  Nuclear hyperfine mixing in 229Th89+, Page 9 of 12  Not logged in ELOG logo
Entry  Thu Mar 17 17:56:33 2022, Kristian, Calibration, HV dividers Bericht_PTB_verbesserter_Teiler_2018.pdf
PTB HV dividers arrived. Divider ratios are:

PT120:          12003,341 
Entry  Thu Mar 31 22:41:35 2022, Kristian, Calibration, HV divider HV-divider2.png
We noticed a difference in the electron cooler high voltage:

The set voltage is ~205 kV, the Ohmlabs divider measures 207 kV and the PTB200 shows a value of ~212 kV. The latter is doing some crazy jumps of up to
Entry  Fri Apr 1 18:19:58 2022, Kristian, Calibration, HV divider 
We checked the strange behavior of the HV dividers again. The Ohmlabs and the PTB divider are in good agreement at 120 kV and 150 kV. There, both dividers
deviated by 0.35% from the value on the HKR multimeter. Ohmlabs divider yields higher voltage.
Entry  Wed Mar 16 12:04:40 2022, Kristian, Detectors, HV cable to particle detector South 
Checked HV cable to particle detector south. Label is HV2 Scraper. 

Connected to Messplatz III HV5
Entry  Tue Apr 5 15:25:47 2022, Sebastian, Kristian, Pierre-Michel, DAQ, Go4 Conditions for "TDC_stacked_Sig/Ref" histogram 
First version of TDC_stacked histograms implemented.

Under folder "Conditions" you can now integrate over user defined Limits (Xmin and Xmax).
Entry  Mon Apr 4 15:32:38 2022, Kristiand & Sebastian, DAQ, Go4 Backup Code 
Due to some code changes (PMT over wavelength historgram) Kristian and I backuped a working Go4 code in the current go4 folder on the atplaser linux account.

It can be found in the subfolder "working go4 code 4.4.2020"
Entry  Sun Apr 3 12:35:17 2022, Max, DAQ, Go 4 manual go4_1.pnggo4_2.pnggo4_3.png
How to start go4?:

- go to console
Entry  Mon Apr 4 00:19:16 2022, Ken Ueberholz, DAQ, FirstTests 44_AM.png06_AM.png
New measurement:
Start: About 00:12
Stop: About 00:43
Entry  Thu Apr 7 02:32:29 2022, all on shift, Runs, First Scans 322-320nm 
Laser scanning from 322-320 steps 0.01nm/s
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
Entry  Fri Apr 8 17:44:25 2022, Konstantin, Calibration, Final Timing 
We measured the timing for the laser bunch synchronization. A summary is given below.

[TABLE border="1"]
Entry  Sun Apr 10 15:29:37 2022, Carsten,Sarper, Accelerator, Ecooler failed 
Ecooler failed @15:23 Carsten is fixing it
Fixed @16:00
Entry  Sat Apr 2 15:54:32 2022, Carsten, Accelerator, ESR setup with primary beam and 1cm Be plate + Schottky data on NTCAP 
MArkus and Bernd have setup the ESR with the thick stripper foil and put the ion beam (238U92+) on the central orbit.

SIS Energy is 536 MeV/u
Entry  Wed Apr 6 16:01:07 2022, BL, Accelerator, ESR settings before adjusting laser: 2022-04-06_14-37-30_ESR_EXP22_02_238U92_Brandau_30March22.C1.txt  
# Version: 1.2
# Comment: ParamModi Export at 2022-04-06T14:37:35.957978+02:00[Europe/Berlin]
# Context: ESR_EXP22_02_238U92_Brandau_30March22.C1
Entry  Wed Apr 6 09:30:53 2022, Danyal, Accelerator, ESR is blue again - errors are fixed 
The ESR is "blue" again. 

The errors (scu related) have been fixed. 
Entry  Wed Apr 6 08:31:54 2022, Danyal & Regina, General, ESR had some errors 
Regina noticed that the ESR was "red" 

This means that there are some errors. 
Entry  Fri Apr 1 02:29:40 2022, Carsten, , ESR Cooler crashed 
ESR cooler crashed. We could not restart the cooler. Since we had several other problems, we stop now.
Entry  Wed Apr 6 22:29:55 2022, Patrick, Accelerator, ECooler voltages 
ECooler voltages:

[B]06.04:[/B]
Entry  Fri Nov 5 17:23:23 2021, Rodolfo, DAQ, Dye trigger out  Cobra_ScanSettings_05-11-2021.PNG
The dye laser offers different scan settings (see pages 65 - 67 of the user manual). We are using the following setting, see attached screenshot, to produce
the status "laser busy" for the DAQ. 
Entry  Thu Mar 3 21:39:21 2022, Rodolfo, Laser, Dye solution, DCM in Ethanol and DMSO FarbstoffVorrat_Thorium.jpg
I have prepared a new dye solution of DCM in Ethanol and DMSO (1:1). The concentration of the "oscillator + preamplifier" is 0,3 g/l. The concentration
of the "main amplifier" is 0,0375 g/l or "1/8" the concentration of the "oscillator".
Entry  Fri Apr 15 20:43:18 2022, Konstantin, Laser, Dye changed 
Dye of the laser system has been changed. UV laser power increased from ~8 mJ to 11 mJ pulse energy.