Fri Apr 25 15:45:39 2025, Jan Glorius, DAQ, DR DAQ - MBS & Go4
|
The DR DAQ is located in AP Messhuette, here is the basic info.
[H2]MBS[/H2]
|
Wed Apr 30 19:59:18 2025, Carsten, DAQ, DR_DAQ
|
enum{TRIG_INJECTION=1,
TRIG_NEWRAMP=2,
TRIG_NEWVOLTAGE=3,
|
Mon May 5 11:29:28 2025, Rodolfo, Simon, Nils, Volker, Detectors, PMTs   
|
These are the PMTs installed at the optical detection region:
[TABLE border="1"]
|
Mon May 5 16:26:07 2025, Carsten, DAQ, DR Computer Names and IPs
|
Computers used for control of HV amplifiers and DR microcontrollers are
[LIST]
[*] main control (big Philips monitor): sadpc058.gsi.de - 140.181.98.68
|
Mon May 5 21:13:29 2025, Rodolfo, Laser, Collimation of laser beam
|
Here some parameters for the collimation of the laser beam
Periscope:
|
Tue May 6 15:35:12 2025, Rodolfo, Konstantin, Laser, MRC Settings  
|
Following settings (P-Factor, Directions and Bandwidth) are used for the laser beam stabilization. Full energy (ca. 150 mJ @ 593 nm).
The triggers for the MRC are generated by AFG3102C. The settings (the same for both CH1 and CH2) are saved under "Abrufen / Setup2" |
Wed May 7 14:02:58 2025, Konstantin, DAQ, DAQ Folder
|
The folder of the data acquisition are:
[B]Go4[/B] (atplaser@lxg1297):[B] ~/go4/drals[/B]
|
Wed May 7 17:29:19 2025, Rodolfo, Simon, DAQ, Fast Photodiodes
|
Signal for the fast photodiodes. The laser beam
[LIST]
[*] enters at the Nord-Ost (NO) Arc and
|
Wed May 7 20:31:52 2025, Carsten, DAQ, DRALS cabling (DR part)
|
[B] Green Corner, patch panel to ESR[/B]
All signals from/ to ESR TTL 5V.
Signals are GND decoupled with grey boxes! Don't use the signals directly.
|
Thu May 8 10:11:37 2025, Carsten, General, Schottky spectrum ( 2 Harmonics) from the earlier 209Bi80+ Libelle run (thesis C. Trageser)
|
|
Thu May 8 10:21:10 2025, Carsten, , Link to thesis of C. Trageser NTCAP/ Automated Schottky analysis of cooler voltage and space charge)
|
The thesis of Christian might be helpful w.r.t. to energy calibration as it shows how to link Schottky frequency to ion energy (might be required since
in DRALS the exact ion energy will be difficult to derive from voltage measurements only).
Cf., chapter 6 (thesis is in German)
|
Thu May 8 13:33:13 2025, Rodolfo, Simon, Laser, Camera @ South-West Screen
|
Network camera to monitor the position of the laser beam after the electron cooler.
[B]
apdev017.gsi.de
|
Thu May 8 17:23:51 2025, Shahab, General, Schottky TV
|
setup Schottky TV:
connect to atppc030 using:
|
Thu May 8 19:35:18 2025, Simon, , PMT count rates
|
I tested the PMTs and tried to reduce the background signal as far as possible.
[TABLE border="1"]
|
Thu May 8 22:50:37 2025, Carsten, DAQ, Preliminary Timing Settings
|
Tentatively:
[B]DAC:[/B]
|
Thu May 8 23:36:19 2025, Konstantin, DAQ, Fileserver
|
To save files, the rfio-server must be started on lxg1297 in the folder /data.local2/BiDrals/. The file structure is
[B]/data.local2/BiDrals/208Bi_XXX.lmd[/B]
|
Fri May 9 00:56:07 2025, Konstantin, DAQ, First Files
|
Files 91 + 92 were recorded with the Trigger logic as explained in [URL=https://elog.gsi.de/esr/E0038/15][COLOR=blue][B]ID 15[/B][/COLOR][/URL]. The laser
was scanned from 593 to 592 in 101 steps. 100 laser steps were accumulated at each step. Bunching HF was synthesized by an AFG with a frequency of 1.787327MHZ
and divided by 89 |
Fri May 9 14:21:18 2025, Rodolfo, Accelerator, T-Scrapers
|
Information about the T-Scrapers at the ESR Electron Cooler. The width of the scraper is approx. 9 mm.
There was some issue about the Scraper GECEDS2VU. Herr Boywitt solved this issue and shortly communicated the procedure about how to move it.
|
Fri May 9 17:47:07 2025, Konstantin, DAQ, HV Divider
|
The HV voltage applied at the high-voltage platform of the electron cooler can be observed at [URL=https://lxyrpc01.gsi.de/grafana/d/bec8zc1424agwc/esr-electron-cooler-voltages?orgId=1&from=now-5m&to=now&refresh=5s]HV-divider[/URL]. |
Fri May 9 18:34:08 2025, Konstantin, DAQ, Important diagnoses of analysis parameters    
|
The filling of some histograms in the go4 analysis framework is controlled by the SetParameters.C and SetConditions.C files.
The WinC conditions are applied to all histograms under the folder [B]"Analysis/Histograms/0_Raw/Scaler"[/B]
|