Sun Apr 10 08:53:26 2022, Kristian, Max, Laser, Timing verified
|
timing is still good.
we checked the relative timing between laser and ions. the time difference between photodiodes, PMTs and particle detectors is still the same as original (see ELOG ID 169). The common stop of changed a slightly so that everything appears one bin later. |
Sun Apr 10 06:39:16 2022, Kristian, Marek, Max, Runs, runs morning shift 10.04.
|
These are the measurement runs of the morning shift (10.04.).
Time | SIS-intensity | cooler-HV | LMD-DAQ1 start-end | LMD-DAQ2 start-end | scan range(nm) | step width(nm) | dwell time(s) | comments
|
06:xx | 3e8 | 2096xx | 0141-0141 | 7876-7876 | 323.2 - 322.8 | 0.002 | 5s | just one loop
|
06:39 | 2e9 | 209621 | 0142-0142 | 7877-7877 | 323.2 - 322.8 | 0.002 | 5s |
|
07:18 | 2.1e9 | 209616 | 0143-0143 | 7878-7878 | 323.2 - 322.8 | 0.002 | 5s | 8000 particles on counter
|
07:57 | 2.2e9 | 209621 | 0144-0144 | 7879-7879 | 323.2 - 322.8 | 0.002 | 5s |
|
08:37 | 1.9e9 | 209620 | 0145-0145 | 7880-7880 | 323.2 - 322.8 | 0.002 | 5s | 7200 particles on counter
|
09:24 | 2.1e9 | 209621 | 0146-0146 | 7881-7881 | 323.2 - 322.8 | 0.002 | 5s | laser pulse energy @ start of run: 488mJ green monitor, 11.6mJ UV, 7700 particles on counte
|
10:07 | 2.1e9 | 209620 | 0147-0147 | 7882-7882 | 323.2 - 322.8 | 0.002 | 5s | optimized WM coupling, the WM data is stretched in this run (buffer problem?), also possible for previous runs
|
11:04 | 2.1e9 | 209620 | 0148-0148 | 7884-7884 | 323.2 - 322.8 | 0.002 | 5s | restarted WM, still wrong WM data, stopped after 1 run
|
11:27 | 2.1e9 | 209616 | 0149-0149 | 7885-7885 | 323.2 - 322.8 | 0.002 | 5s | restarted WM + LV
|
12:11 | 2.0e9 | 209616 | 0150-0150 | 7886-7886 | 323.2 - 322.8 | 0.002 | 5s | wrong WM data
|
12:55 | 1.9e9 | 209616 | 0151-0151 | 7887-7887 | 323.2 - 322.8 | 0.002 | 5s | changed LV code (ID: 198), WM data works now!
|
13:37 | 1.8e9 | 209620 | 0152-0152 | 7888-7888 | 323.2 - 322.8 | 0.002 | 5s |
|
14:16 | 2.1e9 | 209621 | 0153-0153 | 7889-7889 | 323.2 - 322.8 | 0.002 | 5s | 1 loop |
Sun Apr 10 06:34:49 2022, Kristian, Max, DAQ, Wavemeter Readout
|
Problem solved. Ignore this entry
If the wavemeter reading is not shown correctly in Go4 - Step2 - Labview:
Close Labview, close WM software (all windows)
Start WM software, start Labview, restart MBS |
Sun Apr 10 04:37:29 2022, Patrick, DAQ, Next wavelength range
|
Measurements with the next wavelength scan range should be started roughly around 12:00 noon. The plan is to go to lower wavelengths / higher energies.
So the next wavelength scan range would be 322.81 - 322.4 nm, again with a step size of 0.002 nm and 5s dwell time.
In the picture below you find some calculations/estimates for the current scan range. |
Sun Apr 10 04:07:53 2022, Patrick, DAQ, go4 shows wrong frequencies and wavelengths
|
Update (ID:194): In labview "Stop All", close labview, stop python script (ID:197), close wavemeter. Then restart wavemeter, python script, labview an mbs.
Both labview and the wavemeter software are displaying the scanned wavelength correctly (Fig. 1 & 2), but go4 shows something which looks just like a freely drifting wavelength (Fig. 3).
The Cobra set voltages are displayed correctly. |
Sun Apr 10 03:56:56 2022, V. Hannen, Detectors, Low rate in PMT South
|
PMT south shows a much lower count rate than observed in elog entry 170.
Yesterday the voltage on this PMT was increased to 2400 V and the threshold setting of the CAEN CFD to 90 units, (from 70 units before) however the low readout rate was already apparent both for the analog and the discriminated signal then.
Update Sun Apr 10, 21:45: on advice from D. Winters, increased the PMT voltage to 2500 V (max. the HV supply can do, the PMT could handle up to 3000 V) |
Sun Apr 10 02:20:17 2022, Patrick, DAQ, New mbs startup guide
|
Connecting to the rfio server and opening a file is included in '@startup' now. Since mbs crashes reliably every 3 or 4 runs, it may be preferable to use the following procedure (which did not crash yet):
| DAQ1 | DAQ2
|
1 | ssh atplaser@r4l-41 | ssh stoe_exp@r4l-68
|
2 | ThE142_2022 | see ID:179
|
3 | cd mbsrun/th22 | cd mbsrun/th22cd /esr/usr/stoe_exp/E142Brandau/SCA... (auto complete with tab)
|
// loop start | |
|
4 | resl | resl
|
5 | mbs | mbs
|
6 | @startup | @startup
|
7 | // connect go4 | // connect go4
|
8 | sta ac | sta ac
|
9 | // measure | // measure
|
10 | sto ac | sto ac
|
11 | clo fi | clo fi
|
12 | exit | exit
|
13 | // disconnect go4 | // disconnect go4
|
14 | // go to step 4 | // go to step 4
|
Note: At least in DAQ1, step 13 is necessary to be able to reconnect mbs to the rfio server. |
Sat Apr 9 19:50:40 2022, Carsten, , New thresholds for ADC of DAQ2 (starting with file0117)
|
unsigned short caen_V785AH_threshold_mod0 [32] =
{0x0100 , 0x0004 , 0x0003 , 0x0004 , 0x0005 , 0x0004 , 0x0006 , 0x0005 ,
0x0006 , 0x0005 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 }; |
Sat Apr 9 19:09:57 2022, Rodolfo, Laser, Optimization - Resonator of dye laser
|
I have tuned a bit the resonator of the dye laser.
The total output energy after the main amplifier is 86 mJ (@ 645,6 nm)
The laser wavelength read by the wavemeter appears to be more stable (on the windows computer). We can see the histogram once the present scanning is finished. |
Sat Apr 9 16:55:19 2022, Rodolfo, Laser,
|
16:15 I have changed the dye (of both oscillator and amplifier).
16:27 I have increased the pump energy (@ 532 nm) from 476 mJ to 507 mJ.
- UV output energy @ 322,8 nm: 12 mJ. (Before was at 9 mJ).
- New calibration factor is 160 for Quanta Ray monitor signal. |
Sat Apr 9 15:47:21 2022, Carsten, Rodolfo, Sarper, Patrick, Runs, Afternoon and night shift runs
|
After the morning shift, we got new increment (nm) for the laser scanning. We are now scanning 2 nm of wavelength range in 5 big steps (5 x 0.4 nm). Started with 323.2 nm - 322.8 nm range. Every big step (0.4 nm) is scanned with 0.002 nm of wavelength increment.
Every scan takes ~ 18 min (0.4 nm range with 0.002 nm steps), we have scanned 2 loops. Total scanning time ~36 min.
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th
Schottky marker pos: 243.973MHz
Time | SIS-intensity | cooler-HV | LMD-DAQ1 start-end | LMD-DAQ2 start-end | scan range(nm) | step width(nm) | dwell time(s) | comments
|
xx:xx | x.xe9 | 209xxx | 0xxx | 78xx | 32x.x - 32x.x | 0.002 | x | sketch for filling the table
|
15:14 | 1.5e9 | 209624 | 0111-0111 | 7847-7847 | 323.2 - 322.8 | 0.002 | 5s | DAQ 2 no VUPROM TDC Pat south
|
xx:xx | 1.5e9 | 209624 | 0111-0111 | 7847-7847 | 323.2 - 322.8 | 0.002 | 5s | DAQ 2 changed TDC part south from ch 8 to ch 10
|
17:05 | 2.2e9 | 209624 | 0113-0113 | 7848-7848 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
19:04 | 1.8e9 | 209616 | 0116-0116 | 7850-7850 | 323.2 - 322.8 | 0.002 | 5s | Intensity optimization, see below, DAQ 2 TDC part south back to channel 8 - ok, XUV detector and Cu mirror moved correctly, lost the laser beam after the half of the second loop
|
19:49 | 1.8e9 | 209614 | 0117-0117 | 7851-7851 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
20:28 | 2.0e9 | 209622 | 0118-0118 | 7852-7852 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
21:07 | 2.2e9 | 209623 | 0119-119 | 7853-7853 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
22:01 | 1.8e9 | 209621 | 0122-0122 | 7856-7856 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
22:41 | ? | 209620 | 0123 | 7857 | 323.2 - 322.8 | 0.002 | 5s | First scan started by night shift
|
23:21 | 1.1e9 | 209620 | 0124 | 7858 | 323.2 - 322.8 | 0.002 | 5s | mbs crashed at 'sta ac'
|
23:32 | 1.3e9 | 209618 | 0125 | 7859 | 323.2 - 322.8 | 0.002 | 5s | e-cooler turned off
|
23:50 | 1.5e9 | 209xxx | 0126 | 7860 | 323.2 - 322.8 | 0.002 | 5s | the stop point was not set
|
23:53 | 1.5e9 | 209611 | 0127 | 7861 | 323.2 - 322.8 | 0.002 | 5s | DAQ2 must have crashed between 23:30 - 23:50
|
00:07 | 1.9e9 | 209xxx | 0128 | 7862 | 323.2 - 322.8 | 0.002 | 5s | mbs crashed at 'sta ac'
|
00:23 | 1.7e9 | 209623 | 0131 | 7864 | 323.2 - 322.8 | 0.002 | 5s | successful scan. we 'resl' after every measurement from now on. 'open file' is included in '@startup' now (ID:190).
|
01:08 | 1.6e9 | 209623 | 0132 | 7865 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
01:47 | 1.6e9 | 209623 | 0133 | 7866 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
02:35 | 1.3e9 | 209623 | 0134 | 7868 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
03:19 | 2.0e9 | 209622 | 0135 | 7869 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
04:04 | 1.3e9 | 209622 | 0136 | 7870 | 323.2 - 322.8 | 0.002 | 5s | successful scan
|
04:44 | 1.0e9 | 209622 | 0137 | 7871 | 323.2 - 322.8 | 0.002 | 5s | e-cooler turned off
|
05:25 | 1.9e9 | 209622 | 0140 | 7875 | 323.2 - 322.8 | 0.002 | 5s | successful scan. DAQ1 was started ~7s before DAQ2. The files between the last measurement an this one are essentially empty.
|
Carsten (19:00h): I notized that a lot of beam gets lost when starting scraping. I called Sergey Litvinov to help me to introduce some additional cooling/waiting (10s) after the ramp and before scraping and bunching. It increased the number of ions available for the measurement roughly by a factor of 3-5.
Some runs have wrong WM data. Set Wavelength is good.
Following files are saved differently, the command missed the 229Th in the end (/data.local2/BeamTimeTh22/229Thor -auto -rfio)
7850 (19:32)
7855 (21:56)
7856 (22:30) |
Sat Apr 9 14:15:44 2022, Kristian, Max, Konstantin, DAQ, Change in LV code
|
We changed the LV code so that the Cobra Set Wavelength is permanently sent. Before it was only sent once at the beginning of each laser step. This, however, lead to too much data for the LV->MBS server.
This was fixed by inserting a 250 ms wait in LV (in Krypton data VI). Basically, all LV values are now updated every 250 ms, which works fine with MBS |
Sat Apr 9 13:45:42 2022, Max, Kristian, Runs, first scans with LV problem solved!
|
This is the first time with LV buffer problem solved (see Elog ID: 184). Before the PMT signals were out of sync with the wavelength data!
We concluded to scan with step sizes of 0.002nm, scan range of 0.5nm (f
nm), dwelltime = 150(5s), loops = 2 (approx.: 30min measurements).
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th
Shottky marker pos: 243.973MHz
Time | SIS-intensity | cooler-HV | LMD-DAQ1 start-end | LMD-DAQ2 start-end | scan range(nm) | step width(nm) | dwell time(s) | comments
|
11:25 | 1.0e9 | 209622 | 0102 - 102 | 7839-7839 | 323.2 - 322.8 | 0.002 | 3s | LV readout problem solved. WM readout is wrong (Cobra set is good)
|
12:15 | old beam | 209622 | 0103 - 103 | 7840-7840 | 323.2 - 322.8 | 0.002 | 3s | WM readout is wrong
|
12:50 | 2.3e9 | 209617 | 0104 - 104 | 7841-7841 | 323.2 - 322.8 | 0.002 | 3s | couldnt open go4 stream view, stopped after one good scan, WM readout is wrong .
|
13:20 | 2.0e9 | 209617 | 0104 - 105 | 7842-7842 | 323.2 - 322.8 | 0.002 | 3s | first scan with LV correct
|
13:40 | 1.2e9 | 209618 | 0106 - 106 | 7843-7843 | 323.2 - 322.8 | 0.002 | 3s | 3 loops
|
14:20 | 1.5e9 | 209612 | 0107 - 107 | 7844-7844 | 323.2 - 322.8 | 0.002 | 3s | 1 loop
|
Runs after this have the right cobra set value but propably the wrong WM data until 10.04.22 12:44 (see also ID:198) |
Sat Apr 9 06:42:23 2022, Kristian, Laser, Laser step size
|
We changed the laser step size to 0.002 nm which corresponds to 5.75 GHz. |
Sat Apr 9 06:25:22 2022, Max, Kristian, Runs, morning shift runs
|
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th
Shottky marker pos: 243.973MHz
Time | SIS-intensity | cooler-HV | LMD-DAQ1 start-end | LMD-DAQ2 start-end | scan range(nm) | step width(nm) | dwell time(s) | comments
|
xx:xx | x.xe9 | 209xxx | 00xx | 78xx | 32x.x - 32x.x | 0.01 | x | sketch for filling the table
|
06:40 | 1.7e9 | 209619 | 0098-0099 | 7836 | 323.2 - 321.2 | 0.002 | 4s | laser pulse energy is @ ~12mJ after SHG
|
09:35 | 1.4e9 | 209624 | 0101-0101 | 7838-7838 | 323.2 - 322.8 | 0.002 | 3s | only 0.5nm range, WM power + coupling tuned to scan range, Konstantin updated LV code to hold Cobra Set WL value, aborted LV readout problem
|
11:25 | 1.0e9 | 209622 | 0102 - 102 | 7839-7839 | 323.2 - 322.8 | 0.002 | 3s | LV readout problem solved. WM readout is wrong (Cobra set is good)
|
12:15 | old beam | 209622 | 0103 - 103 | 7840-7840 | 323.2 - 322.8 | 0.002 | 3s | WM readout is wrong
|
12:50 | 2.3e9 | 209617 | 0104 - 104 | 7841-7841 | 323.2 - 322.8 | 0.002 | 3s | couldnt open go4 stream view, stopped after one good scan
|
13:20 | 2.0e9 | 209617 | 0104 - 105 | 7842-7842 | 323.2 - 322.8 | 0.002 | 3s | first scan with LV correct
|
|
Sat Apr 9 05:26:25 2022, Patrick, Laser, Wavemeter
|
The wavemeter still shows large spikes from time to time. The exposure time is still at 34ms. |
Sat Apr 9 05:00:11 2022, Carsten, General, Draft Docs and cheat sheet
|
Please check the attached files,
and correct or enhance them.
Print out a copy |
Sat Apr 9 04:40:01 2022, Patrick, DAQ, Bugged Trending plots?
|
The problem from yesterday (ID:147) still persists. All the plots (not only the labview stuff) comes in too late in go4. However, the time axis is correct, see picture. |
Sat Apr 9 04:01:35 2022, Carsten, , 2nd DAQ
|
Switched the CAEN TDC back on and changed the threshoolds for the ADC to:
unsigned short caen_V785AH_threshold_mod0 [32] =
{0x0100 , 0x0004 , 0x0003 , 0x0002 , 0x0003 , 0x0002 , 0x0004 , 0x0002 ,
0x0004 , 0x0003 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 }; |
Sat Apr 9 03:57:10 2022, Carsten, Accelerator, New Cooler voltage after the crash of the cooler
|
After the restart of the cooler the Ohmlabs divider showed a voltage 16V lower than before. This could also be seen as a slight shift in the Schottky lines w.r.t to the bunching frequency and the marker. Corrected
the voltages at injection to DeltaU +3963V and in Sc 5/7 to + 3943V. Now perfect match with the old Schottky frequency.
Note: During the crash we found out that the device GECEKD2D did not show proper values. GECEKD2D is responsible for clearing the cooler of slow trapped ions ("clearing electrode").
Hence, the 16V may be due to partial filling of the space charge potential of te elctron beam. |
|
ELOG V3.1.5-fc6679b |