Mon Mar 23 16:17:58 2020, Laszlo, Calibration, run106 - garbage!,
|
run106 was clicked by accident, please delete! |
Mon Mar 23 16:21:33 2020, Laszlo, Calibration, run107 - Xray90 calib Am241, distance 167.5mm,
|
Detector: 90
Source: Am241
Distance: 167.5mm
|
Mon Mar 23 17:32:42 2020, Laszlo, Calibration, run108 - Xray90 calib Ba133, distance 167.5mm - high rate - deadtime too high!,
|
Detector: 90
Source: Ba133 - strong source
Distance: 167.5mm
|
Mon Mar 23 17:40:33 2020, Laszlo, Calibration, run109 - Xray90 calib Ba133, distance 167.5mm -low rate,
|
Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
|
Mon Mar 23 18:15:55 2020, Laszlo, General, noisy 90degree Xray det,
|
Uwe found out, that there was some common ground effect between 90 and 145 degree detectors. after moving the grounds, we could record much more nicer (Gaus
shaped) peaks in the spectra. Now we continue with this settings |
Mon Mar 23 18:17:27 2020, Laszlo, Calibration, run110 - Xray90 calib Ba133, distance 167.5mm -low rate - LESS NOISE,
|
Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
|
Mon Mar 23 19:32:53 2020, Laszlo, Calibration, run112 - Xray90 calib Am241, distance 200mm - LESS NOISE,
|
Detector: 90
Source: 241Am
Distance: 200mm
|
Mon Mar 23 20:52:37 2020, Laszlo, Calibration, run113 - Xray145 calib Am241, distance 305mm - LESS NOISE,
|
Detector: 145
Source: 241Am
Distance: 305mm
|
Mon Mar 23 21:47:40 2020, Laszlo, Calibration, run114 - Xray145 calib Ba133, distance 305mm - LESS NOISE,
|
Detector: 145
Source: 133Ba - strong source
Distance: 305mm
|
Mon Mar 23 22:25:41 2020, Laszlo, Calibration, run115 - Xray35 Ba133 d=334mm, Xray90 Lead source, d=167.5mm,
|
Simultaneous measurement of Xray35 and Xray90. The dead-time is <3%
Detector: 35
|
Tue Mar 24 00:11:35 2020, Laszlo, Calibration, run116 - Xray35 calib AM241, distance 334mm,
|
Detector: 35
Source: 241Am
Distance: 334mm
|
Tue Mar 24 10:20:26 2020, Laszlo, Calibration, Pictures of the xray calibration,
|
|
Wed Mar 25 14:26:22 2020, Laszlo, Calibration, run111 - Xray90 calib Am241, distance 167.5mm - LESS NOISE,
|
Detector: 90
Source: 241Am
Distance: 167.5mm
|
Wed Mar 25 15:06:09 2020, Laszlo, Analysis, quick summary of the run numbers,
|
|
Thu Mar 26 12:05:22 2020, Laszlo, Calibration, measurement of collimators,
|
The measurements were taken by using a caliper ruler, with precision of +/-0.05mm |
Sun Apr 5 20:34:42 2020, Laszlo, Calibration, Xray energies - 1. calibration, 8x
|
30angle: E[keV] = 0.0158507 *ch-2.07456
90angle: E[keV] = 0.01950*ch-2.76
145angle: E[keV] = 0.017988 *ch-1.809
|
Mon Apr 6 04:39:19 2020, Laszlo, Calibration, 2. Xray energy calibration, 9x
|
30angle: E[keV] = 0.015898 *ch-1.91
90angle: E[keV] = 0.0211075*ch-2.39
145angle: E[keV] = 0.01663 *ch-1.758 |
Mon Apr 6 23:00:51 2020, Laszlo, Calibration, beam energies for 124Xe and 118Te measurements,
|
"The e- cooler settings were the same for the 124Xe and 118Te beams" - Sergey --> speed of the ions is the same --> same MeV/u
E_beam = 10.0606MeV/u
|
Wed Apr 8 20:06:10 2020, Laszlo, Calibration, Counts in the K-REC peaks,
|
3 datasets were investigated:
-124Xe with Scraper (2. Xray calibration parameters needed)
-124Xe without Scraper (2. Xray calibration parameters needed)
|
Wed Apr 8 23:02:24 2020, Laszlo, General, lmd to measurement pairing list,
|
I use the following list of lmd files combined to a single root file for each measurement.
It is very important to highlight that sometimes the trigger==1 data got corrupted,massive number of trigger==2 like events appear on the detector as trigger==1
signal, because the gas jet target didn't switch off after its normal phase (gas jet remains ON during injection. Maybe some gas jet issue, maybe some |