Fri Jun 7 13:52:32 2019, Jan, Runs, prerun002 - MixedAlpha,
|
save point: lxg1275:/datalocal1/e127/predata/prerun002
~20min run
|
Fri Jun 7 14:28:09 2019, Jan, Runs, prerun003 - MixedAlpha,
|
save point: lxg1275:/datalocal1/e127/predata/prerun003
~20min run
|
Fri Jun 7 14:39:36 2019, Jan, Detectors, Pt100 installation & readout,
|
The new detector holder incorporates a Pt100 temperature sensor, which is UHV compatible. The Pt-resistor is read out via 4 wires. These wires are integrated
into the D-SUB connector of the detector. The new wires use the pins as shown in the attached picture
"det_connector.png".
|
Fri Jun 7 15:57:41 2019, Jan, DAQ, DAQ with & without MDPP-16,
|
The DAQ located in
../esrdaq_2018/r4l-58/
|
Tue Jun 11 10:13:56 2019, Jan, Runs, prerun003 - MixedAlpha,
|
save point: lxg1275:/datalocal1/e127/predata/prerun004
~10min run
|
Fri Jun 28 14:06:28 2019, Jan, General, CAD of (p,g) setup,
|
|
Thu Aug 29 10:02:22 2019, Jan, Detectors, bakeout at ESR setup,
|
The bakeout for our detector setup has started on the evening of 27.08.2019
We have a temperature sensor in the vacuum next to the detector.
The goal is to maximize bakeout temperature, while not going above 125°C at the detector/sensor.
|
Wed Oct 23 11:40:29 2019, Jan, Runs, prerun005 - MixedAlpha,
|
save point: lxg1275:/datalocal1/e127/predata/prerun005
start: 23.10.19 - 11:40
|
Wed Oct 23 15:35:29 2019, Jan, DAQ, DAQ with & without MDPP-16,
|
the DAQ located in
../esrdaq_2018/r4l-58_dev/
|
Fri Oct 25 16:34:51 2019, Jan, Runs, prerun007 - MixedAlpha,
|
save point: lxg1275:/datalocal1/e127/predata/prerun007
start: Fr 25.10.19 - 16:35
|
Tue Feb 11 13:20:01 2020, Jan, Detectors, BaF2 - HV settings,
|
with these HV values the 6 channels are roughly gain matched, such that the 7.83 MeV Po-decay is around channel 5000. |
Mon Feb 17 08:14:44 2020, Jan, Simulations, 118Te K-REC X-ray energies,
|
X-ray energies of K-REC for 118Te according to
[URL=http://www-ap.gsi.de/Thomas/ap_html_research/energy/index.php]http://www-ap.gsi.de/Thomas/ap_html_research/energy/index.php[/URL]
|
Fri Feb 21 11:29:59 2020, Jan, Detectors, det. distances at target chamber,
|
This is the lookup table for the distance of each x-ray detector to the target.
See attched foto for explanation.
|
Tue Mar 10 08:52:34 2020, Jan, Detectors, DSSSD positioning,
|
To find a good Si position, we need to follow a procedure similar to this:
1. Si in max-in position
|
Tue Mar 10 09:41:26 2020, Jan, Detectors, x-ray background from BaF,
|
The BaF2 detectors planned to be installed at the target induce an additional background in the x-ray detectors sitting below inside the target chamber.
We made test measurements with an x-ray detector for estimating the background in the region of interest for our K-REC measurements (35 to 50 keV).
|
Wed Mar 11 09:08:21 2020, Jan, DAQ, DAQ & controls - OVERVIEW,
|
For all DAQ & slow control buissines the user
[COLOR=red][SIZE=3]litv-exp[/SIZE][/COLOR]
is used with the password [SIZE=1]ViValasVegas1964[/SIZE]
|
Sun Mar 15 09:42:40 2020, Jan, Calibration, run030 - Xray90 calib Am241,
|
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
|
Sun Mar 15 13:08:38 2020, Jan, Calibration, run031 - Xray90 calib Ba133 hi-rate,
|
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
|
Sun Mar 15 13:40:11 2020, Jan, Calibration, run032 - Xray90 calib Ba133,
|
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
|
Sun Mar 15 13:52:05 2020, Jan, Calibration, run033 - Xray145 calib Am241,
|
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 145
|