ID |
Date |
Author |
Category |
Subject |
53
|
Wed Jun 12 18:48:43 2024 |
Afternoon shift | Runs | First laser scans |
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
19:15 | 38257 | xx e x | 104416 | 10 | 21-23 | 550.1 - 550.9 | 201 | 2s | 8 | 40 SIS injections. Very short lifetime, further investigation is needed. preset_2024-06-12_19-08-23.mps
|
20:22 | 53056 | xx e x | 104416 | 11 | 24-25 | 550.1 - 550.9 | 201 | 1s | 3 | 40 SIS injections. Only PDetector behing cooler, retracted by 8 mm. Still short lifetime.
|
20:58 | 40500 | xx e x | 104416 | 12 | 26-26 | 550.1 - 550.9 | 201 | 1s | 2 | 40 SIS injections.
|
|
68
|
Thu Jun 13 18:03:36 2024 |
Afternoon shift | Runs | Continuation: First scans |
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
00:01 | | 3e5 | 104438 | 13 | 27-29 | 550.1 - 550.9 | 201 | 2s | 4 | 40 injections, there is still a fast decaying component, may be due to the higher electron current of 200 mA.
Going back to 100 mA in next run and also put a "loose" scraper on the inside.
|
ntcap 2024-06-12_21-47-54
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
00:50 | ? | 3e5 | 104438 | 14 | 28- | 550.1 - 550.9 | 201 | 1s | 6 | 30 injections, e-cooler current at 100mA, scraper GE01GD2ID (north arc, inner scraper) now at position +30mm, medusa crashed when started, laser position was lost and retrieved by resetting the beam position stabilization
|
01:43 | 25700 | 3e5 | 104438 | 15 | 31-32 | 550.1 - 550.9 | 201 | 1s | 6 | e-cooler current 100mA, NA position -30 activated, NI +28, SA -28, DS3HG +10
|
02:32 | 21750 | 4e5 | 104417 | 16 | 33 | 550.1 - 550.9 | 201 | 1s | 3 | reduced iterations due to short life time
|
02:54 | 29200 | 4e5 | 104417 | 17 | 34 | 550.1 - 550.9 | 201 | 1s | 3 |
|
03:16 | (?) | 4e5 | 104417 | 18 | 35 | 550.1 - 550.9 | 201 | 1s | 3 |
|
03:40 | 33700 | 5e5 | 104417 | 19 | 36 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:03 | 31600 | 5e5 | 104417 | 20 | 37 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:25 | 29200 | 4e5 | 104417 | 21 | 38 | 550.1 - 550.9 | 201 | 1s | 3 |
|
04:46 | 27500 | 4e5 | 104417 | 22 | 39 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:08 | 30400 | 4e5 | 104417 | 23 | 40 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:31 | 31300 | 4e5 | 104417 | 24 | 41 | 550.1 - 550.9 | 201 | 1s | 3 |
|
05:56 | 30000 | 3e5 | 104417 | 25 | 42 | 550.1 - 550.9 | 201 | 1s | 3 |
|
06:18 | 24500 | 4e5 | 104417 | 26 | 43 | 550.1 - 550.9 | 201 | 1s | 3 | Morning shift took over: Constantine, Peter, Lukas, Jonas, Dounia
|
06:43 | | | | 27 | 44 | 550.1 - 550.9 | 201 | 1s | 3 | Medusa crashed when starting the laser scan
|
06:57 | | 4e5 | 104417 | 28 | 45-46 | 550.1 - 550.9 | 201 | 1s | 3 | DAQ1 started a bit later since we needed to restart and connect the file server again; after starting laser scan we noticed that the beam is gone on glass disk, reason was likely user error as medusa was set to send the laser to 500.9nm instead of 550.9nm
|
07:27 | 11800 | 4e5 | 104417 | 29 | 47 | 550.1 - 550.9 | 201 | 1s | 3 |
|
07:51 | 23500 | 4e5 | 104417 | 30 | 48 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:14 | 15000 | 5e5 | 104417 | 31 | 49 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:37 | 35300 | 5e5 | 104417 | 32 | 50 | 550.1 - 550.9 | 201 | 1s | 3 |
|
08:58 | 32800 | 5e5 | 104417 | 33 | 51 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 17:32 | 23 500 | 5e5 | 104412 | 35 | 52 | 550.1 - 550.9 | 201 | 1s | 3 | First scan of afternoon shift. 30 Injections. Electron current was set to 30 mA. (Note added 2024.06.14 17:55 - Carsten: To what I remember electron current is 80mA) Dye was changed before this scan. Wavemeter overexposed on the second scan for 2-3 steps, not sure if during the stepping or actual measurement. In the middle of the third scan, medusa crashed.
|
13.06.2024 | xx | 5e5 | 104417 | 36 | 53 | 550.1 - 550.9 | 201 | 1s | 3 | 25 Injections 75Hz particle detector south, medusa crashed immediately, laser controller was locked up. Restarting the laser controller solved the issue.
|
13.06.2024 18:30 | 28 200 | 5e5 | 104412 | 37 | 54 | 550.1 - 550.9 | 201 | 1s | 3 | 25 Injections , DAQ1 no data collected (no start acquisition pressed)
|
13.06.2024 18:50 | 26 000 | 5e5 | 104412 | 38 | 55 | 550.1 - 550.9 | 201 | 1s | 3 | 23 of 25 Injections, two were empty injections from SIS
|
13.06.2024 19:13 | 31 600 | 5e5 | 104412 | 39 | 56 | 550.1 - 550.9 | 201 | 1s | 3 | 29 of 30 Injections, one was empty injections from SIS
|
13.06.2024 19:45 | 34 700 | 5e5 | 104411 | 40 | 57 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections
|
13.06.2024 20:07 | 33 300 | 6e5 | 104411 | 41 | 58 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections. Note: before and after run, the laser was exactly in the middle of the screen, but went to its proper position (slightly left of the middle) during the scan.
|
13.06.2024 20:36 | 29 900 | 7e5 | 104411 | 42 | 59 | 550.1 - 550.9 | 201 | 1s | 3 | 30 of 30 Injections. Note: Before the run ad during the first scan, the laser was above the middle. In the third scan, it was back in its proper position.
|
13.06.2024 20:57 | 28 900 | 7e5 | 104411 | 43 | 60 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 21:18 | 26 000 | 7e5 | 104411 | 44 | 61 | 550.1 - 550.9 | 201 | 1s | 3 | 27 of 30 Injections
|
13.06.2024 21:40 | 18 000 | 6e5 | 104411 | 45 | 62 | 550.1 - 550.9 | 201 | 1s | 3 | out of the first injections 4 were with low energy and 4 did not arrive at all, for the rest, pretty much every second injection is missing. Checking with HKR.
|
13.06.2024 22:02 | 24 900 | 7e5 | 104411 | 46 | 63 | 550.1 - 550.9 | 201 | 1s | 3 | Same thing with injection.
|
13.06.2024 22:23 | 31 700 | 7e5 | 104411 | 47 | 64 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 22:46 | 30 900 | 7e5 | 104411 | 48 | 65 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 23:09 | 33 100 | 7e5 | 104411 | 49 | 66 | 550.1 - 550.9 | 201 | 1s | 3 | Laser shifted to the left
|
13.06.2024 23:33 | 25 200 | 7e5 | 104411 | 50 | 67 | 550.1 - 550.9 | 201 | 1s | 3 |
|
13.06.2024 23:55 | 28 600 | 7e5 | 104411 | 51 | 68 | 550.1 - 550.9 | 201 | 1s | 3 | We moved the laser back to the original position using the MRC
|
14.06.2024 00:52 | 33 100 | 8e5 | 104410 | 52 | 69 | 549.85 - 550.9 | 239 | 1s | 3 | Widen scan range to account for the difference between wavelength vac. vs. wavelength air; Laser shifted to the left
|
14.06.2024 01:20 | 29 800 | 8e5 | 104410 | 53 | 70 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 01:44 | -- | 7e5 | 104410 | 54 | 71 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted; measurement interrupted; connection of MBS to Labview server crashed and MBS and go4 were restarted (DAQ1)
|
14.06.2024 02:03 | 34 700 | 8e5 | 104410 | 55 | 72 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 02:26 | 31 400 | 7e5 | 104411 | 56 | 73 | 549.85 - 550.9 | 239 | 1s | 3 |
|
14.06.2024 02:49 | -- | 7e5 | 104411 | 57 | 74 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted; measurement interrupted
|
14.06.2024 02:51 | -- | 7e5 | 104411 | 58 | 75 | 549.85 - 550.9 | 239 | 1s | 3 | Medusa crashed and was restarted;
|
|
The Medusa crash also crashed the Cobra Laser. We tried to restart it but afterwards the power of the Laser was more than one order of magnitude lower than before. It is not clear what caused the problem, Rodolfo is on his way to GSI now to fix the issue. |
73
|
Fri Jun 14 06:38:41 2024 |
Konstantin Mohr, Bernhard Maass, Peter Micke, Lukas Kau, Dounia Boudefla | Runs | Morning runs | Very successful shift, no crashes until 11:10
Previous night shift had problems with the pulse laser starting around 3am.
Something was loose at the FCU. Rodolfo was called in and was able to fix it.
the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here and in the software.
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
14.06.2024 06:35 | 36500 | 7e5 | 104410 | 59 | 76 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 07:00 | 24600 | 8e5 | 104409 | 60 | 77 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 07:23 | 22900 | 8e5 | 104410 | 61 | 78 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 07:44 | 27000 | 8e5 | 104410 | 62 | 79 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 08:06 | 31000 | 7e5 | 104410 | 63 | 80 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 08:29 | 29500 | 8e5 | 104410 | 64 | 81 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 08:52 | 26800 | 8e5 | 104410 | 65 | 82 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 09:13 | 33500 | 8e5 | 104410 | 66 | 83 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 09:35 | 15700 | 8e5 | 104410 | 67 | 84 | 550.1 - 550.9 | 201 | 1s | 3 | only 10 shots accumulation
|
14.06.2024 09:59 | 35000 | 8e5 | 104410 | 68 | 85 | 550.1 - 550.9 | 201 | 1s | 3 | back to 30 shots accumulation
|
14.06.2024 10:24 | 33600 | 8e5 | 104410 | 69 | 86 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 10:48 | 35500 | 9e5 | 104410 | 70 | 87 | 550.1 - 550.9 | 201 | 1s | 3 |
|
14.06.2024 11:09 | | 8e5 | 104410 | 71 | 88 | 550.1 - 550.9 | 201 | 1s | 3 | data crash, labview disconnected
|
2024-06-14 (note added): Intensities from ESR trafo are not reliable at low internisties, offset can change substantially. Use trafo only with ion numbers sginificantly higher than 1e6. |
77
|
Fri Jun 14 14:44:51 2024 |
Afternoon shift | Runs | Afternoon shift | Pizza shift
the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
From hardware side, the laser can only scan in one direction: long wave length to short wave length, independent of what is written in the tables.
Time | Part. Det. Int | ESR-intensity | cooler-HV | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time(s) | n loops | comments
|
14.06.2024 12:45 | 38900 | 2e5 | 104411 | 72 | 89 | 549.6 - 550.4 | 201 | 1s | 3 | starting with 50 accumulation
|
14.06.2024 13:13 | 41400 | 2e5 | 104411 | 73 | 90 | 549.6 - 550.4 | 201 | 1s | 3 | there was a :10 crash during the injection, laser scan started normally.
|
14.06.2024 13:41 | 39500 | 2e5 | 104411 | 74 | 91 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 14:38 | 67750 | 2e5 | 104412 | 75 | 92 | 549.6 - 550.4 | 201 | 1s | 3 | afternoon shift takes over. After short break from hkr to optimize beam, proceeding with 30 injections, particle number detector might not have been reset, check next run
|
14.06.2024 15:02 | 34500 | 2e5 | 104412 | 76 | 93 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 15:38 | 27500 | 1e5 | 104412 | 77 | 94 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 16:01 | 25300 | 1e5 | 104412 | 78 | 95 | 549.6 - 550.4 | 201 | 1s | 3 |
|
14.06.2024 16:23 | 28700 | 1e5 | 104412 | 79 | 96 | 549.6 - 550.4 | 201 | 1s | 3 | At around 550.2 nm the wavemeter is overexposed
|
14.06.2024 16:48 | xxx | xx | xx | 80 | 97 | xx | xx | xx | xx | empty file , no laser scan, work on laser, while trying to adjust the wavemeter, we tried to move the laser using medusa, at some point, this caused the cobra to go into error mode again, onlz a restart could solve it. As a consequence, we still do not know if the wavemeter coupling is optimal
|
14.06.2024 17:17 | 30000 | 1e5 | 104412 | 81 | 98 | 549.6 - 550.4 | 201 | 1s | 3 |
wavemeter coupling looks good now
|
14.06.2024 17:41 | 56900 | 1e5 | 104412 | 82 | 99 | 549.6 - 550.4 | 201 | 1s | 3 | ratemeter was probably not reset
|
14.06.2024 18:04 | 8100 | 1e5 | 104412 | 83 | 100 | 549.6 - 550.4 | 201 | 1s | 3 | really bad injection (looked fine (?)), low ion count from SIS, they exchanged a cathode @UNILAC
|
14.06.2024 18:42 | 3200 | 3e4 | 104412 | 84 | 101&102(automatically opened new file without command after 1GB?) | 549.6 - 550.4 | 201 | 1s | 3 | problems with accelerator: after deacceleration the beam is gone. Scraper position has moved(?). The cooling water from GSI is too warm, so we had to shut down the laser. NTCAP restarted. And NTCAP restarted again @20:39 for test data for Shahan.
|
|
3
|
Thu Apr 25 16:34:05 2024 |
Rodolfo | Laser | Efficiency of Fluorescein27 | The figure below shows the efficiency of Fluorescein27 vs. Wavelength. Each point was averaged over 500 laser shots.
Dye Laser: Sirah, Cobra Strech.
- Solvent: Ethanol + H2O.
- Concentration Oszillator: 0.4 g + 0,2 g/l NaOH.
- Concentration Main Amplifier: (1/8) Concentration Oscillator .
Pump Laser: InnoLas, SpitLight 1500-30
- Seeding energy: 550 mJ / Pulse.
Note. Dye was mixed on 28.04.2022 |
5
|
Wed May 29 09:49:30 2024 |
Rodolfo | Laser | Laser-beam profile | Using a telescope (Plan-concave lens -500 mm and Plan-convex lens +1000 mm) after the pulsed-dye laser system we got the following laser-beam profile at the Electron-Cooler's position. Dye Laser @ 550 nm / 275 nm |
16
|
Mon Jun 3 19:42:06 2024 |
Rodolfo | Laser | Laser-beam through E-Cooler | UV-laser beam is now going through the Electron-Cooler.
Telescope settings:
- Plan concave lens -500 mm
- Plan convex lens + 1000 mm
- Distance between the both linses: 535 mm
- Position at Long Travel Stage: 60,0 mm |
20
|
Thu Jun 6 07:12:49 2024 |
Konstantin Mohr | Laser | T-Scraper | The T-Scrapers can be driven by DeviceControl, which must be started from a console. To do so, first open the app-launcher and then select "Device Control", which is found under "Betrieb & Steuerung". Then select the appropriate context (should be the name of the experiment). After a short time the device icons will be present. The scrapers can be found in the last four columns and the second last row (see screenshot 1). By default, they should be colored like the following: DS1VU: green, DS1HA: yellow, DS2VU: green, DS2HA: yellow.
To test the scrapers, I made a measurement of the current laser position.
motor | laser position (mm) | screenshot |
DS1VU (GECEDS1VU) | 0 | 2 |
DS1HA (GECEDS1HA) | -10 | 3 |
DS2VU (GECEDS2VU) | 2 | 4 |
DS2HA (GECEDS2HA) | -8 | 5
|
Note that the interlock of DS2VU is overridden when DS2HA is at 3mm. The interlock of DS1VU is overridden when DS1HA is at -4mm. The horizontal scrapers can only be moved when the corresponding vertical motors have reached their inner end positions.
Compared to the last position of Bi the horizontal and vertical positions deviate by approximately 5mm at all axes. However, since the electron cooler was completely dis- and reassembled in the meantime, it seems to be a reasonable starting point. |
30
|
Fri Jun 7 11:14:04 2024 |
Rodolfo | Laser | Trigger-signal for MRC | We are using a 2-Channel Function Generator (from Textronik) to generate the trigger signal for the MRC Beam Stabilization.
The triggers are (according to the MRC nomenclature):
- Tmin = 500 us.
- T1 = 40 us.
- T2 = 460 us. |
43
|
Tue Jun 11 21:38:35 2024 |
Julian, Imke, Rodolfo | Laser | Wavemeter HeNe long time stability | The HeNe was running over night and the measured laser frequency of the wavemeter was logged, see picture.
A significant drift of around 300 MHz is visible during the night.
It is possible, that the laser lab got too cold during the night, as the pulsed laser was not running.
For now, we put the wavemeter in a thermoinsulator box... |
51
|
Wed Jun 12 17:12:13 2024 |
Julian, Danyal, Rodolfo | Laser | Timing für 229Th89+ | We did the timing Ion-Beam Laser-beam time overlap.
The ion revolution frequency (and therefore the bunching frequency) is: 1.5429 MHz. |
69
|
Thu Jun 13 18:23:46 2024 |
Julian | Laser | Restarting the laser controller | If the laser controller is locked up again, the little display will show an error message after an attempted medusa scan.
To restart it, follow these steps:
1.) Unlock the beam stabilization.
2.) shut down medusa and the wavemeter software
3.) Turn off the 2 laser stages
4.) Close the shutter of the pumping laser
5.) Turn off and on the power button of the pulsed laser inside of the black box of the pump laser insertion.
To start again go back through the list. |
70
|
Thu Jun 13 18:38:24 2024 |
Rodolfo Sanchez, Julian Palmes, Sebastian Klammes | Laser | Laser Dye change - new MRC settings - scraper checks of laser beam | We changed the dye at ~3pm.
After some fine adjustements of the laser we had to change the MRC settings, because the laser position changed.
Therefore we checked the position of the laser with the T-Scrapers with the settings from post 49 "First Scraper Scan".
New settings of MRC and screen pic is attached.
The position of the laser on the screen is now the 'same' as yesterday (see attachement). |
1
|
Tue Apr 16 14:42:37 2024 |
Jan | General | G-PAC Proposal | |
2
|
Tue Apr 16 16:28:21 2024 |
Danyal | General | Great | Thanks Jan! |
6
|
Wed May 29 10:54:21 2024 |
Shahab, Jan | General | Gas Target Events with new timing system | Using the new timing system, it is possible to grab any events as pulses, but also as long inter event gates under GENESYS. Many instances of GENESYS can be running on ACO computers, so that there is no conflict, if an instance is running in main control room or another in ESR-Messhütte.
GENESYS Settings:
Inter-Event for 162 up and 163 down, signaling the gas target cycle. The signal can either be:
- Controlled automatically by the control system (sub chain patterns)
- Set to manual operation during setup or experiment.
"activating" the condition in GENESYS by clicking on the small "A" button in the condition. (B) us achieved by "deactivating" the condition in GENESYS, by clicking again on the "A", it turns into a "-". After this, it is possible to manually set the level "hi" or "low" on the left box. As a result a orange bar goes up and down as an indicator. Any time you can switch to automatic change by pressing on "-" again.
Using this new timing system, the old hardware in the main control room is not needed anymore.
Additional Information:
In the old experiments, we used the new control system events but still using the old hardware in the main control room. Please see:
https://elog.gsi.de/esr/E121/13 |
Draft
|
Thu Jun 6 07:54:15 2024 |
Danyal | General | pressure gauges and ion pumps | yesterday (05.06.2024), firstly, the two ion pumps (underneath the fluorescence chamber) were switched off by the vacuum group (Björn Mai, Tel: 3438). Secondly, the two pressure gauges (located near the target) were switched off. The pressure gauge at the fluo chamber is called E02VM3x and the pressure was 5.5e-11 mbar. The other gauge is called IG2 and located further south, it measured 3.7e-11 mbar. That was when ion pumps were still switched on. Since ~10:45 (05.06.2024), the two ion pumps and the two ion gauges are switched off. There was no measurable change in the pressure after switching off the ion pumps. |
32
|
Fri Jun 7 17:10:45 2024 |
Sebastian | General | XUV LED | Maximum XUV LED Voltage: 6.5V |
35
|
Sat Jun 8 13:11:52 2024 |
Carsten | General | Put the go4login in atplaser .bashrc | Put the go4login in atplaser .bashrc in order to avoid that one always has to go to the go4 directory to set the environment variable.
go4login64 is now executed automaticlally and the environments are now set every time you open a (bash) shell.
A fixed version 6.3 is used. The login file might be changed back after the beamtime. |
42
|
Tue Jun 11 21:23:11 2024 |
Julian, Imke | General | How to read out voltage | 1.) If not already opened, open the Medusa Laser Control software, see picture:
2.) In the "view" submenu, choose "proteus status".
3.) At the bottom of the list, the K250 shows the scaled voltage. |
|