ID Date Authordown Category Subject
  88   Mon May 30 08:35:41 2022 Kristian, Simon, Ken, RaganDetectorsChanged PMT North
We did not see any signal at PMT North which was the narrowband PMT. Changed it to the old solar blind PMT. Now we see the resonance. It is about 3 times larger as PMT middle, most likely because of the larger window. CFD threshold was set. I also increased the CFD threshold for PMT middle slightly.


HV for  PMT North: 1600 V
  107   Wed Jun 1 08:39:50 2022 Kristian, KonstantinCalibrationWavemeter reading
We recorded files with high wavemeter readout rate and saved it at the atplaser network drive (Q:) - Bismuth - wavemeter folder.

Could be used to calibrate the laser steps. The MBS readout rate is 4 Hz, here 30 Hz, so there are a few more values.
  90   Mon May 30 16:17:39 2022 Kristian, David, Wilfried, Carsten, Danyal, RodolfoRunsAfternoon shift 30.05
These are the first runs after changing to the thick stripper foil.

Time ESR-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
16:15 ? 5.0e6 223439 669-670 8180 601.95 - 601.734 0.005 5s 5 Resonance found in first scan. electron cooling current 100mA
16:45 ? 5.0e6 223439 671-672 8181 601.95 - 601.734 0.005 5s 5 electron cooling current 300mA


Width approximately the same with the higher electron current.
next we go to 208Bi
  81   Fri May 27 10:39:53 2022 KristianDAQPMT North noise
During the last run, PMT North showed some weird noise.

The PMT was replaced by the deep UV PMT with small aceptance range (100-200nm) but there was still a lot of noise on the output channel in the Messhütte, see attached photo. The intensity of the noise signal was as strong as a normal signal count.

I replaced all cables and used other preamplifier channels. The noise is gone and the baseline is flat now. However, the timing of PMT north will be a bit different now, so we should use PMT middle for the laser/ion overlap as we did before.

Graziano opened the valves for the laser and turned off the vacuum gauges and emergency lights. The background of the UV PMT looks normal (Middle 10 Hz, North 1 Hz) so I think we are fine here.
  105   Wed Jun 1 00:03:08 2022 Konstantin, Simon, KristianRuns209-Bi runs
Time SIS-intensity ESR-intensity after orbit change cooler-HV cooler current (mA) LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
23:40 - - 223432 100 781 8246 601.735 - 601.95 0.005 4.95s 2 resonance found but problems with go4
00:02 - 5.5e6 223432 100 783 8248 601.735 - 601.95 0.005 4.95s 2 resonance not in the middle of the scan range
00:14 - 4.5e6 223432 100 784 8249 601.7 - 601.9 0.005 4.95s 2 change the scan range
00:14 - 1.1e7 223432 100 785 8250 601.7 - 601.9 0.005 4.95s 2
00:35 - 9e6 223432 100 786-787 8251 601.7 - 601.9 0.005 4.95s 4
00:50 - 6.5e6 223432 100 788-789 8252 601.7 - 601.9 0.005 4.95s 4
01:07 2.2e8 1.15e7 223432 100 790-791 8253 601.7 - 601.9 0.005 4.95s 4
01:26 2.3e8 1.15e7 223448 200 792 8254 601.7 - 601.9 0.005 4.95s 2 switched to coasting beam, changed ESR current
01:37 2.2e8 1.15e7 223448 200 793 8255 601.7 - 601.9 0.005 4.95s 2
01:48 2.2e8 1.15e7 223448 200 794 8256 601.7 - 601.9 0.005 4.95s 2
01:58 2.2e8 1.15e7 223448 200 795 8257 601.7 - 601.9 0.005 4.95s 2
02:08 2.2e8 1.15e7 223448 200 796 8258 601.7 - 601.9 0.005 4.95s 2
02:19 2.15e8 1.15e7 223432 100 797-799 8259 601.7 - 601.9 0.005 4.95s 6 changed ESR current again
02:45 2.17e8 1.15e7 223432 100 800-802 8260 601.7 - 601.9 0.005 4.95s 6
03:12 2.12e8 1.15e7 223423 50 803-805 8261 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-05-31_19-33-57
03:38 2.2e8 1.15e7 223423 50 806-808 8262 601.7 - 601.9 0.005 4.95s 6
04:05 2.13e8 1.15e7 223427 75 809-811 8263 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-06-01_03-12-03
04:30 2.03e8 1.15e7 223427 75 812-814 8264 601.7 - 601.9 0.005 4.95s 6
04:57 2.08e8 1.15e7 223440 150 815-817 8265 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-06-01_04-04-32
05:21 2.1e8 1.15e7 223440 150 818-820 8266 601.7 - 601.9 0.005 4.95s 6
05:57 2.08e8 1.15e7 223440 150 821-823 8267 601.7 - 601.9 0.005 4.95s 6 laser power reduced from 46.2mJ to 25mJ
06:25 1.5e8 9e6 223440 150 824-826 8268 601.7 - 601.9 0.005 4.95s 6
06:56 2.1e8 1.15e7 223440 150 827-829 8269 601.7 - 601.9 0.005 4.95s 6 Laser power reset to 48mJ. Last ion injection!!!
07:18 - 4.7e6 223439 150 830-832 8270 601.7 - 601.9 0.005 4.95s 6 Last ion injection!!!
7:40 - 2.0e6 223439 150 833-834 8271 601.7 - 601.9 0.005 4.95s 6 old beam
8:02 - 1.0e6 223439 150 835-836 8272 601.7 - 601.9 0.005 4.95s 6 older beam
8:25 - 5e5 223439 150 837-8 8273 601.7 - 601.9 0.005 4.95s 6 very old beam
8:48 - 3e5 223439 150 839-840 8274 601.7 - 601.9 0.005 4.95s 6 very very old beam
9:10 - 2e5 223439 150 841-842 8275 601.7 - 601.9 0.005 4.95s 6 beam close to retirement - no signal anymore


9:33 beam retired. Turned laser off. Stopped NTcap

post beamtime change: replaced "ESR current" by "cooler current" in the comments column. (Wilfried)
  106   Wed Jun 1 00:16:02 2022 Konstantin, SimonDAQTiming 209Bi thick target
Timing for the final measurements on 209Bi was determined for the new frequency 1.93777MHz. Peak positions are

PD NO 106
PD SO 141.5
PMT MIDDLE 45 (due to the broad temporal structure difficult to determine)

Center between PDs is 123.75. Difference to PMT MIDDLE is 78.75. This is ~3 bins larger than the half period of 75. 
  85   Mon May 30 01:13:40 2022 Konstantin, Max, PatrickDetectorsXUV-Detector voltage off
XUV-Detector shows up to 100 kHz and, therefore, was turned off.
  43   Wed May 11 14:48:31 2022 KonstantinDAQVoltage range of the Keysight DMM changed from 10V to 1V
During the setup of the ion beam the voltage range of the Keysight DMM was changed from 10V to 1V. Additionally, the name of the high-voltage divider was adjusted (this changes only the filename of the file, which is stored at the local pc decpc003).
  61   Thu May 12 14:04:09 2022 KonstantinDAQGo4 Laserstep histograms adapted
In order to correctly display the gated PMT signals PMT Middle and PMT North some modifications in the Go4 code were necessary. Affected files are TLdaAnlProc.cxx, TLdaAnlProc.h and TLdaInputDescription.h.
Modifications are indicated according to the scheme "//-- HH:MM mm-dd-yyyy: <description>" (an example is given in the attached screenshot). They were needed to

  1. Create relevant histograms (change of variable MAX_STEP from 10 to 13 and insertion of histogram names "QSWP1MS", "NORTH_GATED" and "MIDDLE_GATED" in the std::string array stepNames.
    Both adaptions were made in TLdaInputDescription.h)

  2. Fill the respective histograms. Changed lines of files TLdaAnlProc.cxx, TLdaAnlProc.h are listed below.

filename modified lines
TLdaAnlProc.h 47
TLdaAnlProc.cxx 258-261, 381-394, 442-443, 484-494, 536-542
  80   Tue May 17 10:38:44 2022 KonstantinDetectorsPMT North changed
PMT located at the northern window was again replaced by the narrow-band PMT (ET Enterprise type 9423 SB, SN: 650).
  89   Mon May 30 09:28:43 2022 Ken, Kristian, Simon, RaganDetectorsXUV-Detector
9:27 XUV-Detector is running again.

-5V cathode, 8A main coil, 2.4A sec.coil
Rate without beam was: 
Park position: ~300Hz 

Pump beneath the detector is now shut down, new rate: <10Hz.

Threshold was adjusted to trigger only on "nice" signals. 

Rate with beam in and laser:
Park position: ~400Hz
147mm: ~1.4kHz
 
with 2e6 ions
  31   Tue May 10 22:29:45 2022 Jan GloriusDAQDAQ2 shaper settings
Attached are the MSCF shaper settings applied on Tue May 10 at 22:30.
  32   Tue May 10 22:37:10 2022 Jan GloriusDAQDAQ2 cabling docu
  68   Fri May 13 16:55:58 2022 Danyal, Carsten, Phillip, WilfriedGeneralEnd of E128 Beamtime
Dear all,

unfortunately, the E128 experiment must now come to an early end. 
The problem(s) with the injection septum are severe and no work around solution (with acceptable risks) could be found. 
Several things were tried, but without hope. 

Many thanks to everybody involved during the preparation time and the shifts during the beamtime. 

Best wishes,

Wilfried
  109   Fri Jun 3 12:01:17 2022 DanyalDetectorsswitched off XUV detector on 1 June ~17:00
I have switched off the XUV detector on 1 June at about 17:00. 
The high voltages of the MCP (in ESR laser lab) were turned off and inside the ESR cave the power supplies (coils and cathode) and the crate were turned off. 
Also the new "control box" for the detector was switched off. 
And I checked (visually) that the XUV detector (and the Cu detector too) were indeed "moved out". 

Danyal
  100   Tue May 31 19:32:56 2022 Carsten, Wilfried, MaxCalibrationelectron cooler calibration voltage - frequency
SetVoltage (DeltaU) MeasVoltage freq Current
+1950V 225247V 245.06725 50 mA
+1950V 225255.4V 245.0595 100 mA
+1960V 225266.4V 245.062125 100 mA
+1970V 225276.8V 245.064185 100 mA
+1980V 225286.6V 245.066 100 mA
+1990V 225296.1V 245.06775 100 mA
+2000V 225305.1V 245.0695 100 mA
+1950V 225255.5V 245.06 100 mA
+1940V 225246.0V 245.058125 100 mA
+1930V 225236.0V 245.05625 100 mA
+1920V 225226.0V 245.0545 100 mA
+1910V 225216.7V 245.052625 100 mA
+1900V 225205.8V 245.050375 100 mA
+1850V 225156.5V 245.041 100 mA
+1800V 225106.6V 245.031375 100 mA
+1950V 225256.1V 245.060125 100 mA
+2000V 225306.0V 245.06975 100 mA
+2050V 225355.8V 245.079375 100 mA
+2100V 225406.0V 245.0890 100 mA
+1950V 225256.2V 245.060125 100 mA
Change electron current and note down High Vltage and revolution frequency
NTCAP was restarted at the beginning and stopped at the end of the measuring period.
The NTCAP folder is 05-31_17-16-49 (system time 19:11)
  8   Wed May 4 22:45:12 2022 Carsten, SebastianDAQQ-Switch Signal (ch15) fixed
The "dead" Q-Switch signal was converted 
two times... This explains why in the 
Thorium experiment the Q-Switch signal 
(ch15) was empty. 

First conversion from TTL to NIM in the 
ESR Laserlab and unfortunately the signal 
was plugged in a second Level Adapter as 
TTL in 
the Messhütte and was "again" converted 
into a NIM for DAQ1/2.

Now the Q-Switch is converted from TTL to 
NIM in the ESR Laserlab. The NIM is 
afterwards guided through a Logic Fan 
In/Out to the 2 DAQs. 
  96   Mon May 30 23:18:06 2022 Carsten, Rodolfo, Konstantin, Patrick, Ragan, Ken, Kristian, Max, DanyalRunsRuns shifts 31.05.
These are the first runs with 208Bi.

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops Particle Detector comments
23:11 2.6e8 - 225259 673-677 8182 548 - 549 0.005 2.97s 4 139e3 Resonance found in first scan.
00:09 ? - 225259 678-679 8183 548 - 549 0.005 2.97s 2 92e3
00:45 2.8e8 - 225260 680-681 8184 548 - 549 0.005 2.97s 2 36e3 In this run, the ions seemed to be not ejected with the new request, compare picture.
01:27 2.7e8 - 225260 684-685 8187 548 - 549 0.005 2.97s 2 88e3 Started delayed due to several issues with both DAQs.
01:53 2.7e8 - 225260 686-687 8188 548 - 549 0.005 2.97s 2 93e3
02:20 2.7e8 - 225260 688-689 8189 548 - 549 0.005 2.97s 2 87e3
02:55 2.1e8 - 225260 692-694 8190 548.335 - 548.535 0.0025 4.95s 4 88e3 Scan range changed
03:30 2.75e8 - 225260 696-697 8191 548.335 - 548.535 0.0025 4.95s 3 95e3 Number of loops reduced to 3
03:59 2.6e8 - 225260 698-699 8192 548.335 - 548.535 0.0025 4.95s 3 89e3
04:30 2.7e8 - 225260 700-702 8194 548.335 - 548.535 0.0025 4.95s 3 89e3 DAQ1 needed to be restarted
04:52 2.55e8 - 225260 703-704 8195 548.335 - 548.535 0.0025 4.95s 3 83e3
05:18 2.62e8 - 225260 706-707 8196 548.335 - 548.535 0.0025 4.95s 3 89e3 DAQ2 needed to be restarted
05:44 ? - 225260 708-710 8197 548.335 - 548.535 0.0025 4.95s 3 ?
06:18 2.1e8 - 225260 711-712 8200 548.36 - 548.51 0.0025 4.95s 3 67e3 we changed the timing by 20 ns. The laser pulse is later now. The file number of DAC1 is correct. I do not know why there is the gap to the previous scan
06:51 2.4e8 6e5 225260 713-714 8201 548.36 - 548.51 0.0025 4.95s 3 67e3 xuv-detector parameters were varied
07:09 2.4e8 6e5 - 715- 8202 548.36 - 548.51 0.0025 4.95s 3 68e3 mbs did not start
07:15 2.4e8 6e5 225260 716-717 8203 548.36 - 548.51 0.0025 4.95s 3 68e3 xuv-detector walklevel and treshhold were lowered
07:35 2.4e8 6e5 225259 718-719 8204 548.36 - 548.51 0.0025 4.95s 3 68e3 we changed the timing by additional 20 ns. The laser pulse is later now. - no fluo. visible
08:01 2.4e8 6e5 225259 720-721 no file opened 548.43 - 548.44 0.00001 ~20s 3 71e3 varying the delay with each laser step to find optimum - It was more a plateau between +-30ns (relative to the night shift's position). We now continue data taking at the same dealy as used by the night shift.
08:23 2.4e8 6e5 225258 722 8205 548.36 - 548.51 0.0025 4.95s 3 68e3 xuv-detector walklevel and treshhold were lowered
08:28 2.4e8 6e5 225258 724-725 8208 548.36 - 548.51 0.0025 4.95s 3 68e3
08:48 2.4e8 6e5 225258 727-728 8209 548.36 - 548.51 0.0025 4.95s 3 ? xuv-detector in Park-position just as a try
09:08 2.4e8 6e5 225259 729-730 8210 548.36 - 548.51 0.0025 4.95s 3 ?
09:28 2.5e8 6e5 225259 731-732 8211 548.36 - 548.51 0.0025 4.95s 3 73e3
09:54 2.5e8 6e5 225259 734-735 8212 548.36 - 548.51 0.0025 4.95s 3 175e3
10:19 2.5e8 6e5 225258 737-738 8214 548.36 - 548.51 0.0025 4.95s 3 175e3
10:38 2.5e8 6e5 225258 741-742 8215 548.36 - 548.51 0.0025 4.95s 3 64e3 DAQ2 started late. Missed the first 25 scan numbers Coasting beam measurement from now on
10:57 2.5e8 6e5 225258 743-744 8216 548.36 - 548.51 0.0025 4.95s 3 87e3
11:17 2.5e8 6e5 225258 745 8217 548.36 - 548.51 0.0025 4.95s 3 e3 MBS crashed
11:18 2.5e8 6e5 225257 746-747 8218 548.36 - 548.51 0.0025 4.95s 3 87e3
11:37 2.5e8 6e5 225257 749-750 8219 548.36 - 548.51 0.0025 4.95s 3 87e3 DAQ2 started late. Missed the first 10 scan numbers
11:56 2.5e8 6e5 225257 751 8220 548.36 - 548.51 0.0025 4.95s 3 78e3
12:59 2.5e8 6e5 225257 752 8221 548.385 - 548.485 0.0025 4.95s 3 78e3 RF off from now on, laser scan settings not correct. stopped
13:06 2.5e8 6e5 225257 753 8223 548.385 - 548.485 0.0025 4.95s 3 87e3
13:26 2.5e8 6e5 225257 754 8224 548.385 - 548.485 0.0025 4.95s 3 - MBS crashed
13:29 2.5e8 6e5 225257 755-756 8226 548.385 - 548.485 0.0025 4.95s 3 93e3 RF now really off (wrong cavity was turned off)
13:43 2.5e8 6e5 225257 757-758 8227 548.385 - 548.485 0.0025 4.95s 3 67e3
14:08 2.5e8 6e5 225257 759 8228 548.385 - 548.485 0.0025 4.95s 3 70e3
14:22 2.2e8 6e5 225257 760 8229 548.385 - 548.485 0.0025 4.95s 3 56e3
14:35 2.2e8 6e5 225257 761 8230 548.385 - 548.485 0.0025 4.95s 3 64e3
14:50 2.2e8 6.5e5 225256 762 8231 548.385 - 548.485 0.0025 4.95s 3 59e3
15:04 2.0e8 7e5 225255 763 8232 548.385 - 548.485 0.0025 4.95s 3 58e3
15:28 1.9e8 ?e5 225272 764 8233 548.385 - 548.485 0.0025 4.95s 2 66e3 changed to 200mA cooler current, NTCAP stopped and started again folder: prob. ~31.5.22 13:??, loops decreased to 2
15:37 2e8 6e5 225272 765 8234 548.385 - 548.485 0.0025 4.95s 2 70e3
15:47 2.5e8 6e5 225272 766 8235 548.385 - 548.485 0.0025 4.95s 2 73e3 laser gone at loop 2 scan 35 (15:56), Ecooler went also off
16:17 ?e8 6e5 225272 767 8236 548.385 - 548.485 0.0025 4.95s 2 48e3 bad shot?, probably ecooler reset
16:27 2.2e8 6e5 225271 768 8237 548.385 - 548.485 0.0025 4.95s 2 77e3
16:36 2.0e8 7e5 225272 769 8238 548.385 - 548.485 0.0025 4.95s 2 65e3
16:52 ?e8 ?e5 225272 770 8239 548.385 - 548.485 0.0025 4.95s 2 61e3 we had to restart go4, it was very slow
17:02 2.0e8 7e5 225272 771 8240 548.385 - 548.485 0.0025 4.95s 2 63e3
17:16 2.3e8 7e5 225248 772-773 8241 548.385 - 548.485 0.0025 4.95s 4 45e3 changed to 50mA cooler current, NTCAP stopped and started again folder: 31.5.22 15:18, loops increased to 4
17:36 2.7e8 7.5e5 225248 774-775 8242 548.385 - 548.485 0.0025 4.95s 6 67e3 loops increased to 6
18:05 2.5e8 7.5e5 225248 776-777 8243 548.385 - 548.485 0.0025 4.95s 6 66e3
18:31 1.7e8 6e5 225247 778-779 8244 548.385 - 548.485 0.0025 4.95s 6 44.6e3 SIS pattern looked strange
  101   Tue May 31 19:37:30 2022 Carsten, Max, WilfriedDetectorsSASZ and NASZ voltages switched off
SASZ and NASZ voltages were switched off bevor we went back to 209Bi. Laser beam stabilization turned off.
  103   Tue May 31 21:08:04 2022 Carsten, MaxDetectorsXUV detector automation off
We deactivated the automation of the XUV detector. It stays out now.