HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018
  proton-capture on 118Te, Page 22 of 29  Not logged in ELOG logo
Entry  Fri May 14 11:17:48 2021, Jan Glorius, Calibration, run0043 - Xray145 calib 210Pb d=305mm, 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 11:17:23 - 14.05.2021
Stop time:  13:47:42 - 14.05.2021

file name: e127b_run0043.lmd
avrg. rate: 40Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Fri May 14 13:58:52 2021, Jan Glorius, Calibration, run0044 - Xray145 calib 210Pb d=305mm, 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

90deg TFA diff set from 50 to 200.
Also threshold is slightly increased.

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 13:58:33 - 14.05.2021
Stop time:  16:46:00 - 14.05.2021

file name: e127b_run0044.lmd
avrg. rate: 40Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Sat May 15 20:21:59 2021, Jan Glorius, Calibration, run0045 - Si pulser run, 2021 
Detector: DSSD
Source: CAEN Detector Emulator

Start time: 20:24:14 - 15.05.2021
Stop time:  21:01:34 - 15.05.2021

file name: e127b_run0045.lmd
avrg. rate: 1000Hz
dead-time:  7.5%


Si bias test cycle:
0V > 50V/200uA > 100V/280uA > 150V/370uA > 120V/320uA > 90V/260uA > 60V/210uA > 30V/160uA > 0V
This cycle is run twice i the first 15 minutes and shall be used for calibration of the scaler channels.

Also ESR trafo is running with beam.
Entry  Sun May 16 11:25:25 2021, Jan Glorius, Calibration, run0046 - Xray E-calib 241Am all dets, 2021 
Detectors: Xray1 (GEM1800), Xray2 (GLP), Xray3 (GEM3000)
Sources: different Am241 at each detector

Start time: 11:25:04 - 16.05.2021
Stop time:  11:29:45 - 16.05.2021

file name: e127b_run0046.lmd
avrg. rate: 6000Hz
dead-time:  30%
Entry  Sun May 16 11:54:12 2021, Jan Glorius, Calibration, Xray performance in lab, 2021 
These are the current Xray detector performances in the lab using data from run0046

det DAQ ch threshold resolution @ 60keV
GEM1800 xray1 5keV 495eV
GLP2000 xray2 5keV 430eV
GEM3000 xray3 7keV 525eV
Entry  Sun May 16 12:55:48 2021, Jan Glorius, DAQ, event server lxg1299:6003, 2021 
new event server is set up at lxg1299 on port 6003
It runs as:
/data.local1/gstore/mrevserv64 lxg1275
Entry  Tue May 18 01:03:59 2021, Jan Glorius, Calibration, run0048 - calib all Xray det., 2021 
Efficiency calibration in ESR

************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detector: all
Source @ 90deg: 210Pb
Source @ 35deg: 133Ba_hi
Source @ 145deg: 241Am_hi
Distance: each
Start time: 01:20:23 - 18.05.2021
Stop time:  11:05:30 - 18.05.2021

file name: e127b_run0048.lmd
avrg. rate: 400Hz
dead-time:  3%
Entry  Tue May 18 01:21:27 2021, Jan Glorius, Calibration, run0047 - crap, 2021 
 
Entry  Tue May 18 11:22:57 2021, Jan Glorius, Calibration, run0049 - calib all Xray det., 2021 
Efficiency calibration in ESR

************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detector: all
Source @ 90deg: 210Pb
Source @ 35deg: 133Ba_hi
Source @ 145deg: 241Am_hi
Distance: each
Start time: 11:22:36 - 18.05.2021
Stop time:  13:36:55 - 18.05.2021

file name: e127b_run0049.lmd
avrg. rate: 400Hz
dead-time:  3%
Entry  Wed May 19 13:47:26 2021, Jan Glorius, General, GasTarget Operation, 2021 IMG_20210519_133923186_HDR.jpgIMG_20210519_133629322_HDR.jpg
We have two tasks to take over for the target operation:

1. We have to refill the blue LN2 bottle on the roof of ESR every morning (see 1st attached foto)
For this we use the small silver dewar on the floor there. In case this dewar is empty we have to refill it in the yard.
No need to stop the measurement for this.

2. We have to monitor the LN2-level warning on Ulis web page: http://ulpc24.gsi.de/mediawiki/index.php
In case there is a warning (~12h remaining), we have to refill the big dewar at the target in ESR.
This means we need ESR on TSG with the gate open.

In case the nozzle is clogging and we do not reach reasonable densities anymore and E1 pressure is going down, we have to heat up the nozzle manually.
This can be done at the control panel on the roof of ESR (see 2nd attached foto).
The proceedure is:
1. push "set point"
2. type new value in Kelvin (e.g. 100K for start of heating)
3. press "enter" to apply or "escape" to correct typo
4. monitor E1 pressure (should go up)

Overview page:
https://popps.org/cgi-bin/jet.py?gasart=Wasserstoff
Entry  Sun May 30 10:24:20 2021, Jan Glorius, Calibration, run0212 - X-calib. 90-pb 145-am, 2021 
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 
Source @ 90deg: 210Pb
Source @ 145deg: 241Am_hi
Distance: each
Start time: 10:24:55 - 30.05.2021
Stop time:  11:19:29 - 30.05.2021

file name: e127b_run0212.lmd
avrg. rate: 250Hz
dead-time:  2%
Entry  Sun May 30 11:32:06 2021, Jan Glorius, Calibration, run0213 - X-calib. 90-am 145-pb 35-ba, 2021 
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 + 35 
Source @ 90deg: 241Am_hi
Source @ 145deg: 210Pb
Source @ 35deg: 133Ba_hi
Distance: each
Start time: 11:31:47 - 30.05.2021
Stop time:  13:10:14 - 30.05.2021

file name: e127b_run0213.lmd
avrg. rate: 700Hz
dead-time:  6%
Entry  Sun May 30 13:23:26 2021, Jan Glorius, Calibration, run0214 - X-calib. 90-ba 145-ba 35-am, 2021 
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 + 35 
Source @ 90deg: 133Ba_lo
Source @ 145deg: 133Ba_hi
Source @ 35deg: 241Am_hi
Distance: each
Start time: 13:22:38 - 30.05.2021
Stop time:  14:28:09 - 30.05.2021

file name: e127b_run0214.lmd
avrg. rate: 1150Hz
dead-time:  8%
Entry  Sun May 30 14:38:09 2021, Jan Glorius, Calibration, run0216 - X-calib. 90-ba 35-pb, 2021 
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 35 
Source @ 90deg: 133Ba_lo
Source @ 35deg: 210Pb
Distance: each
Start time: 14:46:46 - 30.05.2021
Stop time:  20:11:08 - 30.05.2021

file name: e127b_run0216.lmd
avrg. rate: 330Hz
dead-time:  3%
Entry  Thu Jun 24 09:43:01 2021, Jan Glorius, DAQ, Unpacker, 2021 
The unpacker used for the production runs of E127b is located at 
/u/litv-exp/unpacker/unpackexps/e127_TO_2021_with_ECL

It is also aliased for litv-exp user as: E127_unpack

This is for the data structure as documented here:
https://elog.gsi.de/esr/E127/346

As of now, this is compiled for ROOT v6-18-04 located here:
/cvmfs/csee.gsi.de/root/618-04/bin/root
Entry  Thu Jan 17 08:45:43 2019, Jan, Detectors, Micron W1 DSSSD - technical information,  6x
Attached are the spec sheets and design drawings of the new DSSSDs from Micron.
Additionally, the 3D model made by Laszlo is also attached as stp and dwg.
Entry  Fri Apr 5 05:46:20 2019, Jan, DAQ, running DRASI for (p,g),  
We now have a drasi DAQ running for our (p,g) experiments.
The RIO we have it installed right now is: r4l-58
It shall be used with the user <litv-exp>

The DAQ-controle shall be done from lxg1275 inside a SCREEN SESSION named "daq"
tab 0 >> drasi 
tab 1 >> logger
tab 2 >> rate
tab 3 >> stream server
tab 4 >> file writing (optional)

To check and control running daq:

login on lxg1275
> ssh litv-exp@lxg1275

enter screen session
> screen -x daq

DRASI
go to tab 0
> ctrl-a 0 
should be quite, no errors showing, no terminal outputs
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./start.sh (for det. triggered daq)
> ./pulser.sh (for internal pulser trigger)
> ctrl-c (for killing daq)

LOGGER
go to tab 1
> ctrl-a 1
should be quite, no errors showing, no terminal outputs
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./logger.sh

RATE MONITOR
go to tab 2
> ctrl-a 2
rates should update each second
"incr" should be event rate per sec 
"send" column should be reasonable value if data is written
full restart proceedure
> ssh r4l-58
> cd esrdaq_2018/r4l-58
> ./rate.sh

EMPTY STREAM SERVER
go to tab 3
> ctrl-a 3 
empty stream should be running  
number of processed events should keep increasing
you should see the number of clients connected
full restart proceedure (on lxg1275)
> empty stream://r4l-58 --server=stream:6002

WRITE A FILE
can be done by litv-exp from any lgx-machine
ROOT FILE:
> e127_unpack --stream=lxg1275 --ntuple=RAW,/data.local1/e127/pre_data/file.root
LMD FILE:
> e127_unpack --stream=lxg1275 --output=/data.local1/e127/pre_data/test.lmd
 
Entry  Fri Apr 5 06:05:55 2019, Jan, DAQ, making changes to the DAQ ,  
If you need to add or change a module of the VME crate you need to follow this proceedure:

1. stop DRASI
2. add/change module in esrdaq_2018/r4l-58/main.cfg 
   a. add module name and VME address: CAEN_V830 (0x00A00000) {}
   b. use BARRIER keyword between the modules
3. restart DRASI (e.g. ./pulser.sh)
4. check empty unpacker data output
   > empty --stream=r4l-58 --print --data
   should run without error and now have your new module event data
5. adjust unpacker in ~/e127/unpackexps/esr
   a. make sure you have a .spec file defining readout of you module, e.g. "vme_caen_v775.spec" 
   b. edit main_user_esr.spec
      - add module as e.g. "tdc[0] = VME_CAEN_V775(geom=0x2, crate=0);"
      - select "geom=0x1" value according to ascending order of all modules (0x1, 0x2, 0x3, ...)
   c. edit det_mapping.hh 
      - add SIGNAL() keyword for your module
      - e.g.: SIGNAL( TDC_1, esr.tdc[0].data[0],TDC_32, esr.tdc[0].data[31], DATA12); 
      - choose data format, e.g. "DATA12", as specified in module .spec-file
   d. compile your unpacker: 
      > make -j
   e. try it out:
      > ./esr --stream=r4l-58 --print --data
      - should run without errors 
      > ./esr --stream=r4l-58 --ntuple=RAW,/dir/file.root
      - root-file should have your new branches with your module data
Entry  Wed Jun 5 11:29:13 2019, Jan, Detectors, position/channel calib.,  
This is the first position calibration done with the detector installed in the holder. The goal is to assign rough positions (top/bottom/left/right) to PreAmp- and Shaper-channels.


PreAmp Channel polarity position
1-1 (Y-1) neg. bottom
1-16 (Y-16) neg. top
2-1 (X-1) pos. left (close to beam)
2-16 (X-16) pos. right (far from beam)
Entry  Fri Jun 7 10:18:23 2019, Jan, Runs, prerun001 - MixedAlpha,  alpha_centered.jpg
save point: lxg1275:/datalocal1/e127/predata/prerun001-root

~2h run

DAQ setup:
32 ADC channels
 ch. 1-16 (section 0) >> y-strips
 ch. 17-32 (section1) >> x-strips
32 TDC channels
 ch. 1-16 (section 0) >> y-strips
 ch. 17-32 (section1) >> x-strips
32 scaler channels (not used)
16 MDPP channels (not used)

Detector setup:
bias -90V
MixedAlpha source ~in central position
ELOG V3.1.5-fc6679b