Thu May 13 21:22:31 2021, Jan Glorius, Calibration, run0039 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Rate is increasing slowly 200 > 400 Hz, more noise coming in.
Also possible source holder movement!
Resolution still good.
Take this run with care!
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 21:22:00 - 13.05.2021
Stop time: 08:56:31 - 14.05.2021
file name: e127b_run0039.lmd
avrg. rate: 220Hz > 400Hz !!!!
dead-time: 3%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 17:38:13 2021, Jacobus Swartz, Calibration, Runlist for X-ray detector calibrations, 2021
|
Here is a run summary of all the calibration measurements performed so far on the three X-ray detectors. |
Thu May 13 14:16:58 2021, Jan Glorius, Calibration, run0038 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Rate is increasing slowly, from ~70 to ~220 Hz in ~7 hours.
Also the threshold seems to be lower now.
Also possible source holder movement!
Resolution still good.
Take this run with care!
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 14:16:18 - 13.05.2021
Stop time: 21:09:12 - 13.05.2021
file name: e127b_run0038.lmd
avrg. rate: 70Hz > 220Hz!!!!
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 13:24:44 2021, Jan Glorius, Calibration, run0037 - Xray90 calib Pb210 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:24:26 - 13.05.2021
Stop time: 14:06:35 - 13.05.2021
file name: e127b_run0037.lmd
avrg. rate: 70Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 12:21:41 2021, Jan Glorius, Calibration, run0036 - Xray90 calib Ba133_lo d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 133Ba_lo
Distance: 167.5mm
Start time: 12:21:21 - 13.05.2021
Stop time: 13:22:04 - 13.05.2021
file name: e127b_run0036.lmd
avrg. rate: 300Hz
dead-time: 2%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 12:06:51 2021, Jan Glorius, Calibration, run0035 - Xray90 calib Cd109 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 12:06:24 - 13.05.2021
Stop time: 12:16:46 - 13.05.2021
file name: e127b_run0035.lmd
avrg. rate: 1800Hz
dead-time: 12%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 13 11:51:42 2021, Jan Glorius, Calibration, run0034 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************
Detector: GEM1800 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 11:49:53 - 13.05.2021
Stop time: 12:02:24 - 13.05.2021
file name: e127b_run0034.lmd
avrg. rate: 600Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 15:28:55 2021, Jan Glorius, Calibration, run0033 - Xray145 calib Cd109 d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 15:28:01 - 12.05.2021
Stop time: 16:01:57 - 12.05.2021
file name: e127b_run0033.lmd
avrg. rate: 600 Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 14:55:04 2021, Jan Glorius, Calibration, run0032 - Xray90 calib Cd109 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 109Cd
Distance: 167.5mm
Start time: 14:54:36 - 12.05.2021
Stop time: 15:10:18- 12.05.2021
file name: e127b_run0032.lmd
avrg. rate: 1800Hz
dead-time: 12%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 14:16:27 2021, Jan Glorius, Calibration, run0031 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 14:16:07 - 12.05.2021
Stop time: 14:34:00- 12.05.2021
file name: e127b_run0031.lmd
avrg. rate: 600Hz
dead-time: 4%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 13:36:24 2021, Jan Glorius, Calibration, run0030 - Xray90 calib Am241_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
AMPLIFIER JUMPED!
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 13:35:42 - 12.05.2021
Stop time: 14:05:08 - 12.05.2021
file name: e127b_run0030.lmd
avrg. rate: 600Hz
dead-time: 4%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 11:47:41 2021, Jan Glorius, Calibration, run0029 - Xray90 calib Pb210 d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:46:19 - 12.05.2021
Stop time: 13:25:46 - 12.05.2021
file name: e127b_run0029.lmd
avrg. rate: 50Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 11:34:11 2021, Jan Glorius, Calibration, run0027 - crap, 2021
|
|
Wed May 12 11:29:23 2021, Jan Glorius, Calibration, run0028 - Xray90 calib Ba133_hi d=167.5mm, 2021
|
Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM3000 @ 90deg distance
But still operated as Xray3 in DAQ
************************************
Detector: GEM3000 - 90 deg
Source: 133Ba_hi
Distance: 167.5mm
Start time: 11:32:50 - 12.05.2021
Stop time: 11:44:29 - 12.05.2021
file name: e127b_run0028.lmd
avrg. rate: 2500Hz
dead-time: 17%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Wed May 12 10:16:10 2021, Jan Glorius, Calibration, run0026 - Xray145 calib Ba133_hi d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 10:15:51 - 12.05.2021
Stop time: 10:59:35 - 12.05.2021
file name: e127b_run0026.lmd
avrg. rate: 850 Hz
dead-time: 5%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Tue May 11 16:49:23 2021, Jan Glorius, Calibration, run0025 - Xray145 calib Pb210 d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 16:49:03 - 11.05.2021
Stop time: 09:21:13 - 12.05.2021
file name: e127b_run0025.lmd
avrg. rate: 50 Hz
dead-time: 1%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off.
Check for rate spikes |
Tue May 11 14:11:40 2021, Jan Glorius, Calibration, run0024 - Xray145 calib Am241_hi d=305mm, 2021
|
Efficiency calibration in the lab
Detector: GEM3000 - 145 deg
Source: 241Am_hi
Distance: 305mm
Start time: 14:10:55 - 11.05.2021
Stop time: 16:40:27 - 11.05.2021
file name: e127b_run0024.lmd
avrg. rate: 210 Hz
dead-time: 2%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off.
Check for rate spikes |
Thu May 6 21:15:06 2021, Jan Glorius, Calibration, run0022 - Xray35 calib Pb210 d=334mm, 2021
|
Efficiency calibration in the lab
Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:14:44 - 06.05.2021
Stop time: 09:34:10 - 07.05.2021
file name: e127b_run0022.lmd
avrg. rate: 5 Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off.
Check for rate spikes |
Thu May 6 14:30:35 2021, Jan Glorius, Calibration, run0021 - Xray35 calib Pb210 d=334mm, 2021
|
Efficiency calibration in the lab
Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:21:09 - 06.05.2021
Stop time: 21:06:10 - 06.05.2021
file name: e127b_run0021.lmd
avrg. rate: 5 Hz
dead-time: 0%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |
Thu May 6 13:39:02 2021, Jan Glorius, Calibration, efficiency xray 35deg, 2021
|
the attached efficiency curve for the 35 deg. xray detector is made with the runs
e127b_run0015.lmd
e127b_run0016.lmd
e127b_run0017.lmd + e127b_run0019.lmd
All of them were taken in the lab (not in ESR).
This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined. |