ID |
Date |
Author |
Category |
Subject |
88
|
Mon May 30 08:35:41 2022 |
Kristian, Simon, Ken, Ragan | Detectors | Changed PMT North |
We did not see any signal at PMT North which was the narrowband PMT. Changed it to the old solar blind PMT. Now we see the resonance. It is about 3 times larger as PMT middle, most likely because of the larger window. CFD threshold was set. I also increased the CFD threshold for PMT middle slightly.
HV for PMT North: 1600 V |
107
|
Wed Jun 1 08:39:50 2022 |
Kristian, Konstantin | Calibration | Wavemeter reading |
We recorded files with high wavemeter readout rate and saved it at the atplaser network drive (Q:) - Bismuth - wavemeter folder.
Could be used to calibrate the laser steps. The MBS readout rate is 4 Hz, here 30 Hz, so there are a few more values. |
90
|
Mon May 30 16:17:39 2022 |
Kristian, David, Wilfried, Carsten, Danyal, Rodolfo | Runs | Afternoon shift 30.05 |
These are the first runs after changing to the thick stripper foil.
Time | ESR-intensity | ESR-intensity after orbit change | cooler-HV | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step width increment (nm) | dwell time(s) | n loops | comments
|
16:15 | ? | 5.0e6 | 223439 | 669-670 | 8180 | 601.95 - 601.734 | 0.005 | 5s | 5 | Resonance found in first scan. electron cooling current 100mA
|
16:45 | ? | 5.0e6 | 223439 | 671-672 | 8181 | 601.95 - 601.734 | 0.005 | 5s | 5 | electron cooling current 300mA
|
|
Width approximately the same with the higher electron current.
next we go to 208Bi |
81
|
Fri May 27 10:39:53 2022 |
Kristian | DAQ | PMT North noise |
During the last run, PMT North showed some weird noise.
The PMT was replaced by the deep UV PMT with small aceptance range (100-200nm) but there was still a lot of noise on the output channel in the Messhütte, see attached photo. The intensity of the noise signal was as strong as a normal signal count.
I replaced all cables and used other preamplifier channels. The noise is gone and the baseline is flat now. However, the timing of PMT north will be a bit different now, so we should use PMT middle for the laser/ion overlap as we did before.
Graziano opened the valves for the laser and turned off the vacuum gauges and emergency lights. The background of the UV PMT looks normal (Middle 10 Hz, North 1 Hz) so I think we are fine here. |
105
|
Wed Jun 1 00:03:08 2022 |
Konstantin, Simon, Kristian | Runs | 209-Bi runs |
Time | SIS-intensity | ESR-intensity after orbit change | cooler-HV | cooler current (mA) | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step width increment (nm) | dwell time(s) | n loops | comments
|
23:40 | - | - | 223432 | 100 | 781 | 8246 | 601.735 - 601.95 | 0.005 | 4.95s | 2 | resonance found but problems with go4
|
00:02 | - | 5.5e6 | 223432 | 100 | 783 | 8248 | 601.735 - 601.95 | 0.005 | 4.95s | 2 | resonance not in the middle of the scan range
|
00:14 | - | 4.5e6 | 223432 | 100 | 784 | 8249 | 601.7 - 601.9 | 0.005 | 4.95s | 2 | change the scan range
|
00:14 | - | 1.1e7 | 223432 | 100 | 785 | 8250 | 601.7 - 601.9 | 0.005 | 4.95s | 2 |
|
00:35 | - | 9e6 | 223432 | 100 | 786-787 | 8251 | 601.7 - 601.9 | 0.005 | 4.95s | 4 |
|
00:50 | - | 6.5e6 | 223432 | 100 | 788-789 | 8252 | 601.7 - 601.9 | 0.005 | 4.95s | 4 |
|
01:07 | 2.2e8 | 1.15e7 | 223432 | 100 | 790-791 | 8253 | 601.7 - 601.9 | 0.005 | 4.95s | 4 |
|
01:26 | 2.3e8 | 1.15e7 | 223448 | 200 | 792 | 8254 | 601.7 - 601.9 | 0.005 | 4.95s | 2 | switched to coasting beam, changed ESR current
|
01:37 | 2.2e8 | 1.15e7 | 223448 | 200 | 793 | 8255 | 601.7 - 601.9 | 0.005 | 4.95s | 2 |
|
01:48 | 2.2e8 | 1.15e7 | 223448 | 200 | 794 | 8256 | 601.7 - 601.9 | 0.005 | 4.95s | 2 |
|
01:58 | 2.2e8 | 1.15e7 | 223448 | 200 | 795 | 8257 | 601.7 - 601.9 | 0.005 | 4.95s | 2 |
|
02:08 | 2.2e8 | 1.15e7 | 223448 | 200 | 796 | 8258 | 601.7 - 601.9 | 0.005 | 4.95s | 2 |
|
02:19 | 2.15e8 | 1.15e7 | 223432 | 100 | 797-799 | 8259 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | changed ESR current again
|
02:45 | 2.17e8 | 1.15e7 | 223432 | 100 | 800-802 | 8260 | 601.7 - 601.9 | 0.005 | 4.95s | 6 |
|
03:12 | 2.12e8 | 1.15e7 | 223423 | 50 | 803-805 | 8261 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | changed cooler current again, restarted NTCAP folder 2022-05-31_19-33-57
|
03:38 | 2.2e8 | 1.15e7 | 223423 | 50 | 806-808 | 8262 | 601.7 - 601.9 | 0.005 | 4.95s | 6 |
|
04:05 | 2.13e8 | 1.15e7 | 223427 | 75 | 809-811 | 8263 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | changed cooler current again, restarted NTCAP folder 2022-06-01_03-12-03
|
04:30 | 2.03e8 | 1.15e7 | 223427 | 75 | 812-814 | 8264 | 601.7 - 601.9 | 0.005 | 4.95s | 6 |
|
04:57 | 2.08e8 | 1.15e7 | 223440 | 150 | 815-817 | 8265 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | changed cooler current again, restarted NTCAP folder 2022-06-01_04-04-32
|
05:21 | 2.1e8 | 1.15e7 | 223440 | 150 | 818-820 | 8266 | 601.7 - 601.9 | 0.005 | 4.95s | 6 |
|
05:57 | 2.08e8 | 1.15e7 | 223440 | 150 | 821-823 | 8267 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | laser power reduced from 46.2mJ to 25mJ
|
06:25 | 1.5e8 | 9e6 | 223440 | 150 | 824-826 | 8268 | 601.7 - 601.9 | 0.005 | 4.95s | 6 |
|
06:56 | 2.1e8 | 1.15e7 | 223440 | 150 | 827-829 | 8269 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | Laser power reset to 48mJ. Last ion injection!!!
|
07:18 | - | 4.7e6 | 223439 | 150 | 830-832 | 8270 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | Last ion injection!!!
|
7:40 | - | 2.0e6 | 223439 | 150 | 833-834 | 8271 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | old beam
|
8:02 | - | 1.0e6 | 223439 | 150 | 835-836 | 8272 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | older beam
|
8:25 | - | 5e5 | 223439 | 150 | 837-8 | 8273 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | very old beam
|
8:48 | - | 3e5 | 223439 | 150 | 839-840 | 8274 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | very very old beam
|
9:10 | - | 2e5 | 223439 | 150 | 841-842 | 8275 | 601.7 - 601.9 | 0.005 | 4.95s | 6 | beam close to retirement - no signal anymore
|
9:33 beam retired. Turned laser off. Stopped NTcap
post beamtime change: replaced "ESR current" by "cooler current" in the comments column. (Wilfried) |
106
|
Wed Jun 1 00:16:02 2022 |
Konstantin, Simon | DAQ | Timing 209Bi thick target |
Timing for the final measurements on 209Bi was determined for the new frequency 1.93777MHz. Peak positions are
PD NO 106
PD SO 141.5
PMT MIDDLE 45 (due to the broad temporal structure difficult to determine)
Center between PDs is 123.75. Difference to PMT MIDDLE is 78.75. This is ~3 bins larger than the half period of 75. |
85
|
Mon May 30 01:13:40 2022 |
Konstantin, Max, Patrick | Detectors | XUV-Detector voltage off |
XUV-Detector shows up to 100 kHz and, therefore, was turned off. |
43
|
Wed May 11 14:48:31 2022 |
Konstantin | DAQ | Voltage range of the Keysight DMM changed from 10V to 1V |
During the setup of the ion beam the voltage range of the Keysight DMM was changed from 10V to 1V. Additionally, the name of the high-voltage divider was adjusted (this changes only the filename of the file, which is stored at the local pc decpc003). |
61
|
Thu May 12 14:04:09 2022 |
Konstantin | DAQ | Go4 Laserstep histograms adapted |
In order to correctly display the gated PMT signals PMT Middle and PMT North some modifications in the Go4 code were necessary. Affected files are TLdaAnlProc.cxx, TLdaAnlProc.h and TLdaInputDescription.h.
Modifications are indicated according to the scheme "//-- HH:MM mm-dd-yyyy: <description>" (an example is given in the attached screenshot). They were needed to
- Create relevant histograms (change of variable MAX_STEP from 10 to 13 and insertion of histogram names "QSWP1MS", "NORTH_GATED" and "MIDDLE_GATED" in the std::string array stepNames.
Both adaptions were made in TLdaInputDescription.h)
- Fill the respective histograms. Changed lines of files TLdaAnlProc.cxx, TLdaAnlProc.h are listed below.
filename | modified lines
|
TLdaAnlProc.h | 47 |
TLdaAnlProc.cxx | 258-261, 381-394, 442-443, 484-494, 536-542
|
|
80
|
Tue May 17 10:38:44 2022 |
Konstantin | Detectors | PMT North changed |
PMT located at the northern window was again replaced by the narrow-band PMT (ET Enterprise type 9423 SB, SN: 650). |
89
|
Mon May 30 09:28:43 2022 |
Ken, Kristian, Simon, Ragan | Detectors | XUV-Detector |
9:27 XUV-Detector is running again.
-5V cathode, 8A main coil, 2.4A sec.coil
Rate without beam was:
Park position: ~300Hz
Pump beneath the detector is now shut down, new rate: <10Hz.
Threshold was adjusted to trigger only on "nice" signals.
Rate with beam in and laser:
Park position: ~400Hz
147mm: ~1.4kHz
with 2e6 ions |
31
|
Tue May 10 22:29:45 2022 |
Jan Glorius | DAQ | DAQ2 shaper settings |
Attached are the MSCF shaper settings applied on Tue May 10 at 22:30. |
32
|
Tue May 10 22:37:10 2022 |
Jan Glorius | DAQ | DAQ2 cabling docu |
|
68
|
Fri May 13 16:55:58 2022 |
Danyal, Carsten, Phillip, Wilfried | General | End of E128 Beamtime |
Dear all,
unfortunately, the E128 experiment must now come to an early end.
The problem(s) with the injection septum are severe and no work around solution (with acceptable risks) could be found.
Several things were tried, but without hope.
Many thanks to everybody involved during the preparation time and the shifts during the beamtime.
Best wishes,
Wilfried |
109
|
Fri Jun 3 12:01:17 2022 |
Danyal | Detectors | switched off XUV detector on 1 June ~17:00 |
I have switched off the XUV detector on 1 June at about 17:00.
The high voltages of the MCP (in ESR laser lab) were turned off and inside the ESR cave the power supplies (coils and cathode) and the crate were turned off.
Also the new "control box" for the detector was switched off.
And I checked (visually) that the XUV detector (and the Cu detector too) were indeed "moved out".
Danyal |
100
|
Tue May 31 19:32:56 2022 |
Carsten, Wilfried, Max | Calibration | electron cooler calibration voltage - frequency |
SetVoltage (DeltaU) | MeasVoltage | freq | Current
|
+1950V | 225247V | 245.06725 | 50 mA
|
+1950V | 225255.4V | 245.0595 | 100 mA
|
+1960V | 225266.4V | 245.062125 | 100 mA
|
+1970V | 225276.8V | 245.064185 | 100 mA
|
+1980V | 225286.6V | 245.066 | 100 mA
|
+1990V | 225296.1V | 245.06775 | 100 mA
|
+2000V | 225305.1V | 245.0695 | 100 mA
|
+1950V | 225255.5V | 245.06 | 100 mA
|
+1940V | 225246.0V | 245.058125 | 100 mA
|
+1930V | 225236.0V | 245.05625 | 100 mA
|
+1920V | 225226.0V | 245.0545 | 100 mA
|
+1910V | 225216.7V | 245.052625 | 100 mA
|
+1900V | 225205.8V | 245.050375 | 100 mA
|
+1850V | 225156.5V | 245.041 | 100 mA
|
+1800V | 225106.6V | 245.031375 | 100 mA
|
+1950V | 225256.1V | 245.060125 | 100 mA
|
+2000V | 225306.0V | 245.06975 | 100 mA
|
+2050V | 225355.8V | 245.079375 | 100 mA
|
+2100V | 225406.0V | 245.0890 | 100 mA
|
+1950V | 225256.2V | 245.060125 | 100 mA
| Change electron current and note down High Vltage and revolution frequency
NTCAP was restarted at the beginning and stopped at the end of the measuring period.
The NTCAP folder is 05-31_17-16-49 (system time 19:11) |
8
|
Wed May 4 22:45:12 2022 |
Carsten, Sebastian | DAQ | Q-Switch Signal (ch15) fixed |
The "dead" Q-Switch signal was converted
two times... This explains why in the
Thorium experiment the Q-Switch signal
(ch15) was empty.
First conversion from TTL to NIM in the
ESR Laserlab and unfortunately the signal
was plugged in a second Level Adapter as
TTL in
the Messhütte and was "again" converted
into a NIM for DAQ1/2.
Now the Q-Switch is converted from TTL to
NIM in the ESR Laserlab. The NIM is
afterwards guided through a Logic Fan
In/Out to the 2 DAQs. |
96
|
Mon May 30 23:18:06 2022 |
Carsten, Rodolfo, Konstantin, Patrick, Ragan, Ken, Kristian, Max, Danyal | Runs | Runs shifts 31.05. |
These are the first runs with 208Bi.
Time | SIS-intensity | ESR-intensity after orbit change | cooler-HV | LMD-DAQ2 file nr | LMD-DAQ1 file nr | scan range(nm) | step width increment (nm) | dwell time(s) | n loops | Particle Detector | comments
|
23:11 | 2.6e8 | - | 225259 | 673-677 | 8182 | 548 - 549 | 0.005 | 2.97s | 4 | 139e3 | Resonance found in first scan.
|
00:09 | ? | - | 225259 | 678-679 | 8183 | 548 - 549 | 0.005 | 2.97s | 2 | 92e3 |
|
00:45 | 2.8e8 | - | 225260 | 680-681 | 8184 | 548 - 549 | 0.005 | 2.97s | 2 | 36e3 | In this run, the ions seemed to be not ejected with the new request, compare picture.
|
01:27 | 2.7e8 | - | 225260 | 684-685 | 8187 | 548 - 549 | 0.005 | 2.97s | 2 | 88e3 | Started delayed due to several issues with both DAQs.
|
01:53 | 2.7e8 | - | 225260 | 686-687 | 8188 | 548 - 549 | 0.005 | 2.97s | 2 | 93e3 |
|
02:20 | 2.7e8 | - | 225260 | 688-689 | 8189 | 548 - 549 | 0.005 | 2.97s | 2 | 87e3 |
|
02:55 | 2.1e8 | - | 225260 | 692-694 | 8190 | 548.335 - 548.535 | 0.0025 | 4.95s | 4 | 88e3 | Scan range changed
|
03:30 | 2.75e8 | - | 225260 | 696-697 | 8191 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | 95e3 | Number of loops reduced to 3
|
03:59 | 2.6e8 | - | 225260 | 698-699 | 8192 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | 89e3 |
|
04:30 | 2.7e8 | - | 225260 | 700-702 | 8194 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | 89e3 | DAQ1 needed to be restarted
|
04:52 | 2.55e8 | - | 225260 | 703-704 | 8195 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | 83e3 |
|
05:18 | 2.62e8 | - | 225260 | 706-707 | 8196 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | 89e3 | DAQ2 needed to be restarted
|
05:44 | ? | - | 225260 | 708-710 | 8197 | 548.335 - 548.535 | 0.0025 | 4.95s | 3 | ? |
|
06:18 | 2.1e8 | - | 225260 | 711-712 | 8200 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 67e3 | we changed the timing by 20 ns. The laser pulse is later now. The file number of DAC1 is correct. I do not know why there is the gap to the previous scan
|
06:51 | 2.4e8 | 6e5 | 225260 | 713-714 | 8201 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 67e3 | xuv-detector parameters were varied
|
07:09 | 2.4e8 | 6e5 | - | 715- | 8202 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 68e3 | mbs did not start
|
07:15 | 2.4e8 | 6e5 | 225260 | 716-717 | 8203 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 68e3 | xuv-detector walklevel and treshhold were lowered
|
07:35 | 2.4e8 | 6e5 | 225259 | 718-719 | 8204 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 68e3 | we changed the timing by additional 20 ns. The laser pulse is later now. - no fluo. visible
|
08:01 | 2.4e8 | 6e5 | 225259 | 720-721 | no file opened | 548.43 - 548.44 | 0.00001 | ~20s | 3 | 71e3 | varying the delay with each laser step to find optimum - It was more a plateau between +-30ns (relative to the night shift's position). We now continue data taking at the same dealy as used by the night shift.
|
08:23 | 2.4e8 | 6e5 | 225258 | 722 | 8205 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 68e3 | xuv-detector walklevel and treshhold were lowered
|
08:28 | 2.4e8 | 6e5 | 225258 | 724-725 | 8208 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 68e3 |
|
08:48 | 2.4e8 | 6e5 | 225258 | 727-728 | 8209 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | ? | xuv-detector in Park-position just as a try
|
09:08 | 2.4e8 | 6e5 | 225259 | 729-730 | 8210 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | ? |
|
09:28 | 2.5e8 | 6e5 | 225259 | 731-732 | 8211 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 73e3 |
|
09:54 | 2.5e8 | 6e5 | 225259 | 734-735 | 8212 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 175e3 |
|
10:19 | 2.5e8 | 6e5 | 225258 | 737-738 | 8214 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 175e3 |
|
10:38 | 2.5e8 | 6e5 | 225258 | 741-742 | 8215 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 64e3 | DAQ2 started late. Missed the first 25 scan numbers Coasting beam measurement from now on
|
10:57 | 2.5e8 | 6e5 | 225258 | 743-744 | 8216 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 87e3 |
|
11:17 | 2.5e8 | 6e5 | 225258 | 745 | 8217 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | e3 | MBS crashed
|
11:18 | 2.5e8 | 6e5 | 225257 | 746-747 | 8218 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 87e3 |
|
11:37 | 2.5e8 | 6e5 | 225257 | 749-750 | 8219 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 87e3 | DAQ2 started late. Missed the first 10 scan numbers
|
11:56 | 2.5e8 | 6e5 | 225257 | 751 | 8220 | 548.36 - 548.51 | 0.0025 | 4.95s | 3 | 78e3 |
|
12:59 | 2.5e8 | 6e5 | 225257 | 752 | 8221 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 78e3 | RF off from now on, laser scan settings not correct. stopped
|
13:06 | 2.5e8 | 6e5 | 225257 | 753 | 8223 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 87e3 |
|
13:26 | 2.5e8 | 6e5 | 225257 | 754 | 8224 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | - | MBS crashed
|
13:29 | 2.5e8 | 6e5 | 225257 | 755-756 | 8226 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 93e3 | RF now really off (wrong cavity was turned off)
|
13:43 | 2.5e8 | 6e5 | 225257 | 757-758 | 8227 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 67e3 |
|
14:08 | 2.5e8 | 6e5 | 225257 | 759 | 8228 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 70e3 |
|
14:22 | 2.2e8 | 6e5 | 225257 | 760 | 8229 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 56e3 |
|
14:35 | 2.2e8 | 6e5 | 225257 | 761 | 8230 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 64e3 |
|
14:50 | 2.2e8 | 6.5e5 | 225256 | 762 | 8231 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 59e3 |
|
15:04 | 2.0e8 | 7e5 | 225255 | 763 | 8232 | 548.385 - 548.485 | 0.0025 | 4.95s | 3 | 58e3 |
|
15:28 | 1.9e8 | ?e5 | 225272 | 764 | 8233 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 66e3 | changed to 200mA cooler current, NTCAP stopped and started again folder: prob. ~31.5.22 13:??, loops decreased to 2
|
15:37 | 2e8 | 6e5 | 225272 | 765 | 8234 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 70e3 |
|
15:47 | 2.5e8 | 6e5 | 225272 | 766 | 8235 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 73e3 | laser gone at loop 2 scan 35 (15:56), Ecooler went also off
|
16:17 | ?e8 | 6e5 | 225272 | 767 | 8236 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 48e3 | bad shot?, probably ecooler reset
|
16:27 | 2.2e8 | 6e5 | 225271 | 768 | 8237 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 77e3 |
|
16:36 | 2.0e8 | 7e5 | 225272 | 769 | 8238 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 65e3 |
|
16:52 | ?e8 | ?e5 | 225272 | 770 | 8239 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 61e3 | we had to restart go4, it was very slow
|
17:02 | 2.0e8 | 7e5 | 225272 | 771 | 8240 | 548.385 - 548.485 | 0.0025 | 4.95s | 2 | 63e3 |
|
17:16 | 2.3e8 | 7e5 | 225248 | 772-773 | 8241 | 548.385 - 548.485 | 0.0025 | 4.95s | 4 | 45e3 | changed to 50mA cooler current, NTCAP stopped and started again folder: 31.5.22 15:18, loops increased to 4
|
17:36 | 2.7e8 | 7.5e5 | 225248 | 774-775 | 8242 | 548.385 - 548.485 | 0.0025 | 4.95s | 6 | 67e3 | loops increased to 6
|
18:05 | 2.5e8 | 7.5e5 | 225248 | 776-777 | 8243 | 548.385 - 548.485 | 0.0025 | 4.95s | 6 | 66e3 |
|
18:31 | 1.7e8 | 6e5 | 225247 | 778-779 | 8244 | 548.385 - 548.485 | 0.0025 | 4.95s | 6 | 44.6e3 | SIS pattern looked strange
|
|
101
|
Tue May 31 19:37:30 2022 |
Carsten, Max, Wilfried | Detectors | SASZ and NASZ voltages switched off |
SASZ and NASZ voltages were switched off bevor we went back to 209Bi. Laser beam stabilization turned off. |
103
|
Tue May 31 21:08:04 2022 |
Carsten, Max | Detectors | XUV detector automation off |
We deactivated the automation of the XUV detector. It stays out now. |