  62   Thu May 12 15:21:23 2022 Sebastian, SimonLaserFlash lamps changedWe (Sebastian and I) changed the flash lamps
of the Quanta Ray Laser and measured the
output energy of the laser system afterwards
  
  67   Fri May 13 11:46:51 2022 Sebastian, Simon Laser energyWe (Sebastian and I)  measured the output
energy of the laser system again. 
All values in [mJ] and over 500 laser-pulses
  
  33   Tue May 10 23:10:25 2022 Sebastian, Patrick, Volker, GuyAcceleratorT-Scraper ReloadedWe have established a Control to "manually"
move the T-Scrapers via a "remote desktop".
We can now move both scrapers in the horizontal
  
  11   Thu May 5 15:24:34 2022 SebastianLaserlaser busy problemCurrent status: 
Laser busy signal works correctly (5V if
the grating moves, 0V if not) if the laser
  
  12   Fri May 6 10:30:23 2022 SebastianDAQlaser busy / dye out signal fixedSolution for a working laser busy / dye out
signal was to disable the FCU and set the
FCU status to "False" (see attached picture).
 CobraFix_LaserBusy.PNG 
  13   Fri May 6 10:35:06 2022 SebastianDAQlaser step complete cabling changedI changed the laser step complete cabling
from a quad coinc (see attached picture)
to a Logic Fan In / Fan out (same as laser
 stepComplete_beforeChange_QuadCoincCabling.PNG 
  26   Tue May 10 11:40:55 2022 SebastianAcceleratorstill no beam @ESRThere is still no Beam in the ESR.

The beam is currently at TE3 - between SIS
  
  52   Wed May 11 23:36:24 2022 Ruijiu Chen, Carsten, GuyDAQstart NTCAP.Carrier frequency:244MHz
IQ rate: 10M Samples/seconds
Reference level: -25
  
  10   Wed May 4 22:56:22 2022 Rodolfo, Sebastian, KonstantinLaserLaser step complete, scan complete fixed"Laser step complete" and "laser scan complete"
are working again.
  
  9   Wed May 4 22:55:37 2022 Rodolfo, SebastianLaserFCU / Temp Control BugWe found a "Bug" in the control of the Sirah
dye system.
  
  28   Tue May 10 15:26:35 2022 Rodolfo, MaxLaserLaser stabilisation runningIt is now possible to stabilize and move the
laser with both stages.
 MRC_Settings_for_Bismuth_10-05-2022.PNGMRC_hardware_Setting.PNG 
  7   Fri Apr 29 21:18:12 2022 Rodolfo & MaxLaserFluorescein 27, Rh 610 + Rh 640We (Max and Rodolfo) have tested the dyes.
I have prepared the dye-solutions according
to the recipes given in Sirah's User Manual.
  
  75   Mon May 16 21:08:48 2022 Rodolfo & MaxLaserLaser powerWe checked the energy of the laser and optimized
it a bit
  
  92   Mon May 30 19:42:20 2022 Rodolfo & KristianLaserChanging laser wavelengthWe have changed the laser wavelength, e.g.
the dye for the Cobra to Fluorescein 27.
  
  1   Fri Jul 30 10:21:27 2021 RodolfoLaserKinkThe laser beamline in the ESR north arc has
a [B]knik[/B], therefore the laser beam can
not get the "laser tower" in the north east
 Laser_beamline_ESR_NordBogen.jpg 
  2   Fri Apr 22 09:38:24 2022 RodolfoLasergetting ready for Fluorescein548We are going to change dye. Therefore I have
take out the DCM-dye-solution* out of the
dye circulators, rinsed them with Ethanol
  
  3   Mon Apr 25 18:47:12 2022 RodolfoLaserNew dye circulatorsI have put into operation the new circulators
(SN. 182231, 182201) for our dye laser (the
Cobra). I have rinsed them with pure Ethanol
 Neue_Cobra-Farbstoffpumpe.jpg 
  4   Mon Apr 25 21:47:28 2022 RodolfoLasergreen instead UVI am exchanging the optical elements (mirrors
and lenses) for the laser spectroscopy on
bismuth. At some places I am also exchanging
 Libelle_Umtausch_GrüneOptik.jpgUmtausch_Spiegel_2statt3.jpg 
  6   Tue Apr 26 18:15:19 2022 RodolfoDetectorsXUV-Detector and Cu-MirrorConcerning the moveable detectors: 

- I have switched off the power supply and
  
  18   Mon May 9 13:48:20 2022 RodolfoLaserLaser for H-like 209BiI have changed the dye to Rhodamin 610 + Rhodamin
640. The Cobra wavelength can be tuned now
from 600 to 630 nm.
  