Wed May 12 10:15:47 2021, Volker Hannen, DAQ, Unplugged TDC inputs for MCP front and back
|
To reduce the load on the VUPROM TDC, we unplugged the TDC inputs for the front and back signals of the MCP of the XUV detector.
The Anode signal is still plugged in and should be sufficient for the analysis. |
Fri May 14 04:50:46 2021, Team #Night, General, Scraper Tests Ion / Laser Beam
|
laser is 4-6mm below the ion beam.
It could be that the interaction between both beams came only from the interaction between the laser halo with the ion beam ?! |
Fri May 14 06:34:45 2021, Team #Night, Analysis, DSG821 IP
|
140.181.104.108 |
Fri May 14 06:35:55 2021, Team #Night, General, Adjustment of Laser
|
we increased the height of the laser beam
TEM Settings
|
Fri May 14 07:20:18 2021, Team #Night, General, actual bunching frequency
|
25,82556 MHz
h = 20 |
Fri May 14 07:33:42 2021, Team #Night, General, ecooler central frequency shifted relative to bunch central frequency
|
|
Fri May 14 01:34:44 2021, Team, General, strange XUV Anode signal
|
We see a strange behavior of the anode signal when we bunch the ion beam and electron cooler is on.
peak structure... ---> we play now with the bunching amplitude
|
Fri May 14 22:04:25 2021, Team, Accelerator, ESR OPEN ; NO BEAM -> Reset cooling device
|
we had to open the ESR to reset a cooling device |
Fri May 14 22:20:37 2021, Team, Accelerator, ESR Closed - BEAM #ON!
|
|
Sat May 15 03:39:16 2021, Team, General, actual revolution frequency
|
frev = 1,2912735 MHz |
Sat May 15 09:50:31 2021, Team, Runs, Laser cooling of....
|
H=2
H=7
H=10
|
Sun May 16 02:28:21 2021, Team, General, #Beam ON!
|
Laser specs:
pulse length -> 482ps
|
Sun May 16 05:09:04 2021, Team, General, Laser cooling with 482ps
|
we see laser cooling with 482ps pulse length |
Sun May 16 05:09:27 2021, Team, General, Switch to 285ps
|
pulse length -> 285ps
minimum 62mW inside ESR (depending on laser wavelength) |
Sun May 16 05:24:00 2021, Team, General, Laser cooling with 285ps
|
we also see laser cooling with 285ps pulse length |
Sun May 16 06:13:46 2021, Team, General, Switch to 166ps
|
|
Sun May 16 07:03:07 2021, Team, General, Laser cooling with 166ps
|
We saw some indications of laser cooling @ 166ps pulse length |
Sun May 16 21:23:02 2021, Team, General, ESR OPEN -> preparation for Polarisation Tests
|
|
Sun May 16 22:11:49 2021, Team, General, Laser Polarisation values
|
maximal transmission of PBSC (257/514 pbsc was used in combination with xlp12 detector and visual on a screen):
L/4 @ 276° & L/2 @ 348° --> p-polarized light
|
Mon May 17 08:05:42 2021, Team, General, End of beamtime
|
~ 8 a.m. |
Sat May 15 09:22:41 2021, Sebastian , Accelerator, Ekühler
|
Settings für Ekühler an und aus in SC4 und
5 |
Tue May 11 16:07:59 2021, Sebastian, DAQ, DAQ Manual
|
Attached you'll find a short manual for operating the whole DAQ system |
Wed May 12 08:32:04 2021, Sebastian, Analysis, fluorescence & energetic overlap
|
During the night shift we saw several rises of the fluorescence (see another entry) and also energetic overlap in coasting beam modus.
Here the laser wavelength was stabilized @ 1028,967nm and electron cooler was switched off because of some fluctuations which we saw in the schottky.
Left part of the schottky picture shows the c3+ beam, the right part is the o4+ contamination. |
Wed May 12 08:41:36 2021, Sebastian, Analysis, Laser stability
|
During night shift the laser was very stable! Only minor fluctuations occured in the UV |
Wed May 12 08:42:49 2021, Sebastian, Analysis, Laser scans / coastig beam / no ecooler
|
Attached you'll see a first laser scan of ~100GHz in the UV. |
Wed May 12 09:04:48 2021, Sebastian, DAQ, Keysight AUtoCal
|
|
Wed May 12 09:09:15 2021, Sebastian, DAQ, Fluorescence
|
fluorescence with detector in parking position
fixed laser wavelength
electron cooling ON,
|
Wed May 12 19:48:24 2021, Sebastian, General, DSG821 Network
|
To implement the bunch Signal Generator (DSG821 inside ESR) you have to:
1. unplug the Ethernet cable
|
Wed May 12 22:48:25 2021, Sebastian, General, DSG Control SCPI Commands
|
Open Ultra Sigma
Press LAN
Press Search, after a while the whole rigol devices are shown. Press ok
|
Wed May 12 22:57:41 2021, Sebastian, General, Laser Reprate
|
laser reprate is seven times the revoolution frequency: 9.038283 MHz
pulse length: 735ps |
Thu May 13 01:13:06 2021, Sebastian, General, Bunching frequencies
|
relating to Danyals ID 25 post are the current bunching frequencies:
h=2 --> 2,5823668 MHz
|
Thu May 13 01:21:30 2021, Sebastian, General, Current schedule ... so far
|
1.) check h=2,4,6,7,8,10 .... bunching frequencies with different amplitudes (signalgenerator), Ecooler ON, Laser OFF --> record schottky and "normal" DAQ
2.) Ecooler ON, Bunching ON, Laser ON (wavelength fixed and/or scanned)
|
Thu May 13 01:55:15 2021, Sebastian, General, ESR amplifier reset / laser check
|
we had to go into the ESR because the amplifiers hooked up. both amplifiers were resetted!
we also checked why we didnt see the laser at the exit window.
|
Thu May 13 02:42:58 2021, Sebastian, General, Lasershutter ESR Entrance?!
|
attached should be the name of the entrance laser shutter. just in case somebody will close it ;) |
Thu May 13 02:45:29 2021, Sebastian, Analysis, ESR pressure 8x
|
|
Fri May 14 01:28:04 2021, Sebastian, General, "new" laser wavelength
|
Minor changes in electron cooler voltage (~+13V, now 66903V measured by DMM) leads to that we had to shift the laser wavelength to:
1028,965nm
|
Fri May 14 08:55:53 2021, Sebastian, General, ToDO 14.5 ...so far
|
1.) ion lasser overlap and scraper test
2.) change bunching frequency to middle of ecooled coasting beam
|
Fri May 14 09:00:55 2021, Sebastian, General, UNILAC Problems
|
~30 min downtime
--> LASER ADJUSTEMENTS! |
Fri May 14 21:43:40 2021, Sebastian, General, "problems accelerator"
|
There are cooling problems of some accelerator devices
short break, hopefully :)
Laser is switched of for some rest |
Sat May 15 09:23:52 2021, Sebastian, Accelerator, Ekühler
|
Tab "Kühler Elektronenstrahl"
An Gerät schicken |
Sun May 16 02:18:53 2021, Sebastian, General, UV Drop
|
1:25 a.m. huge drop in UV laser power
-> small crystal bump with both y screws of bbo oven |
Sun May 16 02:20:58 2021, Sebastian, General, "Der Laser ist (wieder) ok"
|
|
Sun May 16 02:22:19 2021, Sebastian, General, "Der Laser ist ok"
|
|
Fri May 14 09:35:20 2021, Nightshift, Runs, Coasting beam, scanning laser, ecooler on
|
|
Fri May 14 09:38:04 2021, Nightshift, Runs, Coasting beam, scanning laser, ecooler on
|
|
Fri May 14 09:38:43 2021, Nightshift, Runs, Coasting beam, ecooler on, scanning laser
|
|
Fri May 14 09:40:04 2021, Nightshift, Runs, Ecooling vs laser wavelength shift, coasting beam, laser stronger than Ecooling
|
Last three pictures where taken after
first laser ion beam overlap adjustments
|
Sat May 15 06:51:16 2021, Night Shift Sebastian/Max/Noah, General, Different Laserscantypes
|
We use different scan types (manual scan) or constant laser wavelengths to check the effects of laser cooling in the Schottky diagrams (fluorescence can
not be seen in the detector), for different bunching amplitudes and frequencies:
- Scan from 1028.950 -> 1028.940, stepsize 0.001 every 10 sec
|
Sun May 16 07:18:56 2021, Night Shift Sebastian/Max/Noah, General, Preparation for automated laser scans
|
We shut the laser system down to prepare automated laser scans |
Wed May 12 08:10:17 2021, Lars, General, instabilities in ion frequency
|
looks like the E-Cooler is responsible for this. If one turns off the electron current, the beam stays stable.
M.Steck is informed.
|
Wed May 12 10:14:18 2021, Lars, Accelerator, E-Cooler
|
M.Steck et al. discussed about the E-Cooler and found a possible issue. They asked for enterance to the ESR. Since nobody else was in the HKR and anyway
nothing was running because of some other issues, I told, the experiment would not be disturbed and they could enter and do things. (Sth. like putting
some repeller electrodes onto hand and other magical things).
|
Thu May 13 03:35:25 2021, Lars, General, Status: Current schedule ... so far
|
1.) so far:
h = 2, 4, 6, 7, 8, 10, 12, 14, 16, 18, 20, 30
Amplitudes: (10, 50, 100, 150, 200, 250) mV
|
Wed May 12 09:05:48 2021, Keysight Settings, DAQ,
|
|
Tue May 11 16:33:31 2021, Ken Ueberholz, Detectors, Detector control manual
|
Attached to this elog entry is a short manual for the xuv-detector control. If anything is missing or not understandable I can change anything. |
Wed May 12 18:59:15 2021, Ken Ueberholz, General, Power is back
|
Power is back and most things are running again |
Wed May 12 22:28:55 2021, Ken Ueberholz, General, Bunching succesfull
|
Bunching with 2 and 10 bunches succesfull |
Wed May 12 23:36:18 2021, Ken Ueberholz, DAQ, VI change
|
carbon main vi - kopie is now in operation
here, in keysight.vi the input settings for the new impedence for the multimeter are now hopefully correct --> THIS SHOULD BE CHECKED IF THE RING IS OPEN |
Wed May 12 23:45:23 2021, Ken Ueberholz, General, Laser does not reach the exit window
|
High rates on the detector at 50mm positions, laser does not seem to reach the exit window. |
Thu May 13 18:34:54 2021, Ken Ueberholz, General, What happened scince shift change at 4pm
|
Ion frequency shifts with laser, both manually and scanned, without bunches: successful
Bunching with 7 and 20 bunches: successful
Tried laser scans with bunches: unsuccessful
|
Thu May 13 23:42:37 2021, Ken Ueberholz, General, What was done
|
In this shift a number of successful Schottky measurements were done:
With fixed laser wavelength, while varying the wavelength, with different bunching and more (see logfile 1 for more info: line 53 to at least 107).
The detector software sometimes files to change the current of the power supplies, if this happens a simple restart of the GUI fixes the problem.
|
Tue May 31 07:29:11 2022, Ken, Detectors, XUV-Detektor
|
Rate was very low: ~ 100Hz
Adjusted Walk level and Threshold, new rate: 5kHz
Additionally, parameters of the detector were varied for optimization but were returned to default afterwards. |
Sat May 15 09:57:24 2021, Dayshift, Runs, Currently on schedule
|
Enhance fluorescence rate and perform
laser bunch timing
|
Wed May 12 02:16:25 2021, Danyal, Lars, Accelerator, Ion beam position
|
Laser Position with scrapers verified.
1HA: impossible to scratch laser beam, still there at -1.0 mm
2HA: +3mm --> half the laser power on detector; +4.5mm --> gone |
Mon May 10 07:58:13 2021, Danyal, General, Telephone numbers and Zoom link
|
Telephone numbers:
ESR laser lab: +49 (0)6159 71 - 2882
|
Tue May 11 15:49:30 2021, Danyal, General, setting up ESR
|
The operators are still busy extracting the beam from the SIS18 towards the ESR and injecting it into the ESR. |
Tue May 11 18:29:35 2021, Danyal, General, first ions in the ESR
|
|
Tue May 11 20:14:43 2021, Danyal, General, bad ESR vacuum
|
I just heard that the ESR vacuum is quite bad.
There was a small vacuum problem a while ago, it was sealed, but apparently the pressure is not as low as usual.
|
Wed May 12 02:34:29 2021, Danyal, General, instabilities in ion frequency
|
The ion beam moves around, pretty arbitrary. See picture.
ESR crew believes that it is not due to fluctuations in the HV of the E-Cooler. |
Wed May 12 13:11:58 2021, Danyal, Accelerator, short break between 13:30 and 15:30
|
There will be a short break (2 hours) this afternoon. Between 13:30 and 15:30.
The so-calles "super lense" in the UNILAC needs to be cleaned (rinsed).
After that operation should continue as usual. |
Wed May 12 15:36:04 2021, Danyal, Accelerator, general power failure in ESR! (everything is off)
|
The power in the ESR is off.
(also on the gas-target area on top of ESR)
Electrical engineers are trying to find and fix the problem. |
Wed May 12 22:35:33 2021, Danyal, General, frev and fb
|
Schottky frequency (189th harmonic) = 244.033662 MHz
revolution frequency frev = 1.2911834 MHz
|
Thu May 13 00:58:53 2021, Danyal, DAQ, unbunched ion beam
|
|
Thu May 13 01:01:27 2021, Danyal, DAQ, bunched ion beam
|
|
Thu May 13 01:05:45 2021, Danyal, DAQ, more stability by bunching
|
Because the C3+ part of the ion beam is properly bunched, it is no longer so sensitive to the drifts caused by the e-cooler.
In the O4+ part of the ion beam these drifts can still be seen. |
Mon May 17 21:43:56 2021, Danyal, General, The test beamtime was a success.
|
Dear all,
The test beamtime is now over.
|
Wed May 12 09:59:08 2021, Benedikt, General, Rigol DSG821 signal generator settings
|
|
Thu May 13 20:18:10 2021, Benedikt, General, New f_rev
|
The ions were changed to a different speed today at 2 PM. The new revolution frequency is now 1.291278 MHz. |
Thu May 13 21:00:48 2021, Benedikt, General, New reprate of the UV laser
|
The lasersystem now operates at a reparte of 9.038946 MHz. This is 7 times f_rev (see previous post).
In the future, the laser-bunch-timing can be adjusted. |
Fri May 14 11:01:33 2021, Benedikt, Calibration, Laser Beam adjusted
|
The laser beam was adjusted in his position to overlap better with the ions.
Ion beam is destroyed at scraper position 1VO 6 mm and 2VO 4 mm; the laser beam looses half its power at 1VO 6mm and 2VO 4 mm.
|
Fri May 14 16:37:01 2021, Benedikt, General, Status update
|
Right now, we have to adjust the ion beam a bit to overlap better with the laser beam.
We keep the laser wavelength at resonance with the ions and try to optimize the fluorescence signal on the XUV detector while moving the ion beam. |
Sat May 15 11:06:31 2021, Benedikt, General, New Zoom link for the moment
|
https://tu-darmstadt.zoom.us/j/4948400422?pwd=SEJaUzBuYXM0TTd6UWxoZTFlWEZxUT09
Meeting ID: 494 840 0422
|
Sat May 15 11:26:44 2021, Benedikt, General, TEM parameters
|
We changed the horizontal position of the laser beam a bit, so that we maximized the fluorescence signal on the XUV detector (coasting beam, scanning laser
(0.03 Hz)).
|
Sat May 15 12:11:54 2021, Benedikt, General, Laser-bunch-timing
|
Optimizing the laser bunch timing.
h = 7 (9.0389145 MHz)
Reprate is already set to same frequency.
|
Sun May 16 11:29:22 2021, Benedikt, General, Automated laser scans working
|
The automated laser scans are working now.
We can now scan with the following functions: triangle, sawtooth, cosinus/sinus, arctan, min (linear scan, than stands still)
|