 2021, Jan, Detectors, BaF HV settings, 2021 
The setting below is a start value for final gain matching with a full ADC range of 12MeV:

Det.Nr. ADC/TDC ch HV MSCF Gain
1 20 2400V 6
2 21 1780V 6
3 22 2350V 6
4 23 2130V 6
5 24 2050V 6
6 25 1780V 6
Entry  Wed Mar 31 13:52:04 2021, Jan, Detectors, BaF Na22 Testing, 2021 
On 30. and 31.03.2021 we did some testing with the 6 BaF detectors.

Trigger & threshold:
With a threshold at ~250 keV, each detector has an trigger rate of about 0.5kHz due to internal activity.

Energy spectra
All detectors showed the Na22 lines (511 + 1275 keV) and also the lines from internal activity.
The energy resolution was on the order of 8.5 - 11 keV, depending on the detector.

Time spectra
Raw TDC spectra showed self-stop peak.
TDC diff spectra (between detectors) show one main (sharp) correlation peak about 8ns = 30ch*0.293ns/ch off the zero. 
At very low intensity a time structure around this peak is visible in a region of 120ns around the main peak. This is not understood, the energy spectra look similar for all these time regions (above/on/below the main peak).
Additionally, the coincidence E-specta between detector 1 and 2 (gating on the main time peak), nicely showed only the 511 keV line from Na22 and some gras.

Overnight test run:
Det. 1, 2, 3, 4 with the Na22 source

lxg1275:/data.local3/e127/lmd_2021/test/
e127b_run0021.lmd
e127b_run0022.lmd
e127b_run0023.lmd
e127b_run0024.lmd
e127b_run0025.lmd
e127b_run0026.lmd
e127b_run0027.lmd
e127b_run0028.lmd
e127b_run0029.lmd
e127b_run0030.lmd
e127b_run0031.lmd
e127b_run0032.lmd
e127b_run0033.lmd
e127b_run0034.lmd

Gain matched HV settings: https://elog.gsi.de/esr/E127/289

Issues to solve:
- 2 missing HV channels
- Go4 Analysis not stable
- trigger not on Baf but on Si
Entry  Tue Feb 2 13:59:00 2021, Jan, Detectors, DSSD installation and alignment, 2021 
The DSSD has been exchanged and aligned in 
November 2020. After bakeout at 140°C 
(externally, 120°C at internal temp. Sensor) 
for more than a week, the vacuum in the setup 
is roughly 4.5e-10 mbar.

After the bakeout the detector had to be 
realigned, it was lower by 1-2 mm. Using the 
line laser the realignment was done by 
touching only the screws on the far part of 
the base of the flange. These screws have been 
thightend by about a 1/4 turn after releasing 
the headless positioning screw accordingly.
Entry  Tue Nov 24 15:35:49 2020, Jan, Calibration, DSSD X/Y channel mapping, 2021 
During the detector test test measurements with alpha source the allocation of the 16 X- and 16 Y-channels has been checked.

For the following final allocation, it is always assumed that the (horizontal) y-strips are placed to face the beam directly, while the (vertical) x-strips are on the backside.

Now, all cables from the preamp to the ADC/TDCs are either labeled BLACK (= X-strips, pos. signals) or labeled RED (= Y-strips, neg. signals). These red or black connections should be kept consistently in order to ensure a well known orientation of the DSSD during the experiment. The test run Si2_run006.lmd was taken with this final assignment and serves as reference.

BLACK LABEL > X1 to X16 > pos. MSCF > ADC/TDC ch 0-15 > 50 Ohm resistor at preamps HV-input
RED LABEL > Y1 to Y16 > neg. MSCF > ADC/TDC ch 16-31 > neg. bias voltage at preamps HV-input

Additionally, the RED label indicates the section on the preamp to which negative bias voltage should be applied.
Entry  Fri Oct 23 12:58:12 2020, Jan, Detectors, Test of 2nd DSSSD (gen2), 2021 Si2_run001.jpgSi2_run002.jpg
This is the documentation of the source tests with the 2nd micron DSSSD of 2nd generation (label 3288-17, thickness 529um)

The detector is put into vacuum (~5e-6 mbar) in our test chamber. The source is positioned a few cm above (see fotos).
The Bayard-Alpert Sensor in the chamber has to be deactivated, otherwise the light emission will increase the noise on the DSSSD strongly and reduce its performance.

Additionally the current and voltage from the CAEN HV is monitored with the vulom scalers: ch.13(Icool) = current; ch.14(Ucool) = voltage. 


Source: mixes alpha [239Pu, 241Am, 244Cm]

File directory: lxg1275:/data.local3/test_data_2020/
QuickTest files: si2_test_mixed_source[X].root
lmd files:  Si2_run[XXX].lmd

DAQ Settings:
MADC gate   : 0 delay, 5000ns width
MSCF shaping: 2 us 
trlo config : e127.trlo (trigger=1/tpat=1 for Si, trigger=11 for vulom_scaler)


LMD runs:
-----------------------------
Si2_run001.lmd
Start: Fri 23.10.2020 16:19
Stop:  Mon 26.10.2020  8:36

File-Size: 20GB
Events:   ~53M
comment: 
source roughly centered
CAEN HV scalers not connected

-----------------------------
Si2_run002.lmd
Start: Tue 27.10.2020 16:01
Stop:  Wed 28.10.2020 13:26

File-Size: 12GB
Events:   ~32M
comment: 
source in one corner (x1, y1, see foto)
scalers should be connected now
current monitor range set to LOW
@start det_current=140nA det_voltage=90V
@end det_current=143nA det_voltage=90V

-----------------------------
Si2_run003.lmd
Start: Wed 28.10.2020 13:30
Stop:  Wed 28.10.2020 13:31

File-Size: 
Events:   
comment: 
ramping of det. voltage for scaler/U-F-Converter test
source in one corner (see foto)
scalers should be connected now
current monitor range set to LOW
@start det_current=140nA det_voltage=90V

-----------------------------
Si2_run004.lmd
Start: Tue 24.11.2020 12:50
Stop:  Tue 24.11.2020 

File-Size: 
Events:   
comment: 
source in one corner (x16, y16, see foto)
@start det_current=138nA det_voltage=90V
@end det_current=nA det_voltage=90V

-----------------------------
Si2_run005.lmd
Start: Tue 24.11.2020 14:57
Stop:  Tue 24.11.2020 15:11

File-Size: 
Events:   
comment: 
source in one corner (x1, y16, see foto)
@start det_current=144nA det_voltage=90V
@end det_current=nA det_voltage=90V

-----------------------------
Si2_run006.lmd
Start: Tue 24.11.2020 15:24
Stop:  Tue 24.11.2020 

File-Size: 
Events:   
comment: 
source in one corner (x1, y16, see foto)
Si_X & Si_Y cabling to MADC_0 and TDC_0 exchanged to get right order of channels/orientation
@start det_current=144nA det_voltage=90V
@end det_current=nA det_voltage=90V
    Reply  Fri Nov 20 15:53:37 2020, Jan, Analysis, Test of 2nd DSSSD (gen2), 2021 quick_si_plots.ce127.hSi2_run001.ana.rootSi2_run002.ana.root
Here is the analysis of the test runs with the 2nd DSSD (gen2) using the quick_si_plots.c script attached.

All 32 Si channels are working with acceptable performance. 

In run001, there are some additional low energy peaks in nearly all x-strips, which I do not understand yet. They are around 3 MeV and are not visible in the y-strips. It doesn't look like an electronic problem, because there are at least 4 
peaks, so not a low amplitude copy of the 3 major alpha peaks between 5 - 5.8 MeV.

However, in run002, the peaks have mostly disappeared, only in x6,x7,x8,x9 is a broad structure at somewhat similar energy... maybe this has to do with the small incident angle of the alphas?

Another run to confirm and double check this would be nice.
Entry  Fri Jun 5 14:06:15 2020, Laszlo, Detectors, DSSSD and SCRAPER position estimate for Xe and Te experiments, 2020 SP54120060514100.pdfDSSSD_new1.pdfIMG_20200321_025644.jpg
We don't know the exact absolute positions of the detector (+scraper) and the beam. However, what we have to know is only these two relative positions respect to each other. To get this distance I use two methods:
1, combining the infos from the set position during the beamtime + the measured pg peak position on the detector. The pg peak position is defined only by the eye (because of the low number of counts in every case, it doesnt make much sense to make fits). Since we rely on the detector resolution, we would be never more accurate than ~ +/-1.5mm anyhow. The active area of the detector is 49.5x49.5mm2 with a 45° tilt in y.
2, MOCADI simulation of the beam and the pg peak. This is used only as a crosscheck.
3, The scraper had a small angle in y direction causing ~0.5cm shift to the upper direction. the length of the scraping edge is 7cm


-124Xe with scraper measurement:
  • measurement
    d1 = moved back from beam = 15 +/-.5 mm
    d2 = DSSSD frame width = 8.85 mm
    d3 = pg center on DSSSD = 7-7.5 bin = 21.7-23.2 mm = avg = 22.5 mm

    --> pg from beam in x = -46.4mm +/- 1.5mm
    --> pg on DSSSD from center ~ -3.28mm +/- 1.5mm

  • simulation
    x = -46.5 mm
    y = 0 mm

  • detector active area position
    x = (-73.35mm) - (-23.85mm)
    y = (-14.2195mm) - 23.5125mm


  • SCRAPING: x=-35mm +/-0.5mm away from beam
    y=(-20mm) - (40mm)


-118Te:
  • measurement
    d1 = moved back from beam = 16 +/-.5 mm
    d2 = DSSSD frame width = 8.85 mm
    d3 = pg center on DSSSD = 7.5 bin = 23.2 mm

    --> pg from beam in x = -48.05mm +/- 1.5mm
    --> pg on DSSSD from center ~ -3.28mm +/- 1.5mm

  • simulation
    x = -48 mm
    y = 0 mm

  • detector active area position
    x = (-74.35mm) - (-24.85mm)
    y = (-14.2195mm) - 23.5125mm


  • SCRAPING:
    x=-35mm +/-0.5mm away from beam
    y=(-20mm) - (40mm)



notes during beam-time:
https://elog.gsi.de/esr/E127/97?suppress=1
Entry  Tue May 19 16:48:15 2020, Laszlo, General, E127 beamtime overview,  E127_beamtime_overview.pngE127_beamtime_overview.pdf
Here is a representation how was the time management during E127. The time, what we could spend with measuring the 118Te(pg), was ~20% comparing to the given 6days.
Entry  Wed Apr 29 16:06:28 2020, Laszlo, Calibration, efficiency fit - 90degree, combined dataset,  efficiency_90degree-1an2datasets_phen_lin.pngefficiency_90degree-1an2datasets_phen_lin_bigRange.png
For the efficiency vs E fit of the 90degree Xray detector I have used the following phenomenological funciton:

f(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d)

Here the first exponent member is a saturation curve. This part describes the passing through of the two Be windows (chamber + before detector) and through the dead layer of Ge crystal. One needs a minimum energy to enter to the detecting Ge crystal = C parameter. b parameter = characteristic absorbtion E of these nondetecting layers.
The second exponent is an exponential decrease of the detector efficiency. Photons with higher energy are less detectable by the germaniums. The d parameter is the characteristic E for hard Xray and gamma (>40keV) detectability. 

https://www.amptek.com/internal-products/si-pin-vs-cdte-comparison

//Jan's comment: the tail of this function should more or less follow a linear trend a bit above than 40 keV.


In the attachment there is an example fit for 90 degree with combined 1. and 2. (before and after beamtime) calibration datasets. 
I made the fit with gnuplot:

degrees of freedom    (FIT_NDF)                        : 8
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.419915
variance of residuals (reduced chisquare) = WSSR/ndf   : 0.176329
p-value of the Chisq distribution (FIT_P)              : 0.994094

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
a               = 0.00308376       +/- 0.0005144    (16.68%)
c               = 15.6259          +/- 2.035        (13.03%)
b               = 9.36888          +/- 3.16         (33.73%)
d               = 177.141          +/- 84.95        (47.95%)

Laszlo's out.
    Reply  Thu Apr 30 17:40:57 2020, Laszlo, Calibration, efficiency fits,  90degree_efficiency.zip145degree_efficiency.zip35degree_efficiency.zip
For all 3 detector the calibration data sets were combined to include the systematics in the fit results directly. Combining means not a weighted average, just simple all data points were included into the fit --> doubled efficiency value for most of the energies.
The g(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d) function was used to describe the behavior of the Germanium detectors for the whole range of energies (global behavior). From these fits the 80keV outlier point was excluded. This is very strange that it doesn't follow the trend, it would be nice to find out why not.
Between 40-75keV a linear fit was carried out as well. This can also approximate quite well this local energy range, what we need for the K-REC peaks.
l(x) = m*x+e

All the fits were done by gnuplot, but it was also confirmed that ROOT gives us the same parameters + errors + chisquare. One just need to choose well the starting values :)


90° fits:
degrees of freedom    (FIT_NDF)                        : 8
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.419915
variance of residuals (reduced chisquare) = WSSR/ndf   : 0.176329
p-value of the Chisq distribution (FIT_P)              : 0.994094

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
a               = 0.00308376       +/- 0.0005144    (16.68%)
c               = 15.6259          +/- 2.035        (13.03%)
b               = 9.36888          +/- 3.16         (33.73%)
d               = 177.141          +/- 84.95        (47.95%)

degrees of freedom    (FIT_NDF)                        : 4
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.283372
variance of residuals (reduced chisquare) = WSSR/ndf   : 0.0802998
p-value of the Chisq distribution (FIT_P)              : 0.988405

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
m               = -6.22672e-06     +/- 1.836e-06    (29.48%)
e               = 0.00256426       +/- 9.873e-05    (3.85%)



145° fits:
degrees of freedom    (FIT_NDF)                        : 6
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.379472
variance of residuals (reduced chisquare) = WSSR/ndf   : 0.143999
p-value of the Chisq distribution (FIT_P)              : 0.990247

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
a               = 0.0012346        +/- 0.0006794    (55.03%)
c               = 11.3754          +/- 4.075        (35.82%)
b               = 15.6961          +/- 13.2         (84.1%)
d               = 116.084          +/- 99.83        (86%)

degrees of freedom    (FIT_NDF)                        : 2
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.327728
variance of residuals (reduced chisquare) = WSSR/ndf   : 0.107405
p-value of the Chisq distribution (FIT_P)              : 0.898161

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
m               = -2.39246e-06     +/- 1.278e-06    (53.42%)
e               = 0.000848859      +/- 7.329e-05    (8.634%)



35° fit:
degrees of freedom    (FIT_NDF)                        : 7
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 1.40676
variance of residuals (reduced chisquare) = WSSR/ndf   : 1.97896
p-value of the Chisq distribution (FIT_P)              : 0.0538635

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
a               = 0.000114396      +/- 0.0002367    (206.9%)
c               = 14.3178          +/- 8.322        (58.13%)
b               = 16.7454          +/- 43.21        (258%)
d               = 79.7991          +/- 173.2        (217.1%)

degrees of freedom    (FIT_NDF)                        : 3
rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 1.21233
variance of residuals (reduced chisquare) = WSSR/ndf   : 1.46975
p-value of the Chisq distribution (FIT_P)              : 0.220529

Final set of parameters            Asymptotic Standard Error
=======================            ==========================
m               = -3.7879e-07      +/- 3.143e-07    (82.99%)
e               = 7.31447e-05      +/- 1.788e-05    (24.45%)



> For the efficiency vs E fit of the 90degree Xray detector I have used the following phenomenological funciton:
> 
> f(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d)
> 
> Here the first exponent member is a saturation curve. This part describes the passing through of the two Be windows (chamber + before detector) and through the dead layer of Ge crystal. One needs a minimum energy to enter to the detecting Ge crystal = C parameter. b parameter = characteristic absorbtion E of these nondetecting layers.
> The second exponent is an exponential decrease of the detector efficiency. Photons with higher energy are less detectable by the germaniums. The d parameter is the characteristic E for hard Xray and gamma (>40keV) detectability. 
> 
> https://www.amptek.com/internal-products/si-pin-vs-cdte-comparison
> 
> //Jan's comment: the tail of this function should more or less follow a linear trend a bit above than 40 keV.
> 
> 
> In the attachment there is an example fit for 90 degree with combined 1. and 2. (before and after beamtime) calibration datasets. 
> I made the fit with gnuplot:
> 
> degrees of freedom    (FIT_NDF)                        : 8
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.419915
> variance of residuals (reduced chisquare) = WSSR/ndf   : 0.176329
> p-value of the Chisq distribution (FIT_P)              : 0.994094
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> a               = 0.00308376       +/- 0.0005144    (16.68%)
> c               = 15.6259          +/- 2.035        (13.03%)
> b               = 9.36888          +/- 3.16         (33.73%)
> d               = 177.141          +/- 84.95        (47.95%)
> 
> Laszlo's out.
       Reply  Thu Apr 30 22:55:45 2020, Laszlo, Calibration, inverse square law test,  inverse_square_law.zip
I have made also the inverse square law fits. We have data only for 90degre with the 241Am source, but both for the 1. and 2. calibrations. The 1. and 2. calibration data sets treated separate.
2 peaks were investigated, 59.5keV and 26.3keV, at two distances 184.8mm and 217.3mm. These distances are the sum of 4 distances:
a = width of plastic head. uncert.: +/- 0.05mm, measured with caliper.
b = width of the brass collimator. uncert.: +/- 0.05mm, measured with caliper.
c = width of protector plastic ring. uncert.: +/- 0.05mm, measured with caliper.
d= distance of the paper head. uncert: +/- 0.5mm, judged by the eye :)

The Be window and the dead layer of the Ge detector is not taken into account.

The distances calculated as D=a+b+c+d.

The conclusion is that the uncertainty coming from the distance measurements are negligible compared to the other uncertainties. The data obey the inverse square law.


> For all 3 detector the calibration data sets were combined to include the systematics in the fit results directly. Combining means not a weighted average, just simple all data points were included into the fit --> doubled efficiency value for most of the energies.
> The g(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d) function was used to describe the behavior of the Germanium detectors for the whole range of energies (global behavior). From these fits the 80keV outlier point was excluded. This is very strange that it doesn't follow the trend, it would be nice to find out why not.
> Between 40-75keV a linear fit was carried out as well. This can also approximate quite well this local energy range, what we need for the K-REC peaks.
> l(x) = m*x+e
> 
> All the fits were done by gnuplot, but it was also confirmed that ROOT gives us the same parameters + errors + chisquare. One just need to choose well the starting values :)
> 
> 
> 90° fits:
> degrees of freedom    (FIT_NDF)                        : 8
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.419915
> variance of residuals (reduced chisquare) = WSSR/ndf   : 0.176329
> p-value of the Chisq distribution (FIT_P)              : 0.994094
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> a               = 0.00308376       +/- 0.0005144    (16.68%)
> c               = 15.6259          +/- 2.035        (13.03%)
> b               = 9.36888          +/- 3.16         (33.73%)
> d               = 177.141          +/- 84.95        (47.95%)
> 
> degrees of freedom    (FIT_NDF)                        : 4
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.283372
> variance of residuals (reduced chisquare) = WSSR/ndf   : 0.0802998
> p-value of the Chisq distribution (FIT_P)              : 0.988405
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> m               = -6.22672e-06     +/- 1.836e-06    (29.48%)
> e               = 0.00256426       +/- 9.873e-05    (3.85%)
> 
> 
> 
> 145° fits:
> degrees of freedom    (FIT_NDF)                        : 6
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.379472
> variance of residuals (reduced chisquare) = WSSR/ndf   : 0.143999
> p-value of the Chisq distribution (FIT_P)              : 0.990247
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> a               = 0.0012346        +/- 0.0006794    (55.03%)
> c               = 11.3754          +/- 4.075        (35.82%)
> b               = 15.6961          +/- 13.2         (84.1%)
> d               = 116.084          +/- 99.83        (86%)
> 
> degrees of freedom    (FIT_NDF)                        : 2
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.327728
> variance of residuals (reduced chisquare) = WSSR/ndf   : 0.107405
> p-value of the Chisq distribution (FIT_P)              : 0.898161
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> m               = -2.39246e-06     +/- 1.278e-06    (53.42%)
> e               = 0.000848859      +/- 7.329e-05    (8.634%)
> 
> 
> 
> 35° fit:
> degrees of freedom    (FIT_NDF)                        : 7
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 1.40676
> variance of residuals (reduced chisquare) = WSSR/ndf   : 1.97896
> p-value of the Chisq distribution (FIT_P)              : 0.0538635
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> a               = 0.000114396      +/- 0.0002367    (206.9%)
> c               = 14.3178          +/- 8.322        (58.13%)
> b               = 16.7454          +/- 43.21        (258%)
> d               = 79.7991          +/- 173.2        (217.1%)
> 
> degrees of freedom    (FIT_NDF)                        : 3
> rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 1.21233
> variance of residuals (reduced chisquare) = WSSR/ndf   : 1.46975
> p-value of the Chisq distribution (FIT_P)              : 0.220529
> 
> Final set of parameters            Asymptotic Standard Error
> =======================            ==========================
> m               = -3.7879e-07      +/- 3.143e-07    (82.99%)
> e               = 7.31447e-05      +/- 1.788e-05    (24.45%)
> 
> 
> 
> > For the efficiency vs E fit of the 90degree Xray detector I have used the following phenomenological funciton:
> > 
> > f(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d)
> > 
> > Here the first exponent member is a saturation curve. This part describes the passing through of the two Be windows (chamber + before detector) and through the dead layer of Ge crystal. One needs a minimum energy to enter to the detecting Ge crystal = C parameter. b parameter = characteristic absorbtion E of these nondetecting layers.
> > The second exponent is an exponential decrease of the detector efficiency. Photons with higher energy are less detectable by the germaniums. The d parameter is the characteristic E for hard Xray and gamma (>40keV) detectability. 
> > 
> > https://www.amptek.com/internal-products/si-pin-vs-cdte-comparison
> > 
> > //Jan's comment: the tail of this function should more or less follow a linear trend a bit above than 40 keV.
> > 
> > 
> > In the attachment there is an example fit for 90 degree with combined 1. and 2. (before and after beamtime) calibration datasets. 
> > I made the fit with gnuplot:
> > 
> > degrees of freedom    (FIT_NDF)                        : 8
> > rms of residuals      (FIT_STDFIT) = sqrt(WSSR/ndf)    : 0.419915
> > variance of residuals (reduced chisquare) = WSSR/ndf   : 0.176329
> > p-value of the Chisq distribution (FIT_P)              : 0.994094
> > 
> > Final set of parameters            Asymptotic Standard Error
> > =======================            ==========================
> > a               = 0.00308376       +/- 0.0005144    (16.68%)
> > c               = 15.6259          +/- 2.035        (13.03%)
> > b               = 9.36888          +/- 3.16         (33.73%)
> > d               = 177.141          +/- 84.95        (47.95%)
> > 
> > Laszlo's out.
          Reply  Wed May 6 23:02:42 2020, Laszlo, Calibration, efficiency values,  90degree_efficiency_combined.png90degree_efficiency_combined.ps
I have calculated the efficiency values by putting the energies of the K-REC peaks into the phenomenological (empirical) function and into the linear function (see below). The K-REC peak's position I got from a Gaussian-fit on the peak. There are 5 data sets in total:
With 1. E-calibration parameters:
-124Xe low rate measurement
-118Te 1. data set (before cable-swap)
With 2. E-calibration parameters:
-118Te 2. data set (after cable-swap)
-124Xe with scraper
-124Xe without scraper

When using the 1. E-calibration parameters, the obtained K-REC energies were much offset (124Xe_lowRate E_KREC=4.00577e+01keV, 118Te_1dataset: E_KREC=3.70922e+01keV), even though I tried all possible linear combinations of the parameters for 90° and 145°. Therefore, at the end I used the theoretical energies of the K-REC peaks from Thomas's website: http://www-ap.gsi.de/Thomas/ap_html_research/energy/index.php
Most probably, the problem is not with the 1. E-calibration itself (the source measurement looks consistent), but with the changing gate width during these measurements. These problematic data sets I marked with a " * " in the tables.

90°:
124Xe_wScraper124Xe_woScraper118Te_wScraper_part1118Te_wScraper_part2124Xe_wScraper_lowRate
K-REC E [keV]4,6093E+014,6116E+01*4,3241E+014,3135E+01*4,6336E+01
g(E_KREC) (phenomenology)0,0022850,002285*0,0022890,002289*0,002284
l(E_KREC) (linear)0,0022770,002277*0,0022950,002296*0,002276
diff. betw. f(E_KREC) and g(E_KREC)0,35%0,35%0,26%0,29%0,38%


145°:
124Xe_wScraper124Xe_woScraper
K-REC E [keV]4,1292E+014,1275E+01
g(E_KREC) (phenomenology)0,0007360,000736
l(E_KREC) (linear)0,0007500,000750
diff. betw. f(E_KREC) and g(E_KREC)1,82%1,83%


35°:
124Xe_wScraper124Xe_woScraper
K-REC E [keV]5,2309E+015,2478E+01
g(E_KREC) (phenomenology)0,00005320,0000532
l(E_KREC) (linear)0,00005330,0000533
diff. betw. f(E_KREC) and g(E_KREC)0,15%0,13%



> I have made also the inverse square law fits. We have data only for 90degre with the 241Am source, but both for the 1. and 2. calibrations. The 1. and 2. calibration data sets treated separate.
> 2 peaks were investigated, 59.5keV and 26.3keV, at two distances 184.8mm and 217.3mm. These distances are the sum of 4 distances:
> a = width of plastic head. uncert.: +/- 0.05mm, measured with caliper.
> b = width of the brass collimator. uncert.: +/- 0.05mm, measured with caliper.
> c = width of protector plastic ring. uncert.: +/- 0.05mm, measured with caliper.
> d= distance of the paper head. uncert: +/- 0.5mm, judged by the eye Smile
>
> The Be window and the dead layer of the Ge detector is not taken into account.
>
> The distances calculated as D=a+b+c+d.
>
> The conclusion is that the uncertainty coming from the distance measurements are negligible compared to the other uncertainties. The data obey the inverse square law.
>
>
> > For all 3 detector the calibration data sets were combined to include the systematics in the fit results directly. Combining means not a weighted average, just simple all data points were included into the fit --> doubled efficiency value for most of the energies.
> > The g(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d) function was used to describe the behavior of the Germanium detectors for the whole range of energies (global behavior). From these fits the 80keV outlier point was excluded. This is very strange that it doesn't follow the trend, it would be nice to find out why not.
> > Between 40-75keV a linear fit was carried out as well. This can also approximate quite well this local energy range, what we need for the K-REC peaks.
> > l(x) = m*x+e
> >
> > All the fits were done by gnuplot, but it was also confirmed that ROOT gives us the same parameters + errors + chisquare. One just need to choose well the starting values Smile
> >
> >
> > 90° fits:
> > degrees of freedom (FIT_NDF) : 8
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 0.419915
> > variance of residuals (reduced chisquare) = WSSR/ndf : 0.176329
> > p-value of the Chisq distribution (FIT_P) : 0.994094
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > a = 0.00308376 +/- 0.0005144 (16.68%)
> > c = 15.6259 +/- 2.035 (13.03%)
> > b = 9.36888 +/- 3.16 (33.73%)
> > d = 177.141 +/- 84.95 (47.95%)
> >
> > degrees of freedom (FIT_NDF) : 4
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 0.283372
> > variance of residuals (reduced chisquare) = WSSR/ndf : 0.0802998
> > p-value of the Chisq distribution (FIT_P) : 0.988405
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > m = -6.22672e-06 +/- 1.836e-06 (29.48%)
> > e = 0.00256426 +/- 9.873e-05 (3.85%)
> >
> >
> >
> > 145° fits:
> > degrees of freedom (FIT_NDF) : 6
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 0.379472
> > variance of residuals (reduced chisquare) = WSSR/ndf : 0.143999
> > p-value of the Chisq distribution (FIT_P) : 0.990247
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > a = 0.0012346 +/- 0.0006794 (55.03%)
> > c = 11.3754 +/- 4.075 (35.82%)
> > b = 15.6961 +/- 13.2 (84.1%)
> > d = 116.084 +/- 99.83 (86%)
> >
> > degrees of freedom (FIT_NDF) : 2
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 0.327728
> > variance of residuals (reduced chisquare) = WSSR/ndf : 0.107405
> > p-value of the Chisq distribution (FIT_P) : 0.898161
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > m = -2.39246e-06 +/- 1.278e-06 (53.42%)
> > e = 0.000848859 +/- 7.329e-05 (8.634%)
> >
> >
> >
> > 35° fit:
> > degrees of freedom (FIT_NDF) : 7
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 1.40676
> > variance of residuals (reduced chisquare) = WSSR/ndf : 1.97896
> > p-value of the Chisq distribution (FIT_P) : 0.0538635
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > a = 0.000114396 +/- 0.0002367 (206.9%)
> > c = 14.3178 +/- 8.322 (58.13%)
> > b = 16.7454 +/- 43.21 (258%)
> > d = 79.7991 +/- 173.2 (217.1%)
> >
> > degrees of freedom (FIT_NDF) : 3
> > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 1.21233
> > variance of residuals (reduced chisquare) = WSSR/ndf : 1.46975
> > p-value of the Chisq distribution (FIT_P) : 0.220529
> >
> > Final set of parameters Asymptotic Standard Error
> > ======================= ==========================
> > m = -3.7879e-07 +/- 3.143e-07 (82.99%)
> > e = 7.31447e-05 +/- 1.788e-05 (24.45%)
> >
> >
> >
> > > For the efficiency vs E fit of the 90degree Xray detector I have used the following phenomenological funciton:
> > >
> > > f(x) = a * (1-exp(-(x-c)/b)) * exp(-x/d)
> > >
> > > Here the first exponent member is a saturation curve. This part describes the passing through of the two Be windows (chamber + before detector) and through the dead layer of Ge crystal. One needs a minimum energy to enter to the detecting Ge crystal = C parameter. b parameter = characteristic absorbtion E of these nondetecting layers.
> > > The second exponent is an exponential decrease of the detector efficiency. Photons with higher energy are less detectable by the germaniums. The d parameter is the characteristic E for hard Xray and gamma (>40keV) detectability.
> > >
> > > https://www.amptek.com/internal-products/si-pin-vs-cdte-comparison
> > >
> > > //Jan's comment: the tail of this function should more or less follow a linear trend a bit above than 40 keV.
> > >
> > >
> > > In the attachment there is an example fit for 90 degree with combined 1. and 2. (before and after beamtime) calibration datasets.
> > > I made the fit with gnuplot:
> > >
> > > degrees of freedom (FIT_NDF) : 8
> > > rms of residuals (FIT_STDFIT) = sqrt(WSSR/ndf) : 0.419915
> > > variance of residuals (reduced chisquare) = WSSR/ndf : 0.176329
> > > p-value of the Chisq distribution (FIT_P) : 0.994094
> > >
> > > Final set of parameters Asymptotic Standard Error
> > > ======================= ==========================
> > > a = 0.00308376 +/- 0.0005144 (16.68%)
> > > c = 15.6259 +/- 2.035 (13.03%)
> > > b = 9.36888 +/- 3.16 (33.73%)
> > > d = 177.141 +/- 84.95 (47.95%)
> > >
> > > Laszlo's out.
Entry  Thu Apr 23 19:09:58 2020, Laszlo, Calibration, theoretical K-REC cross sections ,  9x
Find attached Andrey Surzhykov's calculations for the theta angle in respect to the beam direction (in lab. frame) vs. cross section for 124Xe54+ and 118Te52+.
The calculations made for collision with two H atoms with the accuracy of 1%. There are no molecular corrections done, but these corrections are within 1%.

The photon-emission is symmetrical in the azimuthal phi angle, but asymetric in theta. The K-REC cross section is given for each integer theta angle. The problem is that our 90° and 145° xray detectors cover more integer theta-angles --> The disk shaped entrance window of the Xray-detectors is sliced for each covered theta angles, and the CS values are averaged together with the weights of the area of the corresponding slice. The 35° det had a non-disk shaped slit collimator! Was it aligned vertical or horizontal or random? I assumed that it had a vertical position --> only cs at theta=35° needs to taken into account

90°145°35°
weighted cs_Xe [barn/sr]128,44441,96745,550
weighted cs_Te [barn/sr]118,61638,91841,740


For the steradian values I calculated the area of the entrance window of the Xray detectors and I divided it with the area of the sphere with radius = distance between source and det. If r=radius of the entrance window, D=distance between source and det.: covered steradian = r^2*pi/(4*pi*D^2) [%], covered steradian = 4Pi*[r^2*pi/(4*pi*D^2)] [4pi]. This is valid for the 90° and 145° Xray detectors. The area of 35° det. was calculated individually.

90°145°35°
sr [%]0,002370,000760,00002
sr [4pi]0,02980,00960,00030


I converted barn/sr to barn in the following way: steradian[4pi]*weigthed_cs [barn/sr] = weigthed_cs [barn]. I am not 100% sure if I need here the steradian in [4pi] or in [%].

90°145°35°
weighted cs_Xe [barn]3,82470,40120,0137
weighted cs_Te [barn]3,53210,37200,0125
Entry  Tue Apr 21 22:25:00 2020, Jan, Analysis, ,  
 
Entry  Wed Apr 8 23:02:24 2020, Laszlo, General, lmd to measurement pairing list,  gas_jet_mailfunction.png
I use the following list of lmd files combined to a single root file for each measurement.
It is very important to highlight that sometimes the trigger==1 data got corrupted,massive number of trigger==2 like events appear on the detector as trigger==1 signal, because the gas jet target didn't switch off after its normal phase (gas jet remains ON during injection. Maybe some gas jet issue, maybe some bug in the pattern :/ ). To correct for this, one has to cut out these parts from the data. I did these cut outs by hand while looking at the event number vs I_ESR & density gas jet plot (see below). Probably this can be done also in a more automatized way, but I think this is not necessary. In the "124Xe with Scraper" data set it is only 1 time like this, in the 118Te data set 3 times. The other data sets seem fine to me.


-124Xe with scraper:
run090_0001.lmd run091_0001.lmd run092_0001.lmd run094_0001.lmd run095_0001.lmd run096_0001.lmd run098_0001.lmd run099_0001.lmd
if(!(i>1634880 && i<1652240)){do analysis} - (this if condition is valid only if one combines the lmd in numerical order)
(2. Xray calibration parameters)

-124Xe without scraper:

run100_0001.lmd run101_0001.lmd run102_0001.lmd run104_0001.lmd
if(true){do analysis}
(2. Xray calibration parameters)

-118Te with scraper:
run051_0001.lmd run053_0001.lmd run055_0001.lmd run057_0001.lmd run059_0001.lmd run064_0001.lmd run066_0001.lmd
run052_0001.lmd run054_0001.lmd run056_0001.lmd run058_0001.lmd run060_0001.lmd run065_0001.lmd run067_0001.lmd
Changing Xray cables!
run068_0001.lmd run070_0001.lmd run072_0001.lmd run074_0001.lmd run076_0001.lmd run078_0001.lmd run080_0001.lmd run082_0001.lmd run084_0001.lmd run086_0001.lmd
run069_0001.lmd run071_0001.lmd run073_0001.lmd run075_0001.lmd run077_0001.lmd run079_0001.lmd run081_0001.lmd run083_0001.lmd run085_0001.lmd run087_0001.lmd

For Xray analysis:
1. dataset:
Xray[2] = 90 degree Xray[1]=145degree. The timing is switched:
if(!(i>1801830 && i<1807810) && !(i>2348370 && i<2355110)){
if(t_Xray[2]>0) Xray[1] ->Fill();
if(t_Xray[1]>0) Xray[2] ->Fill();
}
(1. Xray calibration parameters --> invalid!)
2. dataset:
if(!(i>5488450-2507171 && i<6125720-2507171)){do analysis}
(2. Xray calibration parameters)

For Si analysis:
if(trigger==1 && !(i>1801830 && i<1807810) && !(i>2348370 && i<2355110) && !(i>5488450 && i<6125720) ){do analysis}
(1. Xray calibration parameters)

-124Xe with scraper - low rate measurement
run046_0001.lmd run047_0001.lmd run047_0003.lmd run048_0002.lmd run049_0002.lmd run050_0001.lmd
run046_0002.lmd run047_0002.lmd run048_0001.lmd run049_0001.lmd run049_0003.lmd run050_0002.lmd
if(true){do analysis}
(1. Xray calibration parameters --> invalid!)


Regarding the gain matching, I assumed that the same 2*16 factors can be used for all data sets, since we didn't change bias voltage (the current remained roughly also the same) and also the detector didn't get any serious radiation damage (this needs to be confirmed!).
A more detailed anaylsis will come on the gain matching after the Easter holiday.

Entry  Wed Apr 8 20:06:10 2020, Laszlo, Calibration, Counts in the K-REC peaks,  124Xe_wScraper_full_spectra_labels_thick.png124Xe_wScraper_90_backgnd.pngKREC-pics.zip
3 datasets were investigated:
-124Xe with Scraper (2. Xray calibration parameters needed)
-124Xe without Scraper (2. Xray calibration parameters needed)
-118Te with Scraper (1. Xray calibration parameters needed)

I looked all the 35°, 90° and 145° detector spectra:
-For both Xe measurements all 3 detector signal can be evaluated
-For Te beam one can see only in the 145° detector spectra the Kalpha and K-REC peaks (with high uncertainty). For the 35° probably the detector was simply not sensitive enough for such low beam intensities. For the 90° case, I am much surprised, the peaks supposed to be there the most prominent of all. In the spectra, I maybe can recognize a peak at ~27,8keV, but this is even in best case only the Kalpha peak. At the range of the expected K-REC (~40keV), there is a bit of increase in the background overlying the peak. This background increase is also in the background spectra. Also probably this twisted cables issue between 90° and 145° didn't help much. I think anyhow, that maybe this must have some noise related origin. I can remember that the cables (despite all of our and Uwe's tries) were not well grounded, the noise level was kind of floating.

In general, I would also remark that we can see some peaks >60keV in the background, but these luckily don't disturb us.


To evaluate the Xray spectra I used the following algorithm:
1, for each type of beams I used the list of event numbers in the next entry (to exclude "bad" events)
To get the Kalpha and K-REC and other peaks I used the condition trigger==1 (jet ON)
To get the background spectra, I used trigger==2 (Jet OFF). The background spectra is only used to see that there is no underlying peak structure below K-REC. To subtract count, the background histo was NOT in use.
2, While using a well-suiting number of bins, I plot the JetON and JetOFF histos.
By eye I choose the range of the K-REC peak and the range for the background fit on the JetON histo. Ofc range_bckgnd > range_peak.
Simultaneously, I check on the JetOFF histo that both, in the fit-range and in the peak-range, there should not be any peak structure visible.
3, For the fit-range in JetON histo, excluding the peak-range, I fit a linear function, m*x+b. For each bin in the fit-range I subtract m*bin_center+b value from the bin content. After the subtraction I check if I got spectra looking like a single peak sitting on a zero
line.
4, To get the K-REC counts, I sum together all the bin values for each bin of the subtracted histo within the peak-range. For the error calculation, I use Gaussian error prop. The uncertainty of the JetON histo counts = sqrt(counts). Also for the subtraction I make the
error like delta(m*bin_center+b)= sqrt(m*bin_center+b) instead using the uncertainty of the fit parameters. This second one wont make much sense, since the slope of the linear fit is usually close to 0 --> the errors grow unrealistically big.


Based on the algorithm above I got the following counts:
-124Xe with Scraper:
35°: 174 +/- 15
90°: 21299 +/- 150
145°: 2104 +/- 52

-124Xe without Scraper:
35°: 65 +/- 9
90°: 7792 +/- 91
145°: 728 +/- 31

-118Te_part1 with Scraper:
35°: -
90°: 427 +/- 40
145°: -

-118Te_part2 with Scraper:
35°: -
90°: 741 +/- 48
145°: -

-124Xe_lowRate with Scraper:
35°: -
90°: 2121 +/- 52
145°: -
Entry  Mon Apr 6 23:00:51 2020, Laszlo, Calibration, beam energies for 124Xe and 118Te measurements,  SP54120040623030.pdf
"The e- cooler settings were the same for the 124Xe and 118Te beams" - Sergey --> speed of the ions is the same --> same MeV/u

E_beam = 10.0606MeV/u

The uncertainty on the cooloer values should be asked from Markus/Regina
ELOG V3.1.5-fc6679b