Sun Jun 22 21:36:48 2025, Carsten, Finn, Zac, Esther, Runs, First laser scans
|
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Note: When doing the accumulation in the ESR the particle number (observed in Lassie Monitor) seems to saturate. This is not actually the case but just an effect of the real signal being buried in other species.
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB | cooler-HV ? | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
22:33 | 10800 | 1.25e9 | 60 | 104513 | 104509 | 106 | 0136 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
22:50 | 10800 | 1.3e9 | 60 | 104513 | 104509 | 107 | 0137 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:07 | 14000 | 1.3e9 | 60 | 104513 | 104509 | 108 | 0138 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:24 | 11500 | 1.3e9 | 60 | 104513 | 104509 | 109 | 0139 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:40 | 13100 | 1.3e9 | 60 | 104513 | 104509 | 110 | 0140 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:57 | 13100 | 1.3e9 | 60 | 104513 | 104509 | 111 | 0141 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:14 | 12700 | 1.4e9 | 60 | 104513 | 104509 | 112 | 0142 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:31 | 13500 | 1.4e9 | 60 | 104513 | 104509 | 113 | 0143 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:48 | 10700 | 1.25e9 | 60 | 104513 | 104509 | 114 | 0144 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
01:05 | - | 1.0e9 | 60 | 104513 | 104509 | 115 | 0145 | 551.91 - 549.41 | 501 | 15 | 1 | yes | BAD FILE(?): beam was killed as some point during the scan, we don't know why; no counts on the particle detector
|
|
Mon Jun 23 10:59:36 2025, Julian, Lukas, Zsolt, Nadine, Konstantin, Runs, Scan routine
|
-Start new injection
->wait for 60 injections to be complete
->note down SIS intensity (at extraction)
->observe Schottky to ensure proper bunching
->check laser position (should be well-centered)
-open file, start Go4
->do '@openfile' in both mbs windows
->start acquisition (green arrow) in both Go4 instances (check indicator 'Current Ev/s' at the bottom, should be green, not red)
->start laser scan
->reset Ch3 of the particle detector (in the electronics hut)
->check laser position (disappears at start of scan then drifts through the image from the left)
-when laser scan is finished:
->stop Go4 acquisition (big red 'STOP' button in the top row)
->do '@closefile' in both mbs windows
->check counts on particle detector
-update elog
-repeat |
Tue Jun 24 18:49:18 2025, Carsten, Rima, Nils, Simon, Runs, Second attempt (Late shift 24.06.)
|
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
19:03 | 14800 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 118 | 147 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -30 +30
|
19:22 | 14900 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 119 | 148 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
19:40 | 14500 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 120 | 149 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -20 +20
|
19:58 | 13800 | 2.0e9 | 60 | 104510 | 104514 | 104514 | 121 | 150 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:18 | 10800 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 122 | 151 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:37 | 10100 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 123 | 152 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:56 | 10500 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 124 | 153 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
21:15 | 9000 | 2.2e9 | 60 | 104510 | 104514 | 104514 | 125 | 154 | 551.91 - 549.41 | 501 | 15 | 1 | yes | after this we change to 45 injections to give more cave access to HHT
|
21:37 | 11500 | 2.3e9 | 45 | 104510 | 104514 | 104514 | 126 | 155 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT
|
22:13 | 13400 | 2.5e9 | 45 | 104510 | 104514 | 104514 | 127 | 156 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT, cooling time in SC7 increased from 20s to 28 s
|
|
Tue Jun 24 22:24:36 2025, Carsten, Finn, Hiroshi, Esther, Rodolfo, Runs, Data taking (night shift 24.06.)
|
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Notes:
1. NTCAP was only started after the fourth scan. The file timestamp is '2025-06-25_00-02-08'.
2. Beam was lost at 03:45 and could not be recovered by the operators. On-call service was called but they have to replace a part, current estimate is 2-3 hours.
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
22:39 | 13500 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 128 | 157 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 60 injections, cooling time (SC 7) 28s -> 32s
|
23:00 | 11500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 129 | 158 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 45 injections
|
23:20 | 10500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 130 | 159 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
23:44 | 9100 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 131 | 160 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
00:05 | 14000 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 132 | 161 | 551.91 - 549.41 | 501 | 15 | 1 | yes | back to 60 injections
|
00:28 | 14300 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 133 | 162 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 75 injections, mesytec channel threshold 0-3 set 255
|
01:26 | 13200 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 134 | 163 | 551.91 - 549.41 | 501 | 15 | 1 | yes | there was an interruption at the source, took some time to get the beam back
|
01:48 | 10500 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 135 | 164 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:10 | 12400 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 136 | 165 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:33 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 137 | 166 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:55 | 10000 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 138 | 167 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
03:18 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 139 | 168 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
|
Thu Jun 26 19:01:46 2025, Carsten, Jan, Simon, Runs, Late shift 26.06.
|
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
19:15 | 23000 | 1.5-3.5 e9 | 33 | 104522 | 104526 | 104526 | 142 | 170 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
19:41 | 19000 | 3.3e9 | 30 | 104522 | 104526 | 104526 | 143 | 171 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -7 and +7
|
20:08 | 17500 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 144 | 172 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | -
|
20:24 | 7000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 145 | 173 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | low Part. Det. Int maybe due to wrond position of part. det.
|
21:19 | 7000 | 3.5e9 | 25 | 104522 | 104526 | 104526 | 146 | 174 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -8 and +8
|
21:40 | 14000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 147 | 175 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | adjusted particle detector position
|
22:04 | 19500 | 3.3e9 | 40 | 104522 | 104526 | 104526 | 148 | 176 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted. |
Thu Jun 26 22:51:04 2025, Hendrik, Finn, Runs, Nightshift data runs
|
After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted.
23:10 Quadrupole magnet stopped working, lost beam in ESR. Contacted on-call ESR service
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
0:15 | 8000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 150 | 177 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
0:28 | 20900 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 151 | 178 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
0:59 | 19000 | 2.3-2.7 e9 | 30 | 104522 | 104526 | 104526 | 152 | 179 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:14 | 14500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 153 | 180 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:30 | 19800 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 154 | 181 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
1:45 | 18000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 155 | 182 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:01 | 6600 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 156 | 183 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:18 | 14500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 157 | 184 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:36 | 14700 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 158 | 185 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
2:45 | 6000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 159 | 186 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:00 | 5500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 160 | 187 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:17 | 5200 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 161 | 188 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:34 | 14300 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 162 | 189 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
3:50 | 17200 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 163 | 190 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:07 | 5000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 164 | 191 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:22 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 165 | 192 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:39 | 14000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 166 | 193 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
4:55 | 17000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 167 | 194 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:12 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 168 | 195 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:26 | 14000 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 169 | 196 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:43 | 13100 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 170 | 197 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
5:56 | 16500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 171 | 198 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
| |
Fri Jun 27 06:13:02 2025, Esther, Konstantin, Runs, Early shift - data taking
|
Notes:
1. TE slits at -10 and +10
2. SIS intensity unstable until about 9:00 after that it became much more stable
3. brief interruption between 09:30 and 10:03 due to interlock issue at SIS
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
06:15 | 16300 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 172 | 199 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
06:30 | 13500 | 1.5-3.5 e9 | 30 | 104522 | 104526 | 104526 | 173 | 200 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
06:44 | 4000 | 1.5-2 e9 | 30 | 104522 | 104526 | 104526 | 174 | 201 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped vertically with scraper 1 after end of laser scan (see entry 66), scraper left in but tilted to allow for injection
|
07:05 | 14800 | 2-3.5 e9 | 30 | 104522 | 104526 | 104526 | 175 | 202 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 1 after end of laser scan (see entry 66), scraper 1 moved back out
|
07:25 | 14000 | 1.1-2.6 e9 | 30 | 104522 | 104526 | 104526 | 176 | 203 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 2 after end of laser scan (see entry 66), scraper 2 left in but tilted to allow for injection
|
07:44 | 4200 | 1-2.5 e9 | 30 | 104522 | 104526 | 104526 | 177 | 204 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | beam scraped horizontally with scraper 2 after end of laser scan (see entry 66), scraper 2 moved back out
|
08:12 | 5700 | 1-2.5 e9 | 30 | 104522 | 104526 | 104526 | 178 | 205 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | !!file in DAQ1 not closed after end of scan!!; ~5 injections missing because nothing came from SIS
|
08:27 | 16000 | 1.5-3 e9 | 30 | 104522 | 104526 | 104526 | 178 | 206 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | !!file in DAQ1 still open from last scan!!
|
08:44 | 11700 | 1.5-3 e9 | 30 | 104522 | 104526 | 104526 | 179 | 207 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
09:00 | 15600 | 2.6-3 e9 | 30 | 104522 | 104526 | 104526 | 180 | 208 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
09:24 | 6500 | 2.8 e9 | 30 | 104522 | 104526 | 104526 | 181 | 209 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:13 | 14600 | 2.2 e9 | 30 | 104522 | 104526 | 104526 | 182 | 210 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:31 | | | | | | | 183 | 211 | | | | | | | BAD FILES: ESR not stopped in SC9, injection during laser scan
|
10:39 | 14500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 184 | 212 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
10:55 | 15400 | 2.3 e9 | 30 | 104522 | 104526 | 104526 | 185 | 213 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:11 | 15500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 186 | 214 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:27 | 15800 | 2.5 e9 | 30 | 104522 | 104526 | 104526 | 187 | 215 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:41 | 5200 | 1.4-2.6 e9 | 30 | 104522 | 104526 | 104526 | 188 | 216 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
11:56 | 16500 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 189 | 217 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:12 | 15000 | 2.9 e9 | 30 | 104522 | 104526 | 104526 | 190 | 218 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:27 | 15000 | 2.8 e9 | 30 | 104522 | 104526 | 104526 | 191 | 219 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
12:44 | 15800 | 2.9 e9 | 30 | 104522 | 104526 | 104526 | 192 | 220 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:21 | 5400 | 2.3 e9 | 30 | 104522 | 104526 | 104526 | 193 | 221 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:42 | 4800 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 194 | 222 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
13:59 | 6200 | 2.6 e9 | 30 | 104522 | 104526 | 104526 | 195 | 223 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| |
Fri Jun 27 14:23:39 2025, Nils, Wilfried, Carsten, Julian, Runs, Runs afternoon shift 27.06.
|
15:30 We have a break due to an issue at UNILAC (quadrupole failed). Also radiation safety interlock was triggered at the fragment separator. This also pushed a faradaycup into the ESR (GE01DC3VP).
Just before the issue during the last bunching, the peak was misaligned to the marker in the Schottky. The marker was sitting on the right flank. So the electron-cooled 229Th did not match with the bunching frequency anymore. We are preparing to adjust the cooler voltage to fix this. For some reason the frequency recovered and is back to its original position. However, our accumulations do not seem to work as well as before, we stagnate much earlier and at a much lower level.
15:50 Beam is back, Carsten is still investigating the drop in intensity we saw before the UNILAC failure.
17:15 TE slits were on -8 and +8, opening to -11 and +11, no change to acquisition pattern, something is still wrong
18:00 We were locked out by HHT people for 30+ mins
18:27 We had to call in ESR support
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
14:24 | 15800 | 2.74 e9 | 30 | 104524 | 104527 | 104528 | 196 | 224 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Carsten optimizing beam at HKR with ESR colleagues
|
14:24 | 6800 | 1.16 e9 | 30 | 104524 | 104527 | 104528 | 197 | 225 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Carsten is back in measurement hut
|
16:30 | 1100 | 1.1 e9 | 40 | 104524 | 104527 | 104528 | 198 | 226 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Searching for reason for small particle number
|
17:00 | 1200 | 0.5 e9 | 30 | 104524 | 104527 | 104528 | 199 | 227 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
19:57 | 7800 | 1e9 | 30 | 104524 | 104527 | 104528 | 200 | 228 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
20:16 | 6800 | 1.5e9 | 30 | 104524 | 104527 | 104528 | 201 | 229 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
20:41 | 7200 | 1.5e9 | 30 | 104524 | 104527 | 104528 | 202 | 230 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
| Restarting some things at ESR
|
21:28 | 6000 | 2.65e9 | 30 | 104524 | 104527 | 104528 | 203 | 231 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
21:47 | 6600 | 2.7e9 | 30 | 104524 | 104527 | 104528 | 204 | 232 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
|
22:08 | 17000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 205 | 233 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7
|
|
| |
Fri Jun 27 22:14:35 2025, Anton, Finn, Carsten, Runs, Runs night shift 27.06. 28.06
|
22:10 SIS Trafo GS09DT scale was changed to 4e8 (not just the lassie monitor)
23:50 Ausfall ESR: GE01 KP01 bis KP05 und KP22 bis KP24; Korrekturelement: GE02KX6 und GE01KX2; beam loss during measurement
01:45 reser press air Scrapers: GE02DD2_G (sc8,11); GE02DS03HG(sc7,8,11); GE01DD2IG(was sc7,10,11)(now7,8,11)
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
22:08 | 17000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 205 | 233 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 |
22:27 | 7200 | 4e8 | 30 | 104524 | 104527 | 104528 | 206 | 234 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 ! small peak on right side, not good scraped |
23:19 | 6000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 207 | 235 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
23:35 | 15000 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 208 | 236 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
23:47 | 00 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 209 | 237 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \ scan not compete, beam loss during measurement |
00:17 | 6200 | 2.7e9 | 30 | 104524 | 104527 | 104528 | 210 | 238 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
00:45 | 6400 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 211 | 239 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:00 | 5100 | 2.75e9 | 30 | 104524 | 104527 | 104528 | 212 | 240 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:15 | 5200 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 213 | 241 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
01:45 | 4900 | 2.3e9 | 30 | 104524 | 104527 | 104528 | 214 | 242 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \bigger peak on right side, badly scraped |
01:58 | 12600 | 2.25e9 | 30 | 104524 | 104527 | 104528 | 215 | 243 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:13 | 12000 | 2.4e9 | 30 | 104524 | 104527 | 104528 | 216 | 244 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:28 | 13900 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 217 | 245 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \small peak on right side, badly scraped |
02:48 | 17000 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 218 | 246 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
02:58 | 13500 | 2.45e9 | 30 | 104524 | 104527 | 104528 | 219 | 247 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
03:13 | 14900 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 220 | 248 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
03:27 | 5200 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 221 | 249 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
03:42 | 13000 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 222 | 250 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
04:08 | 14300 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 223 | 251 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
04:23 | 15400 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 224 | 252 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
04:38 | 13800 | 2.9e9 | 30 | 104524 | 104527 | 104528 | 225 | 253 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:03 | 12800 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 226 | 254 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:23 | 3900 | 2.6e9 | 30 | 104524 | 104527 | 104528 | 227 | 255 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
05:38 | 11200 | 3.e9 | 30 | 104524 | 104527 | 104528 | 228 | 256 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 |
05:53 | 3900 | 2.5e9 | 30 | 104524 | 104527 | 104528 | 229 | 257 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
06:08 | 11700 | low.intensity | 30 | 104524 | 104527 | 104528 | 230 | 258 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 \badly scraped |
|
During whole night shift scraping right peak did not work consistently. Remaining intensity varied from similar peak height compared to torium down to about 20%. |
Sat Jun 28 06:29:58 2025, Stefan, Simon, Konstantin, Runs, Morning shift 28.06.
|
6:30 Problem with ion source, on-call service is called
7:30 There is a short in the ion source and it needs to be changed. Expected down time about 2h
10:30 Ion source was successfully changed, but while ramping up the source a power supply failed. Next on-call service is on their way to GSI.
13:00 ions are back
We changed the dye as the laser intensity was quite low. That didnt help, so Konstantin came in to change the dye cell. That did help.
No scans during morning shift. |
Sat Jun 28 15:28:22 2025, Carsten., Uwe , Runs, Runs with new scan range
|
22:10 SIS Trafo GS09DT scale was changed to 4e8 (not just the lassie monitor)
23:50 Ausfall ESR: GE01 KP01 bis KP05 und KP22 bis KP24; Korrekturelement: GE02KX6 und GE01KX2; beam loss during measurement
01:45 reser press air Scrapers: GE02DD2_G (sc8,11); GE02DS03HG(sc7,8,11); GE01DD2IG(was sc7,10,11)(now7,8,11)
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | TE slits | comments |
|
15:25 | >20000 | 2.5e9 | 30 | - | - | - | 260 | 234 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | detector started too late |
15:52 | 15200 | 2.5e9 | 30 | - | - | - | 261 | 235 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | SIS injection fluctuating |
16:09 | 6000 | 3.3e9 | 30 | - | - | - | 262 | 236 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | accumulation curve in ESR much flatter |
16:31 | 5800 | 3.0e9 | 30 | - | - | - | 264 | 237 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | data not in online DAN; lmd-file: DAQ2_0263.lmd is empty |
16:47 | 10000 | 3.0e9 | 30 | - | - | - | 265 | 238 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | okay |
17:10 | 15500 | 3.0e9 | 35 | - | - | - | 266 | 239 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | okay |
17:45 | 16000 | 3.2e9 | 35 | - | - | - | 267 | 240 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -7 +7 | okay |
18:11 | 4800 | 3.0e9 | 40 | - | - | - | 268 | 241 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -6 +6 | data not in online DAN |
19:00 | 4650 | 3.0e9 | 40 | - | - | - | 269 | 242 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | SIS ions low for first 8 to 10 injections |
19:40 | 14900 | 3.0e9 | 40 | - | - | - | 270 | 243 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | 3-5 empty shots during filling |
19:56 | 14800 | 2.5e9 | 40 | - | - | - | 271 | 244 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | 3-5 empty shots during filling |
20:10 | 5600 | 2.3e9 | 40 | - | - | - | 272 | 245 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | 3-5 empty shots during filling; SIS -> ESR injections 1-3x10**-9 particles variation; accelerator stopped after half filling of ESR; took data for measurement |
21:55 | 14300 | 2.5e9 | 35 | - | - | - | 273 | 246 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
|
17:11 HHT access
17:20 beam back, but problems with SIS intensity
17:35 intensity good, again. 3x10**9 particles in SIS |-
18:30 SIS intensity dropped to 1x10**8 particles
18:45 SIS intensity recovered to 3.75x10**9 particles
Starting from 18:45 slits are set asymmetric:
The dipole magnet TE4MU1 which is in front of our TE slit pair turns to the right. Thus, lower momentum ions are also on the right side. 229Pa89+ has a higher Z=91 than 229Th89+ (Z=90).
Therefore, in the thick target Z=91 , Pa looses more energy than 229Th. With the asymmetric setting we try to further minimize Pa contaminants.
20:05 HHT access
20:15 SIS dipol failure; 1-2h no beam expected; on-call-technical-service is informed
21:10 beam is back. SIS dipole / power supply a fuse needed to be exchanged; we leave the beam for 30min to FRS exclusively
21:35 start taking data again |
Sun Jun 29 03:03:56 2025, Nadine, Finn, Runs, Night shift runs with laser
|
around 22:00 water cooling circuit failure of ESR electron cooler--> ~1h downtime
Wavemeter frequently overexposed around 549,3nm; other wavelengths are fine
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | TE slits | comments |
|
23:25 | 13500 | 2.5e9 | 40 | - | - | - | 274 | 247 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
23:42 | 5000 | 2.5e9 | 40 | - | - | - | 275 | 248 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
00:02 | 5700 | 2.8e9 | 40 | - | - | - | 276 | 249 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
00:26 | 5000 | 3e9 | 30 | - | - | - | 277 | 250 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
00:40 | 5300 | 2.8e9 | 30 | - | - | - | 278 | 251 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
00:55 | 5550 | 3e9 | 30 | - | - | - | 279 | 252 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
01:13 | 16800 | 3e9 | 40 | - | - | - | 280 | 253 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
01:32 | 13300 | 3e9 | 50 | - | - | - | 281 | 254 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
01:50 | 13700 | 3e9 | 40 | - | - | - | 282 | 255 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
02:07 | 14450 | 3e9 | 40 | - | - | - | 283 | 256 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
02:24 | 13800 | 2.8e9 | 40 | - | - | - | 284 | 257 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
02:41 | 12600 | 2.8e9 | 40 | - | - | - | 285 | 258 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | not properly scraped; right peak -100dB |
02:58 | 13200 | 3e9 | 40 | - | - | - | 286 | 259 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
03:15 | 13700 | 2.7e9 | 40 | - | - | - | 287 | 260 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
03:32 | 3400 | 2.7e9 | 40 | - | - | - | 288 | 261 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
03:45 | 5000 | 2.7e9 | 40 | - | - | - | 289 | 262 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
04:12 | 3700 | 3e9 | 50 | - | - | - | 290 | 263 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
04:30 | 4100 | 3e9 | 40 | - | - | - | 291 | 264 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
04:47 | 12800 | 3e9 | 40 | - | - | - | 292 | 265 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
05:10 | 2800 | 2.7e9 | 60 | - | - | - | 293 | 266 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | file 266 was not closed until next file should be opend! Happens at 5:24 in the Morning |
05:28 | 12200 | 2.7e9 | 40 | - | - | - | 294 | 267 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
05:57 | 15000 | 2.7e9 | 40 | - | - | - | 295 | 268 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
06:09 | 00 | 2.7e9 | 40 | - | - | - | 296 | 269 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
| |
Sun Jun 29 06:11:49 2025, Stefan, Konstantin, Runs, Morning Shift 29.06.
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | TE slits | comments |
|
06:09 | 3300 | 2.7e9 | 40 | - | - | - | 296 | 269 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
08:49 | 16000 | 2.5e9 | 40 | - | - | - | 297 | 270 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
09:10 | 4100 | 2.6e9 | 40 | - | - | - | 298 | 271 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | slightly modified the laser position, which was necessary to ensure proper laser-ion superposition after retuning the laser, some bad injections from SIS |
09:2X | 16000 | 2.8e9 | 40 | - | - | - | 299 | 272 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
09:59 | 2200 | 2.4e9 | 40 | - | - | - | 300 | 273 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | One bad injection cycle between the previous measurement and this one. Particle detector reset lately |
10:33 | 3800 | 2.0e9 | 40 | - | - | - | 301 | 274 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
10:52 | 4200 | 2.1e9 | 40 | - | - | - | 302 | 275 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
11:17 | 4100 | 2.1e9 | 30 | - | - | - | 303 | 276 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
11:41 | 4400 | 2.1e9 | 30 | - | - | - | 304 | 277 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
11:53 | 4600 | 2.1e9 | 30 | - | - | - | 305 | 279 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | Bad (empty) injections from SIS |
12:19 | 4600 | 2.5e9 | 30 | - | - | - | 306 | 279 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | Potentially problems with transfer beam line |
12:33 | 3700 | 2.1e9 | 30 | - | - | - | 307 | 280 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
12:50 | 4300 | 2.1e9 | 30 | - | - | - | 308 | 281 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
13:07 | 18300 | 2.5e9 | 30 | - | - | - | 309 | 282 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
13:30 | 2300 | 2.5e9 | 30 | - | - | - | 310 | 283 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | |
14:05 | 1 | 2.5e9 | 30 | - | - | - | 311 | 284 | 550.610 - 548.110 | 501 | 15 | 1 | yes | yes | TE slits -9 +5.5 | Particle detector hasn't moved in |
|
We did a recalibration of the FCU. After this (7:40) a SIS power supply failure was detected, and we are waiting now for the on-call service.
HHT cave access around 10:06 during a short break due to a source failure.
Every once in a while we have empty shots from SIS. |
Sun Jun 29 18:04:11 2025, Carsten, Simon, Runs, Late shift 29.06.
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments |
18:10 | 18000 | 2.8e9 | 30 | 104525 | - | - | 287 | 314 | 548.51 - 548.01 | 201 | 15 | 3 | yes | Part. Det Int. after 3 loops |
18:26 | 14000 | 2.8e9 | 30 | 104525 | - | - | 288 | 315 | 548.51 - 548.01 | 201 | 15 | 2 | yes | Part. Det Int. after 2 loops |
18:41 | 16000 | 2.8e9 | 35 | 104525 | - | - | 289 | 316 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
18:57 | 21000 | 2.0e9 | 40 | 104525 | - | - | 290 | 317 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
19:14 | 13000 | 2.0e9 | 40 | 104525 | - | - | 291 | 318 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
19:53 | 15000 | 2.6e9 | 35 | 104525 | - | - | 292 | 319 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
20:08 | 15000 | 2.0e9 | 40 | 104525 | - | - | 293 | 320 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
20:25 | 19000 | 2.4e9 | 40 | 104525 | - | - | 294 | 321 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
20:44 | 3000 | 2.4e9 | 45 | 104525 | - | - | 295 | 322 | 548.51 - 548.01 | 201 | 15 | 2 | yes | very low intensity, so we kept Go4 off and only ran mbs |
21:04 | 23000 | 2.8e9 | 30 | 104525 | - | - | 296 | 323 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:21 | 24000 | 2.8e9 | 30 | 104525 | - | - | 297 | 324 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:39 | 18000 | 2.8e9 | 30 | 104525 | - | - | 298 | 325 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:57 | 6400 | 2e9 | 40 | 104525 | - | - | 299 | 326 | 548.51 - 548.01 | 201 | 15 | 2 | yes | | low intensity, no Go4
|
All Particle detector values are after the respective number of scans.
AT 1700 Bernd came in since we couldn't properly accumulate. Bernd adjusted the deceleration HF and slightly changed the extraction kicker from 1.9mrad to 1.7 mrad. High intensity of 229/89.
At 19:35 again problems with transmission SIS -> ESR. Put the Kciker (GTE3Mu1/HIKICK) back to 1.9 mrad (orignal value). |
Sun Jun 29 22:06:19 2025, Carsten, Alex, Hendrik, Runs, Night shift 29.06.
|
[TABLE border="1"]
Time | Part. Det. Int | SIS-intensity| accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments |-
22:16 | 4800 | 2e9 | 30 | 104525 | - | - | 300 | 327 | 548.51 - 548.01 | 201 | 15 | 2 | yes | only few ions, did not start Go4 |-
22:28 | 9000 | 2e9 | 15 | 104525 | - | - | 301 | 328 | 548.51 - 548.01 | 201 | 15 | 2 | yes | After this run: Shaky SIS intensity, cathode is being changed |-
22:53 | 7000 | 2e9 | 35 | 104525 | - | - | 302 | 329 | 548.51 - 548.01 | 201 | 15 | 2 | yes |-|-
[/TABLE]
23:20 Failure of GE01KP04, GE01KP05, GE01KP22 (or GE01KP24) devices in the ESR.
Problems with gtemu3-hick (??) , take injections since
01:49 Bernd tried to make things work by switching off and on some of the dipole correction coils. He was even able to get a quite good intensity by leaving one of the coils switched off (although with a slight shift of frequencies visible in Schottky spectrum) but then the ESR electron cooler has failed probably due to problems with the cooling water. So, the decision was taken to stop the measurement at this point. The laser was switched off. |
Fri Jun 6 14:27:10 2025, Rodolfo, Konstantin, Laser, MRC Beam Stabilization Presettings  
|
These are the settings for the Beam Stabilization from MRC.
Laser-beam energy ca. 18 mJ @ 275 nm |
Wed Jun 18 03:12:25 2025, Rodolfo, Laser, Beam Stabilization
|
The laser beam pointing has changed, even thought the MRC settings are the same as given in Entry 5 (this elog). This could be caused by a droping of the laser output energy which is now 13,8 mJ @275 nm |
Wed Jun 18 07:59:00 2025, Julien, Konstantin, Laser, Phase-offset   
|
We recognized that the output power of the pump laser was instable when externally triggered with 200kHz (stand-bye frequency of ESR). Therefore, we changed the coarse timing offset of the pockels cell trigger compared to the flash lamp pulse from 215µs to 219.9µs.
Thereby, the output power increased from ~600mJ to 620mJ, and the PTP stability improved from 15% to 10%. The RMS stability improved from 2% to 1.2 %. With these settings we had UV pulse energies of 20mJ. |
Wed Jun 18 08:07:17 2025, Julien, Konstantin, Laser, new FCU Calibration 
|
UV power was not stable but dropping from ~18 mW to 13 mW after opening COBRA shutters.
New calibration of frequency conversion unit:
Wavelength/nm Position/Steps
554,0001 383029
553,0000 386604
552,0001 390229
551,0000 393954
550,0001 397679
549,0001 401404
548,0001 405104
547,0001 408829
now stable at ~19-20 mJ depending on wavelength
Table of dataset: Attachement 1
Plot of dataset: Attachement 2 |
Fri Jun 20 11:25:32 2025, Julien, Kristian, Konstantin, Laser, Automated laser scans
|
In contrast to entry E12, the cabling for the automated start of the laser scan is now connected to IO3 of sddsc221. |
|